Table of Contents

Advertisement

Quick Links

DEWE-MDAQ SERIES
TECHNICAL
REFERENCE
MANUAL
WELCOME TO
THE WORLD OF
DEWETRON!
Congratulations on your new device! It
will supply you with accurate, complete
and reproducible measurement results
for your decision making.
Look forward to the easy handling and
the flexible and modular use of your
DEWETRON product and draw upon
more than 25 years of DEWETRON
expertise in measurement engineering.
CUSTOMIZED
ISO9001
EN ISO 14001
MODULAR
COMPETENT
COMMITTED
APPROVED

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DEWE-MDAQ Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dewetron DEWE-MDAQ Series

  • Page 1 Look forward to the easy handling and the flexible and modular use of your DEWETRON product and draw upon more than 25 years of DEWETRON expertise in measurement engineering. CUSTOMIZED MODULAR COMPETENT...
  • Page 2 Copyright © DEWETRON GmbH This document contains information which is protected by copyright. All rights are reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. All trademarks and registered trademarks are acknowledged to be the property of their owners.
  • Page 3: Table Of Contents

    Pin assignment of the input connectors ………………………………………………… 25 Amplifier specifications …………………………………………………………………… 27 MDAQ-SUB-ACC-A-BNC Input connection …………………………………………………………………………… 28 Typical sensor connection ………………………………………………………………… 29 Pin assignment of the input connectors ………………………………………………… 29 DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 4 Table of content MDAQ-SUB-ACC-x-MD Specifications ……………………………………………………………………………… 31 Input connection …………………………………………………………………………… 32 Pin assignment …………………………………………………………………………… 32 MDAQ-SUB-BRIDGE-D Specifications ……………………………………………………………………………… 33 Internal signal processing ………………………………………………………………… 34 Pin assignment of the input connectors ………………………………………………… 34 Full bridge signal connection ……………………………………………………………… 35 Half bridge signal connection …………………………………………………………… 35 Typical Sensor connection …………………………………………………………………...
  • Page 5: General Information, Safety Instructions

    Warranty Information A copy of the specific warranty terms applicable to your DEWETRON product and replacement parts can be obtained from your local sales and service office. Support For any support please contact your local distributor first or DEWETRON directly.
  • Page 6: Safety Symbols In The Manual

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the product. DEWETRON Elektronische Messgeraete Ges.m.b.H. assumes no liability for the customer’s failure to comply with these requirements.
  • Page 7: Safety Instructions For All Dewetron Systems

    DO NOT substitute parts or modify equipment: Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to a DEWETRON sales and service office for service and repair to ensure that safety features are maintained.
  • Page 8: General System Information

    General System Information Notes...
  • Page 9: General Module Information

    Calibration certificate according to ÖKD (equivalent to DKD) This manual contains no calibration information. For self calibration, there is a separate Calibration Manual available. DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 10 Notes...
  • Page 11: Mdaq Series Overview

    MDAQ-SUB-V-200-BNC MDAQ-SUB-V-200-BNC MDAQ-SUB-ACC-BNC MDAQ-SUB-ACC-BNC MDAQ-SUB-ACC-A-BNC MDAQ-SUB-ACC-A-BNC MDAQ-SUB-ACC-A-MD MDAQ-SUB-ACC-A-MD MDAQ-SUB-BRIDGE-D MDAQ-SUB-BRIDGE-D MDAQ-SUB-STG-D MDAQ-SUB-STG-D MDAQ-BASE-5 BASE (MDAQ-BASE-10) Filter (optional) MDAQ-FILT-5-xx MDAQ-AAF4-5-BU 1) BU..Butterw orth, BE..Bessel characteristics DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 12: Mdaq-Sub-Module Function Overview

    MDAQ series overview MDAQ Series overview MDAQ-SUB-Module function overview...
  • Page 13: Filter Code Table Of Optional Mdaq-Filters

    For internal usage only Filter connector +5 V DGND DGND n.c. +15 V Power connector RS-485 -15 V n.c. DGND RS-232 FLT (GND) 3 1 4 5 DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 14: Address Settings

    MDAQ series overview MDAQ Series overview Address settings If you use DEWE-ORION-16xx cards, each MDAQ Module is controlled directly via the RS-485 interface of the associated card. When using DEWE-ORION-32xx cards each card controls and converts two MDAQ modules. Set all address selectors to 0. An additional EPAD module has to be connected to the first MDAQ Module.
  • Page 15 A/D converter card module address PCB for CH 240 to CH 255 cross wiring Optional EPAD Modules For special wiring see DEWE-xxxx systems manual appendix B3. DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 16: Mdaq Channel Addressing

    MDAQ series overview MDAQ Series overview MDAQ channel addressing Each MDAQ-V-xx module has 16 analog input channels, which can be independent configured. Since only the base address of the MDAQ-V-xx is changeable the access to the input channel configuration is in the range between (Module address x 16) to (Module address x 16) +15.
  • Page 17: Mdaq-Direct

    CH. 1- CH. 1+ 2 4 6 CH. 0- CH. 0+ RS-485 A RS-485 B +5V DC TRIG DGND MDAQ-DIRECT bottom view Output connector (J1) DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 18 MDAQ-DIRECT Notes...
  • Page 19: Mdaq-Sub-V-200-X

    MDAQ-SUB-V-200-D only! For safety reasons max. 50 V may be applied to modules with BNC input-connectors! Refer to the regulation of maximum allowable touch potential. DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 20: Internal Signal Processing

    MDAQ-SUB-V-200-x Internal signal processing The picture below shows the block diagram of the MDAQ-SUB-V-200-xx MDAQ-SUB-V-200-xx MDAQ-BASE-x opt. MDAQ-FILT-x TEDS TEDS CONTROL divider Differential Amplifier with Offset and 950 k IN + CMR control divider + 15 V 50 k 50 mA Sub+ Sig Out Sub-...
  • Page 21: Typical Sensor Connection

    Output output 50 OHM Shunt Resistor +15V -15V Single ended connection Potentiometric Connectors Floating Connectors Sensor Prog Output 100 % 230 V 50 Hz DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 22: Common Mode Voltage Range

    MDAQ-SUB-V-200-x Common mode voltage range In contrast to isolated amplifiers the input common voltage range is limited at differential amplifiers. The measurement configuration below shows the possibilities to measure the current of a 24 V supplied system. 0.1 Ω 24V Battery Load 0.1 Ω...
  • Page 23: Amplifier Specifications

    Typ. 10 W (max 12 W @ 8 mA sensor excitation) Sensor connection: BNC female Output connector 68-pin Amplimite series (AMP Nr. 174339-6) TEDS DS 2406, DS 2430A, DS 2432, DS2433 DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 24: Input Connection

    MDAQ-SUB-ACC Input connection The picture below shows the principal configuration of the amplifier circuit. If no filter board is installed, this field shows the analog bandwidth MDAQ-SUB-ACC MDAQ-BASE-x MDAQ-FILT-x 24 V PGA: 0.5, 1, 2, 4, 5, 10, 20 and 40; Programmable Programmable offset and common AC/DC couling...
  • Page 25: Typical Sensor Connection

    Input type: Single ended Input type: differential Accelerometer Connectors IEPE® Sensor Measurement: IEPE ® Pin assignment of the input connectors MDAQ-SUB-ACC-BNC Hot: IN + Shield: IN - DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 26 MDAQ-SUB-ACC Notes...
  • Page 27: Amplifier Specifications

    68-pin Amplimite series (AMP Nr. 174339-6) Dimensions (W x D x H) 175 x 61 x 30 mm (6.9 x 2.4 x 1.2 in.) DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 28: Input Connection

    MDAQ-SUB-ACC-A-BNC Input connection The picture below shows the principal configuration of the amplifier circuit. If no filter board is installed, this field shows the analog bandwidth MDAQ-SUB-ACC-A MDAQ-BASE-x MDAQ-FILT-x 24 V 8 mA AC 3.4 Hz 4 mA PGA: 0.5, 1, 2, 4, 5, 10, 20 and 40; Programmable offset 47 nF AC 0.15 Hz...
  • Page 29: Typical Sensor Connection

    Connectors Connectors IEPE® Voltage output Sensor Measurement: IEPE Measurement: Voltage ® Pin assignment of the input connectors MDAQ-SUB-ACC-A-BNC Hot: IN + Shield: IN - DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 30 MDAQ-SUB-ACC-A-BNC Notes...
  • Page 31: Mdaq-Sub-Acc-X-Md

    LED to shine ® green. The sensor check unit isn’t able to check if the sensor measures correct values! DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 32: Input Connection

    MDAQ-SUB-ACC-A-MD Input connection The picture below shows the principal configuration of the input multiplexer and the sensor check function. To MDAQ-ACC-x amplilfier Sensor Input Input + Ref.Sig + Sensor check Input - Ref.Sig - Ref.Volt. Ctrl. Pin assignment The table below shows the pin assignment of the control connector. 1 Ref.
  • Page 33: Mdaq-Sub-Bridge-D

    120 Ohm @ 15 W Standard operating temperature 0 °C to 70 °C ( 32 °F to 158 °F) MDAQ-BASE-10A has the half offset adjustment range DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 34: Internal Signal Processing

    MDAQ-SUB-BRIDGE-D Internal signal processing MDAQ-SUB-BRIDGE MDAQ-BASE-xA MDAQ-FILT-x 50 mA +15V out (optional) Sense+ Differential Amplifier with Offset and 0 .. 7 CMR control +Uref Exc+ (0..6V) AC/DC G 1, G 50 Sub+ 1µF Sig Out Sub- G = (-1) 8 gains Exc- MDAQ-BASE-5: 0.5, 1, 2, 4, 8, 10, 20, 40...
  • Page 35: Full Bridge Signal Connection

    Measurement: Voltage The table on the next page shows how the mV/Vranges are calculated. The ranges depend on the gain and the excitation voltage. DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 36: Typical Sensor Connection

    MDAQ-SUB-BRIDGE-D Excitation 0.25 V 0.5 V 1.00 V 2.5 V 5.00 V 10.00 V Bridge module range [mv/V] Input range ±10000 mV ±5000 mV ±2500 mV ±1250 mV ±1000 mV ±500 mV ±250 mV 1000 ±100 mV ±50 mV ±25 mV ±10 mV ±'2.5 mV 0,25...
  • Page 37: Mdaq-Sub-Stg-D

    350 Ohm 16 Channels typ. 15 W Standard operating temperature 0 °C to 70 °C ( 32 °F to 158 °F) MDAQ-BASE-10A has the half offset adjustment range DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 38: Internal Signal Processing

    MDAQ-SUB-STG-D Internal signal processing PROG MDAQ-SUB-STG-D Sense+ MDAQ-BASE-x MDAQ-FILT-x (optional) Exc+ Differential Amplifier with Offset and CMR control G 1, G 50 Sub+ Sig Out Sub- G = (-1) Exc- Sense- TEDS TEDS Controller...
  • Page 39: Full Bridge Signal Connection

    Sense leads (SUB-D: pin 3 and 6) have to be connected! 'R+' has to be connected only if shunt calibration is required, otherwise it can be left unconnected. DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 40: Typical Sensor Connection

    MDAQ-SUB-STG-D Typical Sensor connection Isolated sensor, battery powered sensor or sensor with differential output MDAQ-SUB-STG Isolated Sensor Battery powered sensor IN (-) Sensor with diff. output Pin assignment of the input connectors MDAQ-SUB-STG-D 1 EXC+ 6 Sense+ 2 IN + 7 IN - 5 4 3 2 1 3 Sense...
  • Page 41: Mdaq-Filt-5-Bx

    ±7.5 V to ±15 V direct via MDAQ-BASE or -V Power consumption Typ. 3 W Dimensions (W x D x H) 175 x 61 x 14 mm (6.9 x 2.4 x 0.9 in.) DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 42: Step Response Of 100 Hz Filter

    MDAQ-FILT-5-Bx Step response of 100 Hz filter Butterworth 2-pole Bessel 2-Pole Output connector CH. 15+ CH. 14+ CH. 13+ CH. 12+ CH. 11+ CH. 10+ CH. 9+ CH. 8+ CH. 7+ CH. 6+ CH. 5+ CH. 4+ CH. 3+ CH. 2+ CH.
  • Page 43: Mdaq-Aaf4-5-Bx

    ±7.5 V to ±15 V direct via MDAQ-BASE or -V Power consumption Typ. 3 W Dimensions (W x D x H) 175 x 61 x 25 mm (6.9 x 2.4 x 1 in.) DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 44: Step Response Of 100 Hz Filter

    MDAQ-AAF4-5-Bx Step response of 100 Hz filter Butterworth 4 order Butterworth - maximally flat in the pass band frequency response - There is some overshoot in the time domain - Butterworth rolls off faster than the Bessel filter...
  • Page 45: Ce-Certificate Of Conformity

    EN 55011 Class B Immunity EN 61000-6-2 Group standard Graz, October 14, 2008 Dipl.-Ing. Roland Jeutter / Managing director Place / Date of the CE-marking DE-M080402E • DEWE-MDAQ series • Technical Reference Manual • Printing version 1.0.2 • January 16, 2011...
  • Page 46 Notes...