Le Pas du Château 85670 SAINT PAUL MONT PENIT FRANCE TEL.: 02-51-98-55-64 FAX: 02-51-98-59-07 EMAIL: info@medicatlantic.fr Website: http://www.medicatlantic.fr USER’S MANUAL VERY WIDE BEDS XXL DIVISYS AND XXL XPRESS CONTENTS IXX1L11/08-120/140 IXL2L11/08-120/140 TRANSPORT CONDITIONS STORAGE CONDITIONS ASSEMBLY CONDITIONS ...
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC TABLE OF CONTENTS Transport conditions……………………………………………………………………3 Transport conditions……………………………………………………………………4 Assembly conditions……………………………………………………………………4 Assembling the bed XXL X Press……..……………………………………………4 3.1. Purpose of bed………………………………………………………………………….9 3.2. 3.3. Dismantling the beds XXL X Press and XXL….……………………………………11 3.4. Setup for use………………….……………………………………………………….13 3.5. List of accessories and compatible boards…....…………...………………14 3.6.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Dear Sir/Madam, You have acquired a MEDICATLANTIC medical bed equipped with its accessories, and we thank you for your custom. Our beds and their accessories are designed and manufactured in compliance with the essential requirements of the European Directive 93/42/EEC.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC STORAGE CONDITIONS The bed, along with the boards and accessories, must be stored at a room temperature of between -10°C and +50°C, and relative humidity of between 30% and 75%. Atmospheric pressure between 700hPa and 1060hPa under the same conditions as for transporting.
Page 6
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC B) Remove the two fastening clips from the foot-half of the bed base. Remove the foot-half of the bed base and place it on the floor. C) Take the jack fastening clip out 1cm. Tilt the base-half and reinsert the clip.
Page 7
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC E) Unfold the whole frame and separate the cross brace from the base until all 6 wheels are placed flat. THE 6 BED WHEELS MUST BE FLAT ON THE GROUND F) Brake the base. G) Place the head-half of the bed base centred and laid on the black brackets, and manually open the back rest so that the handle lies on the end of the base.
Page 8
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC I) Place the foot-half of the bed base centred and laid on the base holders then tilt it so that it is resting on the cross brace bearings. 17/06/2011 - 8 / 33 -...
Page 9
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC J) Ease the foot-half of the bed base and lock it on the bearings. K) Follow operations a, b and c. L) Once the base half is inserted, tighten the Rondo screws fully and put the safety clip...
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC M) Finish by connecting the leg rest jack (according to option). Check that the wires are passed as shown in the photo on the right. 3.2 ASSEMBLING THE BED XXL DIVISYS Brake the wheels and raise the cross braces until the mark is aligned on the base and the bearing.
Page 11
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Place the head-half of the base on the base/cross brace ensemble to lock it with the 2 clips. Place the head-half of the base on the cross brace bearings and, with a second person’s help, connect it in the other base-half.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 3.3 DISMANTLING THE XXL AND XXL XPRESS Brake the wheels and lower the bed base until the mark is aligned on the base and cross brace bearing. Remove the safety clip preventing the base from separating and loosen the two Rondo screws connecting the two half-bases.
Page 13
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Fold the base of the XXL X PRESS Lift the cross brace as far as possible from the base (photo A). While keeping hold of the cross brace, fold the head-half of the base as shown in photo B.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 3.4 SETUP FOR USE Put the bed in the designated room, foreseeing an appropriate perimeter of use for the different functions (variable height), especially if the bed has a lifting pole or side rails. Brake the castors and plug in the power lead, checking that the mains comply with the standards in force and that it is suitable for the supply box voltage.
Bed skirt on wheels 120/140 on XXL only A 553-00 On fixed feet with skirt on XXL only Option a 270 Kg Infrared remote control Option b Only accessories and boards supplied by MEDICATLANTIC guarantee safe use. Compatible boards Reference Abélia 120 cm P126-00 Abélia 140 cm P146-00 Médiform 120 cm...
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Products Characteristics of compatible mattresses 120 cm wide bed Width 116 cm minimum with a high-resilience foam of 34 kg/m³ density minimum, height 14cm mini and 15 to 17cm maxi 140 cm wide bed Width 136 cm minimum with a high-resilience foam of 34 kg/m³...
Page 17
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC ⇒ To lower the side rail. 1. Take the side rail by the top rail. Pull on the locking button. ⇒ To raise the side rail. 1. Take the side rail by the top rail and lift.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Precaution for use. There must be at least 220 mm between the top of the side rail and uncompressed mattress surface. If the side rails are not adapted or are poorly positioned, patient safety may be endangered or a malfunctioning may occur. The side rails must not be used when the patient is a child or if s/he is too small.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 3.8 FLATTENING THE BACK REST ON THE BED In the event of a power cut or failure, flatten the back rest as follows: Disconnect the power supply. Dismantle the headboard and wooden side rail if necessary.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC - The bed must be plugged into a dedicated socket. When use of an adaptor, extension lead or connection plug proves necessary, you must check that its characteristics are suitable for the bed. - The cleaning instructions recommended must be complied with.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 4.3 TECHNICAL CHARACTERISTICS Electrical characteristics Bed code Class II (double insulation) Voltage XXXXXXX Protection level against Current type liquid penetration Frequency Operating time Protection level against electric Absorbed power shocks (type B) 200 VA ...
Page 22
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Weight Normal load in use: 315 kg (Patient 270 kg, Mattress 30kg, Accessories 15kg) Base/cross Head-end Foot-end bed TOTAL brace bed base base 120cm bed XXL XPRESS 110 kg 55 kg 26 kg 29 kg...
Page 23
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Equipotentiality socket Under the head-half of the bed base you will find an equipotentiality socket , identified by the label Label enabling you to connect any electromedical devices. The leads of these devices must pass through the head end and not the sides.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 4.4 USE Carry out a test cycle when the bed is empty to familiarise yourself with the bed functions. Lifting and lowering of back rest Lifting and lowering of variable height Lifting and lowering of electric folding...
Page 25
The wired handset must be hooked to the headboard when not in use. During handling, the spiral lead and/or power lead should be kept away from moving parts such as the cross brace or back rest and should not be knotted. Only use original parts and accessories supplied by MEDICATLANTIC to guarantee safety and maintain product conformity.
Page 26
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Manufacturer’s declaration and guide –electromagnetic emissions The medical bed (see references in contents) has been designed for use in the electromagnetic environment specified below. The user should ensure that it is used in such an environment...
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Note 1 At 80 MHz and 800 MHz, the upper frequency range applies. Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 5.1 MAINTENANCE ECOMMENDATIONS
F OR
P REVENTIVE
M AINTENANCE Preventive maintenance should be carried out in accordance with our specifications and at least once a year by the organisation or person who installed the bed.
Page 29
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Manual leg rest functions properly TR/RTR positions function properly Bed base extension functions properly Castors function properly (pivoting, rolling, etc.) including the steering castor where applicable. Bed immobilises properly (castor brakes, etc.) Verification of tightenings, diverse nuts and bolts, pins, pivot, IV stand...
Page 30
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 20 mm) Electrical safety Electrical safety inspection (Values comply with EN 60601-1) Comments Conclusion Operational (is the safety of the patient, carers and technical staff at risk?) Plan of action (see comments) 3 Recommended date of next quality inspection...
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 5.2 CLEANING AND DISINFECTION High-pressure cleaning is forbidden. Unplug the mains lead. Check that all the electrical parts are connected together. All the sockets of the supply box must be used, otherwise the watertightness of it is not guaranteed.
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC 5. 3 WARRANTIES All of our products carry a warranty against any manufacturing defect, provided the normal conditions for use and maintenance are complied with. Labour costs due to changes in structures or parts under warranty are not taken into account.
Page 33
IXX1L and IXL2L11/08-120/140 MEDICATLANTIC Le Pas du Château 85670 SAINT PAUL MONT PENIT FRANCE TEL.: 02-51-98-55-64 FAX: 02-51-98-59-07 EMAIL: info@medicatlantic.fr Website: http//: www.medicatlantic.fr 17/06/2011 - 33 / 33 -...