The interface module is an accessory that increases the connections options of LAUDA constant temperature equipment. It may only be installed in con- stant temperature equipment that supports the interface provided. Refer to the chapter "Compatibility” in this operating manual for a list of compatible product lines.
This operating manual or parts thereof may not be modified, translated or used in any other capacity without the written consent of LAUDA. Violation of this may obligate the violator to the payment of dam- ages. Other claims reserved.
This operating manual is applicable in combination with the operating manual of the constant temperature equipment in which the interface module is installed. Manuals for LAUDA products are available for download on the LAUDA website: https://www.lauda.de The warnings and safety instructions in this operating manual must be observed without fail.
Use only suitable cables of sufficient length for cable connections. Make sure that the protective screen on the cables and connec- tors complies with EMC regulations. LAUDA recommends using pre- assembled cables. Always lay cables correctly so that they do not pose a tripping hazard.
After installing the equipment, dispose of the packaging materials in line with environmental regulations, see Ä “Packaging” on page 36. If you discover any damage on the interface module, contact LAUDA Service immediately, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Interface module LRZ 917...
The Profibus module is designed for installation in constant temperature equipment that supports the Profibus interface. The Profibus interface allows the user to activate constant temperature equipment via the LAUDA command set (for example, control station). Combining the Profibus interface with an RS 232/485 interface is only permitted in conjunction with the Command remote control, whose RS 232/485 interface can be used independently.
Always observe safety measures against electrostatic dis- charge. The module installation description essentially applies to all LAUDA constant temperature equipment; the example diagrams here show the installation of an analog module in constant temperature equip- ment from the Variocool product line.
Page 12
Secure the cover to the casing using two M3 x 10 screws. The new interface on the constant temperature equipment is ready for operation. Fig. 5: Securing the cover (schematic dia- gram) 12 / 39 Interface module LRZ 917...
Using the module box You can extend LAUDA constant temperature equipment by two additional module slots using the LiBus module box. The module box is designed for interface modules with a large cover and is connected to constant tem- perature equipment via a vacant LiBus socket.
Switch on the constant temperature equipment after installing the new interface. Check whether a software warning appears on the display: Warning SW too old : Please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6. No software warning: Operate the constant temperature equip- ment as normal.
Profibus system. The GSD file provides the Master with device information required for project planning. The bitmap file contains a symbol that shows a stylized plug-in module and the LAUDA logo. The Profibus project planning software can use the file to display a clear graphic representation of the overall system in certain situations.
3 - 6 contain the status value that must be transmitted in the required unit. In responses to write commands, bytes 3 - 6 only contain a value if the response applies to a specific parameter. 16 / 39 Interface module LRZ 917...
Page 17
Analog value not present 0x35 Automatically configured 0x36 Not possible to specify set point value: Programmer is running or paused 0x37 Not possible to start the programmer: Analog set point value input is activated Interface module LRZ 917 17 / 39...
The menu for configuring the interface is integrated in the main menu of the relevant constant temperature equipment: Menu Modules Profibus All operating units except Master Fig. 8: Profibus interface menu The command set used is specified by the Profibus Master. 18 / 39 Interface module LRZ 917...
When using compressed air: Always set a low working pressure to prevent mechanical damage to the connections. If you have any questions about technical modifications, please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Interface module LRZ 917...
The procedure for rectifying a fault depends on the device. Follow the corresponding instructions in the operating manual accompanying the constant temperature equipment. If you are unable to rectify a fault, please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Error messages If an error occurs on the Profibus module, the error number is sent to the Master as an external diagnostic byte.
An additional equipotential bonding line can be used between the con- nected devices to minimize the consequential effects. The integral bus termination of a Profibus device has been set incor- rectly. Interface module LRZ 917 33 / 39...
Page 34
There is no voltage supply to the active bus terminations. A bus termination at the center of the Profibus segment is switched on. The integral bus termination of a Profibus device has been set incor- rectly. 34 / 39 Interface module LRZ 917...
The storage location must meet the ambient conditions specified in the technical data. If you intend to dispose of the module, please read the information in Ä “Old device” on page 36 first. Interface module LRZ 917 35 / 39...
Dispose of the device in accordance with the applicable disposal guidelines in your region. Comply with Directive 2012/19/EU (WEEE Waste of Electrical and Electronic Equipment) if disposing of the product takes place in a member state of the EU. 36 / 39 Interface module LRZ 917...
Page 37
Pin header, 9-pin D-Sub, soldered EQM 042 Connector shell F. 9-pin D-Sub EQG 020 RS 232 cable, shielded, 2 m in length EKS 037 RS 232 cable, shielded, 5 m in length EKS 057 Interface module LRZ 917 37 / 39...
Maximum relative air humidity 80 % at 31 °C and up to 40 °C, 50 % with linear decrease. Ambient temperature range [°C] 5 – 40 Temperature range for storage [°C] 5 – 50 38 / 39 Interface module LRZ 917...