Automatic Test and Calibration Station

Quick Reference Guide

Advertisement

Table of Contents
loading

Summary of Contents for BW Technologies MicroDock II

  • Page 1: Quick Reference Guide

    Automatic Test and Calibration Station Quick Reference Guide...
  • Page 2: Limited Warranty And Limitation Liability

    USA: 188-749-8878 Europe: +44(0) 1295 700300 Email us at: info@gasmonitors.com Visit BW Technologies by Honeywell website at: Limited Warranty and Limitation Liability ’ S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES TORT OR RELIANCE OR ANY OTHER THEORY...
  • Page 3: Safety Information - Read First

    To ensure personal safety, read First before using the MicroDock II base station. The MicroDock II Automatic Test and Calibration Station (“the base station”) provides automated calibration and bump testing for the GasAlert Extreme, GasAlertClip Extreme, GasAlertMi- cro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlert- Max XT detectors.
  • Page 4 MicroDock II Quick Reference Guide Display Elements Item Function AC power Batteries fully charged Batteries half-charged Batteries at low level MultiMediaCard (MMC) MultiMediaCard (MMC) not inserted Test pass and option enabled Test fail and option disabled Cursor and sensor disabled...
  • Page 5: User Options Menu

    Format MMC tiMediaCard (MMC). NOTE: This feature erases all current data. Refer to the MicroDock II User Manual. Inlet Sel selects a gas inlet. If auto is displayed, the base sta- tion automatically selects the correct inlet for the test.
  • Page 6: Installation

    MicroDock II Quick Reference Guide Installation Item Description Inlet filter assembly C-cell batteries (4) Battery cover Philips pan head retaining screws (2) MultiMediaCard (MMC) Battery compartment Charger port USB port Power port Exhaust outlet Gas cylinder Demand flow regulator Calibration gas hose...
  • Page 7 Confirm Inlet Setup. Mounting the MicroDock II Base Station and Docking Modules For wall mounting instructions, refer to the MicroDock II Base Station User Manual. Warning Loosen the retaining screws from the battery cover. Do not remove the screws from the cover.
  • Page 8 Important: If the GasAlertMicro 5/PID/IR detector is fitted with a pump module, the diffusion adapter must be removed from the docking module. Refer to the MicroDock II Base Station User Manual. • Insert the GasAlertMicro 5/PID/IR (LCD facing up) at a 45°...
  • Page 9: Using The Base Station

    • Close the lid and press until the release tabs click. Push the pump connector towards the detector pump. Ensure green is visible in the lower indicator in the pump connector. If it is not visible, the MicroDock II base station will not recognize the detector. Note The detector will briefly alarm when the pump connector is indicated.
  • Page 10 MicroDock II Quick Reference Guide The base station pushbuttons are not labelled. The base station is operated by pressing the C pushbutton that is located directly below the option that displays on the LCD. Activating the Station To activate the station, press and hold C (the left- most button) until the following screen displays.
  • Page 11 Press C exit to return to normal operation. For more information on setting up inlets, gas types, and con- centration levels, refer to Inlet Setup in the MicroDock II Base Station User Manual. Bump Test A bump test is performed to confirm that the detector is responding to gas, and that the audio and visual alarms are operational.
  • Page 12: Data Transfer

    The countdown for time remaining displays beside Purge. = calibration has passed = calibration has failed If any test fails, refer to Troubleshooting in the MicroDock II Base Station User Manual. The detector deactivates. Data Transfer (GasAlert Extreme, GasAlertMicroClip, GasAlertMax XT only) This feature transfers datalogs from the detector to the base station MultiMediaCard (MMC).
  • Page 13: Event Logging

    5 minutes of inactivity. BW Technologies by For more information about the Automatic Datalog Download option, refer to the MicroDock II Base Station User Manual. Eventlogging Bump tests and calibrations are recorded on a MultiMediaCard (MMC). The MMC is located on the base sta- tion in the battery compartment.
  • Page 14: Charging The Battery Pack

    Charging the Battery Pack Warning There is a maximum of six charging docking modules can be installed on the MicroDock II base station (six charging modules plus four non-charging modules for a maximum of ten docking modules per station). To charge successfully, the temperature must be between 50°F to 95°F (10°C to 35°C).
  • Page 15: Maintenance

    Confirm that the inlet filter is free of dirt and replace if required. To order replacement parts, refer to Replacement Parts and Accessories in the MicroDock II Base Station User Manual. To maintain the station in good operating condition, perform the following basic maintenance as required.
  • Page 16: Specifications

    MicroDock II Quick Reference Guide Specifications The MicroDock II base station and docking modules are for indoor use only. Instrument dimensions: (w x l x h) 21.2 x 26.3 x 8.2 cm (8.3 x 10.4 x 3.2 in.) base station and one docking module Weight: 0.98 kg (2.15 lb.)
  • Page 17: Charger Specifications

    The disconnect device should be installed in close proximity to the equipment and the device should be marked as a disconnecting means for the equipment. MicroDock II Charger Specifications...
  • Page 18 MicroDock II Quick Reference Guide...
  • Page 19 iERP: 126102 D5618/7 [English] © BW Technologies 2008. All rights reserved.