5. BASIC OPERATION OF METER UNIT

1.Stopper Button Operation

If the stopper button !3 is pressed
when the stopper button mark !9
, the meter needle !8
is set to
deflects according to brightness.
Then when the button is released,
the needle remains fixed at the
indicated position.
By holding the stopper button de-
pressed and turning it clockwise
so that the mark position becomes
, the needle is released.
At this time, the meter will deflect
freely even if the stopper button
is released. To fix the needle, turn
Black cloth
the stopper button counter-clock-
wise to where the mark is posi-
tioned
.
2.Zero Position Check and
Adjustment
With stopper button in released
L-398A
position, cover light sensor with
hand or black cloth to completely
block light. Check that meter
needle !8 correctly indicates zero
position. If the indication deviates
from zero, while observing meter
indication, use a coin or similar
tool to turn the rear zero adjust
screw @1 and adjust for zero indi-
cation.