Mixed signal oscilloscopes, digital serial analyzers, digital phosphor oscilloscopes, digital phosphor oscilloscopes declassification and security (26 pages)
Page 1
Instruction Manual P6960 and P6980 High Density Logic Analyzer Probes with D-MaxtProbing Technology 071-1528-00 There are no current European directives that ap- ply to this product. This product provides cable and test lead connections to a test object of elec- tronic measuring and test equipment.
Page 3
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
Page 5
......Connecting the P6960 High Density and P6980 Differential Probes ..
Page 7
Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. Provide Proper Ventilation. Refer to the manual’s instructions for details on installing the product so it has proper ventilation. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 8
WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product: CAUTION Refer to Manual P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 9
H The online release notes provide last-minute product and software informa- tion not included in this manual. To access the Probe Manual Release Notes, click Start > Programs > Tektronix Logic Analyzer > TLA Release Notes. H A series of microprocessor support instruction manuals provide operating and service instructions for the individual microprocessor support packages.
Page 10
PCB and the probe input circuitry. Compression Footprint A connectorless, solderless contact between your PCB and the P6960 and P6980 Probes. Connection is obtained by applying pressure between your PCB and the probe through a cLGA c-spring.
Page 11
Trademark name that describes the technology used in the P6960 and P6980 high-density logic analyzer probes. Flying Lead Set A lead set designed to attach to a P6960 Probe to provide general-purpose probing capability. See Figure ii. Figure ii: Flying Lead Set...
Page 12
An interface device for attaching logic analyzer probes to components with a maximum lead diameter of 2.413 mm (0.095 in) and stackable on lead centers of 1.27 mm (0.050 in). viii P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 13
This phone number is toll free in North America. After office hours, please leave a voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 15
Product Description The P6960 and P6980 Logic Analyzer Probes connect TLA7Axx and TLA7NAx Series Logic Analyzer modules to a target system. The P6960 probe consists of 34 single-ended channels in one probe head, while the P6980 probe consists of 34 channels in two probe heads, with each head containing 17 differential channels.
Page 16
P6960 Single-ended probe Figure 1: P6960 High-Density probe with D-Max probing technology The following list details the capabilities and qualities of the P6960 Probe: H Differential or single-ended clock and qualification inputs H Single-ended data inputs H cLGA contact eliminates need for built-in connector H Footprint supports direct signal pass-through H Supports PCB thickness of 1.27 mm to 6.35 mm (0.050 in to 0.250 in)
Page 17
H Any common mode voltage is acceptable so long as the maximum positive voltage does not exceed +5 V and the maximum negative voltage does not exceed - -2.5 V (clock only) NOTE. Refer to Figure 15 on page 22 for P6960 probe routing and pin out information. P6980 High-Density...
Page 18
+5 V and the maximum negative voltage does not exceed - -2.5 V NOTE. Refer to Figure 15 on page 22 for P6980 probe routing and pin out information. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 19
In addition, all P69xx series Logic Analyzer probes connect to the logic analyzer in exactly the same manner. Probe Match color- - coded labels Figure 3: Connecting the probes to the logic analyzer P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 20
Operating Basics Connecting the Probes to the Target System Connecting the P6960 You can connect the P6960 High-Density and P6980 Differential Probes to the High Density and P6980 target system without turning off the power to the target system. The target system must have the probe retention posts installed;...
Page 21
4. Solder the posts to the PCB. The posts can be soldered from the top or bottom of the circuit board. 5. Pull off the carrier from the posts. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 22
2. Secure the screws to the posts. If access is limited, use the adjustment tool that came with your probe. The probe is completely fastened to the PCB when the screws stop in the posts. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 23
Operating Basics Connecting the Flying The flying leadset, Tektronix part number 196-3494-XX, is an optional Leadset accessory for your probe. The flying leadset allows you to connect to individual test points on your PCB. However, for general-purpose probing, the P6810 probe is recommended for best performance.
Page 24
Hang the probe cables so that you relieve the tension on the probes at the retention posts. See Figure 8. Good Cable hanger Logic Analyzer Probe cables Bad contact Figure 8: Proper dressing of the probe cables P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 25
2. Remove any remaining lint using a nitrogen air gun. NOTE. Use alcohol sparingly and be sure that you have removed any remaining lint or residue with the nitrogen air gun. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 26
See Figure 9. Gently slide the probe cover over the probe end and store the probe. Protective cover Figure 9: Protecting the probe heads P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 27
Reference This section provides reference information for the P6960 High-Density and P6980 High-Density Differential Probes with D-Max probing technology. Topics include the following: H Designing an interface between the probe and a target system H Board design H Probe footprint dimensions...
Page 28
Routing several of your design’s clocks and strobes to the analyzer clock inputs will provide you with a greater flexibility in the logic analyzer clocking setup menus. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 29
While all of the modules in a merged set can use their clock inputs independently if needed, it must be remembered that there are a maximum of four clock inputs on a 136 channel wide module. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 30
To see the importance of this reminder we will once again use a microprocessor system as an example. Tektronix logic analyzer processor has a 32 bit address bus and a 64 bit data bus. The data bus is split into four 16 bit subgroups that have independent source synchronous clocks.
Page 31
500 MHz sample rate on one-fourth of the channels. By trading three-fourths of the analyzer’s channels, the remaining channels can be sampled at a 2 GHz rate with quadruple the memory depth. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 32
The above examples assumes a 136-channel LA module. The missing channels in LA modules with fewer than 136 channels are ignored. With merged modules, range recognition extends across the first three modules: the master module contains the most-significant channels. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 33
This section provides information that helps you design your PCB mechanically and electrically for use with the P69XX logic analyzer probes. Probe Dimensions Figure 10 shows the probe dimensions for the P6960 and P6980 probes. 32.56 mm 182.88 mm 60.19 mm (1.282 in)
Page 34
Traces may be routed on inner layers of the board through the keepout area. 3.40 mm 3.51 mm (0.134 in) (0.138 in) Keep out area Figure 12: Keep out area P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 35
(0.370 in) Figure 13: Side-by-side layout Figure 14 shows the dimensions for an end-to-end footprint layout. 34.04 mm 34.04 mm (1.340 in) (1.340 in) 4.70 mm (0.185 in) Figure 14: End-to-end layout P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 36
NOTE. For optimum signal integrity, there should be a continuous, uninterrupted ground return plane along the entire signal path. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 37
Reference Physical Attachment Requirements for the P6960 and P6980 Probes. The P6960 High-Density Probe and P6980 High-Density Differential Probe interconnects are designed to accommodate PCB thickness ranging from 1.27 mm to 6.35 mm (0.050 in to 0.250 in). To accommodate this range, there are two wire lengths in the design: H For board thicknesses of 0.050 in to 0.120 in, use the standard wire that is...
Page 38
Figure 16: High-Density and High-Density Differential probe load model The differential load for the P6960 clock input and the P6980 probe can be modeled by attaching the single line model to each side (+ and - -) of the differential signal. The + and - - sides of the differential signal are well insulated in the probe head up to and including the differential input stage.
Page 39
NOTE. Tektronix recommends using immersion gold surface finish for best performance. Tektronix also recommends that the probe attachment holes float or remain unconnected to a ground plane. This prevents overheating the ground plane and promotes quicker soldering of the retention posts to your PCB. The probe retention posts are designed to allow you to solder the retention posts from either side of your PCB.
Page 40
See detail A Max via dia 0.25 mm (0.010 in) 0.04 mm Dia 0.51 mm (0.002 in) (0.020 in) 0.32 mm (0.013 in) Detail A Scale 50.000 Figure 18: Optional Via-in-Pad placement recommendation P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 41
Figure 19 shows the pad assignments, pad numbers, and signal names for the with D-Max probing P6960 single-ended data, differential clock logic analyzer probe. The P6960 probe has 32 data channels, one clock, and one qualifier for each footprint. technology...
Page 42
Reference Table 4: Channel assignment for a P6960 single-ended data, differential clock logic analyzer probe 136 Channel 68 Channel 102 Channel 34 Channel Pin Number Signal name Probe4 Probe 3 Probe 2 Probe 1 Probe 2 Probe 1 E2:0 A2:0...
Page 43
Reference Table 4: Channel assignment for a P6960 single-ended data, differential clock logic analyzer probe (Cont.) 136 Channel 68 Channel 102 Channel 34 Channel Pin Number Signal name Probe4 Probe 3 Probe 2 Probe 1 Probe 2 Probe 1 E2:2...
Page 44
Figure 20: P6980 differential pinout detail Table 5 on page 31 lists the channel mapping to a 136 channel or 102 channel logic analyzer module for the P6980 differential data and clock logic analyzer probe. P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 49
Specifications Mechanical and Electrical Specifications Table 7 lists the mechanical and electrical specifications for the P6960 and P6980 Probes. The electrical specifications apply when the probe is connected between a compatible logic analyzer and a target system. Refer to the Tektronix Logic Analyzer Family Product Specifications document...
Page 50
45% relative humidity 45% relative humidity Altitude Operating 10,000 ft (3,048 m) 10,000 ft (3,048 m) Non-operating 40,000 ft (12,192 m) 40,000 ft (12,192 m) Electrostatic immunity 6 kV 6 kV P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 51
Service Strategy The P6960 and P6980 Probes use replaceable c-spring cLGA clips. See page 38 for the replacement procedure. If a probe failure other than the cLGA clip occurs, return the entire probe to your Tektronix service center for repair.
Page 52
2. Align the new clip with the probe head and gently snap it into place. 3. Test the probe to confirm that all channels are functional. Pull away Rotate out Figure 21: Replacing the cLGA clip P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 53
Legacy Probe and Attachment Support H Nexus Technology, a Tektronix Partner, sells accessories that allow you to use the P6960 probe with legacy attachment connectors as well as utilize the P6960 probe footprint with select P68XX and P64XX probe products.
Page 55
Using the Replaceable Parts List Replaceable Parts The P6960 and P6980 Probes contain only the elastomer as a replaceable part. If probe failure occurs, return the entire probe to your Tektronix service representa- tive for repair.
Page 56
Items in this section reference figure and index numbers to the exploded view illustrations that follow. Tektronix part number Use this part number when ordering replacement parts from Tektronix. 3 and 4 Serial number Column three indicates the serial number at which the part was first effective. Column four indicates the serial number at which the part was discontinued.
Page 60
Replaceable Parts P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 61
Mechanical specifications, 35 Transmission lines, 23 Merged module, 15 Electrical specifications, 35 Module, viii Flying Lead Set, vii Online Help, v Functional check, vii, 37 Online Release Notes, v Ordering parts information, 41 P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Page 62
URL, Tektronix, ix Using the correct retention post wires, 6 Range recognition, 18 Related documentation, v Release Notes, online, v Repackaging instructions, 39 Replaceable Parts, 41 Web site address, Tektronix, ix P6960 and P6980 High-Density Logic Analyzer Probes Instruction Manual...
Need help?
Do you have a question about the P6960 and is the answer not in the manual?
Questions and answers