pivotal ME2002 User Manual

Elevating table – the bolt

Advertisement

Quick Links

ME2002 Elevating Table – The Bolt
User Manual
Applies to the following Models:
ME2002 (115V)

Product Description

Intended Use: The ME2002 Elevating Table – The Bolt is intended for use in a clinical environment to
support patients during treatment.
Max load rating: 400 lbs. (182 kg)
Important! This User Manual contains important information for the user of the product.
All who use this product should review and completely understand its contents. Remember to
keep this manual in a place where it is always available to those using the product.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ME2002 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for pivotal ME2002

  • Page 1: Product Description

    Applies to the following Models: ME2002 (115V) Product Description Intended Use: The ME2002 Elevating Table – The Bolt is intended for use in a clinical environment to support patients during treatment. Max load rating: 400 lbs. (182 kg) Important! This User Manual contains important information for the user of the product.
  • Page 2: Table Of Contents

    Installation & Unpacking……………………………………………………………………………………………………………. 5 Table Controls and Parts ……………………………………………………………………………………………………………. 6 Technical Data……………………………………………………………………………………………………………………………. 9 Operation…………………………………………………………………………………………………………………………………… 10 Inspection and Maintenance…………………………………………………………………………………………………..11 Upholstery Cleaning and Care……………………………………………………………………………………………………… 12 Troubleshooting…………………….…………………………………………………………………………………………………… 13 Service & Warranty……………………………………………………………………………………………………………………… 14 Manufacturer Information…………………………………………………………………………………………………………… 15 ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...
  • Page 3: Safety Precautions And Definitions

    • Never leave patient unattended on table. • Follow preventative maintenance instructions provided in this manual. • If you need service assistance contact Pivotal Health Solutions at 1-800-743-7738. Never leave patient on table unattended. Powered components used to elevate and lower this table present a potential crush hazard.
  • Page 4 WARNING: DANGEROUS VOLTAGE POTENTIAL CRUSH HAZARD: Indicates the presence of a crush hazard from lowering the table. Product contains electronic materials. Discard in compliance with location regulations. Indicates the device manufacturer. ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...
  • Page 5: Installation & Unpacking

    Lading before signing to accept the shipment, then articles damaged in transit should be reported at once to delivering carrier and a claim must be filed by you. If freight was arranged by Pivotal Health Solutions, note any damage on the Bill of Lading before signing acceptance of product and notify Pivotal Health Solutions right away, the day of arrival where feasible no later than the day after arrival.
  • Page 6 Table Controls and Parts ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...
  • Page 7 ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...
  • Page 8: Memory Switch

    DROP ASSEMBLY S2778 MEMORY SWITCH S2724 FOOT SWITCH 2700 ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...
  • Page 9: Technical Data

    Technical Data Model Number ME2002 Name ME2002 Elevating Table – The Bolt Rated Voltage 115V Rated Frequency 60 Hz Rated Current 4 Amps Fuse Type No user serviceable fuse Lifting Capacity 400 lbs. (182 kg) Duty Cycle 2 min on 18 min off...
  • Page 10: Operation

    Check if the legs are level. If the legs are not level move the table up/down in small increments to level the legs. Contact Pivotal Health Solutions or your dealer with any questions or concerns. ASSISTING PATIENT ON TABLE: 1) Run the elevation up then down to verify legs are synced.
  • Page 11: Inspection And Maintenance

    Failure of all 4 legs to operate in unison may indicate the table has been moved or hit, for example during cleaning. o If legs cannot be returned to level and operated in unison, contact Pivotal Health Solutions service team for assistance.
  • Page 12: Upholstery Cleaning And Care

    WHAT NOT TO USE: Using the wrong cleaning agents you will VOID YOUR WARRANTY and crack, dry out and destroy your vinyl. Do not use any cleaning agents that contain alcohol, harsh chemicals or abrasives. Contact Pivotal Health Solutions with any questions. ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com...
  • Page 13: Troubleshooting

    1. Verify power supply connections. 2. Check that outlet is working and breaker is not tripped. 3. If the problem persists, contact Pivotal Health Solutions. Elevation: Table will not 1. Resync motors by holding down arrow on footswitch until table raise or lower.
  • Page 14: Service & Warranty

    A chiropractic table is considered a medical device by the Food and Drug Administration. Therefore, it is necessary that Pivotal Health Solutions, Inc. be notified if the table is sold, destroyed, or otherwise disposed of. Please notify in writing providing your name and the serial number of your table to:...
  • Page 15: Manufacturer Information

    Manufacturer Information (in US) 800-743-7738 (international) 011-1-605-753-0110 Pivotal Health Solutions, Inc. 3003 9 Ave. SW Fax: 605-882-8398 Watertown, SD 57201 Email: service@pivotalhealthsolutions.com World Wide Web: www.pivotalhealthsolutions.com Rev C 01/22/20 ME2002 Elevating Table – The Bolt PivotalHealthSolutions.com 800-743-7738...