Fluke Victoreen 943-227-15 Operator's Manual page 8

Digital preamplifier for use with model 943-27 current mode beta detector
Table of Contents

Advertisement

943-227-15 Digital Preamplifier
Operator Manual
Table 1-3. Preamplifier Serial Interface Board 977-210-10M Switch Settings
Switch
Normal
Switch Function
Op.
SW1
Open
Microprocessor Reset switch, momentary contact
SW2
N/A
Not supplied
SW3-1
ON
OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)
SW3-2
OFF
OFF • (No function is implemented in the 977-210-10M)
SW3-3
OFF
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for
factory PGA and Auto Zero set-up only)
SW3-4
OFF
OFF • (No function is implemented in the 977-210-10M)
SW3 is a 4-position DIP switch on the 977-210-10M serial
interface board.
Table 1-4. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M
Jumper
Normal Setting Function
JMP1
A to B
JMP2
N/A
Table 1-5. Jumper Positions for Preamplifier Electrometer Board 977-200-15M
Jumper
Normal Setting Function
JMP1
A to B
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment
1-4
NOTE
Enables Victoreen Loop
(normal)
Deleted, Not Used
Normal Operation
Alternate Setting Function
B to C
Enables RS232C
N/A
Deleted, Not Used
Cal. Position
Function
B to C
PGA Offset Adjust *