Diamedica Baby CPAP
INSTRUCTIONS FOR USE MANUAL
Diamedica (UK) Ltd
Grange Hill Industrial Estate
Bratton Fleming
Barnstaple, Devon
EX31 4UH, United Kingdom
Tel: +44 (0)1598 710066 Fax: +44 (0)1598 710055
Email: support@diamedica.co.uk www.diamedica.co.uk
0088
Revision B
31-05-2018
DCN-0012
Page 1 of 13
© Copyright Diamedica (UK) Ltd 2018

Advertisement

loading
Need help?

Need help?

Do you have a question about the Baby CPAP and is the answer not in the manual?

Questions and answers

Summary of Contents for Diamedica Baby CPAP

  • Page 1 Diamedica (UK) Ltd Grange Hill Industrial Estate Bratton Fleming Barnstaple, Devon EX31 4UH, United Kingdom Tel: +44 (0)1598 710066 Fax: +44 (0)1598 710055 Email: support@diamedica.co.uk www.diamedica.co.uk 0088 Revision B 31-05-2018 DCN-0012 Page 1 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 2 FOREWORD This manual is intended to provide guidance on the function, performance and user maintenance of the Diamedica Baby CPAP. The information given in this manual is correct at the date of publication. A Separate IFU manual for the oxygen concentrator is also included and must be reviewed for specific safety and maintenance requirements before use.
  • Page 3 The bubble-CPAP concentrator delivers both oxygen and continuous positive airway pressure, the level of which depends on the distance the tubing is under the water in the bottle. Page 3 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 4 6. Connect the expiratory limb of the circuit to the bottle. 7. Dial up the level of CPAP required: start at 7 cmH O (see figure below) Expiratory limb Bottle with dial-up CPAP level Page 4 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 5 If there are still not continuous bubbles, check for leaks along the circuit, and adjust the oxygen or air flows according to the chart below. Page 5 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 6 Parents and nursing staff should be advised to watch the child to see if he/she develops cyanosis or severe respiratory distress. Page 6 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 7 Brush to clean both inside and outside of circuit. All brushes and cleaning implements • must be properly cleaned after use – soap water and drip dry Gown or waterproof apron, mask and water-proof gloves • Drying rack • Page 7 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 8  Be oxygen compatible,  Be biocompatible,  Comply with the general requirements of the 93/42/EEC European Directive A full list of available spares is available by contacting Diamedica – support@diamedica.co.uk Page 8 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 9 If the product is returned to the manufacturer at the end of its life the company will ensure disposal in line with the Waste Electrical and Electronic Equipment Directive (WEEE) 2012/19/EU Page 9 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 10 Instructions For Use – Additional information This Baby CPAP unit has been fitted with an AVS30 Voltage Stabilizer to automatically protect the on-board circuitry from damage due to unstable voltage. The AVS30 is an Automatic Voltage Switcher rated at 30 Amps. The AVS30 will...
  • Page 11 DIAMEDICA (UK) LTD Grange Hill Industrial Estate, Bratton Fleming Barnstaple, Devon, EX31 4UH, UK Tel : +44(0)1598 710066 Email : support@diamedica.co.uk www.diamedica.co.uk 0088 Page 11 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 12: Symbols Glossary

    Fragile, handle Indicates a medical device that can be with care broken or damaged if not handled carefully. Keep dry Indicates a medical device that needs to be protected from moisture. Page 12 of 13 © Copyright Diamedica (UK) Ltd 2018...
  • Page 13 The battery Chemical symbol for battery type included recycling symbol beneath Does not contain or Have the presence of natural rubber latex Indicates that an object is capable of being recycled Page 13 of 13 © Copyright Diamedica (UK) Ltd 2018...