Owner's Manual
ECG Blood Pressure Monitor
DBP-6679B
Model
Arm Type
Arm Type
Safety Notice
.............................................................................................02
.......................................................................................07
...................................................................09
.............................................................................................................10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2
System Unit Operation .............................................................................................13
.............................................................................................13
BP Unit Operation ...................................................................................................15
..............................................................................15
.....................................................................................16
......................................................................................................18
BP
Memory Check and Last 3 Test Average
........................................................................................19
...................................................................................20
..........................................................................21
ECG Unit Operation ...............................................................................................22
........................................................................22
........................................................................................................22
.......................................................................................25
...................................................................................25
....................................................................................................26
Wireless Unit Operation ..........................................................................................27
Bluetooth requirement and connection .........................................................27
Wifi requirement and connection
Troubleshooting ......................................................................................................30
....................................................................................31
...............................................................................................34
...........................................................................................................35
..........................................................................................................36
................................................................................................................38
......................................................19
..............................................................28
............................................................39
Contact Information
The lay operator or lay responsible or ganization should contact
the manufacturer or the representative of manufacturer.
-for assistance,if needed,in setting up,using or maintaining
the product,or
-to report unexpected operation or events.
Manufactured by JOYTECH Healthcare Co.,Ltd.
No.365, Wuzhou Road,Yuhang Economic Development Zone,
Hangzhou City,311100 Zhejiang,China
Email: info@sejoy.com
Telephone: +86-571-81957767
Fax: +86-571-81957750
1
Safety Notice
Thank you for purchasing the DBP-6673B ECG Blood Pressure Monitor.
Intended Use
ECG Blood Pressure Monitor is intended for non-invasive measuring an adults and adolecents
over 12 years of ageindividual's systolic, diastolic blood pressure and heart rate using the
oscillometric method. By switching to ECG measurement, ECG blood pressure monitor intend to
display average heart rate and real-time heart rate, also can view real-time and historical ECG
waveforms in App.
Contraindications
Product is not intended for people under 12 years of age or individuals who cannot express
their intentions.
Precautions to Ensure Safe, Reliable Operation
1. Do not drop the unit. Protect it from sudden jars or shocks.
2. Do not insert foreign objects into any openings.
3. Do not attempt to disassemble the unit.
4. Do not crush the pressure cuff.
5. If the unit has been stored at temperatures below 0 ℃, leave it in a warm place for
about 15 minutes before using it. Otherwise, the cuff may not inflate properly.
6. If the unit has been stored at temperatures above 40 ℃, leave it in a cool place for about
15 minutes before using it. Otherwise, the cuff may not inflate properly.
7. Do not store the unit in direct sunlight, high humidity or dust.
8. To avoid any possibility of accidental strangulation, keep this unit away from children
and do not drape tubing around your neck.
9. Ensure that children do not use the instrument unsupervised; some parts are small
enough to be swallowed.
10. Some may get a skin irritation from the cuff taking frequent readings over the course
of the day, but this irritation typically goes away on its own after the monitor is removed.
11.Longest expected contact time between patient and cuff:Each blood pressure
measurement should not exceed 3 minutes.
Document No.:
JDBP-7904-154
Version:Z
Date of Issue:2020.06
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DBP-6679B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sejoy DBP-6679B

  • Page 1: Table Of Contents

    Document No.: JDBP-7904-154 Version:Z Owner's Manual Date of Issue:2020.06 ECG Blood Pressure Monitor DBP-6679B Model Arm Type Arm Type Contact Information The lay operator or lay responsible or ganization should contact the manufacturer or the representative of manufacturer. -for assistance,if needed,in setting up,using or maintaining the product,or -to report unexpected operation or events.
  • Page 2 Safety Notice Safety Notice Important Instructions Before Use ”appears on screen. Replace both 23. Replace batteries when Low Battery Indicator“ 1. Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should batteries at the same time. only be interpreted by a health professional who is familiar with your medical history. 24.
  • Page 3: Contents

    If air is leaking from the arm cuff, replace the arm cuff with a new one. It is generally recomm ended to have the cuff replaced timely to ensure correct functioning and accuracy. Please consult your local authorized Sejoy distributor or dealer. 3.Arm Cuff...
  • Page 4: B A T T E R Y I N S T A L L A T I O

    Battery Installation Quick Start Battery Installation ECG Quick Start 1. Install batteries. (See Figure A ) Slide battery cover off as indicated by arrow. Install 3 new AAA alkaline batteries according to polarity. Close battery cover. Figure A 2. Before measuring the device, rinse your hands with water or wipe them with a wet towel to keep the contact points moist.(See Figure B ) Figure B 3.
  • Page 5: Bp Unit Operation

    BP Unit Operation BP Unit Operation Blood pressure test Applying the Arm Cuff 1. Power On 1. Firmly insert air plug into opening located on behind side of monitor unit. Press and hold "START/STOP" button to turn the unit on. Press and hold “START/STOP” button to turn the unit on.
  • Page 6: Bp Memory Deletion

    BP Unit Operation BP Unit Operation Last 3 Tests Average BP Memory Deletion With power off,press the "MEM" button to activate screen display. After the unit performs a self-diagnosis ,the screen will display the average test results from the last 3 readings Memory for a selected group may be deleted while in Memory Check mode.
  • Page 7: Ecg Memory Check

    Unit Operation ECG Unit Operation ECG Unit Operation 2. Testing for ECG Wait for the signal to stabilize,and the 5S countdown starts to count.Device side Atrial Fibrillation Indicator Atrial Fibrillation Indicator display ( See Figure A ), APP side page display ( See Figure B If the monitoring instrument detects the disappearance of P wave in electrocardiogram If the ECG sphygmomanometer detects an abnormality in the RR interval during the...
  • Page 8: Wireless Unit Operation

    Wireless Unit Operation Wireless Unit Operation Bluetooth requirements Wifi requirements The monitor requires a device with: The monitor requires a device with: . Bluetooth 4.0 or later . Android 5.0 or later . Android 5.0 or later . IOS 9.0 or later .
  • Page 9: Blood Pressure Information

    Blood Pressure Information Blood Pressure Information WHO Blood Pressure Classification Indicator Blood Pressure The DBP-6673B is equipped with a classification indicator based on established guidelines from the World Health Organization. The chart below (color coded on monitor unit) Blood pressure is the force of blood pushing against the walls of arteries. It is typically indicates test results.
  • Page 10: Maintenance

    Specifications Maintenance ECG Blood Pressure Monitor Product Description Arm-type Fully Automatic 1. Avoid dropping, slamming, or throwing the unit. DBP-6679B Model Display LCD Digital Display Size:95.5mm×54.5mm (3.76" x 2.15") Oscillometric Method BP Measurement Method BP Pressurization Automatic Pressurization Systolic Pressure 60mmHg~260mmHg...
  • Page 11: Electromagnetic Compatibility Information

    Electromagnetic Compatibility Information Electromagnetic Compatibility Information Table 2 The device satisfies the EMC requirements of the international standard IEC Guidance and declaration of manufacturer-electromagnetic immunity 60601-1-2. The requirements are satisfied under the conditions described in the The device is intended for use in the electromagnetic environment specified below. table below.
  • Page 12 Additional Notes Additional Notes 9.Do not place the ECG Blood Pressure Monitor and cuff at will. It will cause asphyxiation if the Important Instructions Before Use child swallows or twine around his neck. 1. WARNING: Use of this equipment adjacent to or stacked with other equipment should be 10.The cuff and the case of the ECG Blood Pressure Monitor have been tested for biocompatibility avoided because it could result in improper operation.

Table of Contents