Safety Regulations; Special Conditions For Use In Zone; Ex-Data; Technical Details - Ecom Instruments Ex-Pt 720 Instruction Manual

Ecom instruments sensor instruction manual ex-pt 720
Hide thumbs Also See for Ex-Pt 720:
Table of Contents

Advertisement

- obviously incorrect measurements occur.
- measurements with the unit are no longer possible.
- the permissible limiting values were exceed.

4. Safety regulations

In order to exclude incorrect operation of the unit, its use assumes
that the user is aware of and complies with the usual safety regu-
lations.
- the batteries may only be changed outside the Ex-area.
- it is not allowed to take additional batteries into the Ex-area.
- only type approved batteries may be used.
- only sensors which have been approved by ecom may be used.
- It must be ensured that the unit is not taken into Zone 0.
- measurements on electrically live parts are not allowed.
- observe the measuring range of the temperature sensors.
OOverheating will damage the sensors.
- avoid use in aggressive acids or alkalis.
- the unit may not be opened in Ex-areas.

4.1 Special conditions for use in Zone 0

The measurement of temperatures with the Ex-Pt 720 within
Zone 0 is only permitted if the following conditions are complied
with:
- a stainless-steel sensor Type 0628 2232 is used.
- the Ex-Pt 720 itself may not be taken into Zone 0.
- It must be absolutely ensured that the unit does not find its way
a˝accidentally into Zone 0. In order to guarantee this, the carrying
ststrap of the unit must be attached additionally to the belt of the
u˝user or to a loop attached to his belt. The length of the carrying
ststrap must be so adjusted that when the unit hangs freely on the
c˝carrying strap, the Ex-Pt 720 maintains a spacing of 50cm to the
bboundary area of Zone 0.
18
- Should the unit be used in the optionally available leather case,
t˝˝the carrying strap of the unit must still comply with the conditi-
ti˝ons for attachment to the user's body.

5. Ex-Data

EC-test approval-no.: TÜV 01 ATEX 1757 X
Ex-designation:
II 2(1) G EEx ia IIC T4
Approved for Zone 1, equipment group II, gas group C explosions-
endangered gases, vapours or fog, temperature class T4.
Only the sensors specified in the approval certificate may be used.
Sensor no. 0628 2232 may be used in Zone 0 !

6. Technical Details

Ambient temperature Ta: -10 ... +50°C
Storage temperature:
-20 ... +70°C
Batteries:
9V-Block, 6LR61 as per IEC
(only type tested batteries)
Supplier
Varta
Varta
Varta
Duracell
Duracell
Duracell
Everady
(Ralston Energy Systems SA)
Panasonic
Daimon
Type
Alkaline No.4822
Alkaline Universal No.4022
Alkaline Electric Power No.8022
Alkaline
Alkaline Ultra
Professional Alkaline Battery Procell
Alkaline Energizer
Alkaline Power Line Industrial Battery
Alkaline
19