1. GENERAL DESCRIPTIONS ................2 1.1 S ..................... 2 COPE OF APPLICATION 1.2 B ....................2 ASIC PRINCIPLES 1.3 B ................. 2 ASIC SETUP AND NAMES OF PARTS 2. TECHNICAL PARAMETERS ................3 2.1 P ..................... 3 ROPERTY INDEX 2.2 M .....................
Min. area diameter Critical thickness of Type measured (mm) of the base (mm) the base (mm) ∅7 TT220 Note: H -- nominal value. (2) Environment of application: Temperature: 0~40°C Humidity: 20%~75% No strong magnetic field (3) Power source: One 3.6v nickel-cadmium battery (4) Dimension: 150mm X 55.5mm X 23mm...
If the sign “ ” is displayed, it indicates that the voltage of the battery is low and recharge is needed. If the sign “ ” is displayed after switching on and the machine is automatically shut off, it indicates that the voltage has gone down to the limit and immediate recharge is necessary.
Batch method: The method facilitates users to record data taken in batches, with each batch containing 15 values. Each time when a batch of 15 values are stored, the screen displays: Tested 15! The group of data and its statistical value can be printed out by using the mode “PRINT ALL”...
Then, press the “ON/C” button to delete all the data in the memory. (If no deletion is desired, press the key in the “DEL ONE?” mode.) (4) Statistical calculation So long as there are three values, calculation can be done. Repeat the testing at least three times and press the “MODE”...
Only the measured value and statistical value are involved in the conversion between the metric and English systems. (6) Print -Out Single print-out – Corresponding to single measurement method, it prints out the measures value for each measurement. The method of operation is: in the single measurement mode, press “MODE”...
4. Calibration In order to measure the thickness accurately, it is necessary to correct the instrument on the measuring site. 4.1 Standard sheets for calibration (Foil and Base) Foils with known thickness or samples with known thickness of their coatings can be made as standard sheets.
Page 11
(1) Zero point calibration a. To conduct measuring once on the base, the screen displays <X.Xµm>. b. Press the “ON/C” button, the screen displays <0.00µm> and the zero point calibration is finished. c. To calibrate the zero point accurately, the procedure of a and b should be repeated until the measured value of the base is less than 1µm.
After the six values are input, measure the zero point and the instrument automatically shuts off, with the new calibrated values stored in the memory. When the instrument is switched on again next time, it will work according to the calibrated values. 5. Points for Attention...
Measuring should not be done in the curved surface of the test piece. (5) Number of readings As the reading of each time is not entirely the same, it is necessary to obtain several reading for an area measured. The local differences of the thickness of coating also call for a number...
3. If the instrument does not work property without any error messages showing, such as: Unable to shut off automatically; Unable to conduct measuring; Buttons refuse to work; Abnormal reading. The instrument should be forced reset. The method of reset is: a. Press and MODE buttons simultaneously and the screen display the following: TT220.
Page 15
BE SURE4? Release the button and the screen displays: RESET TT220 The reset is thus completed. f. When giving up forced reset in the “RESET” mode, press Attention: After reset, all the previously obtained calibration values are lost and the basic calibration should be re-done (for method, see Chapter Four).
We would be very much grateful if you send back the instrument together with a brief description of the errors. Table of Error Messages Error Code Meaning of the code Causes and solution Instrument trouble Forced reset Probe worn Replace probe Damaged probe or instrument Repair Measuring value not reliable,...