Sejoy DBP-62A4B Owner's Manual

Arm-type fully automatic digital blood pressure monitor

Advertisement

Quick Links

Owner's Manual
Arm-type Fully Automatic
Digital Blood Pressure Monitor
DBP-62A4B
Model
Arm Type
Arm Type
Contents
Safety Notice ...............................................................
Unit Illustration .........................................................
Important Testing Guidelines ..................................... 09
Quick Start ................................................................. 10
Unit Operation ...........................................................
Battery Installation ...........................................
System Settings .................................................. 12
Applying the Arm Cuff ....................................... 14
Testing ................................................................ 15
Power Off .............................................................18
Memory Check and Last 3 Test Average
Memory Deletion ..................................................20
Low Battery Indicator ...........................................20
Bluetooth requirement and connection ...............21
Troubleshooting .....................................................23
Blood Pressure Information ........................................ 24
Maintenance ............................................................... 28
Specifications ............................................................. 30
Warranty .................................................................... 32
........................................................ 37
Contact Information
The lay operator or lay responsible or ganization should contact
the manufacturer or the representative of manufacturer.
-for assistance,if needed,in setting up,using or maintaining
the product,or
-to report unexpected operation or events.
Manufactured by JOYTECH Healthcare Co.,Ltd.
No.365, Wuzhou Road,Yuhang Economic Development Zone,
Hangzhou City,311100 Zhejiang,China
Email: info@sejoy.com
Telephone: +86-571-81957767
Fax: +86-571-81957750
1
Safety Notice
02
Thank you for purchasing the DBP-62A4B Blood Pressure Monitor. The unit has been constructed
using reliable circuitry and durable materials. Used properly, this unit will provide years of
06
satisfactory use.
Indications for use: The Arm-type Fully Automatic Digital Blood Pressure monitors are
intended to measure blood pressure(systolic and diastolic) and pulse rate of adults and
adolescents over 12 years of age.
11
All functions can be used safely and values can be read out in one
LCD DISPLAY. Measurement position is on adult upper arm only.
11
The PATIENT is an intended OPERATOR.
Blood pressure measurement determined with this device are equivalent to those obtained
by a trained observer using the cuff/ stethoscope auscultation method, within the limits
prescribed by the Recognized Consensus Standard (IEC 80601-2-30) for electronic
sphygmomanometers.
.............19
Precautions to Ensure Safe, Reliable Operation
1. Do not drop the unit. Protect it from sudden jars or shocks.
2. Do not insert foreign objects into any openings.
3. Do not attempt to disassemble the unit.
4. Do not crush the pressure cuff.
5. If the unit has been stored at temperatures below 0 ℃, leave it in a warm place for
about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
6. If the unit has been stored at temperatures above 40 ℃, leave it in a cool place for about
15 minutes before using it. Otherwise, the cuff may not inflate properly.
7. Do not store the unit in direct sunlight, high humidity or dust.
8.To avoid any possibility of accidental strangulation, keep this unit away from children
and do not drape tubing around your neck.
9.Ensure that children do not use the instrument unsupervised; some parts are small
enough to be swallowed.
10.Some may get a skin irritation from the cuff taking frequent readings over the course
of the day, but this irritation typically goes away on its own after the monitor is removed.
Document No. : JDBP-A404-054
Version: Z
Date of Issue: 2022.3.11
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DBP-62A4B and is the answer not in the manual?

Questions and answers

Summary of Contents for Sejoy DBP-62A4B

  • Page 1: Table Of Contents

    Fax: +86-571-81957750 Contents Safety Notice Safety Notice ............... Thank you for purchasing the DBP-62A4B Blood Pressure Monitor. The unit has been constructed using reliable circuitry and durable materials. Used properly, this unit will provide years of Unit Illustration ............satisfactory use.
  • Page 2: Safety Notice

    Safety Notice Safety Notice Important Instructions Before Use ”appears on screen. Replace both batteries at the same time. 23. Replace batteries when Low Battery Indicator“ 1. Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should 24. Do not mix battery types. Long-life alkaline batteries are recommended. only be interpreted by a health professional who is familiar with your medical history.
  • Page 3: Important Testing Guidelines

    Unit Illustration Unit Illustration Arm Cuff Contents Owner's Manual Arm-type Fully Automatic Digital Blood Pressure Monitor Model DBP-62A4B Arm Type 2.Owner's Manual Arm Cuff Air Plug 1.Monitor Unit Air Tube Display Electric Quantity Indicator 4.Storage Bag 3.Arm Cuff Last 3 Tests Average Irregular Heartbeat Indicator 5.2MOPPMedical AC Adapter(DC5.0 V,1000mA)
  • Page 4: Unit Operation

    Unit Operation Unit Operation System Settings Battery Installation With power off, press "SET" button to activate System Settings. The Memory Group icon flashes. Slide battery cover off as indicated by arrow. 1. Memory group Settings Install 4 new AAA alkaline batteries according to polarity. .Press the “GROUP”...
  • Page 5: Testing

    Unit Operation Unit Operation Testing 2. Testing After cuff inflation, air will slowly rise as indicated by the corresponding cuff pressure 1. Power On value. A flashing " " will appear simultaneously on screen signaling heart beat Press and hold "START/STOP" button to turn the unit on. The LCD screen will appear for one second as unit performs a quick diagnosis.
  • Page 6: Memory Deletion

    1. Download the free “JoyHealth” App: On your mobile phone or table -Transfer your readings go to www.sejoy.com. 3. Press the BP [START/STOP] button to turn your monitor off. 1. As soon as your measurement is complete, open the app on your smart phone to transfer 2.Open the App on your phone or tablet.
  • Page 7: Troubleshooting

    WHO Blood Pressure Classification Indicator Health Reminder The DBP-62A4B is equipped with a classification indicator based on established guidelines Hypertension is a dangerous disease that can affect the quality of life. It can lead from the World Health Organization. The chart below (color coded on monitor unit) to a lot of problems including heart failure, kidney failure, and cerebral hemorrhaging.
  • Page 8: Blood Pressure Q&A

    Arm-type Fully Automatic Blood Pressure Monitor a soft cloth to clean cuff's surface. Use a damp cloth (water-based) to wipe clean. Allow cuff to dry naturally at room temperature. The cuff must be cleaned and disinfected DBP-62A4B Model before use between different users.
  • Page 9: Warranty

    Specifications Warranty Fits arm circumference 22-36 cm Safety Standard( included but not limited): Fits arm circumference 22-42 cm Cuff Circumference 1. IEC 80601-2-30, medical electrical equipment - part 2-30: particular requirements Fits arm circumference 32-48 cm for the basic safety and essential performance of automated noninvasive Temperature sphygmomanometers.
  • Page 10: Additional Notes

    Electromagnetic Compatibility Information Electromagnetic Compatibility Information Table 3 Table 4 Guidance and declaration of manufacturer-electromagnetic immunity Recommended separation distances between portable and mobile RF Nowadays, many RF wireless equipments have being used in various healthcare communications equipment and the device locations where medical equipment and/or systems are used.