Agilent 8494/95/96G/H Attenuators Operating and Service Manual
Table 1
on
page
11,
show the typical switching arrangement to increase the amount of
attenuation in an 8494G/H, 8495G/H, or 8496G/H in a linear
manner. To ensure specified performance, it is recommended that
the attenuator sections that are shown in the following tables be
used. With the attenuator programmed for 0 dB attenuation, the
resultant attenuation is the insertion loss (residual attenuation).
The 8494G/H has a minimum selectable step of 1 dB, while the
8495G/H and 8496G/H have a minimum selectable step of 10 dB.
The accuracy of the attenuators is within the limits given in
"Specifications"
on
• The 8494G/H are four- section attenuators with a maximum
attenuation of 11 dB.
• The 8495G/H are three- section attenuators with a maximum
attenuation of 70 dB.
• The 8496G/H are four- section attenuators with a maximum
attenuation of 110 dB.
Table 1
8494G/H Attenuator Switching Order
Atten
(dB)
1 dB
0
1
2
3
4
5
6
7
8
9
10
11
Table 2
on
page
12, and
page
15.
8494G/H Attenuator Sections
1
2
2 dB
×
×
×
×
×
×
×
×
×
×
×
×
Introduction
Table 3
on
page 12
3
4
4 dB
4 dB
×
×
×
×
×
×
×
×
×
×
×
×
1
11