Section 3 - Operation

3.1 Power Requirements

Caution! Be sure that the power supply is the same voltage as specified on the nameplate
With the chiller plugged in (before pressing the P
standby (....). If there is no response, check that the circuit breaker on the rear of the chiller is in the
ON position. An extension cord is not recommended.
Front Face Plate

3.2 Temperature Setting and Adjustments

After setting up and filling the chiller:
1.
Press the P
chiller is turned on, there is a 60 second time-out period. During this time the controller is
performing a self-diagnostic and a balancing of pressure across the compressor inlet and outlet.
(a balanced system reduces the AC current required to start up the compressor and will result in
an extended compressor life.) The pump is active during the start delay period. When the time
out period reaches zero, the compressor starts and the chiller is then fully operational.
Upon completion of the self test, the actual fluid temperature is displayed. Your desired
operating temperature can now be set.
2.
Press the S
changed. Turn the S
after pressing the S
keypad inactivity. The degree light stops flashing and the display shows the actual temperature.
3.
The S
button. Allow sufficient time for the chiller to stabilize at the desired temperature.

3.3 Selection of Celsius or Fahrenheit

To change the readout to °C or °F:
1.
Press and hold the Set/Menu button until the display reads "UNITS".
2.
Press Set/Menu again then turn the Setting Adjustment knob and select °C or °F.
3.
Press Set/Menu or the setting will be accepted after a few seconds.
optional heater only)
P
G
RESSURE
(On Positive Displacement
Pump models only)
O
button. The LED display indicates (8888), the power up self test. When the
OWER
N
/M
button. The degree lights flash indicating the temperature can be set or
ET
ENU
A
ETTING
DJUSTMENT
/M
button or will be automatically accepted after a few seconds of
ET
ENU
T
may be checked at any time by pressing and releasing the S
ET
EMPERATURE
O
button), the display responds by showing
OWER
N
(For models with
LED
D
ISPLAY
S
/M
,
ET
ENU
AUGE
B
UTTON
knob to the desired setting. This setting is accepted
9
°C
°F
S
OR
ETTING
D
A
K
ISPLAY
DJUSTMENT
NOB
P
OWER
B
UTTONS
/M
ET
ENU