This manual describes the features, operation principles, maintenance and other as- pects of the particle counter KC-31/KC-32. For information regarding the operation of other equipment in the case of incorporating the KC-31/KC-32 into a measurement system with other equipment, always make sure refer to the documentation of the other equipment.
Page 4
Audit trail Describes the audit trail function. Serial interface Describes how to use the internal serial interface. Ethernet connection Describes how to connect the unit to a computer via Ethernet and commu- nication procedure. Measurement parameter file Describes how to use a computer to set measurement parameters. D/A Converter Interface Describes how to use the factory option D/A converter interface.
FOR SAFETY In this manual, important safety instructions are specially marked as shown below. To prevent the risk of death or injury to persons and severe damage to the unit or peripheral equipment, make sure that all instructions are fully understood and observed.
WARNING This particle counter uses laser sources. This unit is designated as a class 1 laser product according to IEC 60825-1:2014. The internal particle detector mechanism uses lasers, but they are fully shielded so that the beam cannot exit from the enclosure. The identification label shown below is affixed to the top of the unit.
“Specifications”. - With compressed or depressed gases. - In environmental conditions which exceed the temperature and humidity limits listed in the section “Specifications”. When intending to use the KC-31/KC-32 to measure other gases besides air, please consult the supplier.
Page 8
Caution Precautions • Operate the unit only as described in this manual. • Do not exert strong force to the inlet. Be especially careful when con- necting or disconnecting the supplied isokinetic probe. • When connecting the supplied isokinetic probe, make the connection firmly by hand.
Page 9
• If the unit is used in a method not specified by the manufacturer, the protection provided to the unit may be impaired. • Do not pass reactive gases through the unit, to prevent the risk of an internal explosion. •...
Page 10
• Insert the plug of the cable from the AC adapter firmly into the power connector on the unit. • Do not use the AC adapter for any other product besides the KC-31/KC-32. • Do not use the AC adapter outdoors or in an environment with high tem- perature and humidity.
Battery precautions • To avoid the risk of overheating, explosion, or ignition, observe the fol- lowing precautions regarding usage and storage of the battery. Do not insert or remove the battery while power to the unit is on. Avoid corrosion buildup on the electrodes. Do not drop the battery or otherwise subject it to strong shocks or pressure.
• The battery is a lithium-ion type. For disposal, it must be handled as industrial waste. Please return the battery to the supplier when it is no longer needed. • Do not use any battery other than the specified type by Rion. Printer precautions •...
Contents FOR SAFETY ................iii Outline ....................1 Light source service life .............2 Particle size detection ...............3 Air flow control ................3 Panel Explanation................4 Front panel .................4 Top panel ...................5 Left side panel ................6 Rear panel .................7 Preparations ..................9 Place of installation ..............9 Connecting the AC adapter ............10 Battery ..................
Page 14
Measurement .................78 Relation between Average (averaging measurement count) and Period Time (periodic measurement interval) when in automatic measurement mode ........80 Measurement method ...............84 Measurement screen ..............86 Stopping the measurement ............90 Last Data Display screen ............91 Relation between error messages and error classes ....92 Storing and displaying data ............93 Displaying stored past data ............93 Folder Selection screen and File Selection screen ....97...
Page 15
Command messages ............... 149 Information request messages ..........160 Information messages ............161 Response messages ..............170 Response example ..............171 Ethernet connection ..............172 Preparations for connection to computer ........ 172 Ethernet communication settings ........... 175 Ethernet communication ............176 Measurement parameter file ............
The unit can display a particle count for six size ranges: ≥0.3 µm, ≥0.5 µm, ≥1.0 µm, ≥2.0 µm, ≥5.0 µm, and ≥10.0 µm. The flow rate is 28.3 L/min for the KC-31 and 50 L/min for the KC-32, ad- justed by an automatic flow rate control function.
Outline While the power is off, the measurement parameter settings is memorized automatically. Measurement to be continued with the same settings the next time power is on. An alarm level can be set to sound a warning buzzer when the particle count exceeds a preset threshold.
The unit is intended for use with connected tubes under reduced pressure (KC-31 to about −1 kPa at the lowest, KC-32 to about −2 kPa at the lowest). The maximum tube length for automatic flow control is 10 meters. Collection efficiency especially for large particles is not assured in this condition.
Panel Explanation Front panel START button and START LED STOP button and STOP LED POWER button Printer Touch panel section display “START” button and “START” LED Pressing this button starts the measurement. When the button is pressed while a measurement is in progress, the current measurement data are abandoned and a new measurement starts.
Panel Explanation Printer section Prints measurement results and parameter setting information (see page 103). Touch panel display The touch panel is used for entering measurement parameters, displaying measurement results, and other functions (see page 24). Buttons shown on various display screens can be operated directly by touching them.
Panel Explanation Left side panel USB (A) connector USB (B) connector Ethernet connector Active LED Link LED USB (A) connector The optional USB flash drive is inserted here. And the stored data is copied to the USB flash drive. USB (B) connector Serves for connection to a computer.
Panel Explanation Rear panel D/A converter interface space Battery compartment ALARM terminals Terminal hole for outlet Power connector Model plate Charging status LED D/A converter interface space This is used when the D/A converter (factory option) is installed. Normally it is covered with a blind panel. Terminal hole for outlet This is used when the outlet (factory option) is installed.
Panel Explanation Battery compartment The internal battery is housed in this space. Opening the cover allows removing and inserting the battery (see page 11). ALARM terminals When the count in a selected particle size range has exceeded the alarm level, these two terminals short-circuit (see page 45). The M3 terminal block accepts either electric wire with a 1.25 mm cross section or spade terminals (Y-type).
Preparations Place of installation Select an installation location for the unit, taking into account the points listed in the “Precautions” section (see page vi). Also observe the following points. When transporting the unit, proceed carefully and do not swing it or put it down roughly.
KC-31/KC-32, with the arrow facing up. 3. Plug the other end of the power cord into an AC outlet. To remove the AC adapter, turn the KC-31/KC-32 off and then perform the above steps in reverse order.
1. Turn the battery compartment fastening screw on the rear of the KC-31/KC-32 and remove the cover. 2. Insert the battery into a slot, as shown in the illustration below. The battery is designed so that it will only slide all the way into the slot if the orientation is correct.
3. Verify that the battery has gone all the way into the slot. Then close the cover and reattach it with the fastening screw. 4. Turn the KC-31/KC-32 on and verify that the battery status is shown on the touch panel on the front of the unit (see page 24).
The charging time for one battery is approximately 3 hours (while no measurement is being carried out). The operation time with one battery is approximately 6 hours for the KC-31 and 3.5 hours for the KC-32 (new battery, continuous repeated measurement without op- erator intervention, at room temperature).
Page 30
Preparations • The relationship between the battery status on the touch panel display of the KC-31/KC-32 (see page 25) and the remaining battery level LCD indication on the rear of the battery is as shown below. Lit in blue Lit in blue...
KC-31/KC-32 as described in “Inserting the battery” on page 11. 6. Turn the KC-31/KC-32 on and verify that the battery status on the touch panel display on the front of the unit is lit in blue (see page 24).
Isokinetic probe and sampling tube Attach the supplied isokinetic probe or sampling tube (plastic tube with 12 mm internal diameter) to the inlet on top of the KC-31/KC-32. Before doing so, remove the tube joint that is connected to the isokinetic probe in the factory default condition.
Page 33
Preparations Note The shape of the isokinetic probe differs for the KC-31 and KC-32. The inlet cap will be required when storing or trans- porting the unit again. Be sure not to lose it. If measuring the sample air using the isokinetic probe at a point distant from the unit, connect the isokinetic probe to the sampling tube with the supplied tube joint.
Preparations Zero count filter To perform purging or similar, attach the supplied zero count filter to the inlet on top of the unit. To inlet of the unit Important Do not disassemble the zero count filter. Do not drop the zero count filter or otherwise subject it to shocks.
Preparations Purging Before using the unit in a cleanroom or similar environment, use the supplied zero count filter to perform purging as follows. 1. Mount the supplied zero count filter on the inlet. 2. Press the “POWER” button to turn the unit on. 3.
Settings Turning the unit on Important Before turning on the power, make sure that all cords and tubes are correctly connected. Note The default display language is Japanese. Refer the “System Configuration screen” on page 55 to change the display language. Hold down the “POWER”...
Settings When password function is on (Password ON) After the startup screen was shown, the login screen appears. The login screen is shown on the display until a se- lection is made: “Administrator”, “User”, or “Guest”. When “Administrator” is selected, the Password Entry screen appears (see page 22).
Settings When password function is off (Password OFF) After the startup screen was shown, the measurement screen appears (see page 24). If password function is off, measurement parameter settings can be changed. Note In the factory default condition, password function is on (default: 0000).
Settings Turning the unit off To turn off the unit, hold down the “POWER” button for at least one second. The Shutdown screen appears and the power is turned off. Shutdown screen Note The shutdown screen may be shown for up to about 20 seconds, depending on the immediately preceding operation, the measurement settings, etc.
Settings Screen display Besides measurement values, the Measurement screen also shows infor- mation about the operation status of the unit. The illustration below is for explanation purposes only, showing all available indications. The display mode is for all particle sizes. Power Login authority connection...
Settings Battery status Shows the state of charge of the inserted battery. If a single battery is in- serted, one indication is shown. If two batteries are inserted, two indications are shown. When the AC adapter is being used, no indication is shown. The blue segment area becomes smaller as the battery discharges.
Page 42
Settings ISO Classification This indication is shown when the [ISO Classification] is set to [Enable] (see page 117). Popup icon Pressing this icon will redisplay a popup screen that has been closed (see page 204). Data area This is the area for displaying particle counts during or after measure- ment, the measurement conditions, and other information.
Settings Status area Indicates the status of the unit by the color and operation of various icons. The table below shows the relation between icon color and operation status. Icon name Color Operation State of the unit Remote mode has been specified by an external equipment Black through the serial interface, or an Ethernet communica- tion is set up...
Settings Menu tree Login Administrator Password entry User Guest All-size Counts / /L / /28.3L / /1000L CUMU/ DIFF MENU LIST LAST DATA Label INFO CONFIG 0.3 µm 0.5 µm Press the [Particle Size] 1.0 µm button to change 2.0 µm the single-size display 5.0 µm 10.0 µm...
Page 49
Settings BACK ALL COPY ALL DELETE Volume / Time Measurement Mode Volume 10L / 28.3L / 100L / 283L / 1000L Time 00:00:00 (Manual) / 00:00:10 to 01:00:00 Average 1 / 2 to 99 Period Time 00:00:00 / 00:00:10 to 24:00:00 ISO Classification Enable / Disable 0.3 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0...
Settings Menu bar Pressing the “MENU” button in the button area of the measurement screen brings up the menu bar. Using the buttons on the menu bar, you can bring up other screens or execute functions. Pressing the “MENU” button again turns the menu bar off.
Settings Information bar Press the “INFO” button on the Measurement screen. The information bar is displayed. Information Information bar display Information bar display screen screen (Measurement progress) (Measurement waiting) “Label” button Changes to the Label Selection screen (see page 37). Average The averaging measurements are indicated as current times / total number of times.
Page 52
Settings Remained Time When in manual measurement mode, --:--:-- is displayed. When in automatic measurement mode, the remaining measurement time is displayed in the format hh:mm:ss ([24-hour] hours : minutes : seconds). When measurement is paused, the display changes to [Next Start] and the remaining time before the next measurement start is displayed in the format hh:mm:ss ([24-hour] hours : minutes : seconds).
Page 53
Settings Selection and setting of label Press the “Label” button on the information bar. The Label Selection screen is displayed. Assigning different label names according to measurement location and measurement date/time will make data management easier. The label names that can be set are up to 128. “Label”...
Settings Label name setting Press the “EDIT” button on the Label Selection screen. The Label Input screen for changing the name of selected label is displayed. The label names that can be input are up to eight characters in the combi- nation of the alphabet of the capital letter and the numerical digit of 0 to 9.
Page 55
Settings Label Input screen “Alphanumeric input” button Serve for entering a label name. Except for the “0” button, each button has multiple letters and numerals assigned to it. “BACK” button Back to the Label Selection screen. “CLR” button Press this button to delete the character displayed in the input column. All displayed characters are deleted and the input column becomes a blank.
Settings Measurement Parameter Configuration screen Use the Measurement Parameter Configuration screen to set, measure- ment mode, measurement time, volume, number of measurement for average, period for measurement interval, alarm, ISO classification, delay time and other parameters. Press the “CONFIG” button on the Menu bar. The Measurement Parameter Configuration screen is displayed.
Page 57
Settings “MEMORY CONFIG” button Changes to the Memory Selection screen (see page 52). “PRINT” button Prints the set measurement parameter configuration. Measurement parameter setting procedure 1. On the measurement parameter configuration screen, enter or select the value for the respective item. To input a value: Press the parameter section of the item to set.
Page 58
Settings List of parameter settings Setting items Settings Measurement mode Volume, Time (default) Volume 10L (default), 28.3L, 100L, 283L, 1000L 00:00:00 (manual: default), Time 00:00:10 (10 seconds) to 01:00:00 (1 hour) Average 1 (no average calculation: default), 2 to 99 00:00:00 (no periodic measurement: default), Period Time 00:00:10 (10 seconds) to 24:00:00 (24 hours)
[10L], [28.3L], [100L], [283L] and [1000L]. A selected [Volume] and an actual measurement volume are different. The table below shows the relation between the [Volume] selection and actual measurement time and measurement volume. KC-31 [Volume] Measurement time Measurement volume 21 sec 10.0 L 28.3L...
Page 60
The measurement volume is calculated using the following equation. (The first place of the decimal point is rounded off.) KC-31: Measurement volume (mL) = 28320 (mL/min) × [Time] (sec) / 60 KC-32: Measurement volume (mL) = 50000 (mL/min) × [Time] (sec) / 60...
Page 61
Settings [ISO classification] Select [Enable] or [Disable] for measurement of air clean- liness evaluation based on ISO 14644-1:2015. [Alarm 1] [Alarm Size 1]: Select particle size range of [Alarm 1] which trigger alarm. An alarm is triggered when the particle count in the range specified by [Alarm Size 1] exceeds the level specified by [Alarm Level 1].
Page 62
Settings Example of setting measurement parameters Setting [Time], [Period Time] and [Delay Time] • To set [Time] (measurement time) to 00:01:00 (one minute) 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed. 2. Press the “CONFIG” button on the menu bar. The Measurement Parameter Configuration screen 1/3 is displayed.
Page 63
Settings Measurement Parameter Measurement screen Configuration screen Press the “MENU” button Set the [Measurement Mode] to [Time], and press the pa- rameter part of [Time] Menu bar display screen Press the “CONFIG” button Time Entry screen...
Page 64
Settings Setting [Average], [Alarm Level 1] and [Alarm Level 2] • To set [Average] to 99 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed. 2. Press the “CONFIG” button on the menu bar. The Measurement Parameter Configuration screen 1/3 is displayed.
Page 65
Settings Measurement Parameter Measurement screen Configuration screen Press the “MENU” button Press the parameter part of [Average] Menu bar display screen Press the “CONFIG” button Numerical Entry screen...
Page 66
Settings Setting [Volume], [Alarm Size 1] and [Alarm Size 2] • To set [Volume] (measurement volume) to 10L 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed. 2. Press the “CONFIG” button on the menu bar. The Measurement Parameter Configuration screen 1/3 is displayed.
Page 67
Settings Measurement Parameter Measurement screen Configuration screen Press the “MENU” button Set the [Measurement Mode] to [Volume], and press the “ ” button on the parameter part right edge of [Volume] Menu bar display screen Press the “CONFIG” button Measurement Volume Select screen Select the measurement volume...
Settings Memory Config screen This is the screen for preparing multiple patterns of setting parameters in advance. These parameters are copied to the Measurement Parameter Configuration screen and used for measurement. Up to three patterns of measurement parameters can be set (stored). On the Measurement Parameter Configuration screen, press the “MEM- ORY CONFIG”...
Page 69
Settings Memory Config Memory Config Memory Config screen 3/3 screen 2/3 screen 1/3 “Measurement Parameter Configuration” button The setting is changed by pressing the parameter part of a set each item. The item that can be set is the same as the Measurement Parameter Configuration screen (see page 42).
Page 70
Settings Note The parameters on the Memory Config screen cannot be used for measurement until the parameters have been reflected to the Measurement Parameter Configuration screen. ISO classification enable/disable cannot be set from the Memory Config screen. If one or more of the settings on the Memory Config screen do not match the ISO classification settings, the settings will not be reflected on a Measurement Parameter Configuration screen where ISO clas-...
Settings System Configuration screen Use the System Configuration screen to set systems, such as current date and time, printer setting and password setting. Press the “SYSTEM” button on the Menu bar. The System Configuration screen is displayed. System configuration button System Configuration screen 1/4 System Configuration screen 2/4 System Configuration screen 3/4...
Page 72
Settings Setting procedure 1. On the system configuration screen, enter or select the value for the respective item. To input a value: Press the parameter section of the item to set. A numeric input screen appears. Enter the value and press the “ENT” button (see page 64).
Page 73
Settings List of parameter settings Setting items Settings Date yyyy-mm-dd (year-month-day) Date and Time Time hh:mm:ss (hours:minutes:seconds) Serial Number Fixed (impossible to change) Info Auto Print ON (default), OFF Type All (default), Average Unit Counts (default), /L, /28.3L, /1000L Printer Decimal Point ON (default), OFF Statistics...
Page 74
Settings Description of parameter settings [Date and Time]: [Date]: Enter the current date (year-month-day) [Time]: Enter the current time (hours:minutes:seconds). [Info]: [Serial Number]: Displays the unique serial number of the unit. It is impossible to change. [Printer]: Make printer-related settings. It is possible to print as a setting besides the screen display.
Page 75
Settings [Statistics]: Specifies whether to print statistic informations (Standard deviation, maximum value, minimum value): [On] (print data) or [Off] (do not print data). Maximum and minimum values are taken from the first measurement run of the specified measurement. Standard deviation is calculated as follows. The average ( ) is requested by the expression below.
Page 76
However, even after the end of the expiration period, login will be possible if the password is correct. If you have forgotten the password, perform the procedure described on the RION Corporation web site (https://www.rion.co.jp/english/).
Page 77
Select when using a USB (B) connector. Allows serial communication with a computer on which software that can use the RS-232C proto- col of the KC-31/KC-32 (such as the optional RP Monitor EVO Ver. 2) is installed. [Ethernet]: Select when communicating using an Ethernet connector.
Page 78
Settings [Display]: [Power Save]: Specifies whether the power save mode that controls the backlight of the unit’s touch panel display is used [ON] or whether the backlight stays on all the time [OFF]. When [ON] is selected, the backlight intensity is automatically reduced after 15 seconds of inactivity and the backlight is turned off com- pletely after 60 seconds of inactivity.
Page 79
Settings [Digital signature]: [Time to logout]: Select the no-operation time interval after which a repeated login will be necessary. The setting range is 00:05:00 (default) to 01:00:00. The 00:00:00 setting turns the feature off. [Operation log]: [Printer]: Select [PRINT] if the audit trail is to be printed. [Media]: Select [COPY] if the audit trail is to be copied to a USB flash drive as a log file.
Settings Example of system setup Date and time setting • To set the date and time to 12:34:56 on April 1st 2012 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed. 2. Press the “SYSTEM” button on the menu bar. The System Con- figuration screen 1/4 is displayed.
Page 81
Settings Measurement screen Date Entry screen Press the “MENU” button Press the “BACK” button System Configuration screen Press the parameter part of [Time] Menu bar display screen Press the “SYSTEM” button Time Entry screen System Configuration screen Press the parameter part of [Date]...
Page 82
Settings Setting [Unit] (Unit of the number concentration when printing) • To set [Unit] (measurement volume) to /L 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed. 2. Press the “SYSTEM” button on the menu bar. The System Con- figuration screen 1/4 is displayed.
Page 83
Settings System Configuration screen Measurement screen Press the “MENU” button Press the “ ” button on the pa- rameter part right edge of [Unit] Menu bar display screen Press the “SYSTEM” button Number Concentration Select screen Select the number concentration...
Microsoft Windows 7 Professional (32 bit/64 bit) Microsoft Windows 8.1 Pro (32 bit/64 bit) Microsoft Windows 10 Pro Connection to a computer Connect the USB (B) connector on the left side of the KC-31/KC-32 to the computer, using the optional USB cable (A to B).
Settings Installation procedure When connecting the KC-31/KC-32 and the computer for the first time, install the USB driver as follows. Note When the KC-31/KC-32 is recognized by the computer, it is shown under [Control Panel] - [System] - [De- vice Manager] as “RION USB to RS232C Converter Virtual COM Port”.
Page 86
Settings Follow the wizard to complete the installation. Screens during installation are as follows.
Page 87
Settings Depending on your environment, [Windows Security] may be dis- played. Click on “Install” or “Continue”. 2. Turn power to the KC-31/KC-32 on, set [I/O] on the System Con- figuration screen 3/4 to [Serial]. Important The above steps must be completed before connecting a USB cable.
Page 88
Settings In case of Windows 7, Windows 8.1, and Windows 10 When the computer detects the KC-31/KC-32, the device driver software installation is started automatically. When the installation has been com- pleted, USB communication is enabled. In case of Windows XP When the computer detects the KC-31/KC-32, [Found New Hardware Wizard] is started up.
Page 89
Settings Click on “Continue Anyway”. Click on “Finish”. The driver installation creates a virtual COM port in the computer. See the section “Checking the virtual COM port” on next page continuously.
Settings Checking the virtual COM port 1. Open the “Control Panel” from the “Start Menu”. 2. Open the “System” in the “Control Panel”. 3. Under “System Properties”, select the “Hardware” tab and open the “Device Manager”. 4. Click on the + at the left of “Ports (COM & LPT)”.
Page 91
The indication “RION USB to RS232C Converter Virtual COM Port” should be shown as COM port name. If this is not shown, check the connection be- tween the KC-31/KC-32 and the computer. If there is an “×” over the icon, the port is not functioning normally. Install the driver again.
Settings Uninstalling the USB driver To remove the installed USB driver, proceed as follows. 1. Open the “Control Panel” from the “Start Menu”. 2. Open the “System” in the “Control Panel”. 3. Under “System Properties”, select the “Hardware” tab and open the “Device Manager”.
Page 93
4. Click on the + at the left of “Ports (COM & LPT)”. 5. Right-click on “RION USB to RS232C Converter Virtual COM Port” and select “Properties”. 6. Select the “Driver” tab on the “RION USB to RS232C Converter Virtual COM Port Properties” screen and click on “Uninstall”.
Measurement This unit has two modes of measurement. Manual measurement mode The unit performs only one measurement, manually. ([Measurement mode] on the Measurement Parameter Configuration screen is set to [Time], and [Time] is set to [00:00:00]) Automatic measurement mode Measures automatically, according to the set measurement parameters. ([Mea- surement mode] on the Measurement Parameter Configuration screen is set to [Volume], or [Measurement mode] is set to [Time], and [Time] is set to [00:00:10] to [01:00:00])
Page 95
Measurement Important Do not switch the power on with the inlet cap in place, or with the inlet blocked in some other way. Otherwise a pressure change inside the sensor can lead to damage and performance degradation. To protect the sensor from damage, ensure that no water, solvents, or other liquids can enter the unit.
Measurement Relation between Average (averaging measurement count) and Period Time (periodic measurement interval) when in automatic measurement mode Note Printing is available when [Auto Print] is [ON] on the System Configuration screen (see page 58). Single automatic measurement Average=1, Period Time=00:00:00 In this case, one measurement for the specified time or volume is carried out, and the result is printed.
Page 97
When using the Volume instead of the Time setting, the measurement time is calculated using the following equation (see page 43). (The first place of the decimal point is rounded off.) KC-31: Measurement time (sec) = [Volume] (mL) / 28320 (mL/min) × 60 KC-32: Measurement time (sec) = [Volume] (mL) / 50000 (mL/min) ×...
Page 98
Measurement (3) Average=n (2 to 99), Period Time=Setting t (00:00:20 to 24:00:00) 1. Measurement is repeated until the measurement count n specified under Average is reached. 2. When measurement n has been completed, average value calculation is carried out and the result is printed. 3.
Page 99
The measurement time is calculated using the fol- lowing equation (see page 43). (The first place of the decimal point is rounded off.) KC-31: Measurement time (sec) = [Volume] (mL) / 28320 (mL/min) × 60 KC-32: Measurement time (sec) = [Volume] (mL) / 50000 (mL/min) ×...
Measurement Measurement method 1. Install the unit in the measurement location. 2. Hold down the “POWER” button for at least one second to turn on the power. 3. Following the startup screen, the login screen appears. Pressing the “Guest” button brings up the Measurement screen. If the “Ad- ministrator”...
Page 101
Measurement Note If the password function is set to OFF, no login screen will appear. If the “START” button is pressed during measure- ment standby (while a countdown is in progress), the countdown is reset to established [Delay Time]. If a stable flow rate is not achieved also when 10 seconds have elapsed since the “START”...
Measurement Measurement screen “Number concentration conversion” button “CUMU/DIFF” button Popup icon “Particle size” Particle count button indication “MENU” button “INFO” button “LAST DATA” button Measurement screen (All size display) “Particle size” button Particle size display can be all-size (6 size) or single-size. Single-size display shows multiple particle size ranges, one at a time.
Page 103
Measurement Particle count indication Particle counts corresponding to particle size ranges are displayed. Even after measurement is finished, the count display is maintained until a new measurement begins. If a particle count exceeds 99,999,999, the particle count space for the corresponding particle size range reads “OVER”...
Page 104
Measurement • Pressing the “LAST DATA” button returns the display to the immediately preceding data indication, allowing you to view the measurement results up to the last data with the same measurement parameter settings (see page 91). • Pressing the “MENU” button to bring up the menu bar and then pressing the “CONFIG”...
Page 105
Measurement Note After closing the popup screen, you can select the popup icon displayed in the upper right of the screen ) to open one high-priority popup screen. The indicated number concentration conversion value includes a rounding error, so it may be not corresponding to calculated value.
Measurement Stopping the measurement When the “STOP” button is pressed during automatic measurement, the STOP selector screen appears. If [ABORT] is selected, the measurement stops. If [CANCEL] is selected, the measurement continues. (When the “STOP” button is pressed and the STOP Selector screen is displayed, measurement still continues at that time.) STOP Selector screen Note...
Measurement Last Data Display screen Pressing the “LAST DATA” button on the Measurement screen brings up this indication which allows viewing of previous measurement data using the same measurement parameter settings. While these data are shown, the display background is blue. Last Data Display screen •...
Measurement Relation between error messages and error classes The relation between error messages about errors that occurred during mea- surement and error classes is listed below. [Information] error classes except for the “LOW BATT.” error message are only shown in the measurement data file and printout. Error message Error class Occurrence condition...
Storing and displaying data Measurement results and the corresponding measurement parameter settings are stored automatically in the internal memory for about 5000 measurements. Displaying stored past data To display stored past data, perform the following steps on the Measurement screen. 1.
Page 110
Storing and displaying data 3. Selecting “ISO14644” brings up a list of available folder label names. Select a label name for data that you want to display. A list of folders ordered by date appears. Press the “BACK” button to return to the previous screen. Selecting “ORDINARY”...
Page 111
Storing and displaying data 4. Select the measurement date for the data to be displayed. The File Selection screen is displayed. The File Selection screen displays a list of measurement data files. A file name is displayed as “hours:minutes:seconds”. Press the “BACK” button to return to the previous screen. If the date of the file you want to display does not appear, press the “<<”...
Page 112
Storing and displaying data 6. Press the “ ” or “ ” button on the Past Data Display screen to view the results for the past data in the file. 7. To check the measurement parameters used for the measurement, press the “Info”...
Storing and displaying data Folder Selection screen and File Selection screen The past measurement data file stored in an internal memory is selected on this screen. When pressing the “LIST” button on the Menu bar, the Folder Selection screen is displayed. When you select a folder on the Folder Selection screen, a File Selection screen for selecting files in that folder appears.
Page 114
Storing and displaying data “BACK” button When pressing this button on the Folder Selection screen, display returns to the Measurement screen. When pressing this button on the File Selection screen, display returns to the Folder Selection screen. “ALL COPY” button All data stored in the internal memory are copied to the inserted USB flash drive.
Storing and displaying data Past Data Display screen This screen shows the data from the file selected on the File Selection screen. While these data are shown, the display background is green. Past Data Display screen • Pressing the “ ” button moves to the previous data and pressing the “ ” button moves to the next data (up to 20 data).
Do not use USB flash drives other than those specified by Rion. Note Keep backups of data stored on USB flash drives, to guard against accidental loss or destruction of data.
Storing and displaying data Data copy Pressing the “ALL COPY” button on the Folder Selection screen or the File Selection screen causes all data to be copied from the internal memory to the USB flash drive. Data on the USB flash drive can then be viewed on a computer.
Printer Caution Before using the printer, be sure to read and follow the “Printer precautions” on page x. Names of printer parts Paper cutter FEED button STATUS LED FEED STATUS Paper outlet Printer cover Open lever Paper cutter Serves for cutting off the printer paper. FEED button The paper is advanced while this button is held down.
Page 120
Printer Open lever Pull the lever towards you to open the printer cover. Printer cover Open this cover to load printer paper. Paper outlet The printed paper emerges here.
Printer Loading printer paper Thermal paper TP-34 (19 m per roll) or lint-free thermal paper TP-33 (15 m per roll) can be used. To load paper into the printer, proceed as follows. 1. Pull the open lever towards you to open the printer cover. 2.
Page 122
Printer Note Always insert the printer paper (do not leave the paper compartment empty). Be sure to use only the thermal paper TP-34 or the lint-free thermal paper TP-33. Otherwise correct printing and paper feed is not assured. When there is not much paper left, a red mark appears along the edge.
Printer Printer maintenance In order to prevent print quality degradation or malfunction due to buildup of paper lint and dust, printer maintenance should be performed as described below about once every six months or every million lines printed (see page 105). Thermal head Moisten a cotton swab with ethanol and use it to clean the heating element section of the print head.
Printer Printing You can print measurement parameter setting information as well as mea- surement results. Print operation • Press the “PRINT” button on each screen to print. Press the “PRINT” button on the Menu bar to print displayed mea- • surement data.
Page 125
Printer Note Even if file print is canceled by selecting [Yes] on the Print Cancel screen, the data already sent to the buffer of the printer will be printed. To cancel the printing operation after closing the Print Cancel screen by selecting [No], press the “FILE PRINT”...
Printer Sample printouts Example of System Configuration screen printout Printout date S y s t e m C o n f i g D a t e 2 0 1 2 - 0 4 - 0 1 Printout start time Serial number T i m e 1 6 : 0 3 : 3 2...
Page 127
Printer Measurement parameter setting screen printout example [Measurement mode] is [Time] Measurement mode C u r r e n t C o n f i g M e a s u r e m e n t M o d e T i m e Measurement time Number of measurements to be...
Page 128
Printer Examples of measurement result printout Model name Time stamp Measurement time or measurement volume 2 0 1 2 - 0 4 - 0 1 1 6 : 0 8 : 3 5 K C - 3 2 T I M E : 6 0 s Number of measurements Title...
Page 129
Printer Example of measurement result printout (Average, Standard deviation, Maximum value, Minimum value) 2 0 1 2 - 0 4 - 0 1 1 1 : 1 3 : 2 3 K C - 3 2 T I M E : 6 0 s Number of measurements P E R I O D :...
Page 130
Printer • If there are measurement data for which a “Warning” condition was detected, the number of times is indicated in the title of average as follows: “AVERAGE: 4 (NG=1)” which means that there were five measurements to be averaged but the “Warning” condition was de- tected for one measurement.
Page 131
Printer Example of measurement result printout (An error has occurred, stop- ping the measurement) 2 0 1 2 - 0 4 - 0 1 1 1 : 1 3 : 2 3 K C - 3 2 T I M E : 6 0 s P E R I O D : 0 0 : 0 2 : 0 0...
Page 132
Printer • If there are measurement data for which the “Warning” or “Info” con- dition exists, the “Warning” or “Info” content is printed in the title. • If the “Protected” condition occurred during the measurement or if the measurement was canceled, the “Protected” content or only the indication “Interrupted”...
Method” (Air Cleanliness Evaluation Measurement [ISO 14644-1 Standard]) and “Sequential Sampling Evaluation Method” (Air Cleanliness Evaluation Measurement [ISO 14644-1 Sequential]). The KC-31/KC-32 incorporates a function to calculate the measurement results required for evaluation of air cleanliness in compliance with the standard evaluation method (ISO 14644-1 Standard).
Air cleanliness evaluation Setting Enable/disable of air cleanliness evaluation are set up on the Measurement Parameter Configuration screen. Measurement Parameter Configuration screen 1/3 Setting the ISO classification 1. Press the “MENU button on the Measurement screen. The menu bar is displayed. 2.
Air cleanliness evaluation Measurement with [ISO Classification] enabled Setting the measurement parameters When [ISO Classification] is enabled, the following restrictions apply for setting measurement parameters. • Only [Time] is available as measurement mode, and the measurement time setting is changed to the value selected when enabling [ISO Classification] (default 00:01:00).
Page 136
Air cleanliness evaluation Measurement 1. Install the unit in the measurement location. 2. Hold down the “POWER” button for at least one second to turn on the power. 3. Following the startup screen, the login screen appears. When the “Administrator” or “User” button is pressed, the Password Entry screen appears.
Page 137
Air cleanliness evaluation 6. When there are multiple measurement points, install the unit at the next measurement point. Then press the “START” button to start the measurement. If there is only one measurement point, or if measurement at the last measurement point was completed, select [End of the Meas].
Page 138
Air cleanliness evaluation Example of printout The example of measurement result printout is shown below. 2 0 1 2 - 0 4 - 0 1 1 1 : 1 3 : 2 3 K C - 3 2 T I M E : 3 0 s P E R I O D : - - : - - : - -...
Air cleanliness evaluation Calculating the measurement parameters for air cleanliness evaluation (ISO 14644-1 Standard) The standard requirement is considered to have been met when the particle concentration x (particles/m ) at each measurement point is the upper limit concentration C (particles/m ) or below in the respective cleanliness classi- fication.
Page 140
Air cleanliness evaluation Cleanliness classification Measurement particle sizes and the upper limit concentration in each cleanliness classification are shown in Table 2. Table 2 Measurement particle sizes and the upper limit concentration Cleanliness Maximum permitted concentration (particle/m classifications ISO 14644-1 0.1 μm 0.2 μm 0.3 μm...
) for the maximum considered particle size D (µm) in ISO classification number N to be the evaluation target, and the flow rate Q (m /s) of the KC-31/KC-32, according to the following equation. (1-1) × 2. The measurement time must be set to a value of at least the above calculated value or higher.
Page 142
Air cleanliness evaluation Evaluation of measurement results The standard requirement is considered to have been met when the particle concentration at each measurement point is the upper limit concentration or below in the cleanliness classification. Note When the number of measurement points is 2 or more, the overall mean of the averages is printed, and when the number of measurement points is 2 to 9, the standard deviation and 95% upper confidence...
“21 CFR Part 11: Electronic Records; Electronic Signatures” by the American FDA. The KC-31/KC-32 continuously records operation history data (when/who/ which operation). These data can be displayed, searched (by time period, operator, class, event), and the display contents can be printed.
Page 144
Audit trail Event class Event Event string Record timing Measurement Measurement When the “START” button was released and measurement has started start START Measurement Measurement When the “STOP” button was released Measure- and measurement has stopped stop STOP ment When the “STOP” button was Measurement Measurement pressed during automatic measure-...
Page 145
Audit trail Event class Event Event string Record timing File export usb_ When coping to USB flash drive File copy memory started Delete all File all delete When file deletion started file When a file was selected for opening File open Example: Data in LOC1 folder measured File open (label) / (date) /...
For power-on recording, the ID is always “Root”. The time stamp for power-on recording is the time when the KC-31/KC-32 system was started. When the clock setting was changed, the new time is applied from the first recording event after the change.
Audit trail Printing the audit trail Printing the audit trail is available from the System Configuration screen. 1. Press the “MENU button on the Measurement screen. The menu bar is displayed. 2. Press the “SYSTEM” button on the menu bar. The System Con- figuration screen 1/4 is displayed.
Page 148
Audit trail Example of printout 2 0 1 2 - 0 4 - 0 1 1 1 : 1 3 : 2 3 K C - 3 2 1 2 3 4 5 6 7 8 1 . 0 . 0 2 0 1 2 - 0 4 - 0 1 1 1 : 1 3 : 2 3 A d m i n i s t...
Copying the audit trail In order to view or search the audit trail on a computer with the LogViewer application supplied with the KC-31/KC-32, you can copy the audit trail to a USB flash drive. The function is available from the System Configuration screen.
Audit trail LogViewer In order to view or search the audit trail on an external computer, the LogViewer application supplied with the KC-31/KC-32 is required. LogViewer has the following functions. Data display • Read “log_XXXX.tsv” file (where XXXX is the serial number) •...
Page 151
Audit trail Installation procedure 1. Insert the CD-ROM supplied with the KC-31/KC-32 into the CD drive of the computer and unzip the file “LogViewer Install.zip” located in the root folder. 2. After unzipping (decompressing) the file, there will be two folders named “Japanese”...
Page 152
Audit trail 5. When the “Confirm Installation” dialog box appears, the preparations for installation are complete. If there is no problem, click the [Next] button to proceed with the installation. 6. When the process is finished, the “Installation Complete” dialog box appears.
Audit trail Uninstalling LogViewer Deleting LogViewer • Uninstall LogViewer by selecting it in “Add/Remove Programs” in the Windows Control Panel. (You can also use the LogViewer installer to uninstall the program.) Deleting the .NET Framework 4 Client profile • Use the same “setup.exe” file as during installation to remove the pro- gram.
Page 154
Audit trail Screen composition Main window Menu bar List view items Model name serial number List view data Log file name Menu Items...
Opening a file From the menu bar, access [File] - [OpenCtrl+O]. In the dialog box that appears, select the audit trail log file copied from the KC-31/KC-32, and click [Open]. Important Log files output by the KC-31/KC-32 have check- sum information to guard against modification.
Page 156
Audit trail Saving the log From the menu bar, access [File] - [Save Ctrl+S]. In the dialog box that appears, select the format for saving the file, either TSV (tab delimited) or CSV (comma delimited). The file name is generated automatically by appending “−1” to the name of the currently open file (without the extension).
Page 157
Audit trail Printing From the menu bar, access [File] - [Print Ctrl+P]. In the dialog box that appears, select the format, either TSV (tab delimited) or CSV (comma delimited). The data to be printed can be specified by selecting an item range in the print dialog box.
Page 158
Audit trail Header Item Explanation Remarks AuditTrail Title Print Page: “Number of page / total number of pages” Model name, version, serial Model: number Date/time search condition Indicate the search condi- range tions for data to be saved. Date: Start date/time and end date/ (“ALL”...
Page 159
Audit trail Copy From the menu bar, access [Edit] - [Copy Ctrl+C]. The currently selected data are copied to the clipboard. When selecting data, hold down the Shift key to select a range, or the Ctrl key so select non-contiguous data. If nothing is selected, all data are copied to the clipboard.
Page 160
Audit trail The items available in the Find dialog box are shown in the table below. Items Explanations Remarks Entries in this field are checked for format validi- By default, the data start Date and end date/time are shown ty, and problems are high- lighted “Root”, “Administrator”, “User”, “Guest”...
When connecting the KC-31/KC-32 to a computer for the first time, perform the steps described in “Installing the USB driver” on page 68. 1. Turn power to the KC-31/KC-32 on and set [I/O] on page 3/4 of the System Configuration screen to [Serial].
For example, when <CR><LF> is chosen as terminator, the message “Q/D” will look as follows: “Q/D<CR><LF>” (5 characters). Note In this manual, the terminator is indicated as <EOL>. Two types of messages can be sent from external equipment to the KC-31/KC-32. Command message: For KC-31/KC-32 operation control Information request message: For getting an information message from the KC-31/KC-32 Two types of messages can be sent from the KC-31/KC-32 to external equipment.
Serial interface Message header The messages for the unit start with a capital letter followed by a slash, such as “J/G0E0M2” and “&X/X1 A10”. In this example, “J/” and “&X/” are message header. The header serves to identify the message type. The following message headers are available: X/, &X/ Command message...
Page 164
Serial interface Message send/receive timing When the unit receives a valid information request message, it returns an appropriate information message. If it receives an invalid message or a command message, it returns a response message to the external equipment. In measurement data auto send mode (S0 mode), the unit automatically sends measurement data information message when a measurement is completed.
Serial interface Command messages By sending command messages to the unit, the external equipment can con- trol its operation. Command messages consist of the header “X/” or “&X/” and one or more com- mands, such as “X/C<EOL>”, “X/V1D3A1<EOL>” or “&X/X1T100<EOL>”. When the unit receives a command message from the external equipment, it returns an appropriate response message.
Message header: Commands Function Remarks Page Set alarm level 0 (alarm is off) 1000 10000 100000 Arbitrary Reset KC-31/KC-32 (initialize) Single Set count display particle size 0.3 µm 0.5 µm 1.0 µm 2.0 µm 5.0 µm 10.0 µm All-size display...
Page 167
Activate S0 mode (data sent automatically after measurement) Activate S1 mode (data sent after the measurement data in- formation request message Q/D is received) Set measurement time KC-31 KC-32 0 (manual measurement mode) 21 seconds 12 seconds 60 seconds 34 seconds...
Page 168
Serial interface Initialize command C The “C” command has the following effects. • Cancel the measurement and measurement data are not sent. • Discard any messages prepared for sending to external equipment (including messages that are currently being sent). • Cancel the light source and pump ON (L1 command) condition in remote mode.
Page 169
Serial interface Set remote mode / local mode command Rn Activate local mode. In local mode, light source and pump on/ off cannot be controlled. Activate remote mode. In remote mode, light source and pump on/off can be controlled. Light source and pump on/off cannot be controlled by the unit. When wishing to control the light source and pump from the external equipment, the unit must first be set to remote mode.
Page 170
Serial interface Turn light source and pump on/off command Ln Turn light source off and stop pump. Turn light source on and start pump. Light source and pump control by external equipment is possible only in remote mode (see “set remote mode / local mode command Rn” on previous page).
Page 171
Serial interface Set alarm level command An Serves to set the alarm level. The “A1” command cancels the alarm. The “A6” command activates the setting made with the [Alarm Level 1] on the Measurement Parameter Configuration screen (see page 45). 0 (alarm is off) 1000 10000...
Page 172
Serves to set the measurement time. The “V7” command activates the setting made with the [Time] on the Mea- surement Parameter Configuration screen (see page 44). 0 (manual measurement mode) 21 seconds (KC-31), 12 seconds (KC-32) 60 seconds (KC-31), 34 seconds (KC-32)
Page 173
Serial interface Measurement start/stop command Gn Stop measurement. This command is only valid during manual measurement mode. In other modes or when measurement is not being carried out, the error response “R/ER3” will be returned (see page 170). Note In manual measurement mode and when S0 mode is set, this unit automatically transmits a measurement data information message “D/”...
Page 174
Serial interface Commands using &X/ in header Set alarm level command “X1 Annnnnnnn” This command sets the alarm level to any value between 1 to 99999999 or “0” (alarm is off). Only integers are valid. Setting is the same condi- tion as the alarm level set with the [Alarm Level 1] on the Measurement Parameter Configuration screen (see page 45).
Page 175
Serial interface Set measurement time command “X1 Tnnnn” This command sets the measurement time for automatic measurement mode to any value between 10 to 3600 seconds or 0 (manual) in 1-second steps. Setting is the same condition as the time set with the [Time] on the Measurement Parameter Configuration screen (see page 44).
Serial interface Information request messages By issuing an information request message, external equipment can request information from the unit. When the unit receives such a message, it outputs the corresponding information message. An information request message starts with the message header “Q/” or “&Q/”...
Serial interface Information messages Information messages serve to indicate the status of the unit, and to transfer measurement data to external equipment when information request messages are normally received (see page 160). Note If there is a problem with a received information request message or communication error occurs, an error response message is returned (see page 170).
Page 178
S0 mode S1 mode Measurement time setting 0 second Report when set to manual measurement mode 21 seconds (KC-31), 12 seconds (KC-32) Report when set the time using command “X/V2” 60 seconds (KC-31), 34 seconds (KC-32) Report when set the time using command “X/V3”...
Page 179
Serial interface Set repeat mode/hold mode setting Repeat mode Hold mode Count display particle size setting 0.3 µm 0.5 µm 1.0 µm 2.0 µm 5.0 µm 10.0 µm All-size display An alarm level setting When alarm is off Report when set using the command “X/A2” 1000 Report when set using the command “X/A3”...
Page 180
Serial interface Current status information message The current status information message contains information about the cur- rent state of the unit. It is sent in response to the current status information request message “Q/J”. The current status information message takes the following format. J/GnEnMn<EOL>...
Page 181
Serial interface Measurement data information message The measurement data information message contains the result for the im- mediately preceding measurement. The timing when this message is sent depends on the measurement data send mode (S0 or S1). S0 mode: When a measurement is completed, the data are automatically sent to the external equipment.
Page 182
Model name This is the model of the unit. Model name: KC-31 or KC-32 Measurement time This is the time taken for one measurement. “t” is a variable in the range 1 to 7200. The statement varies according to the measurement mode.
Page 183
Serial interface Measurement value: The particle count data for each particle size range. A particle count for each particle size (8 digits): cccccccc The initial “E” is a number that has the following meaning: When the measurement was carried out normally: When over-range has occurred (Particle count has exceeded 8 digits: When an error has occurred during measurement:...
Page 184
Serial interface Measurement error information messages The measurement error information message indicates information on the error status of the immediately preceding measurement. The message is output in response to the measurement error information request message “Q/E”. The following messages are used. If two or more errors have occurred, the message for the error with the highest priority is returned.
Page 185
A=nnnnnnnn: nnnnnnnn is 1 to 99999999 or 0 (alarm off) D=nnnUM: nnn is 0.3, 0.5, 1.0, 2.0 5.0 or 10.0 µm C=n: n is 1 to N (N is an integer): KC-31/KC-32 is 1 P=nn:nn:nn: nn:nn:nn is 00:00:10 (10 second) to 24:00:00 (24 hours) or 00:00:00 (no periodic measurement)
Serial interface Response messages When this unit receives a command message normally from the external equipment, it returns a positive acknowledgment. If there is a problem with a received command message or information request message, an error response is returned. Note If an information request message was received nor- mally from the external equipment, the unit returns...
Serial interface Response example For reference, this section shows a sample message exchange. In this example, this unit is set to manual measurement mode, and the mea- surement is carried out under control of the external equipment for a certain amount of time.
Preparations for connection to computer Perform the following steps to prepare the KC-31/KC-32 and the computer for communication. 1. Connect the Ethernet connector on the left side of the KC-31/KC-32 and the LAN port on the computer with an optional LAN cable (cross- wired).
Page 189
5. If the communication between the KC-31/KC-32 and the computer has been established, set the computer from the displayed IP address to the IP address corresponding to Subnet mask. Information about the KC-31/KC-32 will be shown at the right side of the screen.
Page 190
Ethernet connection Important Do not connect the Ethernet connector on the KC-31/KC-32 to any other system except a Rion measurement system. When using the Ethernet communication, the maximum cable length should be 30 m. If the cable length is longer than 30 m, use a hub.
When [Ethernet] is selected, the [REMOTE] icon is shown and the “START” and “STOP” buttons are inactive. Node address The node address of the KC-31/KC-32 is fixed in “A”. Note When [Ethernet] is selected for [I/O], the label setting becomes “1”.
Ethernet connection Ethernet communication Message Format The format of messages passed between the computer and the KC-31/KC-32 is as shown below. <SOH><source><target><STX><message string><ETX><check code><EOT> <SOH>: Char (01 ) (ASCII control code “Start of Heading”) <source>: Node address (“@”, “A”) <target>: Node address <STX>:...
Ethernet connection Check Code The check code serves to detect any errors that occurred during the transmis- sion. The check code consists of the check sum based on the content of the message. It is generated by summing the <source>, <target>, and <message string>...
Ethernet connection Message type Two types of messages can be sent from the computer to the KC-31/KC-32. Command message: For KC-31/KC-32 operation control Information request message: For getting an information message from the KC-31/KC-32 One type of messages can be sent from the KC-31/KC-32 to the computer.
Page 195
Ethernet connection Measurement control commands basics The KC-31/KC-32 starts and stops measurement when it receives the appro- priate start or stop commands from a computer. The basic operation flow of measurement control commands is as follows. C/R=1 Reset the unit...
Page 196
Ethernet connection Command message The computer specifies an action of the KC-31/KC-32 with a command message. Command message consists of the message header “C/” and one to four commands, such as “C/L=1”. In the following explanation, only the parts after the “C/” message header are explained.
Page 197
Identify command “I=n” This command allows checking whether the KC-31/KC-32 can be con- trolled by the computer. Before starting to control a KC-31/KC-32, the computer should send this command. Checking the <identify flag> of the <status response> returned by the KC-31/KC-32 then allows the computer to determine if power was turned off or if a reset was performed.
Page 198
186. Information messages Information messages serve to indicate the status of the KC-31/KC-32 and to transfer measurement data to the computer when information request messages are normally received.
Page 199
Ethernet connection Parameter information message Gives information about the type of the KC-31/KC-32. Sent by the KC-31/ KC-32 in response to the request parameter information “A/P” received from the computer. The parameter information message of KC-32 has the following format.
Page 200
Ethernet connection Status information message Indicates the current status of the KC-31/KC-32. Sent by the KC-31/KC-32 in response to the request status information “A/S” received from the computer. The status information message has the following format. S/L=n,E=n,M=n,I=n,C=‘XXXX’ The meaning of the various items is as follows.
Page 201
Ethernet connection Comment “C” Character string such as error information to be conveyed to the computer. When the comment is conveyed, check the message on pop up screen and the condition of icons on status area (see page 27). And then refer to “Troubleshooting”...
Page 202
Ethernet connection Measurement data information message Indicates the results of the previous completed measurement. Sent by the KC-31/KC-32 in response to the request measurement data information “A/D” received from the computer. The measurement data information message has the following format.
Page 203
Ethernet connection Comment “C” Character string such as error information to be conveyed to the computer. When the comment is conveyed, check the message on pop up screen and the condition of icons on status area (see page 27). And then refer to “Troubleshooting”...
The KC-31/KC-32 can utilize measurement parameter files created on a computer with a measurement parameter configuration utility and stored on an USB flash drive. The KC-31/KC-32 reads such a file and loads the settings into a Memory Config screen (see page 52).
Page 205
Measurement parameter file 3. Select the various measurement parameters, make the required input, and press the [Output] when you are finished. 4. The file output selection dialog box appears. Here you can specify where the output file (Config.bin) is to be created. If an USB flash drive is currently available to the computer, the file can be created directly on the drive.
“MEMORY CONFIG” button. The Memory Selection screen appears. 2. Insert the USB flash drive with the measurement parameter file into the USB (A) connector on the left side of the KC-31/KC-32. The “MEDIA” button on the Memory Selection screen becomes available.
D/A converter interface This unit can output the measurement results converted into the current sig- nal with a range of 4 mA to 20 mA using the factory option D/A converter interface, which can be connected to a current level recorder etc. Factory options are products that can be installed in the unit before the unit is shipped.
D/A converter interface Part names The D/A converter interface output terminals are located on the rear panel of the unit. CH 1 CH 2 4-20mA Output terminals (Factory option) ALARM...
D/A converter interface Load connection The output terminals of the D/A converter interface provide an output current of 4 mA to 20 mA. This is an M3 output terminal board, which can be used to connect wires of 1.25 mm² cross section, or spade terminals (Y-type). Output is 4 mA for a particle count of 0, and 20 mA for the maximum par- ticle count in the selected count range (ex.
D/A converter interface Connection example 4-20mA Output terminal strip Shield Shielded cable DC current meter Note The sum of the DC current meter impedance and cable impedance should be 400 Ω or less (maximum permissible cable length is 30 m).
D/A converter interface Voltage output To use the output of the D/A converter interface with a DC voltage meter or DC voltage level recorder, connect the output terminals to the input of the measuring equipment, using two-core shielded cable, and connect a resistor R in parallel with the measuring equipment.
Page 212
D/A converter interface Connection example 4-20mA Output terminal strip Shield Shielded cable Resistor R DC voltage meter The resistance value R (Ω) to be connected in parallel with the measuring equipment is calculated using the full-scale value E (V) of the measuring equipment, as follows.
D/A converter interface Output details Output current Depending on the setting of [D/A Interface CH1] - [Output Range 1] and [D/A Interface CH2] - [Output Range 2] on the System Configuration screen, a DC current is output that corresponds to the count of particles at or above the particle size selected under [Size 1] and [Size 2] (see page 63).
Page 214
D/A converter interface Output current update rate • After the unit has been turned on, until the first measurement is completed, the output current will be 4 mA. • The output current is updated at the end of each measurement, and the value is maintained until the end of the next measurement.
Page 215
D/A converter interface Example of selecting the particle size range and output range • To change the particle size to output to the [D/A Interface CH1] to 0.5 μm, the output range to 0 to 10000. 1. Press the “MENU” button on the Measurement screen. The menu bar is displayed.
Page 216
D/A converter interface 5. Select the particle size (e.g. 0.5). [Size 1] is set to 0.5 and the System Configuration screen is displayed. 6. Press the “ ” button on the parameter part right edge of [D/A In- terface CH1] - [Output Range 1]. The output range selection screen is displayed.
Maintenance Internal contamination of sensor When the internal sensor has become contaminated, performance will decrease and servicing may become necessary. It is advisable to avoid measurement of air containing mist, moisture droplets, coarse particles, fibers and similar, or a very high number of particles. Also take care to always store the unit with the inlet cap in place.
Maintenance Battery service life As the battery is being recharged many times or as it ages, the operation time available with a full charge will gradually become shorter. When the available operation time has dropped to about one half of the time that was available when the unit was first purchased, the battery should be replaced.
Maintenance Backup battery The unit has a built-in clock that is kept running by a backup battery also while power is off. Because the capacity of the backup battery will deteriorate over time, you should have the battery replaced about once every five years. Regarding re- placement of backup battery, please contact the supplier.
Troubleshooting About pop-up screens Error messages and similar information are provided on popup screens that are shown as an overlay to the regular display contents. Message categories Message Popup screen example Note If the popup screen says “Press this window to close”, the popup icon appears in the upper right of the screen when you press the popup screen or...
Troubleshooting Problem list Problem situations are listed below. Specific troubleshooting methods start on page 208. Message Problem description Page categories General – Unit does not operate. problems MESSAGE The message “BUSY” appears. The message “BACKUP ERROR” MESSAGE appears on startup. The message “Sensor Not Calibrated”...
Page 222
Troubleshooting Message Problem description Page categories Laser related The message “LASER FAIL” problems WARNING appears, and the [LASER] icon in the status area flashes red. The [LASER] icon in the status area – flashes green. The [LASER] icon in the status area –...
Page 223
Troubleshooting Message Problem description Page categories Printer problems The STATUS LED of printer lights – in red. The STATUS LED of printer flashes – red and green alternately. – The printer paper comes out white. – Printer does not operate.
Troubleshooting Problem solving This section explains the nature and causes of problems, and methods for their solution. If multiple solutions are explained, the best solutions are given first. If a problem cannot be solved by the steps described in this section, please contact the supplier.
Page 225
Troubleshooting [Problem] The message “Sensor not calibrated” appears. [Cause] You attempted to start measurement with the unit that had not been calibrated. [Solution] The unit must be calibrated. Please contact the supplier. [Problem] The message “Please set date and time.” appears on startup.
Page 226
Sample air with high particle number concentration has been drawn into the unit. [Solution] Ensure that particle number concentration of sample air does not exceed maximum particle number con- centration (KC-31: 28,000,000 particles/m KC-32: 16,000,000 particles/m Otherwise correct measurement is not possible. Circuit related problems [Problem] The message “BOOT ERROR”...
Page 227
Troubleshooting Sample air related problems [Problem] The message “PUMP FAIL” appears. [Cause] The inlet is blocked and sample air is not being supplied to the pump. [Solution] Restart the unit after checking whether sampling tube is broken or crushed. If the above checks are negative, the internal pump or pump circuit problem.
Page 228
Remove the USB flash drive and re-insert it. If the error is not remedied, insert a new drive. (Use USB flash drive provided by Rion.) If an error recurs after the above solution, repair is required. Please contact the supplier.
Page 229
The message “MEMORY FULL” appears. [Cause] The USB flash drive has no free space left. [Solution] Insert a new USB flash drive. (Use USB flash drive provided by Rion.) Alternatively, delete data from the USB flash drive. [Problem] The message “MEMORY ERROR” appears. [Cause] Data cannot be written on the USB flash drive.
Troubleshooting [Problem] The message “NO MEMORY” appears when the USB flash drive has not been inserted in the unit. [Cause] Data cannot be written on the internal memory. [Solution] Please contact the supplier. Battery problems [Problem] The message “LOW BATT.” appears. [Cause] The amount of charge of the inserted battery has decreased.
Troubleshooting USB problems [Problem] Changing the [I/O] setting to “Serial” has no effect. [Cause] USB (B) connector problem. [Solution] Please contact the supplier. Printer problems [Problem] The STATUS LED of printer lights in red. [Cause] Paper out. [Solution] Load a new roll of printer paper (see “Loading printer paper”...
Size resolution 15% or less (in the vicinity of 0.3 µm PSL particles) Responsivity 0.5% or less Maximum particle number concentration KC-31: 28,000,000 particles/m (coincidence loss within 10%) KC-32: 16,000,000 particles/m (coincidence loss within 10%) False count rate...
Page 233
00:00:10 to 01:00:00 (setup at one second bit) Sample volume KC-31: 10 L (21 sec), 28.3 L (60 sec), 100 L (212 sec), 283 L (600 sec), 1000 L (2119 sec) KC-32: 10 L (12 sec), 28.3 L (34 sec), 100 L (120 sec),...
Page 234
Specifications Measurement modes Manual measurement Measurement controlled with “START” and “STOP” buttons Automatic measurement Averaging measurement Repeated measurement up to 99 times of preset time or fixed volume and averaging of results Periodic measurement Averaging measurement carried out at each specified time interval (00:00:10 to 24:00:00, setup at one sec- ond bit) Count display...
Page 235
Specifications D/A converter interface (factory option) Converts the particle count in a selected channel into 4 mA to 20 mA DC current Output range Select one from 0 to 10, 0 to 100, 0 to 1,000, 0 to 10,000, 0 to 100,000 Load impedance 0 Ω...
Page 236
Battery Lithium-ion Two batteries can be inserted in the unit Battery life KC-31: Approx. 6 hours (one battery is inserted in the unit) Approx. 12 hours (two batteries are inserted in the unit) KC-32: Approx. 3.5 hours (one battery is inserted in the unit) Approx.
Page 237
Specifications Environmental Requirements Operation Environments Indoor Use Only Altitude Up to 2000 m Supply Voltage Fluctuations 100 V to 240 V AC ±10% Overvoltage Category Pollution Degree 2 Protection Class Environmental conditions for operation +10˚C to +35˚C, 85% RH or less (no condensation) Environmental conditions for storage −10˚C to +50˚C, 90% RH or less (no condensation) Dimensions...
Page 238
Specifications Options Battery Battery charger USB flash drive USB cable (A to B) Carrying case Thermal paper (10 rolls) TP-34 Lint-free thermal paper (6 rolls) TP-33 RP monitor EVO (monitoring software) K0505 RP monitor Evo10 (monitoring software) K1701 Factory options D/A converter interface Outlet (9 mm ×...
Page 239
Specifications Rear view Top view Left side view Front view Unit: mm Dimensional Drawings...
Need help?
Do you have a question about the KC-31 and is the answer not in the manual?
Questions and answers