A copy of the complete Declaration of Conformity or further declarations of conformity can be › Ensure that only trained staff work with the device� requested at sales@ika.com� › Follow the safety instructions, guidelines, occupational health and safety and accident prevention regulations�...
› Check the device and accessories for damage before each use� Do not use damaged com- › Prevent release of the materials listed above� Take measures to protect staff and the ponents� › Safe operation is guaranteed only with the use of original IKA accessories� environment� › Always disconnect the plug before attaching accessories�...
The laboratory diaphragm vacuum pump IKA VACSTAR lite is used together with other labora- tory devices to create a vacuum� For specific vacuum control, an additional vacuum controller IKA VC 10 lite / pro or IKA rotary evaporator RV 10 auto is required�...
Useful information When combined with the vacuum controller VC 10 lite / pro or the rotary evaporator with in- tegrated vacuum controller RV 10 auto, the pump can also be used for “two-position control” (valve-controlled) or “analogue speed control” of the vacuum� Danger! Two-position control You can prevent the release of solvent vapours into the atmosphere using a downstream...
USB interface Connect the vacuum pump to a PC with a USB A - USB C cable� Any device software updates can be loaded using the IKA Firmware-Update-Tool� INLET: Hose connection for suction line d = 8 mm Connect the suction line to this connection�...
Operation VACSTAR lite with vacuum controller VC 10 lite / pro or rotary evaporator RV 10 auto Notice! Observe the operating instructions for the devices� /// Switching on Automatic setting of the operating mode “Speed regulation” as soon as the analogue connect- Switch the device on using the power switch�...
The data transmission may be interrupted by, for example, a crash in the operating system, a power http://www.ika.com/ika/lws/download/usb-driver.zip Install the driver by running the setup file� Connect the IKA device through the USB data cable failure in the PC or an issue with the connection table between the computer and the device�...
- Restart the device� › For cleaning disconnect the mains plug! › Use only cleaning agents which have been approved by IKA to clean the devices: Error code | Causes | Effect | Solutions Water containing surfactant / isopropyl alcohol�...
Technical data Warranty › In accordance with IKA Terms and Conditions of Sale, the warranty period is 24 months� For Pump rate max� (50/60 Hz) m³/h 1�0 claims under the warranty please contact your local dealer� You may also send the device Pump rate max�...