INSTRUCTION MANUAL
Main Unit:
Automatic
Blood Pressure Monitor
HEM-7112
Model
START/STOP button
5331395-1B
Components:
INTRODUCTION
Thank you for purchasing the OMRON HEM-7112 Automatic Blood Pressure Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood's movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within
the range of the designated arm cuff, following the instructions in this instruction
manual. It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
IMPORTANT SAFETY INFORMATION
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
1. Press the
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet hands.
2. Install 4 "AAA" size batteries so the + (positive) and -
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
3. Replace the battery cover.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their
consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
BATTERY REPLACEMENT
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in
incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when
When the Battery low indicator appears on the display, turn the monitor off and
taking measurements at night. This could result in injury when the air pressure in
remove all the batteries. Replace with four new batteries at the same time. Long-life
the air tube is increased.
alkaline batteries are recommended.
• Do not leave the cuff wrapped on the arm if taking measurements during the night.
This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near
the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to "TAKING A MEASUREMENT". If the cuff
is over inflated, it can cause internal bleeding.
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the
power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty
of clean water.
• Use only four "AAA" alkaline batteries with this unit. Do not use other types of
batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the
• Replace old batteries with new ones immediately. Replace all four batteries at the
same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising or
eating.
2. Remove tight-fitting clothing from your upper
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main
unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
3. Apply the cuff to your left upper arm so the
4. Wrap the cuff firmly in place around your arm
APPLYING THE CUFF ON THE RIGHT ARM
KNOW YOUR UNIT
When taking a measurement using the right arm, use this instruction for Step 3.
Apply the cuff to your right upper arm so the
Display
arrow and the air tube runs along the inside of
Battery compartment
your arm. The bottom of the cuff should be
approximately 1 to 2 cm above your elbow.
NOTE:
AC adapter jack
Air jack
1. Sit on a chair with your feet flat on the floor. Place
your arm on a table so the cuff is level with your
heart.
INSTRUCTION MANUAL
Automatic
Air tube
Blood Pressure Monitor
Keep still and do not talk during measurement.
Model
HEM-7112
5331395-1A
INTRODUCTION
Thank you for purchasing the OMRON HEM-7112 Automatic Blood Pressure Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood's movement through your
monitor does not need a stethoscope so the monitor is simple to use.
brachial artery and converts the movements into a digital reading. An oscillometric
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within
2. Press the START/STOP button.
the range of the designated arm cuff, following the instructions in this instruction
manual. It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Air plug
All symbols appear on the display.
Four "AAA" alkaline
Instruction
Standard Adult Arm Cuff
The cuff starts to inflate automatically.
(LR03) batteries
Manual
Remain still and do not move your arm until the entire measurement process is completed.
UNIT DISPLAY
Note: If your systolic or diastolic
pressure is outside the standard
Systolic blood pressure
range (above 135/85 mmHg), the
NOTE: To stop the inflation or measurement, push the START/STOP button. The
Heartbeat symbol ( ) will blink.
Refer to "Taking a Measurement" .
Diastolic blood pressure
3. As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Heartbeat symbol
Memory symbol
Symbol (
(Displayed when viewing
1. Flashes during measurement
2. If flashing after measurement
values stored in memory)
complete indicates blood pressure
Battery low indicator
out of recommended range
Deflation symbol
Pulse display
BATTERY INSTALLATION
4. When the measurement is complete, the arm cuff completely
CAUTION
deflates. Your blood pressure and pulse rate are displayed.
Use only 1.5V "AAA" alkaline batteries with this device.
indicator on the battery cover and slide
the cover off in the direction of the arrow.
5. Press the START/STOP button to turn the monitor off.
(negative) polarities match the polarities of the battery
compartment as indicated.
NOTE: The monitor will automatically turn off after five minutes.
Important:
• If your systolic or diastolic pressure is outside the standard range,
Battery low indicator
SPECIAL CONDITIONS
WARNING
If battery fluid should get in your eyes, immediately rinse with plenty of
clean water. Contact a doctor immediately.
If your systolic pressure is known to be more than 140 mmHg,
CAUTION
1.
Press the START/STOP button to turn the unit on.
Dispose of the device, components and optional accessories according to
Measurement starts.
applicable local regulations. Unlawful disposal may cause environmental
2. When the cuff starts to inflate, press and hold the START/STOP button until the
pollution.
monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
APPLYING THE ARM CUFF
CAUTION
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and
NOTES: • Do not apply more pressure than necessary.
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
main unit.
Every time you complete a measurement the monitor stores the blood pressure and pulse
rate in the memory. The monitor automatically stores the last set of measurement values
(systolic pressure, diastolic pressure, and pulse rate).
1. To view your last measurement values, press and hold the START/STOP button.
The most recent set of measurement values appears on the display.
arm.
Left Arm
arrow is centered on the inside of your arm and
2. Press the START/STOP button to turn the unit off.
1 to 2 cm
aligned with your middle finger.
The air tube runs down the inside of your arm.
TO DELETE THE VALUES STORED IN MEMORY
The bottom of the cuff should be approximately
1. Press and hold the START/STOP button until the measurement value is displayed.
1 to 2 cm above your elbow.
Then keep the button pressed down for about 10 seconds until the display appears as
Arrow
Air Tube
illustrated below.
using the cloth fastening.
2. Press the START/STOP button to turn the unit off.
Right Arm
Maintenance
To protect your unit from damage, please observe the following:
Arrow
Be careful not to rest your arm on the air tube. This will restrict the flow of
air to the cuff.
TAKING A MEASUREMENT
monitor will stop inflating, start deflating, and will turn off.
) flashes.
Calibration and Service
Storage
Keep the unit in its storage case when not in use.
1 .
2 .
the heartbeat symbol will blink when the measurement result is
displayed. Recent research suggests that the following values can
be used as a guide to high blood pressure for measurements taken
at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
CAUTION
1 .
Inflating to a higher pressure than necessary may result in bruising where
the cuff is applied.
2 .
• The monitor will not inflate above 299 mmHg.
USING THE MEMORY FUNCTION
No power.
No display appears on the unit.
Measurement values appear too
high or too low.
NOTE: The monitor will automatically turn off after five minutes.
CARE AND MAINTENANCE
Model:
Display:
Measurement Range:
• Do not subject the main unit and the cuff to extreme temperatures, humidity,
Accuracy:
moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
Inflation:
• Do not inflate the arm cuff over 299 mmHg.
Deflation:
• Do not disassemble the unit.
Rapid Air Release:
• Do not subject the unit to strong shocks or vibrations (for example, dropping
the unit on the floor).
Measurement Method:
• Do not use volatile liquids to clean the main unit. The unit should be cleaned
Rating:
with a soft, dry cloth.
Power Source:
• Use a soft, moistened cloth and soap to clean the arm cuff.
Battery Life:
• Do not wash the arm cuff or immerse it in water.
Applied Part:
• Do not use petrol, thinners or similar solvents to clean the arm cuff.
Protection against
Electric Shock:
Operating Temperature
/Humidity:
Storage Temperature
/Humidity/Air Pressure:
Main Unit Weight:
Cuff Weight:
• Do not carry out repairs of any kind yourself. If a defect occurs, consult
Main Unit Dimensions:
your local authorised OMRON distributor or dealer.
Cuff Size:
• The accuracy of this blood pressure monitor has been carefully tested and is
Cuff Circumference:
designed for a long service life.
Memory:
• It is generally recommended to have the unit inspected every two years to
Contents:
ensure correct functioning and accuracy. Please consult your local
authorised OMRON distributor or dealer.
Optional Accessories:
Notes:
• Subject to technical modification without prior notice.
• Disposal of this product and used batteries should be carried out in accordance
Unplug the air tube from the air connector.
with the national regulations for the disposal of electronic products.
Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
Do not store the unit in the following situations:
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
• If the unit is wet.
Directive).
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust
• This blood pressure monitor is designed according to the European Standard EN1060,
or corrosive vapours.
Noninvasive sphygmomanometers Part 1: General Requirements and Part 3:
• Locations exposed to vibrations, shocks or where it will be at an angle.
Supplementary requirements for electromechanical blood pressure measuring systems.
OPTIONAL PARTS
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
Medium Arm Cuff
Large Arm Cuff
AC Adapter "R"
(cellular) telephones, medical devices in use may be susceptible to
Arm circumference
Arm circumference
electromagnetic interference from other devices.
22 - 32 cm
32 - 42 cm
Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
Using the Optional AC Adapter
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
Insert the AC adapter plug
This medical device manufactured by OMRON HEALTHCARE conforms to this
into the AC adapter jack on
EN60601-1-2:2007 standard for both immunity and emissions.
the side of the main unit.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
Plug the AC adapter into an
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
electrical outlet.
operation of the device in case the distance is shorter.
To disconnect the AC adapter, unplug the AC adapter from the
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
electrical outlet first and then remove the AC adapter plug from the
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
main unit.
Documentation is also available at www.omron-healthcare.com.
ERROR INDICATORS
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
SYMBOL
CAUSE
Remedy
should not be disposed of, with other household wastes at the end
Take another measurement and
of its working life. To prevent possible harm to the environment or
Monitor could not
remain still until the measurement
human health from uncontrolled waste disposal, please separate
detect pulse wave.
is complete.
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Cuff under-inflated.
Household users should contact either the retailer where they purchased this
Cuff not applied
Remove the arm cuff.
product, or their local government office, for details of where and how they can
correctly.
Read "Taking a Measurement".
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
Wait 2-3 minutes.
of the purchase contract. This product should not be mixed with other commercial
Cuff over-inflated.
Take another measurement.
wastes for disposal.
This product does not contain any hazardous substances.
Replace the four batteries. Refer to
Batteries are worn.
"Battery Installation".
Device error.
Contact Customer Service.
Asia Pacific HQ
TROUBLESHOOTING TIPS
Production Facility
PROBLEM
CAUSES AND SOLUTIONS
Replace all four batteries with new ones.
This OMRON product is manufactured under the strict quality system of
Check the battery installation for proper placement
OMRON HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood
of the battery polarities.
pressure monitors, which is the Pressure Sensor, is produced in Japan for assembly.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
cuff, may affect your blood pressure.
Review the section "Taking a Measurement".
HEM-7112-AP_A_M03_110217.pdf
SPECIFICATIONS
HEM-7112
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180 beats/min.
Pressure: ±3mmHg
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
DC6V
4W
1.5V 4 "AAA" batteries or AC adapter (optional)
Approx. 300 uses with 4 new alkaline batteries
= Type B
Internally powered ME equipment
(When using only the batteries)
= Class II ME equipment
(Optional AC adapter)
10ºC to 40ºC /30 to 85% RH
-20ºC to 60ºC /10 to 95% RH /
700-1060 hPa
Approximately 245g
not including batteries
Approximately 120g
Approximately 128 mm × 104 mm × 64 mm
Approximately 146 mm (w) × 446 mm (l)
Cuff tube (600 mm)
Fits arm circumferences 220 to 320 mm
The last reading
Main Unit, Standard Adult Arm Cuff, Battery Set,
Instruction Manual
Large Cuff 320 to 420 mm, AC adapter
Please read this instruction manual carefully before using the device.
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08, Alexandra Technopark
Singapore 119967