Source container radiometric level measurement

Advertisement

Quick Links

SD00292F/00/EN/15.15
71290820
Supplementary safety instructions for radioactive sources and source
containers approved for use in Canada
Products
Safety Instructions
Source Container
FQG60, FQG61/62, FQG63, FQG66
Radiometric level measurement
Solutions
Services

Advertisement

loading
Need help?

Need help?

Do you have a question about the FQG60 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser FQG60

  • Page 1 Products Solutions Services SD00292F/00/EN/15.15 71290820 Safety Instructions Source Container FQG60, FQG61/62, FQG63, FQG66 Radiometric level measurement Supplementary safety instructions for radioactive sources and source containers approved for use in Canada...
  • Page 2 Source Container FQG60, FQG61/62, FQG63, FQG66 Table of contents Introduction ........3 Scope of application .
  • Page 3: Leak Test Procedure

    A0018519 Wiping surface for the leak test along the edge of the nameplate FQG60; Ordering feature 240 "Emission Angle; Application", option 3 "20deg; limit switch + density" FQG60; Ordering feature 240 "Emission Angle; Application", option 5 "40deg; level" FQG60; Ordering feature 240 "Emission Angle; Application", option 4 "20deg; density 30deg diagonal radiation"...
  • Page 4 Source Container FQG60, FQG61/62, FQG63, FQG66 A0018425 FQG61/62; Ordering feature 020 "Version", option A "Cylinder lock fixation ON/OFF + covering cap" FQG61/62; Ordering feature 020 "Version", option B "Padlock fixation OFF + locking bolt ON + rotary bracket" FQG61/62; Ordering feature 020 "Version", option C "Padlock fixation ON/OFF + rotary bracket"...
  • Page 5 Source Container FQG60, FQG61/62, FQG63, FQG66 1. Proceed as follows for FQG66, ordering feature 020 "Version", option A "Manual operation": The wipe test can be performed when the source holder is in either the "EIN/ON" or "AUS/OFF" position. A0027388 Wiping surface at border between source holder and housing block Wiping surface in the bore hole of the locking bolt 2.
  • Page 6 - notify CNSC that a leaking source has been detected Visual check If considerable corrosion is visible at the housing (FQG60: especially in the area of the shutter plate; FQG61/62: especially at the cover plate of the radiation channel) measure the radiation level around the device.
  • Page 7: Emergency Procedure

    Source Container FQG60, FQG61/62, FQG63, FQG66 Emergency procedure Objective and overview This emergency procedure should be put into effect immediately to secure an area in the interests of protecting personnel where an exposed source is known, or suspected, to exist.
  • Page 8 This is to be done as soon as possible and not later than 24 hours of the incident being discovered. 5. After thorough assessment of the damage, the CNSC Inspector, in conjunction with Endress+Hauser, will agree a remedy to the specific problem. Endress + Hauser...