APPENDIX B: Nebulizer Assembly & Maintenance
ORANGE COVER
10
O-RING
To Disassemble Nebulizer: Remove Inlet (8) with hose from Housing (2). Remove Housing with the
Reservoir from the Manifold on the front panel with a pulling and twisting motion. Separate Reservoir (1)
and Housing with a breaking and pulling motion. Remove the exposed aerosol generating parts by
grasping the Tab (3) with thumb and forefinger and pull Target Piece (4) straight out of the Housing.
Remove Cap (5) and Dip Tube (7). Leave the Jet (6) and the O-ring (8) in place (DO NOT take the jet out
from the nebulizer housing). See Figure 5-E & F p27.
To clean or remove obstruction: Soap or detergent may be used to clean all of the above parts. Flush
parts with water. Be careful not to enlarge any of the holes in the parts. Air-dry only before reassemble.
To Assemble Nebulizer:
Place BOTH of the o-rings over the Jet(6). Using dip Tube(7) as handle, insert the Cap(5) into the Jet(6)
inside the Housing(2). Hold the Housing(2) with your left hand as shown by the above picture, while
bending the Dip Tube(7), push the Cap(5) with your right index finger into the underside of the O-ring until
it snaps. Grab the Tap(3) and slide the Target Piece(4) between the guides of Housing(2) and Cap(5) as
far as it will go. A few attempts may be needed for it to work. Snap Reservoir(1) back into Housing(2).
Note:
Always empty oil from the nebulizer and cover the opening with the provided orange cap (10) when it is
not being used to prevent dirt from getting in through the inlet opening.
O-RING
HOUSING
2
RESERVOIR
1
Note: Nebulizer Assembly consists of part #3,4,5,6,7,9, 11
only.
TARGET
PIECE
4
TAB
3
19
O-RINGS
9
JET
6
CAP
5
INLET
8
DIP TUBE
7
FILTER SCREEN
11