Replacing A Fuse - Gilson Quad-Z 215 User Manual

Liquid handler
Table of Contents

Advertisement

Maintenance

Replacing a Fuse

A blown fuse may indicate the existence of another
problem in the instrument. If the replacement fuses
blow, don't try others. Contact your local
representative or Gilson. See
page 5-8.
To change a fuse, follow these steps.
1
Disconnect the power cord from the power
outlet and from the rear panel receptacle.
2
Locate the fuse drawer on the rear panel. See
page
2-16
if necessary.
3
Insert a small screwdriver into the notch next to
the fuse drawer.
4
Twist the screwdriver to open and remove the
fuse drawer. The fuse drawer contains one
5.0A "T" Slo-Blo fuse (5 x 20 mm size) for a
100/120 voltage selection. It contains two 5.0A
fuses for a 220/240 voltage selection.
5
Remove the old fuse(s) and insert the new
fuse(s).
6
Insert the fuse drawer into its receptacle in the
liquid handler.
Fuse drawer for 100/120
voltage selection
Before calling us
on
Fuse drawer for 220/240
voltage selection