Page 5
Source Container FQG60, FQG61/62, FQG63, FQG66 1. Proceed as follows for FQG66, ordering feature 020 "Version", option A "Manual operation": The wipe test can be performed when the source holder is in either the "EIN/ON" or "AUS/OFF" position. A0027388 Wiping surface at border between source holder and housing block Wiping surface in the bore hole of the locking bolt 2.
Page 6
- notify CNSC that a leaking source has been detected Visual check If considerable corrosion is visible at the housing (FQG60: especially in the area of the shutter plate; FQG61/62: especially at the cover plate of the radiation channel) measure the radiation level around the device.
Source Container FQG60, FQG61/62, FQG63, FQG66 Emergency procedure Objective and overview This emergency procedure should be put into effect immediately to secure an area in the interests of protecting personnel where an exposed source is known, or suspected, to exist.
Page 8
This is to be done as soon as possible and not later than 24 hours of the incident being discovered. 5. After thorough assessment of the damage, the CNSC Inspector, in conjunction with Endress+Hauser, will agree a remedy to the specific problem. Endress + Hauser...