P. R. China Zástupce pro EU: Shanghai International Holding Corp. GmbH (EUROPE) Eiffestrasse 80, 20537 Hamburg, GERMANY Dovozce pro ČR, Celimed s.r.o. autorizovaný servis: Sociální péče 3487/5a, 400 11 Ústí nad Labem, Česká republika Pozáruční servis – tel.: 475 208 180 e-mail: info@celimed.cz www.celimed.cz...
Page 40
Development Area, 226 009 Nantong City, Jiangsu, P. R. China Zástupca pre EU: Shanghai International Holding Corp. GmbH (EUROPE) Eiffestrasse 80, 20537 Hamburg, GERMANY Distribútor pre SR, Celimed s.r.o. autorizovaný servis Pri majeri 22 831 06 Bratislava, Slovensko Servis: 02 4468 1249 e-mail: info@celimed.sk www.celimed.sk...
Page 60
Development Area, 226 009 Nantong City, Jiangsu, P. R. China EU-képviselet: Shanghai International Holding Corp. GmbH (EUROPE) Eiffestrasse 80, 20537 Hamburg, GERMANY Forgalmazó / HU Celimed s.r.o. importer: Sociální péče 3487/5a, 400 11 Ústí nad Labem, Česká republika Pozáruční servis – tel.: 475 208 180 e-mail: info@celimed.cz www.celimed.cz...
INTENDED USE Thank you for purchasing the CELIMED C-715 (NB-216C) compressor nebulizer. Intended purpose: This compressor nebulizer is a medical device that used to aerosolize medications for inhalation by the patient to achieve the expected therapeutic effect. The device can be used with adult or pediatric patients.
SAFETY INFORMATION Read all instructions carefully before use! WARNING Use the equipment only for its intended use as described in this manual. Use the equipment with medications only under the instruction of your physician The device, including accessories, should not be shared by two different persons. Do not use the equipment if it has any damaged parts, or it has fallen into water.
Page 65
Do not insert any object into the compressor. the air tube is correctly connected to the compressor and the nebulizer kit. Air may leak from the air tube during use if not securely connected. Inspect the compressor (main unit) and the nebulizer parts each time before using the device. Make sure no parts are damaged, the nozzle and air tube are not blocked and the compres- sor operates normally.
Page 66
(a) Every time before use, please check the adaptor, once damaged, never to use. (b) Please clean the plug of adapter plug at least once a year. Too much dust on plug may Do not plug or unplug the power cord into the electrical outlet with wet hands. Do not immerse the compressor (main unit) in water or other liquid.
SAFETY INFORMATION The followings are maintenance and repair which can be taken by operator, or which must be operated by manufacturer or distributor. service and maintemance Responsible change the inhalation tube operater change the applied part operater operater clean the surface of the device operater daily cleaning and disinfecting operater...
PREPARING THE NEBULIZER FOR USE CAUTION Clean and disinfect set of spray chamber, mouthpiece and inhalation mask before using If the device has not been used for a long period of time please cleans and disinfects the set of spray chamber, mouthpiece and inhalation mask before using them. To prevent the risk of cross infection, different persons should never use the same nebulizer kit.
Page 70
9. Pressure adjust function: In standby mode of the device, the initial pressure of the device is the small gear, and pressing the pressure adjust button can adjust the pressure. There are three gears in total (gear changes are shown in Figure 4). The pressure level can also be adjusted during the device is running (as shown in Figure 5).
CLEANING WARNING Ensure that the remove the plug before the clean and disinfection. Do not place or attempt to dry the device, components or any of the nebulizer parts in a microwave oven. DURABLE PERIOD As with any plastic parts, nebulizer kit and their accessories are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent basis.
3. Whether disinfect immediately or not, dry all parts with a new medical gauze. You can remove excess water more quickly by shaking them. CAUTION Inadequate disinfection encourages the growth of bacteria and thus increases the risk of infection. Adequate cleaning with disinfection can only be assured if application time are adhered to, and if all individual parts are completely immersed in the solution for the entire application time.
CHANGE THE AIR FILTER immediately. CAUTION: CAUTION: CAUTION: 5. Do not maintain or service the device while it is in using. STORING THE DEVICE 1. Put the nebulizer kit and the inhalation accessory (mouthpiece, or mask) in a dry, sealed bag.
WARNING: Do not leave the device or its parts where it will be exposed to extreme temperatures or changes in humidity, such as leaving the device in a vehicle during warm or hot months, or where it will be exposed to direct sunlight. Keep the device out of the reach of unsupervised infants and children.
SPECIFICATIONS Product Category Aerosol Therapy Equipment Product Description Compressor nebulizer Model (code) CELIMED C-715 (NB-216C) Dimensions 166 × 136 × 104 mm USB adapter/ IN 100–240 V AC, 50/60 Hz, 700 mA Max USB adapter/ OUT 12V DC, 1,35 A...
essential performance. The Emissions characteristics of this equipment make it suitable for use in home healthcare environment (CISPR 11 Class B). COMPLIANCE INFORMATION FOR EACH EMC TEST Electromagnetic Emission (Home Healthcare Environment) Emission test (IEC 60601-1-2:2014) Compliance Conducted and radiated RF emissions CISPR 11 Group 1 Class B Harmonic emissions IEC 61000-3-2 Class A...
Page 80
Development Area, 226 009 Nantong City, Jiangsu, P. R. China EU-representative: Shanghai International Holding Corp. GmbH (EUROPE) Eiffestrasse 80, 20537 Hamburg, GERMANY Importer for the Celimed s.r.o. Czech Republic Sociální péče 3487/5a, and authorized 400 11 Ústí nad Labem, Česká republika service: Pozáruční servis – tel.: 475 208 180 e-mail: info@celimed.cz...