Specifications; General Specifications - Fluke 568 EX User Manual

Hide thumbs Also See for 568 EX:
Table of Contents

Advertisement

568 EX
Users Manual

Specifications

General Specifications

IR Temperature
Range
Accuracy
Repeatability
Display Resolution
Spectral Response
Response Time
KTC Input Range
KTC Input
Accuracy
KTC Resolution
Distance: Spot
Laser sighting
Emissivity
Data storage
Operating Altitude
Storage Altitude
Relative Humidity
Operating
Temperature
20
-40 °C to 800 °C (-40 °F to 1472 °F)
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
≥0 °C: ±1 % or ± 1.0 °C, whichever is greater
<32 °F: ±2 °F ±0.1 °/1 °F
≥32 °F: ±1 % or ±2 °F, whichever is greater
±0.5 % of reading or ±0.5 °C (1 °F), whichever is
greater.
0.1 °C / 0.1 °F
8 µm to 14 µm
<500 ms (95 %)
-270 °C to 1372 °C (-454 °F to 2501 °F)
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
0.1 °C/0.1 °F
50:1 (90 % energy)
Single laser, output <1 mW Class II,
wavelength 630 to 670 nm
Digitally adjustable from 0.10 to 1.00 by 0.01 or via
built-in table of common materials
99 points
2000 meters above mean sea level
12,000 meters above mean sea level
10 % to 90 % RH non-condensing up to 30 °C
(86 °F)
0 °C to 50 °C (32 °F to 122 °F)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents