568 EX
Users Manual
Specifications
General Specifications
IR Temperature
Range
Accuracy
Repeatability
Display Resolution
Spectral Response
Response Time
KTC Input Range
KTC Input
Accuracy
KTC Resolution
Distance: Spot
Laser sighting
Emissivity
Data storage
Operating Altitude
Storage Altitude
Relative Humidity
Operating
Temperature
20
-40 °C to 800 °C (-40 °F to 1472 °F)
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
≥0 °C: ±1 % or ± 1.0 °C, whichever is greater
<32 °F: ±2 °F ±0.1 °/1 °F
≥32 °F: ±1 % or ±2 °F, whichever is greater
±0.5 % of reading or ±0.5 °C (1 °F), whichever is
greater.
0.1 °C / 0.1 °F
8 µm to 14 µm
<500 ms (95 %)
-270 °C to 1372 °C (-454 °F to 2501 °F)
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
0.1 °C/0.1 °F
50:1 (90 % energy)
Single laser, output <1 mW Class II,
wavelength 630 to 670 nm
Digitally adjustable from 0.10 to 1.00 by 0.01 or via
built-in table of common materials
99 points
2000 meters above mean sea level
12,000 meters above mean sea level
10 % to 90 % RH non-condensing up to 30 °C
(86 °F)
0 °C to 50 °C (32 °F to 122 °F)