8845A/8846A
Users Manual

Diode Test

Test Current ............................................................ 100 A or 1 mA
Response Time....................................................... 300 samples/sec with audible tone.
Accuracy is given as
Range
5.0000 V
10.0000 V

Measurement Rates (IEEE488[4])

Function
DC Volts, DC Current, and
Resistance
AC Voltage and AC Current
Frequency and Period
Capacitance
[1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
[2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.
[3] For remote operation or external trigger using default settling delay
[4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the
baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus
has a maximum measurement rate of 963 measurement/s.
1-22
(% measurements + % of range)
24 Hour
90 Days
(23 1 C)
(23 5 C)
0.002 + 0.002
0.008 + 0.002
0.002 + 0.001
0.008 + 0.002
Digits
100 NPLC
0.2 NPLC
0.02 NPLC
[2]
1 Year
(23 5 C)
0.01 + 0.002
0.01 + 0.002
Integration Time
Setting
60 Hz (50 Hz)
1.67 (2) s
10 NPLC
167 (200) ms
1 NPLC
16.7 (20) ms
3.3 ms
500 us
3 Hz
20 Hz
[3]
200 Hz
1 s
100 ms
10 ms
Temperature
Coefficient/ C
Outside 18 to 28 C
0.001 + 0.002
0.001 + 0.002
[1]
Measurements/Second
8845A
8846A
0.6 (0.5)
0.6 (0.5)
6 (5)
6 (5)
60 (50)
60 (50)
270
270
995
995
0.47
0.47
1.64
1.64
4.5
4.5
1
1
9.8
9.8
80
80
NA
2