  • Page 1 3:20.06.2024...
  • Page 4: Product Description

    designed to be used on top of It is NOT designed to be placed directly on The intended use of the OSKA Air Overlay system foam underlay with a minimum depth of a base mattress or underlay. It consists of a single the bed base and using it without a suitable is to reduce the risk of pressure related tissue 50mm.
  • Page 5 Contraindications • Ensure the mains power cord and air • Do not place blankets, bedding or other mattresses and bed rails, the customer / • Do not use the OSKA Air Overlay system for hoses are: items over the pump unit – Risk of fire. care provider is responsible for performing patients with unstable spinal fractures.
  • Page 6: Parts Identification

    CAUTIONS • Note to the user and/or patient: any serious • When the OSKA Air Overlay has been incident that has occurred specifically in kept in conditions close to the minimum relation to this device should be reported to / maximum storage temperatures, a...
  • Page 7: Installation

    To install the system correctly, follow the steps below. 4. Using the four fixing straps (two on each side of the overlay), attach the overlay to the bed 7.1 System preparation frame. Remove the OSKA Air Overlay system from its packaging. The unboxed system consists of the following items: NOTE • Air Overlay with integral umbilical (air hose). If the bed frame is a multi-section profiling bed •...
  • Page 8 7.3 Pump installation the umbilical away from the overlay pump (2) 1. Position the air overlay pump either by using until it fully detaches from the pump. the two brackets on the back of the pump 7.4 Connecting / disconnecting the Pump and...
  • Page 9 UNLOCK STATIC 184.17 mm MUTE With the pump unlocked (see 8.3 above) push Under normal operation the OSKA Air Overlay 8.2 Comfort Control buttons LOW PRESSURE the STATIC button once to select static mode. System delivers alternating pressure area care...
  • Page 10: System Operation

    The Cable Tidy feature runs down each side of indicator light will flash. NOTE for example during a cardiac arrest, use the CPR the Air Overlay and allows the mains cable to be 4. Once the system reaches the correct pressure, tag at the head of the overlay. Optimise the rate neatly positioned off the floor and away from any the flashing green ‘POWER’ indicator light will...
  • Page 11 Failure to return the pump simultaneously displaying a flashing red ‘LOW PRESSURE’ indicator light on the control Air Overlay System to active therapy at the If the alarm cannot be resolved, contact OSKA for panel and generating a persistent, beeping audible earliest opportunity could increase the risk of further advice / support.
  • Page 12 / cross for damage not visible to the naked eye, unzip breach in the cover integrity and will require infection from the Air Overlay system during use the cover and examine the white underside of Always turn the Air Overlay System off and a replacement top cover).
  • Page 13: Warranty

    The warranty is based on statutory regulations contamination and local policy / guidelines. Alternating mode Modes and does not apply if the goods supplied by OSKA Static mode are processed, handled or modified in any way by Class II Electrical protection...
  • Page 14: Troubleshooting

    Issue Possible Cause Solution No mains power to Connect overlay pump to mains power supply When placing the OSKA Air Overlay on top Decontaminate the OSKA Air Overlay system the pump. and turn on. of a mattress, this will increase the height of prior to disposal to reduce any risk of cross Push the ‘POWER’...
  • Page 15: Electromagnetic Compatibility (Emc)

    The OSKA Air Overlay is intended for use in the electromagnetic environment specified below. The stacked with other equipment. If adjacent or capacity to generate / radiate or be affected by customer or the user of the OSKA Air Overlay should assure it is used in such an environment. stacked use is necessary, the device should be radio frequency (RF) energy, the system has been observed to verify normal operation in the...
  • Page 16 18.2 Declaration – Electromagnetic Immunity Guidance & Declaration: electromagnetic immunity The OSKA Air Overlay is intended for use in the electromagnetic environment specified below. The customer or the user of the OSKA Air Overlay should assure it is used in such an environment. Guidance & Declaration: electromagnetic immunity The OSKA Air Overlay is intended for use in the electromagnetic environment specified below. The Immunity Test...