Digital Color Doppler Ultrasound System SonoAir 10/ SonoAir 20/ SonoAir 30/ SonoAir 40/ SonoAir 50/ SonoAir 60/ SonoAir 70/ SonoAir 80/ SonoAir 90 Operation Manual V1.1 Feb.6 2024 CHUM SonoAir -001 CHISON Medical Technologies Co., Ltd. We reserve the rights to make changes to this manual without prior notice.
Essential Requirements of the Medical Device Regulation This manual is a reference for the SonoAir 10/ SonoAir 20/ SonoAir 30/ SonoAir 40/ SonoAir 50/ SonoAir 60/ SonoAir 70/ SonoAir 80/ SonoAir 90. Please verify that you are using the latest revision of this document.
Digital Color Doppler Ultrasound System Meaning of the signal words In this manual, the warning words NOTE, CAUTION and WARNING are used for regarding safety and other important instructions. Please understand their meanings clearly before reading this manual. The words and their meanings are defined as follows: Warning words Meaning...
Page 4
This manual contains warnings regarding foreseeable potential dangers, but user shall always be alert to dangers other than those indicated as well. CHISON shall not be liable for damage or loss that results from negligence or from ignoring the precautions and operating instructions described in this operation manual.
Digital Color Doppler Ultrasound System Content Digital Color Doppler Ultrasound System ................1 Chapter 1 Introduction ......................1 1.1 System Overview ......................1 1.2 Contact Information ....................... 1 Chapter 2 System Safety ......................3 2.1 Safety Overview ......................3 2.2 Electrical Safety ......................4 2.3 Labels ...........................
Page 6
Digital Color Doppler Ultrasound System 5.1 Select Scan Mode ....................... 43 5.2 Functional Description of Parameter Adjustment ............48 5.3 After Capturing the Image ................... 53 5.4 Image Browse ......................56 5.5 Archive Management ....................56 5.6 Report ......................... 58 5.7 DICOM ........................
Page 7
Digital Color Doppler Ultrasound System 8.8 Net Setup ........................109 8.9 System Setup ......................110 Chapter 9 System Maintenance ..................112 9.1 Cleaning ........................112 9.2 Maintenance ......................113 9.3 Safety Check ......................113 9.4 Troubleshooting ......................114 9.5 Service Responsibility ....................115 Chapter 10 Probes ........................
1.2 Contact Information For additional information or assistance, please contact your local distributor or the appropriate support resource shown below: CHISON www.chison.com website Service CHISON Medical Technologies Co., Ltd. Support Tel:0086-0510-85311707 Fax: 0086-0510-85310726 E-mail: service@chison.com.cn Placing an CHISON Medical Technologies Co., Ltd.
Page 9
Digital Color Doppler Ultrasound System US Agent Mr. Marco Mu, 2219 Rimland Drive, Suite 301, Bellingham, Barkley Villiage Bellingham, Washington, 98226, UNITED STATES Phone: 360-3257028, Fax: 360-9253199 Email: us.agent@mid-link.net MID-LINK INTERNATIONAL CO., LTD CAUTION Federal law restricts the device to sale by or on the order of a licensed practitioner or therapist.
Digital Color Doppler Ultrasound System Chapter 2 System Safety 2.1 Safety Overview This section discusses the measures to ensure the safety of both the operator and patient. To ensure the safety of both operator and patient, please read the relevant details in this chapter carefully before operating this system.
Digital Color Doppler Ultrasound System NOTE The system has built-in screen saver to avoid the tic mark on the display. It is not recommended to constantly turn on and off the unit. To dispose of this product properly, please call your local service department. 2.2 Electrical Safety Type of protection against electric shock ...
Page 12
Digital Color Doppler Ultrasound System The Equipment is not suitable for use in the environment with FLAMMABLE ANAESTHETIC MIXED WITH AIR (or WITH OXYGEN or WITH NITROUS OXIDE). Conduction Interference Image quality will effect by conducted disturbance. Please do not use the effected image. And put system in a simpler electromagnetic compatibility environment to start working.
Page 13
Locate the equipment as far away as possible from other electronic equipment. Be sure to use only the cables provided by or designated by CHISON. Connect these cables following the installation procedures (e.g. wire power cables separately from signal cables).
Digital Color Doppler Ultrasound System 2.3 Labels Labels on main unit Fig. 2-1 SonoAir Label...
Page 15
Digital Color Doppler Ultrasound System Labels on transducers Fig. 2-10 7L Label Fig. 2-11 3S Label Fig. 2-12 3C Label Fig. 2-13 6MC Label Fig. 2-14 6E Label...
Page 16
Digital Color Doppler Ultrasound System 2.3.1 Symbols on Label This mark indicates that this product contains WASTE OF ELECTRICAL AND ELECTRONIC a limited amount of hazardous substances in EQUIPMENT (WEEE): This symbol is used for the Chinese Standard GB/T 26572-2011 Environment Protection, it indicates that the "Limited Requirements for Restricted waste of electrical and electronic equipment...
Page 17
Digital Color Doppler Ultrasound System AUTHORIZED REPRESENTATIVE IN THE This symbol is followed by the serial number of EUROPEAN COMMUNITY: This symbol is the device. accompanied by the name and the address of the authorized representative in the European Community. MANUFACTURER: This symbol is This symbol indicates the country of accompanied by the name and the address of...
Page 18
Digital Color Doppler Ultrasound System 2.3.2 Other Device Labels The following table describes the purpose of safety labels and other important information provided on the equipment. Table 2-1: Symbol Icons Icon Meaning Type-BF applied part...
The Patient Environmental devices shown above are specified to be suitable for use within the PATIENT ENVIRONMENT. CAUTION DO NOT connect any probes or accessories without approval by CHISON within the PATIENT ENVIRONMENT. DO NOT touch patient and devices without IEC/EN 60601-1 approval to avoid the leakage current risk within the PATIENT ENVIRONMENT.
Page 20
Such malfunction may impact the stability, effectiveness and safety of the system and probes, and users should immediately stop using the system and probes until CHISON authorized engineer has checked the system and confirm the effectiveness and safety of the system and probes.
Digital Color Doppler Ultrasound System 2.5 Biological Safety This product, as with all diagnostic ultrasound equipment, should be used only for valid reasons and should be used both for the shortest period of time and at the lowest power settings necessary (ALARA - As Low As Reasonably Achievable) to produce diagnostically acceptable images.
Digital Color Doppler Ultrasound System 2.6 Scanning Patients and Education The Track-3 or IEC60601-2-37 output display standard allows users to share the responsibility for the safe use of this ultrasound system. Follow these usage guidelines for safe operation: In order to maintain proper cleanliness of the probes, always clean them between patients. ...
Digital Color Doppler Ultrasound System (cranial bone thermal index) should be monitored. MI> 0.3 there is a possibility of minor damage to neonatal lung or intestine. If such exposure is necessary, reduce the exposure time as much as possible. MI> 0.7 there is a risk of cavitations if an ultrasound contrast agent containing gas micro-spheres is being used.
Page 24
Digital Color Doppler Ultrasound System indices, MI and TI. The MI describes the likelihood of cavitations, and the TI offers the predicted maximum temperature rise in tissue as a result of the diagnostic examination. In general, a temperature increase of 2.5°...
Page 25
Digital Color Doppler Ultrasound System Thermal Composition Mode Specification Application Models Soft tissue Unscanned Large aperture Liver PW (>1cm Soft tissue Unscanned Small aperture Pencil Probe (<1cm Soft tissue Scanned Evaluated Breast color surface Soft tissue and Scanned Soft tissue at Muscle color bone surface...
Page 26
Digital Color Doppler Ultrasound System peak pressure to the square root of the frequency. MI=Pr‟/sqrt(fc) Pr‟ is the de-rated (0.3) peak rare-fractional pressure in Mpa at the point where PII is the maximum, and fc is the center frequency in MHz. PII is the Pulse Intensity Integral that the total energy per unit area carried by the wave during the time duration of the pulse.
Page 27
Digital Color Doppler Ultrasound System HINTS: Raise the Acoustic Output only after attempting image optimization with controls that have no effect on Acoustic Output, such as Gain and STC. WARNING Be sure to have read and understood control explanations for each mode used before attempting to adjust the Acoustic Output control or any control that can affect Acoustic Output.
Digital Color Doppler Ultrasound System 2.7 Device Instructions of Cybersecurity Controls 2.7.1 Software Integrity Control The following controls are in place to assure that the device software will maintain its integrity from the point of origin to the point at which that device leaves the control of the manufacturer. ...
Digital Color Doppler Ultrasound System secure this network using WPA (Wi-Fi Protected Access) or WPA2 (Wi-Fi Protected Access II) as security protocol. 2.8 Insturctions to the Intruded Device The following are instructions to protect your device prior to the device leaving your control. ...
Digital Color Doppler Ultrasound System Chapter 3 Preparing the System for Use 3.1 Site Requirement 3.1.1 Operation Environment Requirement The following environmental conditions are within system tolerances for operation: Strong radiation sources or powerful electromagnetic waves (e.g. Electro-magnetic waves from radio broadcasting) may result in image ghosting or noise.
Digital Color Doppler Ultrasound System possible flammable sources. If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. , ~ In storage battery must be charged every three months to 40% 60% capacity.
Digital Color Doppler Ultrasound System Do not place the battery in a microwave oven or pressurized container. Do not use the battery in combination with primary batteries (such as dry-cell batteries) or batteries of different capacity, type or brand. ...
Digital Color Doppler Ultrasound System 3.2 System Specifications 3.2.1 Console View Fig. 3-1 SonoAir 10/ SonoAir 20/ SonoAir 30/ SonoAir 40/ SonoAir 50/ SonoAir 60/ SonoAir 70/ SonoAir 80/ SonoAir 90 Console View Fig. 3-2 System Left Side View 1. DC IN 2. LAN Port 3. HDMI 4. USB Port Fig.
Fig. 3-4 System Bottom Panel View 3.2.2 Physical Specification Dimensions of main unit (approx.): 330mm (Length) × 220mm (Width) × 30mm (Height) Net weight of main unit (approx.): 2.1kg (No probe included) 3.2.3 System Components SonoAir 10 SonoAir 20 SonoAir 30 SonoAir 40 Component...
Digital Color Doppler Ultrasound System 3.2.4 System Configuration SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir Function B Mode Standard Standard Standard Standard Standard Standard Standard Standard Standard B/B Mode Option Standard Standard Standard Standard Standard Standard Standard Standard 4B Mode Option Option...
Page 36
Digital Color Doppler Ultrasound System SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir Function Freeze/ Standard Standard Standard Standard Standard Standard Standard Standard Standard Unfreeze Focus Standard Standard Standard Standard Standard Standard Standard Standard Standard Parameters Standard Standard Standard Standard Standard Standard...
Page 37
Digital Color Doppler Ultrasound System SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir SonoAir Function Cine Loop Standard Standard Standard Standard Standard Standard Standard Standard Standard Save and Standard Standard Standard Standard Standard Standard Standard Standard Standard Recall Print Standard Standard Standard Standard...
Digital Color Doppler Ultrasound System Micro convex probe: 6MC Abdominal, Urology (4.3-9.5MHz) Intracavity convex probe: Trans-vaginal, Urology 6E (4.3-9.5MHz) ② AC/DC adapter Adapter specifications: Adapter Model MANGO150S-19AB Input 100-240V~, 50/60Hz, 2.0-1.0A Output 7.9A 3.3 System Positioning & Transporting Moving the System When moving or transporting the system, take the precautions described below to ensure maximum safety for personnel, the system and other equipment.
Digital Color Doppler Ultrasound System Transporting the System Use extra care when transporting the system in a vehicle. After preparing the system as described above, take the following additional precautions: Only use vehicles that are suitable for transport of the system. ...
Page 40
Digital Color Doppler Ultrasound System , and system pops up dialog. Select “Standby” to enter into the standby status. Press...
Digital Color Doppler Ultrasound System 3.5 Probe Installation and Disassembly CAUTION Only power supply at “turn off” state, can install/take down the probe, otherwise, it will damage the main unit and the probe. When installing and disassembling probe, please keep the probe in a safe environment to prevent it from accidentally falling and thus cause damage.
Digital Color Doppler Ultrasound System Chapter 4 Control Panel 4.1 Overview of Display Area Fig. 4-1 Display Interface Logo Patient ID Hospital Name Current Preset Date and Time Current Probe Type and Frequency Focus Mode Menu Area Image Area Mode Image Parameter Area Thumbnail Area Image Parameter Control Area System Status Information Area...
Digital Color Doppler Ultrasound System 4.2 Start a New Exam Press to display the patient screen. Function Buttons on Patient screen: [Archive]: Operation on the patient information which has already existed. [Worklist]: Recall patient information in worklist. And need to activate the DICOM function. [Query/Retrieve]: Recall patient information for examination.
Digital Color Doppler Ultrasound System 5. To recall information of the previous patient, you can use the Archive or Worklist to recall patient‟s information to exam. CAUTION Creating a diagnostic record, you should check the accuracy of the patient’s information before saving measurement or image;...
Page 45
Digital Color Doppler Ultrasound System Key’s name Key icon Function description File management of system, you can view and 【EasyView】Key edit the patient‟s data. 【SETUP】Key Get in or out the system setting page. 【REPORT】Key Produce/Save/Recall an examination report. 【END】Key Finish the current exam. PRINT1: print the screen image by video printer connected to the system.
Page 46
Digital Color Doppler Ultrasound System Key’s name Key icon Function description 【AIO】Key Optimize the image automatically. 【CINE】Key Save the current cine loop. 【IMAGE】Key Save the current image. 【FREEZE】key Freeze or unfreeze the current image. 【BODYMARK】key Add bodymark to the image. 【ARROW】key Add arrows icon to the image area.
Page 47
Digital Color Doppler Ultrasound System Key’s name Key icon Function description This key‟s function is the same as 【B】Key. 【Single B】Key Display single B mode image. 【2B】Key Display 2B mode. 【L/R INVERT】Key Invert the image from left and right. 【U/D INVERT】Key Invert the image from up and down.
Digital Color Doppler Ultrasound System Key’s name Key icon Function description Interest (CROI). Position and re-size the Doppler Sample Volume Gate. Control digital cine review frames. 4.5 Image Parameter Area Display information about application, frequency, mode, depth, gain and etc. 4.6 Cine Control Item Description...
Page 49
Digital Color Doppler Ultrasound System 2. Click “Enable Wi-Fi” to scan for Wi-Fi signal. 3. Choose the needed Wi-Fi and click. 4. “Wi-Fi Network Authentication Required” dialog prompt displays. Check the Wi-Fi name and input the password. And then click “OK” on SonoAir. The connection is enabled. 5.
Digital Color Doppler Ultrasound System Chapter 5 Imaging This chapter will introduce image display modes and the operation of image control and adjustment. 5.1 Select Scan Mode 5.1.1 The Probe Identification The system default automatically identifies the current probe type when the probe is inserted. Press to switch the probe.
Page 51
Digital Color Doppler Ultrasound System 5.1.6 CFM Mode CFM is a Doppler mode intended to add color-coded qualitative information concerning the relative velocity and direction of fluid motion within the B mode image. CFM is useful to see flow in a broad area. It allows visualization of flow in the CROI, whereas Doppler mode provides spectral information in a smaller area.
Page 52
Digital Color Doppler Ultrasound System To switch from DPD to PD mode, press【PD】key directly in the control panel, or [CPA] option in the left side menu column. 5.1.9 PW Mode Doppler is intended to provide measurement data concerning the velocity of moving tissues and fluids. PW Doppler lets you examine blood flow data selectively from a small region called the Sample Volume.
Page 53
Digital Color Doppler Ultrasound System data simultaneously. PRF: adjust the velocity scale to accommodate faster/slower blood flow velocity. Velocity scale determines pulse repetition frequency. Wall Filter: remove the noise caused by vessel or heart wall motion at the expense of low flow sensitivity. Baseline: adjust the baseline to accommodate faster or slower blood flows to eliminate aliasing.
Page 54
“Biopsy” option to activate this function to show or hide the biopsy line. CAUTION Please confirm the biopsy line before using biopsy function. The biopsy tool provided by CHISON must be used for this diagnostic ultrasound instrument. Otherwise, the accuracy of biopsy cannot be guaranteed.
Digital Color Doppler Ultrasound System 5.2 Functional Description of Parameter Adjustment 5.2.1 Adjust Parameters by Keys on Control Panel Parameters’ Function Description Name L/R Invert Press , the displayed image can be reversed in left/right direction. U/D Invert Press , the displayed image is reversed in up/down direction. Press the【AIO】key to automatically optimize the image.
Page 56
Digital Color Doppler Ultrasound System Parameters’ Function Description Name Scan Width Adjust scan width of B image; frame rate will be fast if scan width is small. Density Adjust the image density, high or low. Change focus numbers. Change focus numbers will affect the frame rate. Focus Num Frame rate will be slow if focus numbers are increased.
Page 57
Digital Color Doppler Ultrasound System Parameters’ Function Description Name Gamma: Adjust B image Gamma parameters. B Rejection: Adjust B image gray scale inhibition parameters. Enter into slide show function. Touch the screen, the slide control bar Slide Show appears in the lower right corner of the screen. Click "×" to exit the function. Keyboard Light Turn on/off the keyboard light.
Page 58
Digital Color Doppler Ultrasound System Parameters’ Function Description Name Adjust PRF value, adjustment range depends on probe. Press the corresponding parameter control key of [PRF] or directly click it on the touch screen to adjust. Improve the current color. Press the corresponding parameter control key of Persistence [Persistence] or directly click it on the touch screen to adjust.
Page 59
Digital Color Doppler Ultrasound System Parameters’ Function Description Name D 2D Map Adjust the color of spectrum. Spectrum Enhance Adjust the brightness of the spectrum. Dynamic Range Adjust the dynamic range of the spectrum. Quick Angle Adjust the angle of the sample volume gate. Utility For details about this section, please refer to Chapter 5.6.2.2.
Digital Color Doppler Ultrasound System Parameters’ Function Description Name Adjust Parameters by DPD Menu Gain Adjust DPD image gain. Wall Thre. Adjust wall threshold. Blood Effection Choose different blood effection. Density Change images density. Switch between DPD mode and PD mode. Utility For details about this section, please refer to Chapter 5.6.2.2.
Page 61
Digital Color Doppler Ultrasound System 3. Input characters at cursor position by soft keyboard then press【ENTER】key to confirm. 4. Press【ANNOTATION】key again or【FREEZE】key to exit. Input Comment Library Characters 1. In comment status, move the cursor by touchpad to image area to edit. 2.
Page 62
Digital Color Doppler Ultrasound System Don‟t activate the single comment; press【CLEAR】key to delete all characters that have been input. CAUTION 【 】 Press CLEAR key will delete the measurement and body mark at the same time. 5.3.1.5 Set the Position of Default Comment Operation: 1.
Digital Color Doppler Ultrasound System 5.4 Image Browse Press【EasyView】Key to enter image information browsing interface and press【ENTER】key to use the functions in the image. ID: ID of current patient. Name: Current patient‟s name. Layout: Select the format to display the images. ...
Page 64
Digital Color Doppler Ultrasound System 【EasyView】key and click [Archive] to archive management interface. All process can be opened up by moving cursor. Item: Type selection, select Patient‟s ID or Name. Keyword: Search for key words. Period: Time filter, select today, one week, one month, three months, six months, recent one year and all.
Digital Color Doppler Ultrasound System 5.6 Report Press 【REPORT】key to enter into report interface. Move cursor to the images and press【ENTER】key to add the image into the report page. The report can be saved and printed. It is convenient for the doctor to view and edit the patient‟s information.
Digital Color Doppler Ultrasound System The image on the right side: Press [ENTER] key on the image to add the image into the report. Select All: Press this key to add all the images on the right side into the report. ...
Page 67
Digital Color Doppler Ultrasound System Select DICOM Storage in left, choose DICOM server and press Export button to DICOM storage. Enter Task Queue and watch or edit DICOM process. DICOM Print DICOM Print operation is the same as DICOM storage. 5.7.3 DICOM SR Press [Send DICOM SR] button in report interface, this task is added into Task Queue.
Digital Color Doppler Ultrasound System Chapter 6 Measurement and Calculation Main contents of this chapter: Normal calculation and measurement on B mode image, B/M mode image, PW mode image. The system can enter into corresponding measurement mode depends on current exam mode, and enter into the corresponding report depends on the measurement mode.
Digital Color Doppler Ultrasound System 3. During the Ellipse measurement, when fix the long axis, but the short axis is not fixed, press 【UPDATE】key to switch the long axis and short one. 【CLEAR】key Press【CLEAR】key to delete all the measurement results, comments and traces. 【Parameter Control】key Press the corresponding parameter control key to activate the function and use the function.
Page 70
Digital Color Doppler Ultrasound System 3. Fix another line D2 as step 2. 4. After finishing the measurement, the result of the angle will be displayed on the screen. Ratio (Distance) Calculate the ratio of the 2 distances through 2 distances measurement. Refer to distance measure above.
Page 71
Digital Color Doppler Ultrasound System Volume (2 Distances) Press corresponding parameter control key for Volume on the screen or directly click [Volume(3Distances)] to switch the item to [Volume(2Distances)]. Calculate the volume of the object through 2 distances measurement. Refer to distance measure above. ...
Digital Color Doppler Ultrasound System Heart Rate Heart rate is used to calculate the number of heart beats per minute from cardiac image. Measurement steps: 1. To enter into this item, please refer to the above method. 2. The measurement steps are the same as Time. 3.
Digital Color Doppler Ultrasound System 5. The second cursor “+” will appear on the screen automatically, move the cursor to the end point of current cycle, and press【ENTER】key to fix. 6. The measurement results and the other calculated parameters will appear on the screen automatically. ...
Page 74
Digital Color Doppler Ultrasound System Profile Profile is used to measure the gray distribution of the ultrasound signals in the vertical or horizontal direction on a certain profile (section). This measurement is only available in the frozen mode. Measurement steps: 1.
Page 75
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to Distance D1: First Distance Ratio(Distance) Meas. D2: Second Distance Formula: R=D1/D2 Ratio Refer Ellipse A1: First Area Ratio(Area) Meas. A2: Second Area Formula: R=A1/A2 Line 1 Refer to Distance Angle Range:...
Page 76
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Height Refer to Distance Meas. Width Refer to Distance Meas. D1: Length of Refer to Distance Meas. Normal StD% Formula: ((D1-D2)/D1)*100% D2: Length of Lliac Stenosis (Right/Left) A1: Area of Normal Refer to ellipse Meas.
Page 77
Digital Color Doppler Ultrasound System Meas. Submenu Unit Meas. Method Comment Menu Formula choose: Bessis, CFEF, Campbell, Chitty, Doubilet, Refer to Distance Hadlock, Hansmann, Hohler, Meas. Jeanty, Johnsen, Korean, Merz, Osaka, Shinozuka Formula to choose: Hansmann, Korean Formula: Bessis Formula: CFEF Fetal Refer to Distance FTA(Ellipse)
Page 78
Digital Color Doppler Ultrasound System Meas. Submenu Unit Meas. Method Comment Menu Refer to Distance AFI=AFI1+AFI2+AFI3+AFI4 Meas. Refer to Distance CX_L Meas. 6.3.4.1 Twins and Multiple Births Measurement 1. In the new patient OB page, choose the number of gestations from one to four. 2.
Page 79
Digital Color Doppler Ultrasound System 2. Click on the line image region, and appear one line with "+". Move the line to the target measurement region. 3. Move the line by the touchpad or directly move it on the touch screen to adjust the line angle, press [ENTER] key to fix the line.
Page 80
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to Distance Volume(1Ellipse) Meas. Refer to Distance Volume A: Long Axis and Ellipse Meas. Volume(1Ellip1Dis) B: Short Axis Formula: M: Distance V=(π/6)*A*B*M Refer to Distance D1: First Distance Ratio(Distance) Meas.
Page 81
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula (Left/Right) Refer to Distance Meas. Follicle height (Left/Right) Refer to Distance Meas. Follicle width 2distance A: the longer distance Formula of 2 distance V= B: the shorter (π/6)*A*B Follicle Volume distance...
Page 82
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Kidney Refer to Distance Meas. L: Kidney_L Volume Formula: H: Kidney_H V=(π/6)*L*H*W (Left/Right) W: Kidney_w Refer to Distance Meas. L: Bladder _L Vol.(L*W*H) Bladder Formula: H: Bladder _H Vol.(Biplane) V=0.497*L*H*W W: Bladder _w...
Page 83
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Comment Meas. Formula SV=|EDV-ESV| EDV(A2C) CO=SV*HR/1000 ESV(A2C) Simpson BP EF=SV/EDV*100 EDV(A4C) SI=SV/BSA ESV(A4C) CI=CO/BSA Refer to Distance Meas. & Ellipse Mea. Formula: EDV= LVLd LVLs ESV= LVAMd Modi Simpson LVAMs SV=|EDV-ESV| LVAPd CO=SV*HR/1000...
Page 84
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Comment Meas. Formula Refer to Distance Meas. Formula: EDV=π/6*(0.98*LVIDd+5.9)*LVIDd*LVIDd ESV=π/6*(0.98*LVIDs+5.9)*LVIDs*LVIDs LVIDd SV=|EDV-ESV| Gibson LVIDs CO=SV*HR/1000 EF=SV/EDV*100 FS=(LVIDd-LVIDs)/LVIDd*100 SI=SV/BSA CI=CO/BSA MV Diam Refer to Distance Meas. & Ellipse Mea. Mitral Valve MV Area Formula:LA/AO=LAD/AOD LA/AO...
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Comment Meas. Formula Refer to Distance Meas. & Ellipse Mea. Formula: A-L(LVAd Epi, LVAd sax Endo, LVLd apical) Refer to Distance Meas. & Ellipse Mea. Formula: T-E(LVAd Epi, LVAd sax Endo, a, d) AV Diam Qp/Qs Refer to Distance Meas.
Page 86
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Double Refer to M Time Meas. Cycles 6.4.3 B/M Cardiology Measurement In cardiology preset, press【CAL】key in B/M mode to enter into cardiology measurement. Meas. Meas. Method/ Submenu Comment Menu...
Digital Color Doppler Ultrasound System Meas. Meas. Method/ Submenu Comment Menu Meas. Formula IVSd/LVPWd EPSS Refer to M Distance Meas. MV E Amp MV A Amp MV D-E Exc Dist Mitral MV E-F Slope Valve Refer to M Velocity Meas. MV D-E Slope A-C Int Slope E Duration...
Page 88
Digital Color Doppler Ultrasound System In order to get accurate result, the PW image must be clear and high quality. Insure that you have fixed the cursor at the exact place of cardiac systole and diastole. 6.5.1 General Measurement Methods ...
Page 89
Digital Color Doppler Ultrasound System 【ENTER】key to fix it. 4. The measurement results will be displayed on the screen and calculate other values of parameters. Other Measurement Items These listed items share the same measure steps as those in the corresponding mode fast measurement except for the way to enter into the measure item: Distance: Please refer to the Distance measurement in B mode fast measurement.
Page 90
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula D1:Distance Formula: StD% StD%=((D1-D2)/D1)*100% D2:Distance Area1 Refer to B Ellipse Meas. Area2 Refer to B Ellipse Meas. StA% A1:Area1,A StA% Formula: StA%=((A1-A2)/A1)*100% 2:Area2 cm/s Refer to PW Velocity Meas. Formula: Pressure(ICA) mmHg...
Page 91
Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Artery, MCA, Formula: SD=Vs/Vd Duct Veno, Formula: SD=(Vs-Vd)/Vs Formula: SD=(Vs-Vd)/VMean FHeart Rate 6.5.4 PW GYN Measurement In GYN preset, press【CAL】key in PW mode to enter into GYN measurement. Meas.
Page 92
Digital Color Doppler Ultrasound System Refer to PW Velocity Meas. MV E Vel cm/s Formular:MV E PG=4*MV E Vel*MV E Vel/10000 E/Ea=MV E Vel/Ea Refer to PW Velocity Meas. Formular:MV A PG=4*MV A Vel*MV A Vel/10000 MV A Vel cm/s E/A=MV E Vel/MV A Vel A/E=MV A Vel/MV E Vel Refer to PW Trace Meas.
Page 93
Digital Color Doppler Ultrasound System Refer to PW Trace Meas. Formular: LVOT VTI LVOT PGmax=4*LVOT Vmax*LVOT Vmax/10000 LVOT PGmax=4*LVOT Vmean*LVOT Vmean/10000 LVPEP Refer to PW Time Meas. LVET Refer to PW Velocity Meas. AR Vmax cm/s Formular:AR PGmax=4*AR Vmax*AR Vmax/10000 Refer to PW Trace Meas.
Page 94
Digital Color Doppler Ultrasound System Refer to PW Velocity Meas. MPA Vmax cm/s Formular: MPA PGmax=4*MPA Vmax*MPA Vmax/10000 Refer to PW Velocity Meas. RPA Vmax cm/s Formular: RPA PGmax=4*RPA Vmax*RPA Vmax/10000 Refer to PW Velocity Meas. LPA Vmax cm/s Formular: LPA PGmax=4*LPA Vmax*LPA Vmax/10000 RVPEP Refer to PW Time Meas.
Page 95
Digital Color Doppler Ultrasound System MV lateral (Sa,Ea,Aa,ARa) 6.5.6 PW Vascular Measurement In vascular preset, press【CAL】key in PW mode to enter into vascular measurement. Meas. Method/ Meas. Menu Submenu Unit Meas. Formula cm/s Refer to PW Slope Meas. Pressure(Vs) mmHg Formula: Pressure=4*Vs*Vs/10000 cm/s Refer to PW Slope Meas.
Digital Color Doppler Ultrasound System Chapter 7 Cine-Memory This chapter introduces the theory of saving images in Cine-Memory and the operation of image playback in Cine-Memory. 7.1 The Principle of Cine Storage In real image status, the image can be stored in the movie memory in chronological order, maximum frames can be set.
Digital Color Doppler Ultrasound System recall it; or you can recall archived patient‟s information to recall image, please refer to archive chapter. 7.6 Save and Recall Cine In freeze status, press to save cine, then it will be displayed in the upper right corner of the screen, move cursor to needed cine, press【ENTER】key to recall cine.
Digital Color Doppler Ultrasound System Chapter 8 Setup This chapter introduces the operation to make settings to the system. The setting function is used to set up working environment and status, parameters of each examination mode. The settings are stored in the memory of the system and will not get lost even after the system is switched off.
Page 99
Digital Color Doppler Ultrasound System Screen Controller: Enable screen saver, setup the wait minute. User can custom screen saver images. A custom picture named "screensaver" in JPG, PNG, BMP format with size not exceed 512 * 384 pixels can be imported into the system. ...
Digital Color Doppler Ultrasound System Print De-Indentification: Set up not to show the identification of the patient when printing the image. This dialog box can be set to pop up every time before printing. 8.1.3 Keyboard Interface Users can set the different functions for number keys and other optional buttons according to the using habits.
Digital Color Doppler Ultrasound System Fig. 8-4 Unit: Display the unit of all measurement items. Measure Option: Select clear results upon unfreeze, freeze upon measure, delete results upon 【CLEAR】key. Measure Result: The color of the result font is alternative, including yellow/white/orange/green. ...
Page 102
Digital Color Doppler Ultrasound System Fig. 8-5 Region: Pull down and select needed measurement menu. Manage: Pop up measurement software edition interface, add, modify, delete, change marshalling sequence in measurement menu. B, M, D: Display measurement of each Exam mode. ...
Page 103
Digital Color Doppler Ultrasound System Measure Order None: Disable rule, Repeat: Repeat this item, Next: measure by sequence. Default Item After choosing the Repeat and Sequential, choose one measurement or calculation to activate the measurement rule. Has Left Right Display Left and Right in the mode measurement package. Has Prox Mid Distal Display Prox, Mid, Distal in the mode measurement package.
Page 104
Digital Color Doppler Ultrasound System Display In Report Whether display in the report or not. Average Method The average rule of data. Press this button to pop up interface to add measurement operation. Remove Press this button to delete selected measurement operation. 8.2.2.2 Create Measurement Operation Press [Add] button in measurement operation interface, pop up the following dialog box.
Page 105
Digital Color Doppler Ultrasound System Unit Precision Set unit precision. Method The average rule of data. Display In Result Whether display in the result or not. Display In Report Whether display in the report or not. 8.2.2.3 Formula Edit It is necessary to enter into the following interface when creating measurement operation except OB. Fig.
Page 106
Digital Color Doppler Ultrasound System 8.2.3 OB Table Interface Fig. 8-9 Measure Item: Display OB measurement Item. GA Tables: Gestational list for the current measuring project. Growth Table: Growth table for the current measuring. Fetal Weight: Fetal weight calculation formula. ...
Digital Color Doppler Ultrasound System 8.3 Comment Setup Fig. 8-10 Comment Lib: The annotation database of the system is classified as: Abdomen, Cardiac, Gynecology, Obstetrics, Pathological Changes, Small Parts, Vessel. Annotation can be made by inputting characters from the soft keyboard or recalling the terms saved in annotation database. ...
Digital Color Doppler Ultrasound System 1. Press [Edit Comment Type] item, edit box pops up. 2. Input new type name into the new created comment box. Press [Create] button, the new comment will be created and appear in selected comment list. 3.
Digital Color Doppler Ultrasound System 8.5 Exam Mode Setup 8.5.1 Exam Mode Selection Interface Fig. 8-12 Probe: Select needed probe and ExamMode fit for it. ExamModeLib: Show all existing Exam modes. ExamMode Selected: Show Exam modes in probe column. ...
Page 110
Digital Color Doppler Ultrasound System Fig. 8-13 ExamMode: Show all existed exam modes in system. Comment: Double press 【ENTER】 key to activate widget box, can choose existed annotation name. After setting, the default of the exam mode is user-selected. ...
Page 111
Digital Color Doppler Ultrasound System Fig. 8-14 Status After Freeze: Set the status after freezing. Status After Real Time saving Cine Frames: Whether unfreezing upon save or not. End Exam Option: Select the working status after finish the exam. ...
Digital Color Doppler Ultrasound System Fig. 8-15 Preset: Display the current preset. Press the corresponding parameter control key to switch to available preset. Rename: Rename the current preset. Load Preset: Load the preset displayed. Save: Save the current preset. ...
Page 113
Digital Color Doppler Ultrasound System Fig. 8-16 Hospital Logo: Set to use logo in report. Report Font Size: set up the font size of the hospital name, title, text. Zoom In:Zoom in the logo. Zoom Out:Zoom out the logo. ...
DICOM has been activated. In the system page of setting interface, you can check whether DICOM is open or not. If you want to activate DICOM, please contact with CHISON. There must be DICOM SCP server which has been installed with PACS or other relative DICOM server...
Page 115
Digital Color Doppler Ultrasound System Fig. 8-18 Local AE Title: Input local DICOM Title to separate the DICOM equipment in local network. Service: Display the local DICOM function worklist. AE Title: Display the name of local DICOM AE title. ...
Digital Color Doppler Ultrasound System Fig. 8-19 Press [Add] button, then system pops up the DICOM Setting interface. Default: check this option, set DICOM function as default. Service: choose DICOM Storage, DICOM Worklist, DICOM Print or DICOM SR. ...
CHISON). Installment Input relevant key to open trial function and the detail please contact CHISON Company. System Activation Select this item to activate the device if the user does not activate after powering on the system.
Page 118
Digital Color Doppler Ultrasound System Choose the video data: NTSC and PAL-D. Video opened: Choose the item to open this function. VGA opened: Choose the item to open this function. Hardware Function Export hardware SN and import hardware key, it only for engineer use. ...
When it is necessary to clean the probe connector and peripheral instrument, please contact Sales office contact customer service or agent of CHISON. Any self-cleaning may result in malfunction or degrading the function of device.
Otherwise there will be a danger of the electricity shock. 9.3 Safety Check CHISON has verified ESSENTIAL PERFORMANCE and BASIC SAFETY testing including details of the means, methods and recommended frequency and test reports are available from our factory.
Digital Color Doppler Ultrasound System 9.4 Troubleshooting According to the most frequently occurred errors and system messages, the list of possible causes and relevant solutions is attached as below: Malfunction Reason Measures Switch button lights but power 1. Battery lose efficacy Check the connector between LCD not 2.
CHISON warranty after the above improper handling. Standard maintenance must be performed by CHISON‟s authorized service engineer during the life time of the product.
Digital Color Doppler Ultrasound System Chapter 10 Probes 10.1 General Description Cable Scanhead (Lens) Transducer Fig.10-1 Convex Probe Overview NOTE The photograph above is 3C probe for reference. It is used only to identify component names. The appearance of different probes varies. Please be subject to the actual product you have purchased. The probes provide high spatial and contrast ultrasound imaging.
Digital Color Doppler Ultrasound System Make sure not to use or deposit the probe in the environment above 50 degree. If any abnormal phenomena of probe is found, immediately stop operation and contact with Sale Office, Customer Service department or Agents of manufacturer. ...
Page 125
Digital Color Doppler Ultrasound System Transducer Components Transducer Components Fig.10-2: 1. Lens 2. Shell 3. Cover 4. Sheath 5. Cable 6. TYPE-C connector To choose a suitable probe disinfection method, you must first determine its category according to the purpose of the probe. For more information about compatible cleaning and disinfecting agents, please refer to the section "Cleaning or disinfecting ".
Page 126
Digital Color Doppler Ultrasound System When you use OPA (ortho-phthalaldehyde)-based disinfectants, if you do not follow the manufacturer's instructions carefully, residual solution may remain on the probe. In order to minimize the impact of residual OPA or any other disinfectant, it is recommended as below: ...
Page 127
Digital Color Doppler Ultrasound System and then follow the appropriate intermediate-level or high-level disinfection procedure. Probe Disinfection Class, Use, and Method Class Method Cleaning followed by Non-Critical Class Touches intact skin intermediate-level disinfection (ILD) Semi-Critical Touches mucous membranes and Cleaning followed by high-level Class non-intact skin disinfection (HLD)
Page 128
Digital Color Doppler Ultrasound System Fig.10-4: Probe Immersion Levels ® 3. Immerse the probe in Cidex OPA solution up to the immersion line and ensure no air or bubbles are trapped. Allow soaking according to the manufacturer‟s instructions. 4. Thoroughly rinse the probe (up to the immersion line) by immersing it in a large volume of room temperature purified water for a minimum of one (1) minute.
Digital Color Doppler Ultrasound System 10.2.2.3 Approved and compatible cleaning and disinfectant Trade name Manufacturer Origin Qualified Use Remarks 80% Ethanol Wipe 70% Isopropyl alcohol Wipe United Universal wipes clinell Wipe Kingdom Schulke&Mayr Perform classic wipes EP Germany Wipe GmbH WIP'ANIOS Laboratoires ANIOS France...
Before inserting the connector into the probe port, inspect the probe connector pins. If a pin is bent, DO NOT use the probe until it has been inspected and repaired/replaced by a CHISON Service Representative. Electrical leakage checks should be performed on a routine basis by CHISON Service or qualified hospital personnel. Mechanical hazard: A defective probe or excess force can cause patient injury or probe damage: ...
Page 131
Digital Color Doppler Ultrasound System contact the patient. Only immerse the probe to its specified level. Ensure that no solution has entered the probe‟s connector before scanning the patient. If sterilant comes into contact with the patient, refer to the sterilant‟s instruction manual.
Page 132
CAUTION Please do not use any gel or other materials which are not provided by CHISON. Un-authorized gel, lubricants and other materials may corrode probes and other parts of the device, for example the keyboard.
Digital Color Doppler Ultrasound System Daily check: inspect the probes. After each use: clean the probes and disinfect the probes. As necessary: inspect the probes, clean the probes, and disinfect the probes. Returning/Shipping probes and repair parts Transportation dept. and our policy require that equipment returned for service MUST be clean and free of blood and other infectious substances.
Page 134
Digital Color Doppler Ultrasound System and 8%-81% for commercial probe covers have been observed in recent studies. For maximum safety, one should therefore perform high-level disinfection of the probe between each use and use a probe cover or condom as an aid in keeping the probe clean. There are four generally recognized categories of disinfection and sterilization.
Digital Color Doppler Ultrasound System 10.3 Probe Operation Instructions This section introduces instruction for details on connecting, activating, deactivating, disconnecting, transporting and storing the probes, see Chapter 3. 10.3.1 Scanning the Patient In order to assure optimal transmission of energy between the patient and probe, a conductive gel must be applied liberally to the patient where scanning will be performed.
Page 136
Digital Color Doppler Ultrasound System 10.3.3 Cleaning and Disinfecting Transvaginal Probe We strongly recommend wearing gloves when cleaning and disinfecting any endocavitary probe. Every time before and after each exam, please clean the probe handle and disinfect the transvaginal probe using liquid chemical germicides.
Page 137
Digital Color Doppler Ultrasound System instructions. 6 . Flush all visible germicide residues from the probe and allow to air dry. REFERENCE: 1) Amis S, Ruddy M, Kibbler CC, Economides DL, MacLean AB. Assessment of condoms as probe covers for transvaginal sonography. J Clin Ultrasound 2000;28:295-8. 2) Rooks VJ, Yancey MK, Elg SA, Brueske L.
Digital Color Doppler Ultrasound System Appendix A: The Information of EC Representative R Sight B.V. Address: Roald Dahllaan 47, 5629 MC, Eindhoven. The Netherlands SRN: NL-AR-000010445 Tel:0031640845545 CCI No.:76704726 E-mail:info@rsight.nl...
Digital Color Doppler Ultrasound System Appendix B: Acoustic Output Report Table Transducer Model: 7L Operation Mode: B Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.88 0.24 0.24 0.24 Index component Value 0.24 0.24 0.24 0.24 r.α (MPa) 1.93 (mW)
Page 140
Digital Color Doppler Ultrasound System Transducer Model: 7L Operation Mode: M Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.02 0.22 0.32 0.28 Index component Value 0.18 0.22 0.18 0.32 r.α (MPa) 2.25 (mW) 6.85 6.85 6.85 (mW) 6.85 6.85...
Page 141
Digital Color Doppler Ultrasound System Transducer Model: 7L Operation Mode: B+C Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.92 0.38 0.38 0.62 Index component Value 0.38 0.38 0.38 0.38 r.α (MPa) 2.14 (mW) 10.74 10.74 7.73 (mW) 10.74 10.74...
Page 142
Digital Color Doppler Ultrasound System Transducer Model: 7L Operation Mode: PW Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.34 0.15 0.54 0.30 Index component Value 0.15 0.10 0.15 0.54 r.α (MPa) 0.77 (mW) 5.91 5.91 5.91 (mW) 5.91 5.91...
Page 143
Digital Color Doppler Ultrasound System Transducer Model: 3S Operation Mode: B Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.50 0.05 0.05 0.12 Index component Value 0.05 0.05 0.05 0.05 r.α (MPa) 2.16 (mW) 8.32 8.32 8.32 (mW) 4.78 4.78...
Page 144
Digital Color Doppler Ultrasound System Transducer Model: 3S Operation Mode: B+M Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.46 0.24 0.33 0.28 Index component Value 0.24 0.08 0.24 0.33 r.α (MPa) 2.38 (mW) 18.46 18.46 18.46 (mW) 10.25 10.25...
Page 145
Digital Color Doppler Ultrasound System Transducer Model: 3S Operation Mode: B+C Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.23 0.09 0.09 0.27 Index component Value 0.09 0.09 0.09 0.09 r.α (MPa) 1.51 (mW) 18.64 18.64 18.94 (mW) 9.21 9.21...
Page 146
Digital Color Doppler Ultrasound System Transducer Model: 3S Operation Mode: PW Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.96 0.10 0.42 0.25 Index component Value 0.06 0.10 0.06 0.42 r.α (MPa) 0.77 (mW) 13.45 14.25 17.22 (mW) 7.95 8.89...
Page 147
Digital Color Doppler Ultrasound System Transducer Model: 3C Operation Mode: B Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.05 0.11 0.11 0.25 Index component Value 0.11 0.11 0.11 0.11 r.α (MPa) 1.58 (mW) 10.02 10.02 12.62 (mW) 10.02 10.02...
Page 148
Digital Color Doppler Ultrasound System Transducer Model: 3C Operation Mode: B+M Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.12 0.26 0.58 0.36 Index component Value 0.26 0.18 0.26 0.58 r.α (MPa) 1.65 (mW) 20.12 20.12 20.12 (mW) 14.58 14.58...
Page 149
Digital Color Doppler Ultrasound System Transducer Model: 3C Operation Mode: B+C Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.88 0.22 0.22 0.50 Index component Value 0.22 0.22 0.22 0.22 r.α (MPa) 1.39 (mW) 24.21 24.21 28.90 (mW) 18.68 18.68...
Page 150
Digital Color Doppler Ultrasound System Transducer Model: 3C Operation Mode: PW Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.63 0.20 0.74 0.38 Index component Value 0.15 0.20 0.16 0.74 r.α (MPa) 1.00 (mW) 21.79 18.92 22.59 (mW) 12.61 14.60...
Page 151
Digital Color Doppler Ultrasound System Transducer Model: 6MC Operation Mode: B Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.05 0.11 0.11 0.25 Index component Value 0.11 0.11 0.11 0.11 r.α (MPa) 1.58 (mW) 10.02 10.02 12.62 (mW) 10.02 10.02...
Page 152
Digital Color Doppler Ultrasound System Transducer Model: 6MC Operation Mode: B+M Index Label Below Below Surface Surface Surface Surface Maximum Index Value 1.12 0.26 0.58 0.36 Index component Value 0.26 0.18 0.26 0.58 r.α (MPa) 1.65 (mW) 20.12 20.12 20.12 (mW) 14.58 14.58...
Page 153
Digital Color Doppler Ultrasound System Transducer Model: 6MC Operation Mode: B+CFM Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.88 0.22 0.22 0.50 Index component Value 0.22 0.22 0.22 0.22 r.α (MPa) 1.39 (mW) 24.21 24.21 28.90 (mW) 18.68 18.68...
Page 154
Digital Color Doppler Ultrasound System Transducer Model: 6MC Operation Mode: PW Index Label Below Below Surface Surface Surface Surface 0.63 0.20 0.74 0.38 Maximum Index Value Index component Value 0.15 0.20 0.16 0.74 r.α (MPa) 1.00 (mW) 21.79 18.92 22.59 (mW) 12.61 14.60...
Page 155
Digital Color Doppler Ultrasound System Transducer Model: 6E Operation Mode: B Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.36 0.12 0.12 0.16 Index component Value 0.12 0.12 0.12 0.12 at Z (MPa) 0.78 r.α (mW) 4.14 4.14 4.14 (mW)
Page 156
Digital Color Doppler Ultrasound System Transducer Model: 6E Operation Mode: B+M Index Label Below Below Surface Surface Surface Surface 0.36 0.06 0.06 0.09 Maximum Index Value Index component Value 0.06 0.06 0.06 0.06 at Z (MPa) 0.78 r.α (mW) 2.14 2.14 2.14 (mW)
Page 157
Digital Color Doppler Ultrasound System Transducer Model: 6E Operation Mode: B+CFM Index Label Below Below Surface Surface Surface Surface Maximum Index Value 0.37 0.04 0.04 0.08 Index component Value 0.04 0.04 0.04 0.04 at Z (MPa) 0.73 r.α (mW) 2.06 2.06 2.06 (mW)
Page 158
Digital Color Doppler Ultrasound System Transducer Model: 6E Operation Mode: B+PW Index Label Below Below Surface Surface Surface Surface 0.32 0.11 0.27 0.22 Maximum Index Value Index component Value 0.11 0.08 0.10 0.27 at Z (MPa) 0.64 r.α (mW) 5.82 5.34 5.82 (mW)
Digital Color Doppler Ultrasound System Appendix C: Guidance and Manufacturer’s Declaration 1. Guidance and manufacturer‟s declaration – electromagnetic emissions The SonoAir series is intended for use in the electromagnetic environment specified below. The customer or the user of the SonoAir series should assure that it is used in such an environment. Electromagnetic environment –...
Page 160
Digital Color Doppler Ultrasound System Electrical fast ± 2 kV , 100 kHz, for AC power port Mains power quality transient/burst should be that of a typical IEC 61000-4-4 commercial or hospital environment. Surge ± 1 kV line(s) to line(s) Mains power quality IEC 61000-4-5 ±...
Page 161
Digital Color Doppler Ultrasound System 3. Guidance and manufacturer‟s declaration – electromagnetic immunity The SonoAir series is intended for use in the electromagnetic environment specified below. The customer or the user of the SonoAir series should assure that it is used in such an environment. Electromagnetic environment –...
Page 162
Digital Color Doppler Ultrasound System NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a.
Page 163
Digital Color Doppler Ultrasound System 1 kHz sine Pulse LTE Band 13, 704-787 modulation 217Hz GSM 800/900, TETRA 800, Pulse 800-960 iDEN 820, modulation CDMA 850, 18Hz LTE Band 5 1720 GSM 1800; CDMA 1900; 1845 Pulse GSM 1900; 1700-1990 modulation DECT;...
Page 164
Cable or transducer Manufacturer Model and specification 2m long, 60227IEC 53 (RVV) 3×0.75mm Power cable KING-CORD 300/500V unshielded wire 1.2m long, 2464 18AWG VW-1 80 ℃ Cable of powered MEGMEET supply 300V unshielded wire,1core Cable of probe CHISON 2m,shielded wire...
Digital Color Doppler Ultrasound System Appendix E: Display Accuracy and Acoustic Measurement Uncertainties Overall uncertainties Center Frequency: The accuracy of the center frequency measurement is primarily dependent on the digitizer, and is therefore given as ± 15%. Pressure: Depends on the hydrophone measurement, digitizer, non-linear distortion, and water temperature.
Expression of uncertainty in measurement. Three transducer samples have been tested and the confidence coefficient is at 95%, the value of t.975 is 4.30. The measurement data were obtained under the test conditions employed at CHISON. Maximum surface Maximum surface temperature(℃)
Digital Color Doppler Ultrasound System Appendix H: Procedures of setting network sharing For Windows set up, set up a shared document 1. Choose the file you want to share, as the “test” file. 2. Right click this file, choose “properties”, and click “share”. 3.
Page 171
Digital Color Doppler Ultrasound System 4. Choose “read/write” in the permission level in everyone, then click “share”, after that, confirm. 5. If the windows have not set the code, then turn off the password protected sharing is necessary. Instructions as the image shows. a) Click “network and sharing center”...
Page 172
Digital Color Doppler Ultrasound System Set up in system IP set up 1. First confirm the service address of shared files, you can get the IP address in windows interface. In windows “start”-“run” type in “cmd” and enter, then type in “ ipconfig” and enter, you can see the IP address of local service.
Page 173
Digital Color Doppler Ultrasound System 3. Choose “Storage” interface, type in service name, IP address, user name, password and the name of shared files, click “Add” to add a network storage, you can choose the export route. As shown in the picture.
Page 174
Digital Color Doppler Ultrasound System Ping: test if the IP is connected or not. Clear: clear all the IP address, user name, password and names of shared file Update: update the content to the chosen item. Delete: delete the chosen service item. Default: set the chosen item as the default net route.
Page 175
Digital Color Doppler Ultrasound System 2. Configure Net Printer Operation: a) Input name and IP Address which has installed printer equipment. b) Click [Ping] to confirm whether the net connect normally. c) Select PPD File and click [Add], the net connecting information can be saved on the Printer List.
Page 176
CHISON Medical Technologies Co., Ltd. No.3 Changjiang South Road, Xinwu District, Wuxi, 214028 Jiangsu, P.R. China...