CPAPDVB - MANUAL ENG
Baby CPAP 10 & 20
Instructions for use
Revision D
02/02/2023
© Copyright Diamedica (UK) Ltd 2023
Tel: +44 (0)1598 710066
WhatsApp: +44 (0) 7716 503156
Email:
support@diamedica.co.uk
Web:
www.diamedica.co.uk
DCN-0211
ENG
0120
Page 1 of 26

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Baby CPAP 20 and is the answer not in the manual?

Questions and answers

Summary of Contents for Diamedica Baby CPAP 20

  • Page 1 Baby CPAP 10 & 20 Instructions for use Diamedica (UK) Ltd Tel: +44 (0)1598 710066 Grange Hill Industrial Estate WhatsApp: +44 (0) 7716 503156 Bratton Fleming, Barnstaple, Devon, Email: support@diamedica.co.uk EX31 4UH, United Kingdom Web: www.diamedica.co.uk Alphamed Consulting Ltd, Knock, Barnaderg, Tuam, Co.
  • Page 2 FOREWORD This manual is intended to provide guidance on the function, performance and user maintenance of the Diamedica Baby CPAP. The information given in this manual is correct at the date of publication. A separate user manual for the oxygen concentrator is also included and must be reviewed for specific safety and maintenance requirements before use.
  • Page 3 Please refer to all warnings listed in accompanying Oxygen Concentrator manual before use. Only persons who have read and understood this entire manual and therefore deemed competent, are authorized to operate this equipment CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 3 of 26...
  • Page 4 DIAMEDICA BABY CPAP 10 / 20 MANUAL 1. INTRODUCTION 2. CLEANING, GENERAL MAINTENANCE AND DISPOSAL 3. THE COMPONENT PARTS OF THE BABY CPAP 4. SETTING UP THE BABY CPAP 5. ALARMS AND INDICATORS 6. POWER CIRCUIT PROTECTION 7. SPECIFICATIONS 8. GLOSSARY CPAPDVB - MANUAL ENG ©...
  • Page 5 The introduction of bubble CPAP on a neonatal unit should be an important element of an improved “neonatal care package” adapted to resource limited contexts. The Diamedica Baby CPAPs are self-contained mobile units that are powered by mains electricity. Based on an oxygen concentrator, these units supply their own oxygen and medically filtered air to deliver and maintain a positive airway support at an adjustable set oxygen concentration level.
  • Page 6 Both units can be supplied with an optional UPS (Uninterruptible Power Supply) unit which provides not only voltage stabilisation but also battery backup for power outages. For more information, please contact Diamedica. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 6 of 26...
  • Page 7 5. Check that there is no pooled water in the circuit. Store the circuit and bottle in a clean plastic bag (labelled and dated). Store in the dry and clean area (separate from a soiled equipment area). CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 7 of 26...
  • Page 8 Use sterile water or normal saline. During the change of water/humidification bottle – ensure patient- safety. The circuit should be interrupted only for a short period of time. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 8 of 26...
  • Page 9: Water Trap

    The filter should be removed and cleaned in warm soapy water, completely dried with an absorbent towel and replaced. Have a spare dry filter to replace with so there are minimal interruptions to the concentrator function. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 9 of 26...
  • Page 10 • Be oxygen compatible, • Be biocompatible, • Comply with the general requirements of the 93/42/EEC European Directive A full list of available spares is available by contacting Diamedica – support@diamedica.co.uk CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023...
  • Page 11 Image shown above is 10l version, 20l may differ in appearance. Oxygen and air mixing chart External foam filter Bubble bottle holder Voltage protector Power connection CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 11 of 26...
  • Page 12 6 After a few seconds but up to a few minutes the low oxygen light (yellow) will then go out and the correct oxygen level light (green) will come on. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 12 of 26...
  • Page 13 8.Fill the humidifier bottle with sterile water or N-Saline (do not over fill). Connect the humidifier bottle to the gas port outlet and ensure the grill below is not obstructed. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 13 of 26...
  • Page 14 11. Dial up the level of CPAP required: start at 7 cm H O (see figure below). Expiratory limb Bottle with dial-up CPAP level CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 14 of 26...
  • Page 15 In most cases these values will be enough to deliver CPAP, which you can see by continuous bubbling in the bottle. Baby CPAP 10 oxygen /air mixing chart Baby CPAP 20 oxygen /air mixing chart CPAPDVB - MANUAL ENG ©...
  • Page 16 Secure tubing by nostril. placing elastic strap over tubing and secure to Velcro. Elastic strap is correctly positioned when strap is in line within the securement line indicator. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 16 of 26...
  • Page 17 All tapes used (e.g. for naso or oro-gastric tubes, IV lines) need to be “skin- friendly”. Prongs Change every 3 days or earlier if any significant soiling or obstruction. Different sizes and types are available: Options: 1) Hudson prongs (different sizes) CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 17 of 26...
  • Page 18 3.5 mm N4903 Infant 5 mm 4 mm N4904 Small 5.75 mm 5.25 mm N4905 Medium 6.75 mm 5.75 mm N4906 Large 7.75 mm 6.5 mm CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 18 of 26...
  • Page 19: Patient Monitoring

    Check for bubbles in the pressure regulation bottle. Bubbles should be present during the whole respiratory cycle. Exclude any obstruction in the circuit and the patient’s airway. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 19 of 26...
  • Page 20 Increase oxygen flow, check (including humidifier Increase air flow, for bubbles. and water- trap). check for bubbles. Increase CPAP level Consider weaning Increase CPAP level CPAP pressures & CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 20 of 26...
  • Page 21: Alarms And Indicators

    May be illuminated Check tubing and connections from gas with low light also. outlet for blockage. If no obvious fault, refer to concentrator manual or contact Diamedica support. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 21 of 26...
  • Page 22: Power Circuit Protection

    Circuit breaker reset on front of unit. 3A on CPAP 10 version. 5A on CPAP 20 version. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 22 of 26...
  • Page 23: Specifications

    ± 1.5 cm H₂O to lid set point For EMC and electrical safety compliance data please refer to the accompanying Concentrator manual. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 23 of 26...
  • Page 24: Symbols Glossary

    Indicates a medical device that can be care broken or damaged if not handled carefully. Keep dry Indicates a medical device that needs to be protected from moisture. CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 24 of 26...
  • Page 25 Chemical symbol for battery type symbol included beneath Does not contain or presence of natural rubber latex Indicates that an object is capable of being recycled CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 25 of 26...
  • Page 26 Medical device Indicates the item is a medical device Unique Device Identifier Indicates a carrier that contains Unique Device Identifier information CPAPDVB - MANUAL ENG © Copyright Diamedica (UK) Ltd 2023 Page 26 of 26...