Halogen Monitor
• If HyperTerminal has never been run in association with the monitor, then a new
HyperTerminal connection needs to be made as follows:
a) Click Start; click Run. In the Run box, type "hypertrm.exe" and click OK. The New
Connection dialog box should appear.
b) Type in a New Connection Name. Click OK to select the default icon (if desired, select a
different icon before clicking OK). The "Connect To" dialog box should appear.
c) Click on Connect from the drop-down menu and choose the COM port to which the
analyzer is connected. Then click OK to display the COM Properties dialog box for that
COM port.
d) Set Port Settings to:
- Bits per second: ........ 19200
- Data bits: .......................... 8
- Parity: ........................ None
- Stop bits: .......................... 1
- Flow control: .............. None
Then click OK to display HyperTerminal's main window.
e) Click File; click Save. This creates a file with an .ht extension in the folder C:\Program
Files\Accessories\Hyper Terminal, using the name that was entered in Step b.
3. After HyperTerminal is running: click Transfer, click Capture Text, and then click Start to accept
the default drive\directory\filename to which all received data will be stored as an ASCII text file.
NOTE:
C:\Program
Browse to select a different drive, directory and filename. The next time data is
captured, it will be stored under the same location and filename.
NOTE: Changing the filename extension to "CSV" (comma separated variable) will
allow the file to be directly opened in the spreadsheet program. For example, double
clicking the filename CAPTURE.CSV in Windows Explorer will automatically start
Microsoft Excel and load the file.
4. From the Data Display screen, press both the Keypad Left and Right buttons at the same time
to display the first Function Menu screen. Next, use the Keypad buttons to move the arrow (>)
until it points to PC-DUMP, and then press the ENTER button to begin the download process.
2ndUNIT
PLOT
5. Observe that the monitor's screen goes blank and that HyperTerminal displays the data as it is
being received. After all data has been transmitted, the Refrigerant Monitor re-displays the
Function Menu screen.
6. To stop capturing data and save it, click Transfer, click Capture Text, click Stop.
7. HyperTerminal and the connection to the monitor can now be closed.
Instruction 3015-4584 Rev 9
By default, the captured text file will be stored on the hard drive at
Files\Accessories\HyperTerminal\CAPTURE.TXT.
PRINT
>PC-DUMP
Operation
If
desired,
click
19