Page 1
OPERATORS MANUAL for the ORS-2000 Series Irrigation Solution Warming System O.R. Solutions, Inc. 3901 Centerview Drive, Suite L Chantilly, VA 20151 (800) 343-6771 • (703) 834-0101 (703) 834-0966 Fax 3118.B...
Page 2
OPERATORS MANUAL INTRODUCTION The ORS-2000 Series Solution Warming System is designed to provide controlled warming for irrigation solutions. It is not intended as a blood warmer. Any reference to a model includes all versions of that model (eg. ORS-2066 includes ORS-2066 and ORS-2066D and ORS-2066R).
NOTE: GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN THIS EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTACLE MARKED "HOSPITAL ONLY" OR "HOSPITAL GRADE." CAUTION: POWER MUST BE OFF UNTIL MINIMUM FLUID LEVEL IS POURED INTO DRAPE. TEMPERATURE CONTROLS Upper indicator (PV - Present value) indicates present temperature of solution.
Page 4
OPERATING INSTRUCTIONS Drape solution warmer with O.R. Solutions, Inc. warmer drape. Make sure power switch is OFF before draping. Pour solution into drape, making sure that minimum fluid requirements are maintained. Adjust drape to ensure uniform contact with the basin walls i.e. no air pockets or folds.
2. Temperature Controller verification: The ORS 2000 Series Warmer temperature controller should be verified every (5) years. O.R. Solutions, Inc. requires the use of the following items (or equivalent fluid temperature measuring system with no greater manufactures variance than +/- 2º F) to accurately measure the water temperature for verification: •...
O.R. Solutions, Inc. warrants the ORS-2000 Series Solution Warmer against defects in workmanship and materials for one year from the date of shipment to the user. During the warranty period O.R. Solutions, Inc. will, at its option, either repair or replace defective warmers. Improper use and/or treatment of the warmer voids this warranty at the option of O.R.
Dry the basin with a lint free cloth and isopropyl alcohol as per maintenance instructions on page 4. WARNING: If the Solution Warmer does not operate properly, contact ORS. O.R. Solutions, Inc., can be contacted by telephone at (800) 343-6771 3901 Centerview Drive, Suite L, Chantilly, VA 20151 Page 7 of 9...
SPECIFICATIONS ORS-2000 Series Irrigation Solution Warmer Power: 115 VAC; 2A 50-60 H (230 VAC optional) Fuse: 2X; 3A Slo-Blo type Normal Temperature Range: Room temperature to 120 F (49 Temperature Variance: +/-2 F (1 Temperature Control: Microprocessor temperature controller; LED readout of temperature and Set Point temperature.
Need help?
Do you have a question about the ORS-2066 and is the answer not in the manual?
Questions and answers