Table of Contents

Advertisement

Quick Links

NEED HELP?
SCAN OR CLICK HERE
TO CONTACT US
Air-Met Scientific
www.airmet.com.au
FALCO 2 11.7 VOC Detector
Instrument User Manual V1.0
Register
your instrument
online to receive your
extended warranty
Pioneering Gas Sensing
Technology.
ionscience.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FALCO 2 11.7 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ION FALCO 2 11.7

  • Page 1 NEED HELP? SCAN OR CLICK HERE TO CONTACT US Air-Met Scientific www.airmet.com.au FALCO 2 11.7 VOC Detector Instrument User Manual V1.0 Register your instrument online to receive your extended warranty Pioneering Gas Sensing Technology. ionscience.com...
  • Page 2 FALCO 2 11.7 Instrument User Manual V1.0 Register your instrument online for extended warranty Thank you for purchasing your Ion Science instrument. The standard warranty of your instrument can be extended to two years. To receive your extended warranty, you must register your instrument online within one month of purchase (terms and conditions apply).
  • Page 3: Table Of Contents

    Preparation for Installation ......................14 Dimensions for Installation.......................14 To Install the Housing Module ......................15 To install the FALCO 2 11.7 as a complete assembly..............15 After-Installation Test ........................16 Installation in Zones with Explosive Atmospheres ................16 Configurations of the 4–20 mA Current Loop ................16 Duty Cycling Measurement ......................
  • Page 4 FALCO 2 11.7 Instrument User Manual V1.0 User Interface ...........................19 Keypad ............................19 Status Light ............................19 Start-Up Routine ........................20 Logo Screen ............................20 Info screen 1 .............................20 Info screen 2 .............................20 Warm-up ............................20 Normal Operation ........................21 Display Overview ..........................21 Display Menu Navigation........................21 Lock Screen ............................22...
  • Page 5 FALCO 2 11.7 Instrument User Manual V1.0 Relay 2 Options ..........................30 4-20 mA ............................. 30 4 – 20 mA Enable/Disable.........................30 4 - 20mA Range..........................30 Modbus Address ........................30 Status Light Brightness ....................... 31 Service Mode ..........................31 Password Lock ........................... 32 Servicing and Maintenance ......................
  • Page 6: Safety

    • To the extent permitted by law, ION Science shall not be liable to any person or entity for any loss or damage which may arise from the use of this manual.
  • Page 7 The FALCO 2 11.7 must not be exposed to atmospheres known to have an adverse effect on Thermoplastic Elastomers or Polycarbonate. Outside the items covered in this manual, the FALCO 2 11.7 must be serviced in a Non-Hazardous environment and by ION Science Ltd authorised service centres only.
  • Page 8: Disposal

    Statements Responsibility for Correct Use Ion Science Ltd accepts no responsibility for incorrect adjustments that cause harm or damage to persons or property. The users are responsible to respond appropriately to the readings and alarms given by FALCO 2 11.7.
  • Page 9: Introduction

    VOCs can be dangerous as they are poisonous to humans and there is a risk of explosion. VOCs are detectable using photo ionisation detection (PID) detector. The FALCO 2 11.7 multi coloured LED status display screen can be seen up to 20 metres away in direct sunlight ensuring that personnel are alerted to hazards present.
  • Page 10: Specification

    FALCO 2 11.7 Instrument User Manual V1.0 Specification Detail Specification Variant FALCO 2 11.7 eV (Diffused) Photo-Ionisation Detector Detector principle Lamp 11.7 eV Diffused Sampling Detection Range 0 to 200 ppm Response time T 60 seconds (one full cycle) ± 12 % ±1 digit...
  • Page 11: Unpacking And Inspection

    Ion Science Ltd. Ion Science will not be responsible for discrepancies not reported within ten days of your receipt of the shipment. Every FALCO 2 11.7 (new units and those returned from a Service Centre) must have a certificate of calibration before you install it.
  • Page 12: System Description

    Both relays can be programmed to switch 60 VDC / 2 A or 50 VAC / 2 A maximum load. RS 485 Modbus Interface The FALCO 2 11.7 Modbus interface uses Modbus RTU • 9600 baud, 8 data bits, no parity, 1 stop bit.
  • Page 13: Installation Requirements

    EMI. The cable glands supplied with the FALCO 2 11.7 are EBU2MBNC M25 x 1.5 Ex D (Cable diameter 13 to 18 mm) The manufacture and build of the cable glands are the responsibility of the installer. The cable glands must conform to the certification standards required for the installation site.
  • Page 14: Installation

    Dimensions for Installation • RS 485 Interface Requirements Dimensions for Installation Figure 1 - Dimensional drawing of the FALCO 2 11.7, showing front and side views with key measurements and mounting details Pioneering Gas Sensing Technology. Page 14 of 39...
  • Page 15: To Install The Housing Module

    To install the FALCO 2 11.7 as a complete assembly To ensure secure installation, use two M8 screws to mount the FALCO 2 11.7 (including both the main unit and sensor housing) onto a solid and stable support. Refer to Figure 1 for the device's dimensions and mounting details.
  • Page 16: After-Installation Test

    Installation in Zones with Explosive Atmospheres The wiring diagram for the FALCO 2 11.7 is shown below, that covers in input power, MODBUS and Current Loop. There are four possible configurations for the 4–20 mA current loop depending on the installation site that are following section.
  • Page 17 FALCO 2 11.7 Instrument User Manual V1.0 FALCO 2 11.7 Current Loop Configuration 1 – active current loop, using internal power supply connected to the current source. FALCO 2 11.7 Current Loop configuration 2 – passive current loop, externally powered current source.
  • Page 18: Duty Cycling Measurement

    FALCO 2 11.7 Instrument User Manual V1.0 Duty Cycling Measurement The FALCO 2 11.7 has been designed with prolonging the lifetime of the 11.7 eV lamp in mind. The unit will sample for 20 seconds every minute as indicated by the lamp icon in the top left corner.
  • Page 19: Operating The Falco 2 11.7

    Status Light Yellow Displayed during start-up only, when power is first applied. Green Indicates that FALCO 2 11.7 is operating correctly. Also displayed during the start-up routine. Amber Flashing amber indicates Alarm 1 has been triggered i.e. the measured level of VOC is above the alarm threshold.
  • Page 20: Start-Up Routine

    The FALCO 2 11.7 will then display the following screens, in this order: Logo Screen After the power is turned on, the FALCO 2 11.7 displays the 'Ion Science' logo for 3 seconds and the status light is green. Info screen 1 Info screen 1 then appears for 3 seconds and the status light goes amber.
  • Page 21: Normal Operation

    The screen then appears continuously and shows the PID reading and units. The Status Light colour depends on the status. A progress bar showing how far through the current cycle the FALCO 2 11.7 is through the cycle the lamp on is indicated by the lamp icon in the top left corner.
  • Page 22: Lock Screen

    FALCO 2 11.7 Instrument User Manual V1.0 Lock Screen The Lock screen appears if a password has been set (see menu i6). It prevents unauthorized adjustments, and menus cannot be accessed until the correct password is entered. Press the Enter key.
  • Page 23: Menu I1

    MiniPID2 is on. From here you can also put the FALCO 2 11.7 into test mode. This means the FALCO 2 11.7 will simulate its output behaviours. To configure the FALCO 2 11.7 to display a fixed output level, select .
  • Page 24: Menu I4

    FALCO 2 11.7 Instrument User Manual V1.0 Menu i4 Detection units: Used to change the detection units from the default of ppm to mg/m . The current units are displayed. PID status light mode: Used to switch the status light between steady illumination and slow pulsing when alarming.
  • Page 25: Menu I7

    • Zero air (UHP Air) • Span gas (100ppm Isobutylene) • Suitable tubing to connect gases to FALCO 2 11.7 calibration adapter (part no. A-873201) • Fixed Flow Regulator (part no. 5/RP-04) Pioneering Gas Sensing Technology. Page 25 of 39...
  • Page 26: Zero

    FALCO 2 11.7 Instrument User Manual V1.0 Zero For best practice ION Science recommend using a cylinder of Zero air with a fixed flow regulator (part no. 5/RP-04) attached with suitable tubing to the calibration adapter (part no. A-873201). 1) Enter zero Cal mode by moving the cursor to the zero Cal icon and then press the Enter key.
  • Page 27: Rf (Response Factor)

    VOC’s response factor. If a response factor is applied the concentration displayed on the FALCO 2 11.7 will represent the concentration of the VOC. For example, if the RF is 00.50, and 100ppm is detected based on an isobutylene calibration: 100 ppm x 00.50 = 50 ppm is the value displayed...
  • Page 28: Alarms

    The alarm levels are set in menu i2 The FALCO 2 11.7 has 2 alarms levels, 1 and 2. When alarm level 1 is reached the status bar will go amber and Alarm 2 is reached the status bar will go red.
  • Page 29: Alarm Pulsing

    FALCO 2 11.7 Instrument User Manual V1.0 While in the screen, the Status Light will change to red and will change brightness in response to change to brightness percentage. If a brightness of higher than 100% is entered, when the Enter...
  • Page 30: Relay 2 Options

    FALCO 2 11.7 Instrument User Manual V1.0 Relay 2 Options Used to specify which of 4 conditions will trigger the Relay 2 output. See the description of the Relay 1 options setting for details. 4-20 mA 4-20 mA options are accessed from menu i5 4 –...
  • Page 31: Status Light Brightness

    FALCO 2 11.7 Instrument User Manual V1.0 Status Light Brightness The Status Light brightness is accessed in menu i4 Used to set the brightness of the Status Light during alarm conditions, from 0 to 100% for green, and 50-100% for amber and red. “AUTO” brightness mode is selected by setting the brightness to 0% Change the percentage brightness as required.
  • Page 32: Password Lock

    Instrument leads, adaptors fittings should be checked before use for damage and continuity. To maintain the reliability of your instrument, regular bump testing is recommended ION Science recommends that the FALCO 2 11.7eV device is calibrated on a weekly basis. This includes checking the functionality of individual components.
  • Page 33: Minipid2 Electrode Stack Replacement

    1. There are no user replaceable parts within this instrument, other than the MiniPID2 Electrical stack and lamp. 2. Before carrying out any servicing/maintenance on the FALCO 2 11.7, set the device to Servicing Mode. MiniPID2 stack assembly and removal tools:...
  • Page 34: Minipid2 Reassembly

    CAUTION Irreparable damage will be caused by forcing the MiniPID2 (PN MP6SX7FWXU2) into the Sensor Housing if not correctly aligned. INFORMATION Always calibrate the FALCO 2 11.7 after completing any servicing. Replacing the Lamp CAUTION Never refit a damaged Lamp.
  • Page 35: Cleaning The 11.7 Ev Lamp

    The accuracy of the measurements can be easily checked at any time. ION Science recommends performing a weekly Bump Test on the FALCO 2 11.7 to ensure the instrument responds correctly and that the alarm indicators activate properly when gas is detected at the specified alarm levels. Bump tests should also be conducted whenever these alarm levels are changed.
  • Page 36: Fault Diagnostics

    FALCO 2 11.7 Instrument User Manual V1.0 Using the Span gas to verify the unit is within satisfactory reading (see speciation table for accuracy): Attach the span gas cylinder to the FALCO 2 11.7. Verify the FALCO 2 11.7 display updates If the displayed result is out of specification/satisfactory reading carry out calibration procedure as detailed in Calibration section of this manual.
  • Page 37: Fault Conditions

    FALCO 2 11.7 Instrument User Manual V1.0 Fault Conditions The FALCO 2 11.7 is equipped with diagnostics to ensure instrument faults are detected and communicated. The table gives a fuller description of each fault and lists some possible causes and corrective actions you can try.
  • Page 38: Manual Log

    Quality Assurance The FALCO 2 11.7 instruments are manufactured by ION Science Limited within an ISO 9001 compliant quality management system, which ensures that the equipment supplied to our customers has been designed and assembled reproducibly, and from traceable components.
  • Page 39: Ion Science Contact Details

    FALCO 2 11.7 Instrument User Manual V1.0 ION Science Contact Details ION Science Ltd – UK/Head Office Tel: +44 (0)1763 208 503 Web: www.ionscience.com | Email: info@ionscience.com ION Science Inc – USA Office Tel: +1 877 864 7710 Web: https://ionscience.com/usa/ |Email: info@ionscienceusa.com...

Table of Contents