HMM Diagnostics smart LAB mini User Manual page 29

Self-monitoring blood glucose meter
Table of Contents

Advertisement

Memory error
Display: „005 Error"
Solution: Replace batteries first. If "Err"still oc-
curs, please contact your local supplier.
Test strip already used or wet
Display: „Err" & „Test strip symbol"
Solution: Use a new test strip.
Temperature too high
Display: „Ht" & „Thermometer symbol"
The operating temperature is too high (above
the required temperature range from 10°C -
40°C (50°F - 104°F)). The error is a warning
that a continuation under these conditions
might lead to wrong blood glucose readings.
Solution: Take the meter to a location with
appropriate working temperature and wait for
the next measurement until the error does not
occur again.
29