_&
DANGER
Ultraviolet radiation emitted from this product.
AVOID EXPOSURE TO EYES AND SKIN. W EAR
PROTECTIVE CLOTHING AND EYEWEAR. Never
look directly into the lamp. Exposure can cause eye
and skin allergy and allergic reactions. Medications
or cosmetics may increase your sensitivity to
ultraviolet radiation. Consult physician before
operating this product if you are using medications
or have a history of skin problems or believe
yourself especially sensitive to sunlight.
FOR PROFESSIONAL USE ONLY
INTRODUCTION
The Spectro-UV B-260 UV lamp comes complete with a
built-in rechargeable battery and charger. When used as
directed and maintained properly, it is designed to provide
many years of trouble-free use.
UNPACKING AND INSPECTION
Carefully unpack the lamp and inspect it for possible
damage in shipment. If any damage is noted, immediately
notify the carrier and supplier before attempting to use the
lamp.
DESCRIPTION
3. Power Switch
I
4. Tube guard
5. UV-A tubes
6. Tube guard
screw
Battery-Operated UV Lamp
2. Charging
Indicator
1. Lamp base
(charger cord
inside)
8-260
OPERATION
NOTE: Charge the battery for 20 hours before using the
B-260 lamp for the first time. Charge to full, which will
take 16-24 hours before the next use and at least once
every three months to help extend the battery life.
Never completely drain battery.
• Slide open the bottom tab on the lamp base and take
out the charger cord.
• The lamp has a three-position switch on the top,
covered with a sticker marked "Remove before use".
Remove the tape before use and if desired replace
when not in use to prevent inadvertent operatio n .
• Switch modes are: O-OFF, I-one tube, II-two tubes.
• Push the power switch to the OFF position. Plug the
charger cord into an appropriate AC outlet.
The charging indicator (red LED) will light during
charging. It will remain on as long as the charger cord is
plugged into the AC outlet. The tubes will not light
while the lamp is charging.
MAINTENANCE
To Replace the UV-A Tubes:
Turn OFF the lamp. Remove both tube guard screws, lift
up the tube guard and set aside the guard and screws.
Grasp the metal ends of the tubes. Gently rotate each
tube a quarter turn until it loosens and remove the tube.
Use a clean cloth or gloves when handling the new
tubes to avoid getting fingerprint oil on the tubes.
Reverse the above procedure to insert the new tubes.
Replace the tube guard and tighten both screws.
NOTE: For best performance, both tubes should be
replaced at the same time.
Housing and Tube Maintenance:
• Always turn off the power and unplug the lamp from the
power outlet before performing any maintenance.
• Use a soft cloth to immediately clean spills from the
lamp housing and wipe dry. Do not allow any liquids
to leak into the internal electrical components. Keep
the lamp away from heat sources and out of
high-humidity areas.
• Use a soft cloth with a mild detergent solution to clean
the tubes.
EYE PROTECTION
Do not look directly into the UV light. Wear UV-absorbing
protective eyewear (such as Spectro-UV UVS-30
spectacles or UVG-50 goggles) whe n operating this or any
UV lamp. In addition to providing protection, using the
Spectro-UV eyewear will increase contrast between the
fluorescent area and the background and will eliminate
"blue haze" interference.
Need help?
Do you have a question about the B-260 and is the answer not in the manual?
Questions and answers