  • Page 2: Product Introduction

    Shenzhen Linshang Technology Co., Ltd. I. Product introduction Ultrasonic thickness gauge adopts the principle of the reflected plus ultrasonic measurement, which is specialized in thickness measurement of materials that can transmit ultrasonic waves such as metals (e.g. steel, aluminum, copper, etc.), plastics, ceramics, glass, etc. The instrument employs a professional design of time-chip with a resolution of up to 0.001 mm.
  • Page 3: Operation

    Shenzhen Linshang Technology Co., Ltd. good stability and measurement accuracy. 2. With the function of gain automatic change, the instrument automatically selects the appropriate gain according to the material type and thickness so as to achieve the best measurement effect.
  • Page 4 Shenzhen Linshang Technology Co., Ltd.  The calibration block used is wrong. Please use the standard block in the lower right corner of the instrument to calibrate.  The fluid coupling applied to the standard block is not enough. Please apply enough fluid coupling, press the probe tightly against the standard block and keep it still until the calibration is prompted.
  • Page 5 Shenzhen Linshang Technology Co., Ltd. calibration/Exit/Shutdown]. Short press on the button to confirm your selection. (1) System Setup Language: Short press the button to enter the language selection. These two buttons are available to select language. A short press on the button completes the setting.
  • Page 6 Shenzhen Linshang Technology Co., Ltd. Mode: Short press the button to enter the mode selection. These two buttons available to select measuring modes. A short press on the button completes the setting. Limit Set: The limit setting is only displayed in QC mode. Short press the button to enter the interface of limit settings.
  • Page 7 Shenzhen Linshang Technology Co., Ltd. Set by thickness: Knowing the thickness of the material, the user can measure the sound velocity through the thickness, and then use the back-measured sound velocity to measure the quality of the material and the thickness of the material with similar thickness. Short press the button to enter the prompt interface "Please press the probe tightly against the material with known thickness".
  • Page 8 Shenzhen Linshang Technology Co., Ltd. (3) Reference calibration The function of " Reference Calibration" in the main menu is the same as that of "Benchmark Calibration" when the instrument is powered on. (4) Exit A short press on the button exists the main menu and enters the measurement interface.
  • Page 9 Shenzhen Linshang Technology Co., Ltd. interface of data deletion, and select [Yes] with a short press on the button to delete all recorded data. VII. Aviation plug connection It cannot be rotated or pulled violently because the aviation plug connector has a spring limit. Please refer to the following figure below for correct operation: V.
  • Page 10: Packing List

    Shenzhen Linshang Technology Co., Ltd. VI. Packing List Number Products’ name Quantity Ultrasonic Thickness Gauge Ultrasonic dual-crystal probe(5MHz) 1.5V AA Alkaline battery Fluid Coupling Silicone Case Instructions Certificate/Warranty Card VII. Service 1. The gauge has one-year warranty. If the gauge works abnormally, please send the whole gauge to our company for maintenance.