Summary of Contents for LabCo STIRRER MAGNETIC 1.5L
Page 1
STIRRER MAGNETIC 1.5L USER MANUAL sales@labcoscientific.com.au 1800 052 226 ABN 57 622 896 593 labcoscientific.com.au PO Box 5816, Brendale, QLD 4500 Version 1.1...
INTRODUCTION Welcome to the Magnetic Stirrer User Manual. Users should read this Manual carefully, follow the instructions and procedures, and be aware of all the cautions when using this instrument. PRODUCT CODE DESCRIPTIONS 400.100.002 Stirrer Magnetic Hotplate up to 1.5L WARRANTY This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 12 months from the date of invoice.
1. SAFETY INSTRUCTIONS Warning! • Read the operating instructions carefully before use. • Ensure that only trained staff works with the instrument. Protective ground contact! • Make sure that socket is grounded (protective ground contact) before use. • When working wear personal safety guards avoid the risk from: Splashing and evaporation of liquids Release of toxic or combustible gases •...
2. PROPER USE The instrument is designed for mixing liquids in schools, laboratories or factories. Observe the minimum distances between the devices and the wall and above the assembly (min. 100 mm) This device is not suitable for using in residential areas or other constraints. 3.
5. INSTALLATION INSTRUCTIONS • Unpack and place the magnetic stirrer on a stable and temperature resistant surface. • Ensure good working environment free of explosive, hazardous and inflammable substances. • Place the unit at >10cm distance from the wall and other units if using multiple units together. •...
7. TRIAL RUN • Make sure the required operating voltage and power supply voltage match. • Ensure the socket is properly grounded. • Plug in the power cable and ensure the power is on and begin initialising. • Add the medium into the vessel with an appropriate stirring bar. •...
8. FAULTS • Instruments can’t be power ON Check whether the power line is unplugged. Check whether the fuse is broken or loose. • Fault in power on self-test Switch OFF the unit, then switch ON and reset the instruments to factory default setting. •...
10. MAINTENANCE AND CLEANING • Proper maintenance can keep instruments working properly and lengthen its lifetime. • Do not spray cleanser into the instrument when cleaning. • Unplug the power line when cleaning. • Only use recommended cleansers: Dyes Isopropyl alcohol Construction materials Water containing tenside / Isopropyl alcohol Cosmetics...
11. ASSOCIATED STANDARDS AND REGULATIONS Construction in accordance with the following safety standards: EN 61010-1 UL 3101-1 CAN/CSA C22.2(1010-1) EN 61010-2-10 Construction in accordance with the following EMC standards: EN 61326-1 Associated EU guidelines: EMC-guidelines: 89/336/EWG Instrument guidelines: 73/023/EWG Changes or modifications not approved by the party responsible for compliance could void the user's authority to operate the equipment.
12. SPECIFICATIONS PARAMETER SPECIFICATION Plate Dimension 120x115mm Plate Material Motor Type DC Motor Brushless Motor Rating Output Power 7.2W Voltage 240V/60Hz Stirring Positions Maximum Stirring Volume (H2O) 1.5L Maximum Magnetic Stirrer Bar 40mm Speed Range 300 to 2,000rpm Speed Setting Scale Protection Class IP21...
Need help?
Do you have a question about the STIRRER MAGNETIC 1.5L and is the answer not in the manual?
Questions and answers