Novametrix Medical Systems 8100 User Manual page 45

Respiratory profile monitor
Table of Contents

Advertisement

SpO
Sensors
2
If using the first and third holes on the foam wrap it may be necessary to cut the middle strip
2.
Face the blue side of the wrap toward the skin and wrap around the site (cut excess foam if necessary on
small version). Secure with the Velcro
Do not wrap the tape around the limb so tightly that circulation is restricted.
Inspect the site often for adequate circulation—at least once every four hours. When applying
sensors take note of patient's physiological condition. For example, burn patients may exhibit
more sensitivity to heat and pressure and therefore additional consideration such as more
frequent site checks may be appropriate.
3.
Ensure the sensor heads are directly opposite each other through the tissue. This prevents the sensor from
being placed on a site too thick for proper operation.
4.
Position the sensor so that the foam wrap does not extend over the spaces between the fingers or toes.
This ensures no light transmission through this space. To maximize sensor life, secure the cable along the
limb with tape as shown in the following illustrations.
Treat foam wrap in accordance with hospital protocol for single-patient use. Check site
Rev. 02
®
Pediatric toe
Adult/pediatric finger
regularly to ensure adequate circulation and proper sensor positioning.
Sensors and Patient Connections
NOTE
off the Y-sensor.
tab.
WARNING
Neonatal hand
WARNING
CO
Neonatal/pediatric foot
Adult toe
SMO Plus! User's Manual
2
39