  • Page 1 LogTag ® UTRIX-16 Users Guide...
  • Page 2: Table Of Contents

    TABLE OF CONTEXT GENERAL INFO ..................INTRODUCTION .................. CONFIGURATION ................ALARMS ....................ALERTS/STOPPING ................ACCESSING THE FILES ............... TIPS AND TRICKS ................INTERPRETING THE DATA ..............ALARM EVALUATION REPORT ............DATA LIST EVALUATION REPORT ..........UTRIX - 16 FACTORY PRESETS ............
  • Page 3: Safety Information

    Safety Information SAFETY INFORMATION The USRIC-8(M) PDF USB temperature recorder contains a non-replaceable Lithium Battery. When the The UTRIX-16 PDF USB temperature recorder contains a non-replaceable Lithium battery indicates “LOW”, the recorder should be replaced, and the battery recycled or disposed of according Battery.
  • Page 4: Introduction

    Follow the instruction to Green indicator • LogTag UTRIX-16 temperature recorder. It is the need to install proprietary software or hardware at the Blinks if the unit is operating without alerts being present. install and start the software. destination. The recorder can be ordered with pre-set profiles ready...
  • Page 5: Advanced Configuration Options

    recorders for comparison. You can also download data directly into LogTag ® Analyzer without accessing the recorder‘s USB memory storage. CONFIGURING THE UTRIX - 16 ADVANCED CONFIGURATION OPTIONS Click Advanced Options for additional configuration UTRIX-16 products can be ordered pre-configured, Configuring the USRIC-8(M) ready to start.
  • Page 6: Finalizing The Configuration

    CONFIGURING MULTIPLE ALARMS and lower conditions, or only upper, or only lower conditions, or none at all. You can, however, not LogTag Analyzer allows up to 6 different alarm trigger make all alarms upper alarms, and you cannot conditions when configuring a UTRIX-16 for recording, make all alarms lower alarms either.
  • Page 7: Marking A Reading With An Inspection Mark

    the USB port and will continue to do so until you configuration with LogTag Analyzer. Your distributor start the unit. No alarms are processed while can supply more information about this option. pre-start readings are being taken, pre-start readings will not appear in the PDF file, and no Manually PDF file will be generated.
  • Page 8: Accessing The Files

    3. USB Input Device (HID) Driver The file names contain information about the This device is used for communication to LogTag recorded data, which is detailed in the section about Analyzer and its driver will always be installed, Interpreting the Data on page 10. If a recorder has even if LogTag Analyzer is not present on the only taken pre-start readings, PDF and CSV files will computer.
  • Page 9: Tips And Tricks

    LIGHT PATTERNS ON USB RECORDERS TIPS AND TRICKS Configuration During the trip When configuring a USB recorder, allow enough time Always replace the USB protective cap before placing for the unit to acclimatise to the target environment, the UTRIX-16 with the goods. While recording, make particularly if you have configured an upper alarm.
  • Page 10: Interpreting The Data

    Interpreting the Data INTERPRETING THE DATA Figure 5: Sample report page DATA EVALUATION REPORT Recorded Data Chart Data Evaluation - Report The chart shows a graphical representation of Alarm Status the data during the trip. As part of the UTRIX-16 Alarm Status Recorded Data Chart This shows at a glance if the UTRIX-16 recorded alarm...
  • Page 11: Alarm Evaluation Report

    Figure 6: Sample report page with Multiple Alarm Thresholds DATA EVALUATION REPORT WITH MULTIPLE ALARMS Lower Threshold Values A UTRIX-16M shows a different report summary       Data Evaluation - Report with multiple alarms The lower alarm thresholds are shown with blue page to accommodate the extra alarms.
  • Page 12: Data List Evaluation Report

    Figure 7: Sample Data List DATA EVALUATION - DATA LIST List legend File Information Shows the symbols for download marks,inspections The information from the report page is Data Evaluation - Data List   marks and USB paused readings if they appear in repeated here.
  • Page 13: Utrix - 16 Factory Presets

    UTRIX - 16 FACTORY PRESETS LogTag UTRIX-16 recorders can be client configured using LogTag Analyzer. For an out-of-box solution they can also be ordered in volume, pre-configured with a factory profile, ready for starting. If a default configuration does not suit, customers can compile their own profile by specifying the parameters in the tables below when ordering.
  • Page 14: Factory Presets Continued

    For rules regarding setting up multiple alarms please see Alarm Configuration Options on page 1. 3. Upper value must be above lower value. 4. • Instant = one temperature reading is above (below) the threshold. • Consecutive = temperature readings are above (below) the threshold for the time defined in the activation delay without interruption. •...
  • Page 15: Technical Specifications

    TECHNICAL SPECIFICATIONS Environmental Model Numbers IEC 60529: IP64 with USB cap fitted. 1. UTRIX-16 - Single Trip PDF Temperature Recorder with dual alarms. Fits into IP67 Protective Enclosure 200-000020. 2. UTRIX-16M - Single Trip PDF Temperature Recorder with 6 alarms. Case Material Temperature Sensor Measurement Range Polycarbonate.
  • Page 16 AMERICANTHERMAL.COM | 937.429.2114 americanthermal.com | 937.429.2114 | info@americanthermal.com CIM24-004_UTRIX-16_User_Guide © 2024 American Thermal Instruments, Inc. All Rights Reserved. All other trademarks are the properties of their respective owners. AMD_3300-7699 Revision 1 Effective Date: 04/28/2025...