Page 6
Table of Contents 7.3.1.1 Direct entry of transducer characteristics ......7.3.1.2 Calibrating the characteristic curve of the transducer .
Page 7
Table of Contents 8.1.2 Assigning remote control contacts ..........8.2 Limit values (single-channel modules only) .
Page 8
Table of Contents 10.1.3 Set access for operator ............10.1.4 Delete user .
Page 9
Safety instructions SAFETY INSTRUCTIONS Intended use The amplifier system is to be used exclusively for measurement tasks and directly related control tasks. Use for any purpose other than the above is deemed to be non-designated use. In the interests of safety, the device should only be operated as described in the operating manuals.
Page 10
Safety instructions After making settings and carrying out activities that are password- protected, you must make sure that any controls that may be connected remain in a safe condition until the switching performance of the amplifier system has been tested. Working safely The supply voltage connection, as well as the signal and sense leads, must be installed in such a way that electromagnetic interference does not ad...
Page 11
Safety instructions Qualified personnel are persons entrusted with siting, mounting, starting up and operating the product and who possess the appropriate qualifications for their function. This device is only to be installed and used by qualified personnel strictly in accordance with the specifications and with the safety rules and regula tions which follow.
Page 12
Safety instructions Electromagnetic compatibility If an amplifier channel is removed, the module must be sealed with a blind panel. Only operate built-in devices once they are installed in the housing provided. The device complies with the safety requirements of DIN EN 61010 Part 1 (VDE 0411 Part 1);...
Page 13
Markings used Markings on the device MARKINGS USED Markings on the device CE mark By way of the CE mark the manufacturer guarantees that the product com plies with the requirements of the relevant EC directives (the Declaration of Conformity can be found at http://www.hbm.com/HBMdoc). Statutory waste disposal mark In accordance with national and local environmental protection and material recovery and recycling regulations, old devices that can no longer be used...
Page 14
Markings used The markings used in this document Symbol Significance This marking draws your attention to important information about the product or Important about handling the product. This marking indicates application tips or other information that is useful to you. This marking draws your attention to information about the product or about hand...
Page 15
Introduction Degree of protection INTRODUCTION Degree of protection The degree of protection given in the technical data indicates the suitability of the housings for various ambient conditions and also the protection of persons against potential risks when used. The letters IP (International Pro tection) which are always present in the designation, are followed by two digits.
Page 16
Introduction Notes on documentation Notes on documentation The present operating manual explains operation of the device and how to perform measurements with it. Further information and the possibility to download the system CD can be found on the Internet at www.hbm.com Guides Several guides are available to help you:...
Page 17
Introduction System description System description The MGCplus system is structured modularly. Depending on the housing variant, up to 16 slots are available for single and multi-channel amplifier modules. Thus up to 128 measuring points can be measured in an MGCplus. Each amplifier module works independently through its own CPU.
Page 18
Introduction System description Digital signal conditioning Digital Filtering Display signal conditioning Communication Scaling, zero processor CPxx balance, ... Filtering, scaling, control zero balance, ... panel 8-channel Single channel module Serial bus module Fig. 3.1 Block diagram of MGCplus MGCplus...
Page 19
Introduction Layout of the MGCplus device Layout of the MGCplus device Connection boards (AP01i, AP815i, ...) Power CPxx supply Communication processor AB22A display and control unit Amplifier plug−in board (ML30B, ML55B, ML801B...) Fig. 3.2 Device layout with display and control unit AB22A Double-width connection boards (AP03i, AP455i) must be plugged into the odd-numbered slots.
Page 20
Introduction MGCplus housing designs MGCplus housing designs The MGCplus system is available with different housing versions (dimensions in mm; 1 mm = 0.03937 inches): Desktop housing TG 009E (177x161x386) Desktop housing TG 001E (258x161x386) Desktop housing TG 003E (462x161x386) 19” rack frame ER 003E (482x133x365) Desktop housing Rack frame Slots...
Page 23
Introduction MGCplus housing designs Special function modules ML71B ML74B ML77B ML78B AP71 AP74 CANHEAD AP75 AP77 AP78 Digital output Digital input Analog output ProfiBus CANBus HBK hardware CANHEAD Separate Operating Manual available on hbkworld.com MGCplus...
Page 24
Introduction CP52 communication processor CP52 communication processor 3.7.1 Mounting For type "D" housings (ER003D or TG001D, etc.) the existing communication processor (CP22/CP42) can be replaced by the CP52 communication pro cessor. Loosen the screws on the old communication processor, the blind panel (only with CP22) and the cover of the NT030 power supply unit.
Page 25
Introduction CP52 communication processor CP42 RS 232 CARDBUS USB DEVICE USB HOST YE SLAVE RD ERROR ETHERNET GN MASTER CTRL I/O GND 1 2 24V 2 Communication processor CP42, power supply Communication processor CP52, power supply unit NT030 unit NT040 Fig.
Page 26
Introduction CP52 communication processor Without CP … View from above With CP … View from above yes (1) yes (1) no (2) no (2) General plan of the interface switches(housing cover open, view from above): Housing Flat ribbon cable CP switch Power supply Interface switch...
Page 27
Introduction CP52 communication processor 3.7.2 Status LED The CP52 communication processor has a status LED with three colors: green, yellow and red. LED lit Meaning Green The device is ready for measurement, no errors present. Yellow The device is in the boot process. There is an error present which could possible interfere with the measurement.
Page 28
Introduction Conditions at the place of installation Conditions at the place of installation CAUTION Protect the devices in a desktop housing from moisture and dampness or weather conditions such as rain, snow, etc. Make sure that you do not cover the ventilation openings at the side, the openings for the power pack fan on the back side of the device and the openings underneath the device.
Page 29
Introduction Maintenance and cleaning Maintenance and cleaning The MGCplus system devices are maintenance-free. Please note the follow ing points when cleaning the housing: CAUTION Disconnect the mains plug from the socket before cleaning. Clean the housing with a soft, slightly damp (not wet!) cloth. You should never use solvents, since this may damage the labeling on the front panel and the display field.
Page 30
Introduction Maintenance and cleaning MGCplus...
Page 31
Connection Connecting the MGCplus in a tabletop housing/rack frame CONNECTION Connecting the MGCplus in a tabletop housing/rack frame 4.1.1 Mains connection Mains connection The NT030 and NT040 power supply units are designed for a 115 - 230 V connector and for a maximum configuration of 16 modules and connection boards.
Page 32
Connection Connecting the MGCplus in a tabletop housing/rack frame 4.1.2 Synchronization of multiple CP52 devices 4.1.2.1 Synchronization of multiple CP52 devices via a synchronization jack Connected devices are automatically detected and synchronized when synchronization jacks are occupied. A distinction is made between the sync master and the sync minor.
Page 33
Connection Connecting the MGCplus in a tabletop housing/rack frame If several MGCplus systems are to be synchronized with each other, each system must be equipped with a CP52 communications processor. To syn chronize MGCplus systems with CP52, you need a synchronization cable with the HBM part number 1-KAB2125-2 (2 m in length).
Page 34
Connection Connecting the MGCplus in a tabletop housing/rack frame Powering up the system When connecting the system, the sync minors must be connected first. Connect the system that will work as the sync master last of all. 4.1.2.2 Synchronization of multiple CP52 devices via Ethernet IEEE1588:2008 (PTPv2) If there is a long distance (>...
Page 35
Connection Connecting the MGCplus in a tabletop housing/rack frame Domain: 0 (default setting) ... 127 With synchronization via PTP, the LEDs on the synchronization jack indicate states as follows: PTP master PTP minor Green LED Sync Out is lit Green LED Sync Out is lit Green LED Sync In is not lit Green LED Sync In is lit MGCplus...
Page 36
Connection Connecting the MGCplus in a tabletop housing/rack frame Ethernet PTV2 switch Fig. 4.2 Two CP52 devices synchronized via PTPv2 It should be noted here that when synchronizing via PTPv2, the clocks and carrier frequencies, etc. of the devices are synchronized with high precision (<...
Page 37
Connection Connecting the MGCplus in a tabletop housing/rack frame 4.1.3 Synchronization of CP52 with CP42 The synchronization jack enables MGCplus systems with CP52 to be syn chronized with MGCplus systems with CP42. To do this note the following points: It is recommended that CP42 is used as synchronization master in systems consisting of CP42 and CP52.
Page 38
Connection Connecting the MGCplus in a tabletop housing/rack frame CP42 RS 232 CARDBUS USB DEVICE USB HOST YE SLAVE RD ERROR ETHERNET GN MASTER SYNC CTRL I/O 24V 2 GND 1 2 Fig. 4.3 Example of synchronizing two MGCplus systems equipped with CP52 and CP42.
Page 39
Connection Shielding design It is also essential to ensure at this point that the setting parameters of all devices in the group all have the same settings (see 4.1.2.2). The LEDs of the CP52 show the same thing as with a simple MGCplus topology. Shielding design Sources of interference can cause electromagnetic fields which can induce interference voltages inductively or capacitively via the connection cable...
Page 40
Connection Shielding design Notice All parts of the measurement chain (including all cable connection points such as plugs and couplings) must be surrounded by a closed EMC-tested shield. Shield junctions must represent a full contact, closed and low-imped ance connection. This is the case for original HBM plug connections. Ground connection and grounding As the signal ground and shielding are separated in EMC-compliant cabling, the shielding can be connected at more than one point to the ground, i.e.
Page 41
This applies to all bridge connections. When using long cables and with a high transducer resistance, HBK additionally recommends soldering resistors in the sense leads. The resistors should be placed as close as possible to the transducer. The value of the resistances must be half that of the bridge resistance /2).
Page 42
Connection Connecting the transducer Measurement Connecting the transducer signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Important Measurement signal (-) Transducers with four-wire configuration Cable shield If you connect a transducer with a 4-wire cable, you must connect the sense Sense lead (+) lead with the corresponding bridge excitation circuit in the transducer plug Feedback...
Page 43
Connection Connecting separate TEDS modules View of the mating connector AP402i (solder side) Hsg. TEDS TEDS data MGCplus...
Page 44
Connection SG full bridges, inductive full bridges 4.3.2 SG full bridges, inductive full bridges AP01i AP03i AP14 AP455i Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Measurement signal (-) Hsg. Hsg. Cable shield Sense lead (+) Sense lead (-) Cable color code: wh= white;...
Page 45
Connection Full bridge circuits on AP810i/AP815i 4.3.3 Full bridge circuits on AP810i/AP815i AP810i/AP815i Subchannel Subchannel Subchannel Subchannel Subchannel 1...4 Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Measurement signal (-) Hsg. Hsg. Hsg. Cable shield Hsg. Sense lead (+) Subchannel 5...8 Sense lead (-) Cable color code: wh= white;...
Page 46
Connection Strain gage half bridges, inductive half bridge circuits 4.3.4 Strain gage half bridges, inductive half bridge circuits AP01i AP03i AP14 AP455i Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Cable shield Hsg. Hsg. Sense lead (+) Sense lead (-) Cable color code: wh= white;...
Page 47
Connection LVDT transducers 4.3.5 LVDT transducers AP455i LVDT transducers Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Measurement signal (-) Cable shield Hsg. Sense lead (+) Sense lead (-) MGCplus...
Page 48
Connection Strain gage half bridges on AP810i 4.3.6 Strain gage half bridges on AP810i AP810i Subchannel Subchannel Subchannel Subchannel Subchannel 1...4 Sense lead (-) Bridge excitation voltage (-) wh/rd Measurement signal (+) wh/gn Cable shield Hsg. Hsg. Hsg. Hsg. Subchannel 5...8 wh/br Bridge excitation voltage (+)
Page 49
Connection Strain gage half bridges on AP815i 4.3.7 Strain gage half bridges on AP815i AP815i Subchannel Subchannel Subchannel Subchannel Subchannel 1...4 Sense lead (-) Bridge excitation voltage (-) wh/rd wh/gn Measurement signal (+) Cable shield Hsg. Hsg. Hsg. Hsg. Subchannel Measurement signal (-) 5...8 Bridge excitation voltage (+)
Page 50
Connection Single strain gage 4.3.8 Single strain gage 4.3.8.1 Single strain gage on AP14 AP14 Sense lead (-) Excitation voltage (-) Hsg. Excitation voltage (+) Measurement signal (+), sense lead (+) Three-wire connection Four-wire connection Important If the strain gage is in a three-wire connection, symmetrical cables must be used to ensure that cable resistance is correctly compensated.
Page 51
Connection Single strain gage 4.3.8.2 Single strain gage on AP814Bi AP814Bi Subchannels Subchannel 1...8 Measurement signal (+), excitation voltage (+) Cable shield Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Hsg. Excitation voltage (-) Sense lead (-) Three-wire connection...
Page 52
Connection Single strain gage 4.3.8.3 Single strain gage on AP815i AP815i Subchannel Subchannel Subchannel Subchannel Subchannel 1...4 wh/rd Excitation voltage (-) Sense lead (-) Cable shield Hsg. Hsg. Hsg. Hsg. Subchannel Measurement signal (+), 5...8 wh/gn sense lead (+) Excitation voltage (+) Cable color code: wh= white;...
Page 53
Connection SG chains and strain gage rosettes on AP815i 4.3.9 SG chains and strain gage rosettes on AP815i Subchannel AP815i Excitation voltage (-) Subchannel Sense lead (-) Subchannel Subchannel Cable shield Hsg. 1...4 Measurement signal (+), sense lead (+) Subchannel Excitation voltage (+) Subchannel Subchannel...
Page 54
Connection SG chains and strain gage rosettes on AP815i Subchannel AP815i Excitation voltage (-) Subchannel Sense lead (-) Cable shield Hsg. Subchannel 1...4 Measurement signal (+), sense lead (+) Excitation voltage (+) Subchannel Subchannel 5...8 Excitation voltage (-) Subchannel Sense lead (-) Cable shield Hsg.
Page 58
Connection Torque shaft (T4A, T5, TB1A) 4.3.11 Torque shaft (T4A, T5, TB1A) 4.3.11.1 Torque measurement (slip rings or direct cable connection) AP01i AP03i AP14 AP455i Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Measurement signal (-) Cable shield Hsg.
Page 59
Connection Thermocouples 4.3.12 Thermocouples Universal sockets made of copper are used beginning with hardware revi sion 3.00 to make it possible to connect different thermocouple types. Cop per is softer than the miniature thermo connector of thermocouple type K, for example. The following points should therefore be considered: Proceed carefully when inserting the plug in the socket Insert it at an angle as close as possible to perpendicular If you can feel resistance, file the edges of the plug slightly to reduce...
Page 60
Connection DC voltage sources Type Thermal material 1 (+) Thermal material 2 (-) Iron Copper-nickel Nickel‐chrome (color code Nickel‐aluminum (color code green) white) Copper Copper-nickel 4.3.13 DC voltage sources AP01i AP03i Maximum input voltage against ground = ±12V Supply voltage zero Cable shield Hsg.
Page 61
Connection DC voltage sources AP402i Maximum input voltage against ground = " 100V Cable shield Hsg. MGCplus...
Page 62
Connection DC voltage sources AP801 AP801S6 Maximum input voltage against ground = +50V Hsg. Supply voltage 8V/16V Power supply 0V Hsg. *) For information on switching the supply voltage see next page MGCplus...
Page 63
Connection DC voltage sources Supply voltage +16V No function (for special versions only) Jumper No function (for special versions only) Supply voltage +8V Connection board AP 801S6 (side view) MGCplus...
Page 64
Connection DC voltage sources AP402i (−) Supply voltage 5V/8V/16V Power supply 0V Hsg. For information on switching the supply voltage see next illustration Supply voltage +5V Supply voltage +8V Supply voltage +16V No function (factory setting) Jumper Connection board AP402i (side view) MGCplus...
Page 65
Connection DC voltage sources AP836i Subchannels Subchannel 1...4 Bridge excitation voltage (+) Sense lead (+) Measurement signal (+) Measurement signal (-) Cable shield Hsg. Hsg. Hsg. Hsg. Sense lead (-) Subchannel 5...8 Bridge excitation voltage (-) Common signal and supply voltage zero, power line not corrected on one side. Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground, the design of the active transducer must without exception be potential-free! AP836i...
Page 66
Connection DC voltage sources AP836i Subchannels Subchannel 1...4 Bridge excitation voltage (+) Sense lead (+) Measurement signal (+) Measurement signal (-) Cable shield Hsg. Hsg. Hsg. Hsg. Sense lead (-) Subchannel 5...8 Bridge excitation voltage (-) Common signal and supply voltage zero, power lines fully corrected. Since the bridge excitation voltage used to supply the active transducer is symmetrical to GND/ground, the de...
Page 67
Connection DC current sources 4.3.14 DC current sources AP01i AP03i Maximum input voltage against ground = ±12V Supply voltage zero Cable shield Hsg. Hsg. MGCplus...
Page 68
Connection DC current sources AP402i Maximum input voltage against ground = ±100V Cable shield Hsg. MGCplus...
Page 69
Connection Resistors, Pt100 4.3.15 Resistors, Pt100 AP835 Excitation voltage (-) Measurement signal (-) Θ Cable shield Hsg. Measurement signal (+) Excitation voltage (+) MGCplus...
Page 70
Connection Frequency measurement without directional signal 4.3.16 Frequency measurement without directional signal AP01i AP03i Supply voltage zero Frequency generator/ pulse Cable shield Hsg. Hsg. generator Rotational speed/pulse signal 1 (frequency f Deactivate the analysis of the f signal in this mode (factory setting: Off), see page 152. MGCplus...
Page 71
Connection Frequency measurement with directional signal 4.3.17 Frequency measurement with directional signal AP01i AP03i Supply voltage zero Cable shield Hsg. Hsg. Frequency generator / Rotational speed/pulse pulse signal 1 (frequency f generator Pulse signal 2 (frequency f Activate the analysis of the f_2 signal in this mode (factory setting: Off), see page 152 MGCplus...
Page 73
Connection Pulse counting, differential 4.3.19 Pulse counting, differential AP17 Zero index signal + Zero index signal - Industrial pulse generators Transducer error MGCplus...
Page 74
Connection Active piezoelectric transducers 4.3.20 Active piezoelectric transducers AP418i * Use special coaxial cable Shield Piezoelectric transducer with preamplifier (with T‐ID/TEDS for AP418i) Input * Information Information about AP418i connection boards: When laying transducer cables outside of enclosed areas or with cable lengths greater than 30 m between the connection board and transducer, the sensor cables must be designed with an additional, separately grounded shield to ensure overvoltage protection.
Page 75
Connection Piezoresistive transducers 4.3.21 Piezoresistive transducers AP01i AP03i Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Measurement signal (-) Hsg. Hsg. Cable shield Sense lead (+) Sense lead (-) MGCplus...
Page 76
Connection Potentiometric transducers 4.3.22 Potentiometric transducers AP01i AP03i Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Cable shield Hsg. Hsg. Sense lead (+) Sense lead (-) AP836 Subchannels Subchannel 1...4 Measurement signal (+) Bridge excitation voltage (-) Bridge excitation voltage (+) Cable shield Hsg.
Page 77
Connection Connection via the distributor board VT810/815i 4.3.23 Connection via the distributor board VT810/815i Strain gage half bridge on (AP810i) on Single SG; four-wire connection, AP815i only VT810/815i RJ45 socket RJ45 socket wh/rd Excitation voltage (-) Sense lead (-) Bridge excitation voltage (-) wh/rd Sense lead (-) wh/gn...
Page 78
Connection Connection via the distributor board VT810/815i Connecting diagram Measuring point Distributor board VT810/815i Connection board AP810i/815i D‐Sub25 SG 1 D‐Sub25 RJ45 1-Kab156-3 1-Kab263-3 SG 2 1-Kab156-3 D‐Sub25 D‐Sub25 1-Kab263-3 1-AP810i 1-AP815i SG 8 1-Kab156-3 Extensive connection instructions are enclosed with the distributor board. MGCplus...
Page 79
Connection Connecting CANHEAD modules Connecting CANHEAD modules To connect CANHEAD modules to the MGCplus system you need the ML74B communication card and the AP74 connection board. In combina tion with the NT030 power pack, you can connect a maximum of 12 mod ules per board and a maximum of 25 modules per MGCplus device (a maxi...
Page 80
Connection Communication card ML74B 4.4.1 Communication card ML74B ML74B Labeling Color Meaning CHAN. Yellow Channel is selected. ERROR/WARN. Error/Warning Yellow CAN protocol being received Yellow CAN protocol being sent OVRN Overrun occurred BUS-/ERR Bus error CONFIG Yellow The assigned CANHEADS are being set up AP74 Yellow...
Page 81
Connection AP74 connection board 4.4.2 AP74 connection board AP74 LED color Meaning Green Normal status in operation Short circuit or force overshoot None Power supply turned off MGCplus...
Page 82
Connection Inputs and outputs, remote controls Inputs and outputs, remote controls 4.5.1 Inputs/outputs of the CP52 The CP52 communication processor has two digital inputs and outputs each. The digital inputs and outputs must be operated with an external cur rent feed (12 V ... 24 V). Shielded cables are used for EMC reasons. Inputs (0 V ...
Page 83
Connection Inputs and outputs, remote controls cable shield 24V DC Fig. 4.6 Wiring example for the "Start Trigger" function at the CP52 control inputs MGCplus...
Page 84
Connection Inputs and outputs, remote controls 4.5.2 Analog output on the front panel On the front panel of a single-channel module there is a BNC socket for the analog output signal V . (This socket is used for test purposes. Stationary wiring should always be done with the connection boards, since there is no noise voltage with this method).
Page 85
Connection Inputs and outputs, remote controls 4.5.3.1 Socket assignment AP01i/AP03i/AP14/AP17 Function Digital Remote control 1 Input Remote control 2 Input Remote control 3 Input Remote control 4 Input Remote control 5 Input Remote control 6 Input Sct 2 Outputs Remote control 7 Input OUTPUT Remote control 8...
Page 86
Connection Inputs and outputs, remote controls On pin 13 the analog output signal of V is present (and also on the BNC socket on the front panel). The connected load resistance must be greater than 5 kohms. The following signals can be assigned to outputs VO1 and VO2: S1: Gross S2: Net S3: Peak value 1...
Page 87
Connection Inputs and outputs, remote controls With AP01i and AP03i, jumpers can be used to adapt the filter settings of the two analog outputs follows: Analog output filtering With AP01i and AP03i, jumpers can be used to adapt the filter settings of the two analog outputs follows: Filter approx.
Page 88
Connection Inputs and outputs, remote controls Function F2- rotational speed 90°, calibration signal trigger Input/output F2+ rotational speed 90°, calibration signal trigger ground Input/output The current information is for the maximum permitted continuous currents of the AP17. The number of connection boards per housing is not limited, but a maximum of three connection boards can be used for transducer supply (16 V, for example for torque flange T10 series, T12/T12HP, T40 series).
Page 89
Connection Inputs and outputs, remote controls Profibus 4.5.3.2 AP77 The pin assignment of the 9‐pin Sub‐D socket complies with Profibus stan dards IEC 61158/61784. Function RS485-B RS485-RTS RS485-A Information Further information can be found in the ML77B operating manual. MGCplus...
Page 90
Connection Inputs and outputs, remote controls 4.5.4 Inputs and outputs of AP75 The AP75 connection board has 8 digital inputs, 8 digital outputs and 2 ana log outputs. The digital outputs must be operated with an external current feed (12 V...24 V). The AP75 connection board can be operated together with the special function module ML78B.
Page 91
Connection Inputs and outputs, remote controls Analog outputs V and V have a common ground system that is sepa rated from the ground systems of the digital inputs and outputs. Wiring example for using analog output MGCplus...
Page 92
Connection Inputs and outputs, remote controls 4.5.5 Analog outputs on the AP78 The AP78 connection board has 10 analog outputs. The analog outputs out puts designated A03...A10 are electrically isolated, while outputs VO1 and VO2 can be digitally filtered. The AP78 connection board can be used GND A03 together with the ML78B.
Page 93
Starting up Devices in the desktop housing and rack frame STARTING UP This section shows you the necessary operating steps to place your mea surement chain (measuring amplifier system and transducer) in operation. This will enable you to perform a functional test of all the components. The steps are deliberately described in very general terms so that there is no need to go into the details of specific transducers or amplifier modules.
Page 94
Starting up Devices in the desktop housing and rack frame For safety reasons all free slots (amplifiers or connection boards) must be covered by blind panels. Check to make certain the amplifiers and connection boards are securely plugged in. Connect the device to the mains with the mains cable provided. Connect your transducer to the applicable socket on the back of the connection board (designation BU01).
Page 95
Starting up Devices in the desktop housing and rack frame We recommend first setting the Language if you want to use a language other than German. System Display Amplifier Options Password Save/Load Data acquisition Interface Synchronization Language Time Auto assignment LANGUAGE Language: Deutsch ↓...
Page 96
Starting up Devices in the desktop housing and rack frame MGCplus...
Page 97
Functions and symbols of the AB22A Control elements of the AB22A FUNCTIONS AND SYMBOLS OF THE AB22A Control elements of the AB22A You can make all the settings on your device with the keys on the AB22A/AB32 display and control unit. You can use the Switch key select the operating states "Measurement"...
Page 98
Functions and symbols of the AB22A Display Display 6.2.1 The first display After the mains voltage is turned on, the initialization of the device is in dicated in the display by a horizontal progress bar. You are also informed of the current software version.
Page 99
Functions and symbols of the AB22A Display 6.2.2 Display in measuring mode Display with image type "1 measured value" è Settings for different image types are described in detail in section 9 "Display", page 207. The image types set by default in the factory are pre sented here as an overview.
Page 100
Functions and symbols of the AB22A Display Display with image type "3 measured values" Gross Gross To the next image type with Gross Measure Gross Gross Display with image type "6 measured values" Gross Gross To the next image type with Gross Gross Measure...
Page 101
Functions and symbols of the AB22A Display Symbols in the display The status line provides you with information about the current status of the measuring device: Gross Channel Measure Status line Measure Measure, Zero, Calibrate Status of the amplifier input Status display of the parameter set memory Number 1...8 Number of the current parameter set...
Page 102
Functions and symbols of the AB22A Display Channel Channel number of a single-channel module Channel Channel number of a multi-channel module First number is the slot number Second number is the number of the subchannel MGCplus...
Page 103
Functions and symbols of the AB22A AB22A in Setup mode 6.2.3 Messages of AB22A/AB32 Message in the display Cause Remedy The CP42/CP52 hard disk The capacity of the storage medium is Insert a new storage medium or delete is full! exceeded.
Page 104
Functions and symbols of the AB22A AB22A in Setup mode Display in Setup mode Options Display Amplifier Display Amplifier Options Selection bar System-related settings User-oriented settings of Amplifier settings for Additional settings for that are usually only the display such as the each channel, for exam...
Page 105
Functions and symbols of the AB22A AB22A in Setup mode Selection level 1 Selection level 2 Pull‐up menu Password Password Setting level 1 Save/load Data acquisition Display PASSWORD Interface delete User: Synchronization Setting level 2 Language Password: chang Display Selection bar Add user Access: User:...
Page 106
Functions and symbols of the AB22A AB22A in Setup mode 6.3.1 Call menus Press the Switch key . At first only the selection bar appears. Now if you press the function keys F1...F4, the associated pull-up menu appears above the corresponding term (in this example System). Move the light bar to the desired term in the pull‐up menu with the cursor keys (in this case Language) and press the confirmation key...
Page 107
Functions and symbols of the AB22A AB22A in Setup mode 6.3.2 Exit menus If you want to leave a menu level press the Switch key Result: Return to measuring mode: or one of the function keys F1...F4 Result: Return to selection level 2 or (if present) the Cancel button symbol.
Page 108
Functions and symbols of the AB22A AB22A in Setup mode 6.3.3 Channel selection in measuring mode There are two ways to select a channel in measuring mode: 1. With the channel keys 2. With direct input of the channel number using the alphanumeric keypad (recommended with multi-channel modules).
Page 109
Functions and symbols of the AB22A AB22A in Setup mode Channel input field 6.3.4 Channel selection in Setup mode CHANNEL NAME Channel In most setup windows of the AB22A the name of the currently selected set ting menu and the selected channel appear in the header. Channel name: 2-ML55 There are two ways to select a channel in setup mode:...
Page 110
Functions and symbols of the AB22A AB22A in Setup mode 6.3.6 Drop-down menus In the first two selection levels you can select terms. On the first level (menu bar) by pressing the corresponding function key, on the second level by making a selection in pop‐up menus.
Page 111
Functions and symbols of the AB22A AB22A in Setup mode Channel-related activation fields SIGNAL SELECTION Cancel All channels All signals 1 2 3 4 5 6 7 8 10 111213 141516 Channel Gross PV12 LV status Fig. 6.3 Example of a channel-related activation field Multi-channel modules are identified by two dots under the channel number (channel number 8 in the example).
Page 112
Functions and symbols of the AB22A AB22A in Setup mode Information The "Unit" selection field in the "Transducer" setup window can be edited by double clicking with . Then you can enter up to four characters (for example min , with as a character that is placed on the key).
Page 113
Functions and symbols of the AB22A AB22A in Setup mode The minus sign of the key can also be used as a separator in text fields. key also has special characters °, W, m assigned to it. 2, 3, key also has the superscript numbers and special character @ assigned to it.
Page 114
Functions and symbols of the AB22A AB22A in Setup mode MGCplus...
Page 115
Measuring General information MEASURING General information This section describes the steps you will need to follow to perform a mea surement with MGCplus. At the beginning of each section an example ex plains the transducer-specific details for adapting the amplifier modules. After you have adapted the amplifier modules you can start with the mea...
Page 116
Measuring General principles for adjusting a measurement channel General principles for adjusting a measurement channel The "Amplifier" pull-up menu is adapted to the signal flow of a measurement chain. It is the basis for making adjustment to the entire measurement chain.
Page 117
Measuring General principles for adjusting a measurement channel CONDITIONING CHANNEL 1 Make settings for signal conditioning 0.0000 ... Zero reference: Zero balance (zero drift) −>0<− Zero offset: 0.0000 ... −>T<− Tare: 0.0000 ... Taring Disable zeroing: Disable taring: Filters 100 ↓ Bessel ↓...
Page 118
Measuring General principles for adjusting a measurement channel 7.2.1 Adapting to the transducer TRANSDUCER CHANNEL1 Type Type: SG full bridge Selection field for transducers that can be connected. The selection de Excitation: 5V↓ pends on the amplifier type and connection board. kg ↓...
Page 119
Measuring General principles for adjusting a measurement channel 7.2.1.1 Extended functions of the ML38B TRANSDUCER CHANNEL1 Nominal (rated) value Type: SG full bridge First enter the nominal (rated) value. See the table for the linear interpola Ext.fct. 5V↓ Excitation: tion equation (in the calibration certificate) for the value. If it is not available, kg ↓...
Page 120
Measuring General principles for adjusting a measurement channel Nominal (rated) electrical unit (right input field) C + A @ Conversion: Nominal (rated) physical unit (left input field) Example: Enter the coefficients from an HBM DAkkS calibration certificate for Force. Note the following correspondences: DAkkS calibration certificate Polynomial equation Notice...
Page 121
Measuring General principles for adjusting a measurement channel Loading TEDS data into the amplifier In the default setting, the TEDS key F2 is in level 2 of the function keys. Press it in Measuring mode Press the TEDS key The actual amplifier will then be parameterized with the transducer settings from the TEDS chip if it supports this functionality.
Page 122
Measuring General principles for adjusting a measurement channel 7.2.3 Signal conditioning CONDITIONING CHANNEL 1 Zero reference 0.0000 ... Zero reference: The amount by which the relative zero point is offset relative to absolute −>0<− Zero offset: 0.0000 ... zero. −>T<− Tare: 0.0000 ...
Page 123
Measuring General principles for adjusting a measurement channel A platform is standing on three load cells and will be used for tank weighing. Tank Two different components are poured into the tank and their individual Platform weight (net) will be displayed. The table summarizes the states before and after zero balance or taring is Load cells triggered.
Page 124
Measuring General principles for adjusting a measurement channel Low pass Step response Low-pass filters are used to suppress unwanted higher frequency interference that lies above a specific cut-off frequency. Amplitude response, runtime and step response are dependent on the filter characteristic.
Page 125
Measuring General principles for adjusting a measurement channel 7.2.4 Display Unit1 Select the unit in the display. You can choose between the basic unit (mV/ V), the user unit (kg, etc.) and the unit of the analog output (V). Then the cor System Display Amplifier...
Page 126
Measuring General principles for adjusting a measurement channel 7.2.5 Analog outputs (single-channel modules only) Output V Signal selection for analog output 1 (socket Bu2 on the back side of the de vice and BNC socket on the front panel for test purposes). System Display Amplifier...
Page 127
Measuring General principles for adjusting a measurement channel Example: Minimum setting for first measurement The transducer is a load cell with the following nominal (rated) data: System Display Amplifier Options Maximum capacity 50kg Nominal (rated) sensitivity 2mV/V è 1. Use the channel selection keys or a direct entry ( see page 108) to Transducer...
Page 128
Measuring General principles for adjusting a measurement channel 10.Use to switch to the button and confirm with (the mea measure... sured value appears in the right edit field of Zero point) System Display Amplifier Options 11.Use to switch to the Nominal (rated) value edit field and enter the value 50.
Page 129
Measuring Adapting to the transducer Adapting to the transducer 7.3.1 SG transducers SG transducers (load cells and force transducer from HBM) are passive transducers that have the following features: They must be supplied with an excitation voltage (carrier frequency or direct voltage) They are SG full bridge circuits The transducer is identified by the following data:...
Page 130
Measuring Adapting to the transducer 7.3.1.1 Direct entry of transducer characteristics The settings are explained with the following example: Load cells with characteristics: Maximum capacity 20 kg, excitation voltage 5 V, nominal (rated) sensitivity 2 mV/V System Display Amplifier Options Measurements will be made up to 10kg, i.e.
Page 131
Measuring Adapting to the transducer 11.Use to switch to the measure button and confirm with (the measured value appears in the right edit field of Zero point) System Display Amplifier Options 12.In the left edit field Nominal (rated) value enter the value 20. 13.In the right edit field Nominal (rated) value enter the value 2 (under the unit mV/V).
Page 132
Measuring Adapting to the transducer 19.Go back to the pull-up menu with 20.In the pull-up menu select Analog outputs and confirm with System Display Amplifier Options 21.In the Output Vo1 selection field select the desired signal and confirm with Transducer Conditioning 22.In the Output Vo2 selection field select the desired signal and confirm Display...
Page 133
Measuring Adapting to the transducer 7.3.1.2 Calibrating the characteristic curve of the transducer Transferring the signals supplied by the transducer at a defined loading Example: A calibration weight of 4 kg is used to calibrate a 10 kg load cell. System Display Amplifier...
Page 134
Measuring Adapting to the transducer 11.Use to select the button and confirm with Adjust amplifier measure... 12.Use the cursor keys to select the button and confirm with 13.Unload the transducer. 14.In the left edit field for characteristic curve point 1 enter the value 0 and confirm with 15.Use the cursor key to select the...
Page 135
Measuring Adapting to the transducer The resistance of the SG changes due to strain. Since this change lies in the mΩ and μΩ range, a "Wheatstone bridge" must be used with great accuracy. = Bridge excitation voltage = bridge output voltage + relative change of the excitation voltage ML10B ML30B...
Page 136
Measuring Adapting to the transducer The measurement system always measures the total strain (e) of the indi vidual SGs active in the Wheatstone bridge. Individual SGs are usually used for mechanical stress analysis in practical applications. The half bridge circuit and full bridge circuit are used on account of temper ature compensation, the larger measurement signal, better cable resistance compensation and for some stress states (measuring on a bending beam, etc.).
Page 137
Measuring Adapting to the transducer 7.3.2.1 Direct entry of transducer characteristics Example: Using 120 W individual strain gages, up to 1000 mm/will be mea sured. The gage factor of the strain gage is 2.05. The excitation voltage System Display Amplifier Options plays no role in this example.
Page 138
Measuring Adapting to the transducer to switch to the right edit field Zero point and enter the value 0. 8. Use Confirm with System Display Amplifier Options 9. Use to switch to the right edit field Nominal (rated) value and enter the value 1000.
Page 139
Measuring Adapting to the transducer 7.3.3 Inductive transducers Inductive transducers (for example force transducer from HBM) are passive transducers that have the following features: They must be supplied with an excitation voltage (carrier frequency) They are inductive half or full bridge circuits The transducer is identified by the following data: Nominal (rated) displacement (for example 20 mm) Nominal (rated) sensitivity (for example 10 mV/V)
Page 140
Measuring Adapting to the transducer 7.3.3.1 Direct entry of transducer characteristics The example below illustrates the settings for a displacement transducer: System Display Amplifier Options Nominal (rated) displacement 20 mm, excitation voltage 2.5 V, nominal (rated) value 10 mV/V. 1. Use the channel selection keys or a direct entry to choose the desired Transducer Signal conditioning...
Page 141
Measuring Adapting to the transducer 12.In the left edit field Nominal (rated) value enter the value 20. 13.In the right edit field Nominal (rated) value enter the value 10 (under the unit mV/V). System Display Amplifier Options 14.Use to select the button and confirm with .
Page 142
Measuring Adapting to the transducer 7.3.3.2 Calibrating the characteristic curve of the transducer By transferring the signals supplied by the transducer at a defined deflection Example: A displacement transducer with a nominal (rated) displacement of System Display Amplifier Options 20 mm (nominal (rated) value 10 mV/V) is set with a gaging block of 10 mm, but the measuring range should be 15 mm.
Page 143
Measuring Adapting to the transducer 10.Use to switch to the button and confirm with (the mea measure sured value appears in the right edit field of Zero point) 11.In the left edit field Nominal (rated) value enter the value 20 (under the System Display Amplifier...
Page 144
Measuring Adapting to the transducer 20.Use to select the OK button and confirm with (the amplifier con verts the nominal (rated) value to 20 mm. The calibration data for 10 mm remains intact). System Display Amplifier Options 21.Go back to the pull-up menu with 22.In the pull-up menu select Analog outputs and confirm with Transducer Conditioning...
Page 145
Measuring Adapting to the transducer 7.3.4 Torque transducer Torque transducers with different measurement principles require different amplifier modules in the MGCplus system and therefore different operating steps. The HBM torque transducers in the type series T4A, T5 and TB1A work with è...
Page 146
Measuring Adapting to the transducer 7.3.4.1 Direct entry of torque characteristics The following example explains the operating steps for adjusting the torque, rotational speed and power channel. Explanation of the extended functions è can be found on page 151. System Display Amplifier Options...
Page 147
Measuring Adapting to the transducer 8. In the left edit field Zero point enter the value 0. 9. In the right edit field Zero point enter the value 10 (or if the transducer is measure unloaded activate the button). System Display Amplifier Options...
Page 148
Measuring Adapting to the transducer 7.3.4.2 Calibration with shunt installed Information If the zero point and nominal (rated) torque are not changed (for example in a System Display Amplifier Options recalibration), you can skip steps 1.-9. or a direct entry to choose the desired channel (for example 3.2 1.
Page 149
Measuring Adapting to the transducer 9. Use to select the Adjust amplifier button and confirm with System Display Amplifier Options 10.Use to select the button and confirm with measure... 11.Unload the transducer. 12.In the left edit field for characteristic curve point 1 enter the value 0 and Transducer confirm with Conditioning...
Page 150
Measuring Adapting to the transducer Example: Entry 50.0 N⋅m → Step 1 means display jumps every 0.1N⋅m → Step 5 means display jumps every 0.5N⋅m Entry 50.000 N⋅m → Step 1 means display jumps every 0.001N⋅m → Step 5 means display jumps every 0.005N⋅m 21.Use the Switch key to go to measuring mode and confirm the confir...
Page 151
Measuring Adapting to the transducer "Extended functions" setup window: Glitch filter (ML60B only) System Display Amplifier Options When this filter is turned on interference signals with pulse widths less than 1.6ms are suppressed. Transducer Switch output LV1 (ML60B only) Conditioning Display The frequency signal F1 or the counting signal can be applied to the switch Analog outputs...
Page 152
Measuring Adapting to the transducer 7.3.5 Adjusting the rotational speed channel, frequency measurement The following calculation is required to adjust the torque channel: Nominal (rated) rotational speed: n = 3000 rpm Number of pulses/revolution: i = 360 (see table below, type T10F) + Pulse frequency In this example: 3000 x 360...
Page 153
Measuring Adapting to the transducer 1. Use or a direct entry to choose the desired channel (for example 3.2 System Display Amplifier Options 2. Use the Switch key to switch to Setup mode. Transducer 3. Press the function key. Conditioning Display Analog outputs Switch...
Page 154
Measuring Adapting to the transducer 14.In the Decimal places edit field enter the desired number of decimal places and confirm with 15.In the Step selection field select the desired step and confirm with Information The step refers to the last decimal place of the display value. Example: Entry 1000.0 rpm →...
Page 155
Measuring Adapting to the transducer 7.3.6 Adjusting the rotational speed channel, power measurement 1. Use or a direct entry to choose the desired channel (for example 3.2 System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key.
Page 156
Measuring Adapting to the transducer 12.In the Decimal places edit field enter the desired number of decimal places and confirm with System Display Amplifier Options 13.In the Step selection field select the desired step and confirm with Information Transducer Conditioning The step refers to the last decimal place of the display value.
Page 157
Measuring Adapting to the transducer 7.3.7 Thermocouples Thermocouples are active transducers. To perform measurements with thermocouples you will need the ML801B amplifier module as well as con nection board AP809. The temperature cold junction is built into the AP809. The amplifier module performs cold-spot compensation and linearization for the selected thermocouple type (beginning with hardware revision 3.00).
Page 158
Measuring Adapting to the transducer 7.3.7.1 Direct entry of transducer characteristics or a direct entry to choose the desired channel (for example 3.2 1. Use System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key.
Page 159
Measuring Adapting to the transducer 9. In the Decimal places edit field enter the desired number of decimal places and confirm with 10.In the Step selection field select the desired step and confirm with Information The step refers to the last decimal place of the display value. Example: Entry 10.0°C →...
Page 160
Measuring Adapting to the transducer 7.3.8.1 Direct entry of transducer characteristics 1. Use or a direct entry to choose the desired channel (for example 3.2 System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key.
Page 161
Measuring Adapting to the transducer 15.In the Step selection field select the desired step and confirm with System Display Amplifier Options Information The step refers to the last decimal place of the display value. Transducer Example: Conditioning Display Entry 10.0 Nm Analog outputs Switch →...
Page 162
Measuring Adapting to the transducer 7.3.9 Resistance temperature sensors Resistance temperature sensors are passive transducers. For this trans ducer you will need the ML801B multi-channel module with connection board AP835. It performs linearization automatically and indicates the cor rect temperature. If you selected 'Resistance temperature sensor' as the operating mode and °C or °F as the unit, the corresponding temperature display will be in degrees of the selected unit.
Page 163
Measuring Adapting to the transducer 7.3.9.1 Direct entry of transducer characteristics 1. Use or a direct entry to choose the desired channel (for example 3.2 System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key.
Page 164
Measuring Adapting to the transducer 13.In the Decimal places edit field enter the desired number of decimal places and confirm with System Display Amplifier Options 14.In the Step selection field select the desired step and confirm with Information Transducer Conditioning The step refers to the last decimal place of the display value.
Page 165
Measuring Adapting to the transducer 7.3.10 Resistors Resistors are passive transducers. For these transducers you will need the ML801B amplifier module with AP835, which will provide you with a coarse measuring range (0...500 Ω). The settings are explained with the following example: Resistance value 400 Ω, Display full scale 400.00 Ω...
Page 166
Measuring Adapting to the transducer 11.Go back to the pull-up menu with 12.In the pull-up menu select Display and confirm with 13.In the Decimal places edit field enter the desired number of decimal System Display Amplifier Options places and confirm with 14.In the Step selection field select the desired step and confirm with Transducer Conditioning...
Page 167
Measuring Adapting to the transducer 7.3.11 Pulse counting To count pulses you will need the ML60B amplifier. This amplifier is able to process a maximum pulse string frequency of 1 MHz. Further information è about possible settings of this amplifier can be found in section 7.3.4 "Torque transducer", page 151.
Page 168
Measuring Adapting to the transducer 7.3.11.1 Direct entry of transducer characteristics or a direct entry to choose the desired channel (for example 3.2 1. Use System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key.
Page 169
Measuring Adapting to the transducer 15.Use to select the button and confirm with . If Adjust amplifier you do not want to make any more changes to the display, you can now continue with step 20. 16.Go back to the pull-up menu with 17.In the pull-up menu select Display and confirm with System Display...
Page 170
Measuring Adapting to the transducer "Extended functions" setup window of the ML60B Glitch filter (only with ML60B) System Display Amplifier Options When this filter is turned on interference signals with pulse widths less than 1.6 ms are suppressed. Transducer Frequency quadrupling Conditioning Display Frequency quadrupling is equivalent to increasing the signal resolution.
Page 171
Measuring Adapting to the transducer Counting signal to LV1 with quadruple analysis activated MGCplus...
Page 172
Measuring Current-fed piezoelectric transducer Current-fed piezoelectric transducer Piezoelectric transducers with a built-in preamplifier are often used for ac celeration and force measurements. These transducers require a constant current for supply. The measurement signal corresponds to the modulated voltage on the power line. The DeltaTronT acceleration transducer from Brüel&Kjaer is a typical representative of this product family.
Page 173
Measuring Current-fed piezoelectric transducer 7.4.1 Direct entry of transducer characteristics or a direct entry to choose the desired channel (for example 3.2 1. Use System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key. Conditioning Display Analog outputs...
Page 174
Measuring Current-fed piezoelectric transducer 12.In the pull-up menu select Display and confirm with 13.In the Decimal places edit field enter the desired number of decimal System Display Amplifier Options places and confirm with 14.In the Step selection field select the desired step and confirm with Transducer Conditioning Display...
Page 175
Measuring Piezoresistive transducers Piezoresistive transducers Piezoresistive transducers are passive transducers. The settings are ex plained with the following example: Pressure transducer with characteristics: Nominal (rated) pressure 300 bar, current feed 10 V, rated output (nominal) 200 mV/V, display range 300 bar, test pressure 250 bar (partial load) Due to the rated output you must use the ML10B measuring amplifier.
Page 176
Measuring Piezoresistive transducers 10.In the right edit field Nominal (rated) value enter the value 200. 11.Use to select the button and confirm with . If Adjust amplifier you do not want to make any more changes to the display or output char acteristics, you can now continue with step 16.
Page 177
Measuring Piezoresistive transducers 7.5.1.1 Calibrating the characteristic curve of the transducer Transferring the signals supplied by the transducer at a defined test pres sure. System Display Amplifier Options Information If the zero point and nominal (rated) value are not changed (for example in a recalibration), you can skip steps 1-10.
Page 178
Measuring Piezoresistive transducers 10.Use to select the button and confirm with Adjust amplifier measure... measure... System Display Amplifier Options 11.Use the cursor keys to select the button and confirm with 12.Unload the transducer. Transducer Conditioning 13.In the left edit field for characteristic curve point 1 enter the value 0 and Display Analog outputs confirm with...
Page 179
Measuring Potentiometric transducers Potentiometric transducers Potentiometric transducers are passive transducers that must be supplied with an excitation voltage. To work with a potentiometric transducer you will need the ML10B amplifier with AP01i or AP03i, or ML801B with AP836i. Please note that the maximum resistance value is 5 kΩ. The settings are explained with the following example: The measurement will be performed with a potentiometric displacement transducer with a nominal (rated) displacement of 10mm.
Page 180
Measuring Potentiometric transducers 7.6.1 Direct entry of transducer characteristics or a direct entry to choose the desired channel (for example 3.2 1. Use System Display Amplifier Options 2. Use to switch to Setup mode. Transducer 3. Press the function key. Conditioning Display Analog outputs...
Page 181
Measuring Potentiometric transducers 15.In the Step selection field select the desired step and confirm with 16.Use to go to measuring mode and confirm the confirmation prompt with Information The step refers to the last decimal place of the display value. Example: Entry 10.0mm →...
Page 182
Measuring Potentiometric transducers 7.6.1.1 Calibrating the characteristic curve of the transducer Transfer of the signals supplied by the transducer at a defined deflection. Information System Display Amplifier Options If the zero point and nominal (rated) value are not changed (for example in a recalibration), you can skip steps 1-10.
Page 183
Measuring Potentiometric transducers 10.Use to switch to the right edit field Nominal (rated) value and enter the value 1000. Confirm with System Display Amplifier Options 11.Use to select the button and confirm with measure... measure... 12.Slide the core into the transducer until the display indicates zero. Transducer 13.In the left edit field for characteristic curve point 1 enter the value 0 and Conditioning...
Page 185
Additional functions Remote control (single-channel modules only) ADDITIONAL FUNCTIONS Remote control (single-channel modules only) System Display Amplifier Options 8.1.1 Turning on remote control The remote control contacts are used to control important functions of the Remote contacts amplifiers via digital inputs. These contacts are only active if remote control is turned on There are three ways to turn remote control on or off.
Page 186
Additional functions Remote control (single-channel modules only) 8.1.2 Assigning remote control contacts The assignment of control contacts is freely selectable. They are not as signed in the factory settings. The table below shows the possible functions with a description of the function. Function Level 5V Level 0V...
Page 187
Additional functions Remote control (single-channel modules only) Parameter set selection PSEL1 PSEL2 PSEL4 Selected parameter set Tab. 8.1 0: Remote control contacts not activated; level 5 V 1: Remote control contacts activated; level 0 V; connected to digital ground MGCplus...
Page 188
Additional functions Limit values (single-channel modules only) Limit values (single-channel modules only) To assess dimensional or weight tolerances, or when monitoring forces, pressures etc., it is often necessary to comply with specific setpoint or limit values. Four limit switches are available for this in each amplifier module (they are turned off in the factory settings).
Page 189
Additional functions Limit values (single-channel modules only) 8.2.1 Turning on limit switches The limit switches are not enabled in the factory settings (Enable "Off"). 1. Use the Switch key to switch to Setup mode. System Display Amplifier Options 2. Press 3.
Page 190
Additional functions Limit values (single-channel modules only) 8.2.2 Adjusting limit values LIMIT SWITCHES 1 Channel 1 Setup window for limit switches 1 ... Limit switches 1-LV1 ... Adjusting and turning on limit switches (Enable) Name No ↓ Enable Gross ↓ Input signal Limit switches Level...
Page 191
Additional functions Limit values (single-channel modules only) Switching level Hysteresis Amplifier output signal Limit switch on Limit switch off Fig. 8.1 Switch when the switching level is exceeded Hysteresis Switching level Limit switch on Limit switch off Fig. 8.2 Switch when the switching level is undershot Limit value delay You can enter a Limit value delay time of 0 - 99999 ms.
Page 192
Additional functions Limit values (single-channel modules only) Output logic You can change the output logic of the remote controls as required. Positive logic Negative logic Turned on =High Turned on =Low Turned off =Low Turned off =High Input Enable or disable entry of the limit value level via function keys. Message when On Edit field for a message in the display value when switching on (for example è...
Page 193
Additional functions Limit value combination (single-channel modules only) Limit value combination (single-channel modules only) You can use this function to logically combine selected limit switches with a limit value output. System Display Amplifier Options limit value output Entry for the desired limit value output (---, 1, 2, 3, 4). Combine LIV Combination A logical operation on input signals (AND, OR, EXOR, NAND, NOR, NEXOR).
Page 194
Additional functions Limit value combination (single-channel modules only) Limit switch 1 Channel 1 Example Limit switch Name 1−LV1 Task: The range between 10 kN and 20 kN will be monitored and assessed Yes ↓ Enable as "OK". The assessment will be represented in the display by "OK" or "NOK". Gross ↓...
Page 195
Additional functions Limit value combination (single-channel modules only) COMBINE LIV CHANNEL 1 With the settings in this example, the following display is produced when limit value 1 is undershot (image type "2 Status limit switch"): LIV output AND ↓ Combi.: LIV1 ↓...
Page 196
Additional functions Set peak values Set peak values Measured peak 8.4.1 Peak-value memory value The "Peak values" function can be used to record and save isolated signal peaks and minimum/maximum signal amplitudes. Each amplifier contains two peak-value memories. You can use these to save: 26 μsec Maxima Minima or...
Page 197
Additional functions Set peak values 8.4.2 Combining peak-value memories Combining peak-value memories System Display Amplifier Options Subtraction PV1 - PV2 (used as peak-to-peak). Peak value stores PEAK VALUE STORES (PVS) Channel 1 ↓ Maximum gross Store 1 Function Signal curve Off ↓...
Page 198
Additional functions Set peak values Mean-value calculation ( PV1)PV2 ) System Display Amplifier Options Peak value stores Mean value PEAK VALUE STORES (PVS) Channel 1 ↓ Maximum gross Store 1 Function Off ↓ Store 1 Envelope curve 0 ... Maximum gross ↓ Store 2 Function Fig.
Page 199
Additional functions Set peak values 8.4.3 Control of peak-value memory Three remote control contacts affect peak-value memory: CPV: is used to clear peak-value memory HLD: freezes the current value of memory or releases it again INT: starts and stops integration over a certain time interval You can implement other functions with these remote control elements, for example instantaneous value memory.
Page 200
Additional functions Set peak values 8.4.5 "Instantaneous value" operating mode Measurement signal In Instantaneous value operating mode memory is continuously updated (Run function). Use the "Hold" function to hold the contents of the store. Output You can switch peak-value memory to the Instantaneous value operating mode with the remote control contacts.
Page 201
Additional functions Set peak values 8.4.6 Envelope curve operating mode Output signal Peak-value memories can also be used to realize an envelope curve. The 100% envelope function is suitable for measuring amplitude-modulated vibration. Entering a time constant determines how quickly peak-value memory will discharge to 30% of the peak value again when the peak value is no longer Time constant Input signal...
Page 202
Additional functions Set peak values 8.4.7 Clear peak-value memory Due to the abbreviated notation, the peak-value memories in the setup win dows are referred to as PV1 and PV2. System Display Amplifier Options Clear peak-value memory There are three ways to clear the peak-value memory: Peak value stores 1.
Page 203
Additional functions Version Version You can use the channel selection keys to show information about the ver sion of your device components one after the other. The name of the device component (AB, CP, channel1, etc.) whose version will be displayed appears at the right of the header.
Page 204
Additional functions Switching Switching The Switching setup window contains first of all three buttons you can use to ->T<- Tare System Display Amplifier Options ->0<- Perform a zero balance Clear peak-value memory. Switch In the selection fields underneath you can switch the possible states of functions Autocalibrate, Parameter set, Amplifier input, Remote control and LED display.
Page 205
Additional functions Switching SWITCH CHANNEL 1 Remote control ->T<- ->0<- Turning remote control on/off (activates remote control contacts) Autocal Off ↓ Off ↓ Remote control 1-internal ↓ Param.set LED display Status ↓ Null ↓ Ampl. input LED display Switches the function of the LEDs on the front panel of the amplifier. Status Status display (amplifier active, error, limit values) 1-internal...
Page 207
Display Display format DISPLAY 1 one value Display format Gross Channel The settings affect how the selectable signals in the display are realized. Measure Measure Normally you can select four different signals (gross, net, limit values and 3 values peak values) for each channel. Gross The measured values can be realized as numeric values or as a graph.
Page 208
Display Display format 9.1.1 Select setup window 1. Use the Switch key to switch to Setup mode. 2. Press System Display Amplifier Options 3. In the pull-up menu select Display format and confirm with You will then be in the Display format setup window. Display format F-keys Channel names...
Page 209
Display Display format Setup window Display format 9.1.2 DISPLAY FORMAT One value The layout of the Display format setup window depends on the image type 0 ... Number: selected. The window mask changes depending on the image type selected. one value ↓ Type: For example the Status line selection field is only present with the one value All ↓...
Page 210
Display Display format 9.1.3 Setup window components DISPLAY FORMAT One value Image no. 0 ... Number: You can enter the numbers 1...9 in this edit field. This allows you to save one value ↓ Type: your current display settings as a number or call up factory default settings. All ↓...
Page 211
Display Display format 9.1.3.1 Numeric value display DISPLAY FORMAT 3 values Channels/signals 0 ... Number: You can define here which channels will be realized with which signals in 3 values ↓ Type: the first display line. You can define your settings for all channels or only for Value 1 (base): select ↓...
Page 212
Display Display format Absolute The measured value and the signal of the selected channel are displayed independently of the basic value. The number that is entered corresponds to the actual channel number. Switching channels in measuring mode has no effect on this display value. Relative to basic value The number entered here refers to the basic channel (display value 1).
Page 213
Display Display format Setup window DISPLAY FORMAT 0 ... Number: SELECT CHANNE/SIGNAL 3 values ↓ Type: Cancel All channels All signals Value 1 (base): Channel select ↓ define... 11 12 Channels/signals: Gross Relative to base ↓ Value 2: Channel Store1 Store2 Gross ↓...
Page 214
Display Display format DISPLAY FORMAT 3 values Example1: 6‐channel device; image type 3 values 0 ... Number: The setting in the Channels/signals setting field and the relative channel 3 values ↓ Type: number affect the order of the measured values displayed in measuring Value 1 (base): mode.
Page 215
Display Display format DISPLAY FORMAT 3 values b) Channels/signals: + selection 0 ... Number: 3 values ↓ Type: When the channel selection buttons are activated in measuring mode, Value 1 (base): the selected channels of the basic channel (in this example 1, 3, 5) are acti select↓...
Page 216
Display Display format DISPLAY FORMAT 3 values Example2: The net weights of three containers will be displayed simultane 0 ... Number: ously. The following assignment will apply: 3 values ↓ Type: Container 1 → Channel 1 Value 1 (base): Container 2 → Channel 2 select ↓...
Page 217
Display Display format 15.Use to select edit field Absolute/relative, select Absolute and con firm with 16.Use to select edit field Signal, select Net and confirm with to select edit field Value 3:Channel and enter 3. 17.Use 18.Use to select edit field Absolute/relative, select Absolute and confirm with 19.Use to select edit field Signal, select Net and confirm with...
Page 218
Display Display format Example3: Force and displacement will be measured on each of the two presses. The following assignment will apply: → Channel 1 Force - press 1 → Channel 2 Displacement of press 1 → Channel 3 Force - press 2 →...
Page 219
Display Display format CHANNEL NAMES 7. In edit field Channel 4 enter DISPLACEMENT P2 and confirm with Channel 1: Force P1 Channel 2: Displacement P1 8. Press Channel 3: Force P1 Channel 4: Displacement P2 9. In the pull-up menu select Display format and confirm with DISPLAY FORMAT 3 values 10.In edit field Image no.
Page 220
Display Display format DISPLAY FORMAT 3 values in the Absolute/relative selection field to select Relative and 18.Use Number: 3 values ↓ Type: confirm with Value 1 (base): select ↓ define... Channels/signals: in the Signal selection field to select PV1 and confirm with 19.Use relative ↓...
Page 221
Display Display format 9.1.3.2 Graphic display Image type y-t realization y-t realization This realization allows you to consider the progression of measured values over time. Sampling rate Time interval of signal sampling. Channels/signals Setup window You can define here which channels will be realized with which signals in DISPLAY FORMAT YT-Diagram the display.
Page 222
Display Display format Image type xy realization x‐y realization Sampling rate Time interval of signal sampling. Channels/signals You can define here which channels will be realized with which signals in the display. You can define your settings for all channels or only for certain define...
Page 223
Display Display format Setting menu DISPLAY FORMAT XY-Diagram Maximum value of the vertical axis relative to the current measuring range 0 ... Number: (%). XY-Diagram ↓ Type: Measuring interval: 1s↓ Value 1 (base): All↓ define... Channels/signals: Minimum value of the vertical axis relative to the current measuring range 100...
Page 224
Display Display format 9.1.4 Status limit switch Image type: Status limit switch Title Title Any user-specific name; appears in the header (factory setting "limit value"). LIMIT VALUE 1−ML30 Gross 8.483 kN Display value 1 (base): Channel 1−LV1 Below 10kN You can define here which channels will be displayed with which signals in 1−LV2 Below 20kN the first status line.
Page 225
Display Display format 9.1.5 Data acquisition Data acquisition comments Bar display Bar display This realization shows the current status of the data acquisition. Test series 3 SAMPLING RATE: 50Hz TIME: 00: 00: 00 PERIODS: 100 Data acquisition comments −4.0S 16.0S User-specific comment (for example the test series number).
Page 226
Display F keys F keys F keys level 1 9.2.1 F keys in measuring mode Gross Channel Function keys F1...F4 are active both in measuring mode and in Setup mode. Measure You can switch a total of 9 functions on three levels in measuring mode: −>...
Page 227
Display F keys Function Effect Tare Taring is triggered. Zero balance A zero balance is being performed. Delete memory Peak-value memory is deleted. Zero/Cal/Measure Switching between zero signal, calibration signal and measurement signal. Autocal Autocalibration is turned on/off. LED Status/Level Switching between LED display "Status"...
Page 228
Display F keys Setup window Select setup window F-KEYS SETUP Level 1 1. Use the Switch key to switch to Setup mode. 1 ... level F1: Tare ↓ F2: Zero/Cal/Measure↓ One channel 2. Press F3: Autocal ↓ One channel ↓ One channel F4: ...F-level 3.
Page 229
Display Channel names Channel names Setup window Select setup window CHANNEL NAME Channel 3.1 1. Use the Switch key to switch to Setup mode. Force ... 2. Press Channel name: Display Amplifier Options 3. In the pull-up menu select Channel names and confirm with All channels present in the setup window have the channel names assigned in the factory at first.
Page 231
System Password SYSTEM 10.1 Password You can protect all the settings of your device with a password. This password protection is turned off in the factory settings. As soon as the password protection is activated, the password must be entered every time Setup mode is called (and every time the device is turned on again).
Page 232
System Password 10.1.1 Define new user 1. Use the Switch key to switch to Setup mode System Display Amplifier Options 2. Press 3. In the pull-up menu select Password and confirm with Password Save/load 4. Use to select the new... button symbol and confirm with Data acquisition Interface 5.
Page 233
System Password 10.1.2 Password protection activation PASSWORD If you are still in the Password setup window, continue with item 4. new... delete... User: 1. Use the Switch key to switch to Setup mode. change... Password: set... Access: 2. Press 3. In the pull-up menu select Password and confirm with 4.
Page 234
System Password 10.1.3 Set access for operator PASSWORD If you are still in the Password setup window, continue with item 4. new... delete... User: 1. Use the Switch key to switch to Setup mode change... Password: set... Access: 2. Press 3.
Page 235
System Password 10.1.4 Delete user If you are still in the Password setup window, continue with item 4. 1. Use the Switch key to switch to Setup mode. System Display Amplifier Options 2. Press Password 3. In the pull-up menu select Password and confirm with Save/load Data acquisition You will then be in the Passwordsetup window.
Page 236
System Password 10.1.5 Change password If you are still in the Password setup window, continue with item 4. 1. Use the Switch key to switch to Setup mode System Display Amplifier Options 2. Press Password 3. In the pull-up menu select Password and confirm with Save/Load Data acquisition You will then be in the Passwordsetup window.
Page 237
System Save/load 10.2 Save/load You can use the Save/load function to permanently save the current settings of the AB22A, the CP42/CP52 or the amplifier modules (up to 8 parameter sets per channel) or to load previously saved settings. You can System Display Amplifier...
Page 238
System Save/load Save parameters 1. Use the Switch key to switch to Setup mode 2. Press 3. In the pull‐up menu select Save/ and confirm with You will then be in the Save/ settings setup window. 4. Use to select the Save button and confirm with 5.
Page 239
System Save/load Load parameters 1. Use the Switch key to switch to Setup mode. System Display Amplifier Options 2. Press 3. In the pull‐up menu select Save/load and confirm with Password You will then be in the Save/load settings setup window. Save/Load Data acquisition 4.
Page 240
System Save/load Factory settings A new window opens under Factory settings where you can define whether all or only certain amplifiers should be set to factory settings. You can also load the factory settings of the display and control unit (AB) and the System Display Amplifier...
Page 241
System Save/load Copy amplifier All or only certain settings can be transferred from one channel to another (or to multiple channels) as defined by the user. Use the Copy button to do this. System Display Amplifier Options 1. Use the Switch key to switch to Setup mode.
Page 242
System Recording series of tests 10.3 Recording series of tests USB mass storage With the MGCplus you can configure up to 17 programs for recording series of tests (16 on the hard disk) and save the configuration. The recording is saved in the RAM of the communication processor or to in storage medium (optional).
Page 243
System Recording series of tests 10.3.1 Setting parameters of test series Select setup window 1. Use the Switch key to switch to Setup mode. System Display Amplifier Options 2. Press the function key, use the cursor keys to select Data acqui sition and confirm with Data acquisition Measuring interval/sampling rate...
Page 244
System Recording series of tests Example: 50 measured values 1sec 1000 measured values 20sec 20sec 20sec Measurement period 1 Measurement period 2 MGCplus...
Page 245
System Recording series of tests Time channel If the time channel is selected ( ), two additional time columns appear in the data acquisition files. System Display Amplifier Options Example: You selected: Channel 2, Gross; Channel 4 Net, PV1 and Time channel. Password Save/Load Data acquisition information file:...
Page 246
System Recording series of tests Trigger function DATA ACQUISITION PARAMETERS To use the trigger function you must first define the trigger conditions. ↓ Measuring rate/interval: 300Hz There are three trigger types available: ↓ Values: 1200 ... Time: s... 1. Start trigger (with maximum four trigger conditions) Periods: 1 ...
Page 247
System Recording series of tests Start Trigger 1 è In the example on page 246 the following could have been defined as a Start Trigger 2 start condition: Start Trigger 3 Start Trigger 4 Stop Trigger 1 Start measurement when force greater than 5 kN Stop Trigger 2 Stop Trigger 3 Stop Trigger 4...
Page 248
System Recording series of tests Mode Start Trigger 1 Configure start/stop conditions. Start Trigger 2 Start Trigger 3 Off: Trigger inactive Start Trigger 4 Stop Trigger 1 Measurement level: The trigger condition is triggered when the Stop Trigger 2 Stop Trigger 3 measurement signal exceeds/undershoots the level value.
Page 249
System Recording series of tests Start Trigger 1 Measured value band: The trigger condition is triggered when the Start Trigger 2 measurement signal exceeds/undershoots the band limits. The band lies Start Trigger 3 Start Trigger 4 between "Level 1" and "Level 2". Stop Trigger 1 Stop Trigger 2 Example:...
Page 250
System Recording series of tests Start Trigger 1 LV1-LV4: The trigger is in response to a limit value (High: LED on, data Start Trigger 2 acquisition program is started; Low: LED off) Start Trigger 3 Start Trigger 4 Stop Trigger 1 External trigger: The trigger is in response to an external signal (remote Stop Trigger 2 control 7 on the connection board).
Page 251
System Recording series of tests Mode Gross You can use this mode to define when a trigger condition is met. You can set different modes depending on which trigger type is set: Type Mode Measured value level higher; lower; lower higher;...
Page 252
System Recording series of tests Pre‐Trigger You can define in the Pre‐trigger input box how many measured values to System Display Amplifier Options record before the trigger event (measurement acquisition is divided into the pre‐trigger range and the post‐trigger range ). The definition refers to the "Number"...
Page 253
System Recording series of tests Stop condition Number of measured values: The measurement ends when the number of System Display Amplifier Options measured values specified (entered under "Number") is reached. AND‐combined: All stop conditions must be met to stop Password recording data.
Page 254
System Recording series of tests Information Measured value files larger than 2 Gbyte cannot be processed on computers with Windows operating systems. The communication processor will therefore close the file if it has a size of 2 Gbyte and will continue recording in System Display Amplifier...
Page 255
System Recording series of tests Compression factor When you activate data compression ( ), an additional data set is created on the storage medium while measured values are being recorded. In addition to all measured values, extreme values (Min/Max) are also saved in System Display Amplifier...
Page 256
System Recording series of tests . . . Recording format Reading a recorded file To be able to read and interpret the data in the file, you must be familiar with the internal format of the file. The data is recorded in a binary file structured as follows: The following header data items come before the actual data area: File ID (4‐byte LONG) // currently = 6001...
Page 257
System Recording series of tests This channel block is followed by information about the date and time: TimeDate (30‐byte CHARACTER) The bytes that follow to the end of the header are reserved for future use and may be left out. They are followed by the values, data line for data line (the example shown here is based on bit 0 (Gross) and bit 2 (peak value 1) being set): C1, 1 (Gross)
Page 258
System Recording series of tests Information The 1253 LONG format contains status information in the least significant byte. To access this status, you should check this byte first (!) Scale, since this will cause the least significant byte to be lost due to division by 256. The status byte is structured as follows: Bit 0: Limit value output 1 active if set Bit 1: Limit value output 2 active if set...
Page 259
System Recording series of tests Autostart System Display Amplifier Options Information You can save all the settings (start/stop conditions, etc.) in a recording parameter set and reload it. To use a parameter set, you must assign two Password function keys: Save/Load Data acquisition Interface...
Page 260
System Recording series of tests Save The recording parameters are saved in the selected parameter set. System Display Amplifier Options You can save up to 16 measured value recording programs on the hard disk of the MGCplus. Password Load Save/Load Data acquisition A recording parameter set - saved under a ParaSetNo.
Page 261
System Recording series of tests 10.3.2 Format of the MGCplus measurement files When measurements are recorded on a storage medium, files with the extensions *.mea, *.me1 and *.me2 are created. All the channels recorded at sampling rate no. 1 are stored in the file with the extension mea, those at measuring rate no.
Page 262
System Recording series of tests Byte 1 Byte 2 Byte 0 Byte 3 Status information Measured value (24 bits) Bit no. This status information is displayed as an integer in the range s {0 ... 255}. When measured values are imported from an MEA file (4-byte INT format) the status is checked to see whether it is s>15.
Page 263
System Recording series of tests and c represent the scaling information which is contained in the scale offset relevant amplifier and also in the measurement files on the PC card hard disk. 10.3.2.2 Time channels The "MGCplus Device Time" time channel (called the "HBM" time channel in the AB22A) is used to count 76.8 kHz periods.
Page 264
System Recording series of tests 10.3.3 MEA format in detail (MGC binary format 2) MGCplus measurement files consist of a file header, a section with channel parameters for all channels, a section with information on the amplifier and a block with the actual measurement data (not shown in this table). Header Offset Type...
Page 265
System Recording series of tests Offset Type Name Content Offset+0 char chn_code channel_code Offset+1 char amp_code amplifier_code Offset+2 char para_flg parameterflag (1..8: internal, 9: XM001) Offset+3 char tab_quelle where's the table from ? Offset+4 char acal_flg autocal used ? Offset+5 char loca_flg local/remote used ?
Page 266
System Recording series of tests Offset Type Name Content Offset+64 char unit_range[4] String for Unit at Range setting Offset+68 char ind1point indication adaptation dec. point (range 1) Offset+69 char ind1step indication adaptation increment (range 1) Offset+70 char unit_txt1[4] default display units (range 1) Offset+74 char unit_anz[4]...
Page 267
System Recording series of tests Offset Type Name Content Offset+145 char remote4 Offset+146 char remote5 Offset+147 char remote6 Offset+148 char remote7 Offset+149 char remote8 Offset+150 char group_kenn group assignment Offset+151 char ver_flg Offset+152 ver_sig1 Offset+154 ver_sig2 Offset+156 ver_sig3 Offset+158 ver_sig4 Offset+160 char ver_type...
Page 268
System Recording series of tests Slot specific data Offset Type Content Offset+0 INT32 Number of used slots For each used slot static and dynamic slotparameter Offset Type Content Offset+0 INT32 Slotnumber (Channel number) Offset+4 INT32 No. of Bytes of the Block with static slotparameter Offset+8 INT8[15] serial number of amplifier...
Page 269
System Interface 10.4 Interface You can use the Interface function to adjust the characteristics of the interfaces: Device address Subnet mask DHCP Yes or No Baud rate Format (word length, parity, stop bit) The following interfaces are available with the CP42 and CP52 communication processor: CP52: 2 x independent Ethernet (DHCP capability), upper interface (X4) addi...
Page 270
System Interface If you access the MGCplus measuring amplifier system from your PC through a router, you must specify the address of the router in MGCplus. System Display Amplifier Options Please note that the CP52’s two Ethernet interfaces are not in the same sub net.
Page 271
System Interface Direct Ethernet connection between PC and MGCplus with permanent IP addresses For example, if your computer has the IP address 172.34.24.13, your MGCplus should have IP address 172.34.24.x (x≠13, x≠255) if you are using a subnet mask 255.255.255.0. If you connect the MGCplus to a PC via a hub, use a patch cable.
Page 272
System Interface Router IP address Subnet mask Router A address Computer Computer Ethernet Router A Worldwide Computer Computer Computer Ethernet Router B If data packets are sent via node points, for example from the company computer to the Internet or a WAN / LAN, routers are needed to distribute the data packets in different directions based on their address.
Page 273
System Interface 10.4.1 Port usage Certain Ethernet ports must be shared for communication between PC and The following TCP ports must be shared for CP52: Port 7 for communication with the device Port 80 for the web server and firmware updates The following UDP ports must also be shared to be able to find CP52 with the HBM Device Manager or change interface settings: Port 31,416 to find CP52...
Page 274
System Interface While the measured values of one client are being recorded, none of the other clients can make changes to the recording of measured data. External clients can only reset the time channel. All the other measured data recording parameters, such as sampling rate groups, triggers, channel selection, etc., are divided between the client processes.
Page 275
System Language 10.5 Language You can use this function to select the language in the display, menu and texts. System Display Amplifier Options Password Save/Load Data acquisition Interface Synchronization Language Time Auto assignment LANGUAGE ↓ Language: Deutsch Deutsch English Francais MGCplus...
Page 276
System Time 10.6 Time You can use this function to set the date, day of the week and time. System Display Amplifier Options Password Save/Load Data acquisition Interface Synchronization Language Time Auto assignment DATE/TIME SET−UP ↓ Day ... Month Year ... Date: ↓...
Page 277
Menu structure MENU STRUCTURE The menu structure shown below will help you find the required setting menus more rapidly. The necessary key sequence is also shown here. Structural elements F‐keys assignment in measuring mode (original state) Key required to go to the first drop-down menu Selection bar The lines of the pull‐up menu are...
Page 278
Menu structure Symbols mV/V ↓ Selection field 0.000000... Edit field measure Button (triggers action) measure ... Button (opens new setup window) Activation fields Cursor keys Direction arrows, pointing in the working direction of the keys MGCplus...
Page 279
Menu structure System Display Amplifier Options Password PASSWORD new... delete... User: change... Password: set... Access: Access rights for operator Off ↓ Password protection System Password Save/load Add user Data acquisition Change password Interfaces User: Synchronization New password: Delete user Password: Language Miller delete...
Page 282
Menu structure System Display Amplifier Options Data acquisition DATA ACQUISITION PARAMETERS Measuring rate/interval: ↓ ↓ ↓ 1: 2400 Hz 2: 2400 Hz 3: 100 Hz Values: 1000 ... Time: 20.00 ↓ s... 10 ... Periods: Time Channels/signals: define... stamd: Trigger: define...
Page 283
Menu structure CP52 System Display Amplifier Options Interface Interface settings Ethernet 2 ... Ethernet 1… Info Info RS232… Ethernet 1 settings Information on Ethernet 2 settings AB disabled in computer mode Address: 172.20.14.182 ... Ipv4 act: Subnet mask: 255.255.0.0 ... Mask act: 0.0.0.0 ...
Page 284
Menu structure CP42 System Display Amplifier Options Interface Interface settings IEEE 488 settings Address: RS232 ... Ethernet ... NI PCMCIA − GPIB Card type USB ... IEC ... Cancel AB disabled in computer mode Ethernet settings IEEE488USB settings Address: 172.20.14.182 ... Address: CP42p 400000 ...
Page 285
Menu structure CP52 System Display Amplifierer Options oder Synchronization Synchronization Status Status: PTP Sync: ok Offset: 0,000171 ms ↓ Interface: Domain: ↓ Transport: UDPv4 PTP OFF ↓ Delay: end 2 end Eth1 Ipv6 Multicast: UDPv4 UDPv6 Server: IEEE802.3 end 2 end peer 2 peer Auto MGCplus...
Page 286
Menu structure CP42 System Display Amplifier Options oder Synchronization Synchronization Status Status: MASTER Server: MGCplus...
Page 287
Menu structure System Display Amplifier Options Language Time LANGUAGE DATE/TIME SET−UP ↓ Language: German ↓ Day ... Month Year ... Date: ↓ Friday Day: Hour ... Min ... Sec ... Time Deutsch English Francais MGCplus...
Page 288
Menu structure System Display Amplifier Options Channel names Display format F−keys 0...9 DISPLAY FORMAT One value F-KEYS SETUP Level 1 0 ... Number: CHANNEL NAMES Channel 1 level 1 ... One value ↓ Type: Free F1: Tare ↓ All ↓ define...
Page 289
Menu structure System Display Amplifier Options Transducer Signal conditioning Display DISPLAY CHANNEL1 CONDITIONING CHANNEL 1 TRANSDUCER CHANNEL1 V ↓ 0.0000 ... Unit Type: Zero reference: SG full bridge −>0<− Decimal places Excitation: 5V↓ Zero offset: 0.0000 ... kg ↓ −>T<− Display range from −10.000 V...
Page 290
Menu structure System Display Amplifier Options Analog outputs Switching ANALOG OUTPUTS CHANNEL1 SWITCH Channel 1 −>T<− −>0<− Gross ↓ Net ↓ Output Vo1: Output Vo2: Output characteristics Off ↓ Off ↓ Autocal Remote Point1.: 0.0000 0.0000 1-internal modified ↓ Param.set Point 2.: 100.0000 0.0000...
Page 291
Menu structure System Display Amplifier Options Remote contacts Limit switches Combine LIV REMOTE CONTACTS Channel 1 COMBINE LIV Channel 1 LIMIT SWITCHES 1 Channel 1 On ↓ Remote control 1 ..Limit switch −−− ↓ LIV output: Input 1 ↓...
Page 292
Menu structure System Display Amplifier Options Peak value stores Version PEAK VALUE STORES (PVS) Channel 1 VERSION Channel 1 ↓ Maximum gross Store 1 Function Identification: HBM,RD002−ML55,Y,P4.10 Off ↓ Comments: Store 1 Envelope curve 0.000... s Store 2 Function Maximum gross Serial and revision number...
Page 293
Index INDEX AB22A, Control elements, 97 Call menus, 106 AB22A/AB32 CANHEAD, connect, 79 Display, Measuring mode, 99 Channel selection in measuring mode, 108 Drop-down menus, 110 Channel selection in setting mode, 109 Menus, 106 Communication card ML74B, 80 Save setting, 109 Compression factor, 255 Setting mode, 103 Conditions at the place of installation, 28...
Page 294
Index Data acquisition, 100, 207, 225 Glitch filter, 151, 170 Data acquisition comments, 225 Grounding switch, 31 Data acquisition parameters, 246, 252, 253, 254, 255, Guides, 16 259, 260, 281, 282 Data acquisition program, 248 Data storage, 109 Housing, MGCplus, 20 DC current source, connect, 67 Hysteresis, 190 DC voltage source, connect, 60...
Page 295
Index Measuring with Recording file, 253 Inductive transducers, 139 - 144 Recording format, 254 - 258 piezoresistive transducers, 172 - 178 Recording parameter set, 259 Strain gages, 134 - 138 Recording parameters, 260 Measurment periods, 243 Recording program, 246 Menu, Pull-up menu, 106 Recording series of tests, 242 Menu structure, 277 Remote controls, 82...
Need help?
Do you have a question about the MCGplus and is the answer not in the manual?
Questions and answers