Summary of Contents for Avtech BioSafe-ECO/DUAL Series
Page 1
Ultra Low Temperature Freezer U ser Manual Version: UTFM-202411 BioSafe-ECO/DUAL Series ○ Please read this manual carefully before use. ○ The appearance of the product is subject to the actual product. ○ In case of product technology or software upgrade, without prior notice.
Page 2
BioSafe-ECO/DUAL Series ULT Freezer Manual Ultra Low Temperature Freezer Qingdao Antech Scientific Co., Ltd. No. 182 Haier Road, Qingdao, China +86 532 8789 0321 | info@antechscientific.com | www.antechscientific.com This manual reviews the operation and maintenance of Antech BioSafe-ECO/DUAL series ULT Freezer.
BioSafe-ECO/DUAL Series ULT Freezer Manual Contents Safety Precautions ................................3 Unpacking ....................................5 Packing List .....................................5 General Recommendations .............................5 a) Temperature Monitoring ............................5 b) General Usage ............................... 5 c) Initial Loading ................................6 Operating Standards ................................6 Installation ....................................6 a) Location ..................................6 b) Wiring ..................................7...
BioSafe-ECO/DUAL Series ULT Freezer Manual Safety Precautions In this manual, the following symbols and conventions are use This symbol used alone indicates important operating instructions which reduce the risk of injury or poor performance of the unit. CAUTION: This symbol, in the context of a CAUTION, indicates a potentially hazardous situation which if not avoided could result in minor to moderate injury or damage to the equipment.
Page 5
BioSafe-ECO/DUAL Series ULT Freezer Manual Below are important safety precautions that apply to this product: Use this product only in the way described in the product literature and in this manual. Before using it, verify that this product is suitable for its intended use.
BioSafe-ECO/DUAL Series ULT Freezer Manual Unpacking At delivery, examine the exterior for physical damage while the carrier’s representative is present. If exterior damage is present, please take pictures of damaged packages and send to the seller. Carefully unpack and inspect the unit and all accessories for damage.
BioSafe-ECO/DUAL Series ULT Freezer Manual Initial Loading Allow the freezer to operate at the desired temperature for a minimum of 12 hours before loading. Load the freezer one shelf at a time, beginning with the top shelf. After loading each shelf, allow the freezer to recover to the desired set point before loading the next shelf.
BioSafe-ECO/DUAL Series ULT Freezer Manual Wiring CAUTION: Connect the equipment to the correct power source. Incorrect voltage can result in severe damage to the equipment. CAUTION: For personal safety and trouble-free operation, this unit must be properly grounded before it is used. Failure to ground the equipment may cause personal injury or damage to the equipment.
BioSafe-ECO/DUAL Series ULT Freezer Manual • Durable construction for reliable operation and safe product storage CAUTION: When moving the freezer, always grasp cabinet surfaces; never pull the freezer by the latch handle. Opening the Door • Remove the padlock if installed.
BioSafe-ECO/DUAL Series ULT Freezer Manual Intended Use The -86°C freezer (refer to Section 1 for the specific model series) described in this manual are high performance units for professional use. These products are intended for use as cold storage in research use and as a general-purpose laboratory freezer, storing samples or inventory at operating temperatures between -40°C and -80°C.
BioSafe-ECO/DUAL Series ULT Freezer Manual Components Fig.1 Control panel (on the wall of the external door): Display the temperature setting and indication of operating status. External door: To open the door, grip the handle. When closing the door, please completely lock the external door handle.
BioSafe-ECO/DUAL Series ULT Freezer Manual System settings Enter the language selection interface when the device is powered on After selecting the language interface, the selected background color is green, click to enter the next step. After selecting the language, enter the initial setting of the system. The initial setting includes ...
Page 13
BioSafe-ECO/DUAL Series ULT Freezer Manual Then enter the current time and click OK to complete the current time setting. The screen returns to the initial setup interface, if the current time is correct can directly carry out the system temperature unit setting, directly click on the interface of "℃" and "℉"for temperature...
Page 14
BioSafe-ECO/DUAL Series ULT Freezer Manual Access mode settings In the authorization mode, there are two modes: Open Access and Authorization Mode. After selecting the authorization mode, it cannot be changed to open access without password. Select Open Access and click “Next” to enter the system; Select the Authorization mode and ...
BioSafe-ECO/DUAL Series ULT Freezer Manual Interface of Touch Screen Controller Main page function introduction: Fig 1. Home page Home page middle area: The center of the home page shows inner temperature, the lower left shows ambient temperature, the lower right shows current power voltage, the upper-left corner shows the system working mode (high-performance or energy-saving), the upper-right corner shows system time.
Page 16
BioSafe-ECO/DUAL Series ULT Freezer Manual Silence: Press the“mute”button when alarming, and the buzzer stops. History curve: Enter the history curve page. Setting page function introduction: Fig. 2 Setting page Parameter setting: Set the temperature and alarm in the cabinet ...
Page 17
BioSafe-ECO/DUAL Series ULT Freezer Manual Note : When using licensing mode to enter the system, click on the parameter settings to enter the authorization password, the correct password can enter the parameter settings ; When entering the system in open access mode, click enter parameter setting to enter the parameter setting page directly.
Page 18
BioSafe-ECO/DUAL Series ULT Freezer Manual The left area of the status module mainly displays the current information of the devices. Door Open: Records the number of door opening alarm. Open door alarm information refresh: Click the refresh button to count the number of ...
Page 19
BioSafe-ECO/DUAL Series ULT Freezer Manual is normally, please do not press this button) Return. Environment Refresh the environment information. Alarm test Alarm test can enter the device alarm test interface, understand and test equipment alarm ...
Page 20
BioSafe-ECO/DUAL Series ULT Freezer Manual Fig. 7 Status page The right area of the status bar mainly displays some device alarm information. In the fault list, the 6 most recent alarm messages will be displayed. The fault bar shows the ...
Page 21
BioSafe-ECO/DUAL Series ULT Freezer Manual Login interface introduction. Fig. 9 Note: Access to the system requires an authorization password to enter the login interface regardless of the mode of use. Full access: Click to enter normal mode. Authorization mode: Click to enter the authorization mode.
BioSafe-ECO/DUAL Series ULT Freezer Manual Alarm and safety features Alarm and safety Safe status Instruction buzzer features operation When the main probe in Main sensor alarm the cabinet is short-circuited or broken Probe alarm Ring When the ring Ring temperature...
Page 23
BioSafe-ECO/DUAL Series ULT Freezer Manual The power supply voltage is too high, please check the High voltage alarm Ring power supply voltage is normal The power supply voltage is too low, please check the Low voltage alarm Ring power supply voltage...
BioSafe-ECO/DUAL Series ULT Freezer Manual Maintenance and Troubleshooting WARNING: Unauthorized repair of your freezer will invalidate your warranty. CAUTION: Maintenance should only be performed by trained personnel. Cleaning the Condenser Filter Clean the condenser filters every two or three months. To clean the filters, complete the...
BioSafe-ECO/DUAL Series ULT Freezer Manual Warranty The Warranty for products covered by this manual is ONE YEAR for whole product parts(except expendable items), THREE YEARS for compressor(s). Period starts THREE months from date your equipment is shipped from our facility. This allows for shipping time so the warranty will go into effect at approximately the same time your equipment is delivered.
Need help?
Do you have a question about the BioSafe-ECO/DUAL Series and is the answer not in the manual?
Questions and answers