  • Page 1 P60074000 Issue C Support Instructions CDX/IDX Pump motors...
  • Page 2: General Description

    Motors may be suitable for inverter operation between 25 and 60 Hz, although derating of the flameproof  classification may apply. Refer to manufacturers’ instructions or Edwards for more details. For details of the electrical specification of the motor refer to the nameplate.
  • Page 3: Technical Data

    Starting current / time Figure 1 Recommended electrical supply cable size Direct on-line (DOL) starting 16 mm (6 AWG) Inverter starting 10mm (8 AWG) © Edwards Limited 2020. All rights reserved. Page 2 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 4: Installation

    CDX pump is part of an ATEX system, refer to the ATEX System Instruction Manual before you make any of the electrical connections. There may be additional electrical requirements mandated by the ATEX Directive. © Edwards Limited 2020. All rights reserved. Page 3 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 5 The pump motor can be started either Direct On- Line (DOL, see Figure 1) or through a suitable inverter, as described below. If required, contact your supplier or Edwards for advice on starting. 5.1.2 Direct on-line (DOL) starting with variable overload Your system must incorporate a programmable variable overload relay to allow for the high current drawn (over a prolonged period) during DOL starting, and to provide overload protection under normal pump operating conditions.
  • Page 6 3. Connect the wires in the control cable to the thermistor output terminals in the terminal box: refer to the motor manual supplied with your pump. 4. Tighten the cable-gland nut strain-relief screws. 5. Refit and secure the terminal-box cover. © Edwards Limited 2020. All rights reserved. Page 5 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 7: Maintenance

    If you are not suitably trained, you may damage the motor and it may not operate correctly or safely. © Edwards Limited 2020. All rights reserved. Page 6 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 8 You are fully accredited/authorised to do so by the motor manufacturer.  You have been suitably trained in all of the necessary procedures.  © Edwards Limited 2020. All rights reserved. Page 7 Edwards and the Edwards logo are trademarks of Edwards Limited.