MicroDock II Quick Reference Guide Display Elements Pushbuttons Item Function Pushbutton Description To bump test a detector, press AC power C BUMP CHECK Batteries fully charged When connecting a new docking module, press and hold C BUMP CHECK BUMP CHECK to send a Batteries half-charged confirmation signal back to the...
MicroDock II User Options Menu User Options Menu Description Display Description Display Time/Date modifies the time and About displays the firmware date of the base station. revision for the base station and the docking module(s). Format MMC Inlet Setup modifies the gas formats the Mul- type, the gas concentration lev- tiMediaCard (MMC).
MicroDock II Quick Reference Guide Installation Item Description Inlet filter assembly C-cell batteries (4) Battery cover Philips pan head retaining screws (2) MultiMediaCard (MMC) Battery compartment Charger port USB port Power port Exhaust outlet Gas cylinder Demand flow regulator Calibration gas hose...
Page 7
MicroDock II Installation Warning The disconnect device must be installed in close proximity to the base station and must be marked as a disconnecting The atmosphere must be free of background gas. Do not method for the base station. use the base station in a hazardous area. All required National Electrical Codes (NEC) and safety Battery Installation standards must be followed.
Page 8
MicroDock II Quick Reference Guide Inserting the Detector station LCD displays the docking module number, and the type and serial number of the detector. Caution GasAlertMicro Infrared or intense ambient light (sun or halogen) may • Insert the bottom of the detector (serial number face up) interfere with the base station/detector communication.
Page 9
MicroDock II Inserting the Detector • When the detector has been inserted correctly, the RUN GasAlertMax XT LEDs light yellow, Unit Inserted displays on the base • Insert the bottom of the detector (serial number face up) station LCD, and MicroDock displays on the detector at a 30°...
MicroDock II Quick Reference Guide ual. To charge the battery pack separately, refer to the Activating the Station MicroDock II Base Station User Manual. To activate the station, press and hold C (the left- most button) until the following screen displays. •...
Page 11
MicroDock II Bump Test Bump Test • Inlet 1—default connection for ambient air. • Inlet 2—default connection for four-gas mix (unless oth- A bump test is performed to confirm that the detector is erwise specified when purchased). responding to gas, and that the audio and visual alarms are •...
Page 12
MicroDock II Quick Reference Guide The gas is automatically applied. All calibration cylinders must be used with demand flow regulators and must meet the following maximum inlet pressure specifications • Disposable cylinders 0-1000 psig/70 bar • Refillable cylinders 0-3000 psig/207 bar The LCD displays the results of the bump test.
Unsuccessful Data Transfer: The FAIL LED lights red on the station MultiMediaCard (MMC). docking module. Attempt to transfer the data again. If the sec- ond data transfer is unsuccessful, ensure the detector is acti- vated. If the problem persists, contact BW Technologies by Honeywell.
To replace the transferred every time a bump test or calibration is performed. MMC, contact BW Technologies by Honeywell. After performing a bump test or calibration on the Access Test Results GasAlertMicroClip, GasAlertMax XT, and GasAlertQuattro To access and view test results, refer to Importing Data into docking module, the PASS LED lights green.
MicroDock II Charging the Battery Pack (Optional) • GasAlertMicroClip Note • GasAlertMax XT The detector will not charge while bump tests or • GasAlertQuattro calibrations are being performed. If the battery is low, charge the detector for 30 minutes, then begin a test. The detector can be configured prior to performing a bump test When the test is complete, resume charging the and/or calibration to...
MicroDock II Quick Reference Guide Maintenance The CHARGE LED lights red (all charging modules). Allow the battery to obtain a full charge (2 to 4 hours, Warning depending upon how many docking modules are con- No user-serviceable parts inside. nected to this station). Caution If a docking module is in queue waiting to charge, the CHARGE LED flashes red.
MicroDock II Specifications Specifications Configuration recognition: Automatic (instrument and sensor) Alarm/calibration parameters: User defined The MicroDock II base station and docking modules are for Calibration gas connections: Built-in (base station) indoor use only. Gas connection: 1/8” SMC connect sub-miniature coupling Instrument dimensions: (w x l x h) 21.2 x 26.3 x 8.2 cm Solenoid: Built-in (docking modules) (8.3 x 10.4 x 3.2 in.) base station and one docking module...
MicroDock II Quick Reference Guide Charger Specifications This equipment generates, uses, and can radiate radio fre- quency energy and, if not installed and used in accordance with Size: 8.6 x 8.2 x 7.8 cm (3.4 x 3.2 x 3.1 in.) the instruction manual, may cause harmful interference to radio Weight: 97 g (3.4 oz.) per model communications.