Omron HEM­790IT Blood Pressure Monitor Troubleshooting Guide 
This guide will help you solve problems that may occur when you're using your Omron Blood Pressure 
Monitor or uploading data from it to your HealthVault account. You may find it useful also to refer to 
the printed instructions that came with your device.  
Troubleshooting Tips 
Problem 
Blood Pressure Monitor 
status in Connection Center 
shown as 'Your device isn't 
connected' 
There are measurements 
on your blood pressure 
device but they are not 
uploading when the device 
is connected 
Device display is blank. 
Blood pressure 
measurements are 
unexpectedly high or low. 
 
Possible Causes 
Connection is not well 
established 
Computer and device might 
be out of sync 
Power off 
No battery power 
Error in taking 
measurements 
 
Solutions 
Ensure the USB cable is properly plugged 
into the computer and the device. 
Disconnect the USB cable and the 
reconnect it, then wait several seconds 
1. Disconnect the USB cable. 
2. Make sure the device is turned off 
by pressing start/stop.  
3. Reconnect the USB cable 
If you connect your cable more than 20 
seconds after you take a reading, 
Connection Center may not detect it.  If 
Connection Center does not detect it, 
please disconnect the cable and reconnect 
it again. 
Make sure the Blood Pressure Monitor 
selector is switched to the correct user (A 
or B).  NOTE: You can configure Connection 
Center to upload data from A or B or both. 
Check that the device is turned on.  
If running from a wall outlet, check that the 
cord is attached to both the device and 
outlet. 
Replace batteries, referring to the battery 
installation instructions. 
 
Check for proper placement of the battery 
polarities (+ and – symbols). 
Review the instructions for taking a 
measurement. 
Blood pressure can fluctuate throughout 
the day for a number of reasons. If you 
observe fluctuations outside the range your 
doctor has defined as normal for you, you 
may require prompt attention. Contact 
your doctor. 

Advertisement

loading

Summary of Contents for Omron HEM­-790IT

  • Page 1 Omron HEM­790IT Blood Pressure Monitor Troubleshooting Guide  This guide will help you solve problems that may occur when you're using your Omron Blood Pressure  Monitor or uploading data from it to your HealthVault account. You may find it useful also to refer to  the printed instructions that came with your device.   Troubleshooting Tips  Problem  Possible Causes  Solutions  Blood Pressure Monitor  Ensure the USB cable is properly plugged  status in Connection Center  Connection is not well  into the computer and the device.  shown as ‘Your device isn’t  established  Disconnect the USB cable and the  connected’  reconnect it, then wait several seconds  1. Disconnect the USB cable.  2. Make sure the device is turned off  by pressing start/stop.   3. Reconnect the USB cable  If you connect your cable more than 20  There are measurements  seconds after you take a reading,  on your blood pressure  Computer and device might  Connection Center may not detect it.  If  device but they are not  be out of sync  Connection Center does not detect it,  uploading when the device  please disconnect the cable and reconnect  is connected  it again.  Make sure the Blood Pressure Monitor  selector is switched to the correct user (A ...
  • Page 2   Device Errors  Error or symbol  Cause  Resolution  Device shows “EE”   Cuff is under‐inflated, or has  Single Mode:  not been applied correctly.  The monitor will attempt 1 reading.    TrueRead™ Mode:  Device shows   Cuff is over‐inflated.   The monitor will attempt up to 5 times   "E      E" vertically in two lines  After  the automatic attempts, remove  the arm cuff.  Read, “Taking a  Measurement” in the instruction manual.   Take another measurement.  Monitor could not detect  pulse wave.  Check air plug connection.  Take another  Device shows ‘E’  measurement and remain still until the  Air plug is not connected  measurement is complete.  Device shows “Battery  Batteries are worn  Replace the four batteries.  picture with a cross mark”  Refer to “Battery Installation” in the  instructionj manual.  Check that the device is turned on.     Check the battery installation for proper  No Power  placement of the battery polarities.   ...
  • Page 3   If you have a problem with your device that this information does not solve, please:  Refer to the printed instructions that came with your device.  Phone Omron Customer Service: 1 (800) 634‐4350  Go to http://www.omronhealthcare.com  Trouble with Microsoft® HealthVault Connection Center or your  HealthVault™ account?  If your device is working correctly and you are having problems using your HealthVault Connection  Center software or connecting to your HealthVault account, go to:  http://healthvault.com/help.html Here you'll be able to access HealthVault Connection Help and FAQs and HealthVault account Help.     For device troubleshooting tips, go to: http://support.microsoft.com/kb/943236    If you have questions that are not answered by HealthVault Help, go to:  http://www.healthvault.com/support Here you'll be able to submit a request for support to the HealthVault team.    For general Microsoft support, go to http://support.microsoft.com/       Microsoft, HealthVault and the HealthVault logo are trademarks of the Microsoft group of companies.  All other trademarks are property of their respective owners. Portions of this material are © 2009  Microsoft Corporation. All rights reserved. ...