Maintenance

Replacing a Level Sensing Cable

Removing the level sensing cable
1
Detach the transfer tubing
from the top of the isolation
probe holder by using the
supplied headless nut extender
(part number 49041032) to
loosen and remove the 1/4"-28
nut and ferrule.
2
Unplug the right-angle
connector for the liquid level
sensing cable (LLD cable) from
the socket on the back of the
Z-arm.
3
Detach the LLD cable from the
LLD cable guide.
4
Loosen the small hexagonal
nut on the probe holder for
probe one. Detach the split-
tongue connector from the nut.
5
Remove the transfer tubing
from the LLD cable
casing.
4
4-11