Lauda LRZ 917 Operation Manual

Profibus interface module

Advertisement

Quick Links

Operation manual
Interface module LRZ 917
Profibus module
Read this manual prior to performing any task!
V1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LRZ 917 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lauda LRZ 917

  • Page 1 Operation manual Interface module LRZ 917 Profibus module Read this manual prior to performing any task!
  • Page 2 Manufacturer: LAUDA DR. R. WOBSER GMBH & CO. KG Laudaplatz 1 97922 Lauda-Königshofen Germany Telephone: +49 (0)9343 503-0 Fax: +49 (0)9343 503-222 E-mail: info@lauda.de Internet: https://www.lauda.de Translation of the original operation manual Q4DA-E_13-014, 1, en_US 10/1 /2021 © LAUDA 2021...
  • Page 3: Table Of Contents

    Read commands (Profibus Master input data)..................19 7.4.2 Write commands (Profibus Master output data)................. 24 7.4.3 Availability of the interface functions....................27 Maintenance..................................31 Faults....................................32 Error messages................................32 Troubleshooting................................33 Decommissioning................................35 Disposal....................................36 Interface module LRZ 917 3 / 39...
  • Page 4 Accessories................................... 37 Technical data..................................38 Index..................................... 39 4 / 39 Interface module LRZ 917...
  • Page 5: General

    The interface module is an accessory that increases the connections options of LAUDA constant temperature equipment. It may only be installed in con- stant temperature equipment that supports the interface provided. Refer to the chapter "Compatibility” in this operating manual for a list of compatible product lines.
  • Page 6: Technical Changes

    This operating manual or parts thereof may not be modified, translated or used in any other capacity without the written consent of LAUDA. Violation of this may obligate the violator to the payment of dam- ages. Other claims reserved.
  • Page 7: Safety

    This operating manual is applicable in combination with the operating manual of the constant temperature equipment in which the interface module is installed.  Manuals for LAUDA products are available for download on the LAUDA website: https://www.lauda.de  The warnings and safety instructions in this operating manual must be observed without fail.
  • Page 8: Information About The Interface Module

    Use only suitable cables of sufficient length for cable connections.  Make sure that the protective screen on the cables and connec- tors complies with EMC regulations. LAUDA recommends using pre- assembled cables.  Always lay cables correctly so that they do not pose a tripping hazard.
  • Page 9: Unpacking

    After installing the equipment, dispose of the packaging materials in line with environmental regulations, see Ä “Packaging” on page 36. If you discover any damage on the interface module, contact LAUDA Service immediately, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Interface module LRZ 917...
  • Page 10: Device Description

    The Profibus module is designed for installation in constant temperature equipment that supports the Profibus interface. The Profibus interface allows the user to activate constant temperature equipment via the LAUDA command set (for example, control station). Combining the Profibus interface with an RS 232/485 interface is only permitted in conjunction with the Command remote control, whose RS 232/485 interface can be used independently.
  • Page 11: Before Starting Up

     Always observe safety measures against electrostatic dis- charge. The module installation description essentially applies to all LAUDA constant temperature equipment; the example diagrams here show the installation of an analog module in constant temperature equip- ment from the Variocool product line.
  • Page 12 Secure the cover to the casing using two M3 x 10 screws.  The new interface on the constant temperature equipment is ready for operation. Fig. 5: Securing the cover (schematic dia- gram) 12 / 39 Interface module LRZ 917...
  • Page 13: Using The Module Box

    Using the module box You can extend LAUDA constant temperature equipment by two additional module slots using the LiBus module box. The module box is designed for interface modules with a large cover and is connected to constant tem- perature equipment via a vacant LiBus socket.
  • Page 14: Commissioning

    Switch on the constant temperature equipment after installing the new interface. Check whether a software warning appears on the display:  Warning SW too old : Please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6.  No software warning: Operate the constant temperature equip- ment as normal.
  • Page 15: Operation

    Profibus system. The GSD file provides the Master with device information required for project planning. The bitmap file contains a symbol that shows a stylized plug-in module and the LAUDA logo. The Profibus project planning software can use the file to display a clear graphic representation of the overall system in certain situations.
  • Page 16: Large Protocol

    3 - 6 contain the status value that must be transmitted in the required unit. In responses to write commands, bytes 3 - 6 only contain a value if the response applies to a specific parameter. 16 / 39 Interface module LRZ 917...
  • Page 17 Analog value not present 0x35 Automatically configured 0x36 Not possible to specify set point value: Programmer is running or paused 0x37 Not possible to start the programmer: Analog set point value input is activated Interface module LRZ 917 17 / 39...
  • Page 18: 7.2.2 Short Protocol

    The menu for configuring the interface is integrated in the main menu of the relevant constant temperature equipment: Menu  Modules  Profibus All operating units except Master Fig. 8: Profibus interface menu The command set used is specified by the Profibus Master. 18 / 39 Interface module LRZ 917...
  • Page 19: Interface Functions

    [°C] – (internal / external Pt / external analog / external serial) 7 External temperature TE (Pt) [°C] 18 – 23 ASCII: XXX.XX / -XX.XX 8 External temperature TE (analog [°C] – input) Interface module LRZ 917 19 / 39...
  • Page 20 Large protocol Short protocol resolu- Cmd No. Byte Meaning tion (byte 2) (byte 3) 11 Controller actuating signal [%], – 0.1 % 136 Controller actuating signal – 12 – 17 ASCII: 00-100 = 000100 20 / 39 Interface module LRZ 917...
  • Page 21 47 Control parameter KpE [–] – 49 Control parameter TnE – 51 Control parameter TvE – 53 Control parameter TdE – 55 Correction limitation – 57 Control parameter XpF – 61 Control parameter Prop_E – Interface module LRZ 917 21 / 39...
  • Page 22 (byte 2) (byte 3) 63 Status of keyboard Master: [–] – 0 = free / 1 = blocked 65 Status of keyboard (remote [–] – control): 0 = free / 1 = blocked 22 / 39 Interface module LRZ 917...
  • Page 23 118 Solenoid valve for cooling water [–] – 119 Solenoid valve for automatic [–] – filling device 121 Solenoid valve, shut off valve 1 [–] – 122 Solenoid valve, shut off valve 2 [–] – Interface module LRZ 917 23 / 39...
  • Page 24: Write Commands (Profibus Master Output Data)

    Large protocol Short protocol Cmd No. Byte Meaning (byte 2) (byte 3) 17 Pump power stage (1 – 6 or 1 – [–] – 30 Set pressure (for pressure con- [bar] – trol settings) 24 / 39 Interface module LRZ 917...
  • Page 25 46 Control parameter KpE [–] – 48 Control parameter TnE – 50 Control parameter TvE – 52 Control parameter TdE – 54 Correction limitation – 56 Control parameter XpF – 60 Control parameter Prop_E – Interface module LRZ 917 25 / 39...
  • Page 26 (byte 2) (byte 3) 62 Keyboard Master (equivalent to [–] – "KEY"): 0 = unlock / 1 = lock 64 Keyboard remote control unit [–] – command: 0 = unlock / 1 = lock 26 / 39 Interface module LRZ 917...
  • Page 27: 7.4.3 Availability Of The Interface Functions

    ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü * Equipment type as per rating label Interface module LRZ 917 27 / 39...
  • Page 28 ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü ü * Equipment type as per rating label 28 / 39 Interface module LRZ 917...
  • Page 29 – – ü ü ü ü ü ü ü ü ü ü ü ü ü – – – ü ü ü ü ü ü ü ü * Equipment type as per rating label Interface module LRZ 917 29 / 39...
  • Page 30 ü ü ü ü ü ü ü ü ü ü ü – – – – – – – ü ü ü ü ü – – – * Equipment type as per rating label 30 / 39 Interface module LRZ 917...
  • Page 31: Maintenance

    When using compressed air: Always set a low working pressure to prevent mechanical damage to the connections. If you have any questions about technical modifications, please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Interface module LRZ 917...
  • Page 32: Faults

    The procedure for rectifying a fault depends on the device. Follow the corresponding instructions in the operating manual accompanying the constant temperature equipment. If you are unable to rectify a fault, please contact LAUDA Service, see Ä Chapter 1.6 “Contact LAUDA” on page 6. Error messages If an error occurs on the Profibus module, the error number is sent to the Master as an external diagnostic byte.
  • Page 33: Troubleshooting

    An additional equipotential bonding line can be used between the con- nected devices to minimize the consequential effects.  The integral bus termination of a Profibus device has been set incor- rectly. Interface module LRZ 917 33 / 39...
  • Page 34 There is no voltage supply to the active bus terminations.  A bus termination at the center of the Profibus segment is switched on.  The integral bus termination of a Profibus device has been set incor- rectly. 34 / 39 Interface module LRZ 917...
  • Page 35: Decommissioning

    The storage location must meet the ambient conditions specified in the technical data. If you intend to dispose of the module, please read the information in Ä “Old device” on page 36 first. Interface module LRZ 917 35 / 39...
  • Page 36: Disposal

    Dispose of the device in accordance with the applicable disposal guidelines in your region. Comply with Directive 2012/19/EU (WEEE Waste of Electrical and Electronic Equipment) if disposing of the product takes place in a member state of the EU. 36 / 39 Interface module LRZ 917...
  • Page 37 Pin header, 9-pin D-Sub, soldered EQM 042 Connector shell F. 9-pin D-Sub EQG 020 RS 232 cable, shielded, 2 m in length EKS 037 RS 232 cable, shielded, 5 m in length EKS 057 Interface module LRZ 917 37 / 39...
  • Page 38: Technical Data

    Maximum relative air humidity 80 % at 31 °C and up to 40 °C, 50 % with linear decrease. Ambient temperature range [°C] 5 – 40 Temperature range for storage [°C] 5 – 50 38 / 39 Interface module LRZ 917...
  • Page 39 Abbreviations ......15 Addressing ......15 Interface module LRZ 917 39 / 39...
  • Page 40 Manufacturer: LAUDA DR. R. WOBSER GMBH & CO. KG ◦ Laudaplatz 1 ◦ 97922 Lauda-Königshofen Telephone: +49 (0)9343 503-0 ◦ Fax: +49 (0)9343 503-222 E-mail: info@lauda.de ◦ Internet: https://www.lauda.de...