Prior to contacting X-Rite support department for device problems, try the applicable solution(s)
described below.
If the condition persists, contact X-Rite following the official service RMA process.
https://www.xrite.com/service-support/services-offered/repair-certification/rma-request
Please make sure that you send us your White Calibration Tile and Camera Test Target with the
same Serial Number.
Problem
Cannot measure.
Device screen remains dark. Device is turned off.
Screen is instable (turns
on/off periodically) when
PPG Service Troubleshooting Guide for Digi Match
Cause
White Tile is dirty.
Device is in power
down mode.
Battery is very low.
Battery is defective.
Battery is defective.
Solution
Clean White Tile with a microfiber towel.
Make sure to have no fingerprints on the White Tile.
Turn device on by pressing the power on/off button.
The battery pack for your new device or in case of longer
storage time results in a low to medium charge state.
In this case, charge the battery before use (up to 4 hours for
full charge).
This icon at the top of the screen would indicate the battery
is fully charged.
Touch the screen or lift the device.
Please check the battery settings in your device and use the
default settings: Stand by time: 10 min. Power off: 30 min.
Charge the battery min. 1 hour. After max. 5 minutes the
screen will turn on.
This icon at the top of the screen would indicate the battery
is fully charged.
Charge the battery. If the screen does not turn on after max.
5 minutes check if the USB cable is correctly connected to
the computer or power supply. If it still does not work the
battery needs to be replaced.
Battery needs to be replaced.

Advertisement

loading
Need help?

Need help?

Do you have a question about the Digi Match and is the answer not in the manual?

Questions and answers

Summary of Contents for PPG Digi Match

  • Page 1 PPG Service Troubleshooting Guide for Digi Match Prior to contacting X-Rite support department for device problems, try the applicable solution(s) described below. If the condition persists, contact X-Rite following the official service RMA process. https://www.xrite.com/service-support/services-offered/repair-certification/rma-request Please make sure that you send us your White Calibration Tile and Camera Test Target with the same Serial Number.
  • Page 2 device is connected to a Exposure to temperature beyond 50°C can cause computer. irreversible damage to the battery. Safe-boot the device while “long holding” the power on/off Device unable to boot or Firmware is corrupt or firmware crashes. incompletely loaded. button until the LED indicator turns red to restore the previous firmware version.
  • Page 3 Error Message: “Wavelength Device most likely Perform calibration on calibration reference. reference” dropped. If Error message persists, contact technical support. Drop Damage. Device will not measure or Contact sensor(s) do Please clean all three pins. calibrate. not properly contact the Lift the device and place it properly on the sample.