Page 2
Chapter 1 Product Introduction 1.1 Product overview Ultrasonic anemometer is a wind speed and direction measuring instrument developed based on the principle of ultrasonic waves. It uses the transmitted sound pulses to measure the time or frequency difference at the receiving end to calculate wind speed and direction.
Page 3
accuracy Wind direction 0~360° measurement range ±3° Wind direction measurement accuracy Response time Less than 5 seconds 9600 Baud rate RS485/GPRS/4G/4-20mA Communicati on port 12VDC Power supply Working 70mA current Operating -30~80℃ temperature Working 0~100%RH(15~95%RH) humidity environment 1.3 4G parameters Parameters Parameters content name...
Page 4
1.4 Features ◆ No start-up wind speed limit, zero wind speed work, no angle limit, 360 ° full Azimuth, wind speed and wind direction data can be obtained at the same time; ◆ No moving parts, low wear and long service life; ◆...
Page 5
High precision 1set sensor 2G module sensor 1set (Optional) USB to 485 device 1set (Optional) waterproof 1set power supply (Optional) Warranty 1 serving card/certificate 2.2 485 wiring method The power interface is a wide voltage power input of 12V. 第4页...
Page 6
When wiring the 485 signal line, pay attention to the two wires A\B not to be reversed, and the address between multiple set devices on the bus cannot be conflicted. Thre Description colo Power is positive blac Power negative supp yell 485A icati...
Page 7
Precautions: (1) Please be careful not to connect the wrong wiring sequence, the wrong wiring will cause the equipment to burn (2) The voltage/current positive output is an active output. Never connect the voltage/current positive output to the positive position of the power supply, which will definitely cause burnout.
Page 8
2.4Installation method Flange installation is adopted. The threaded flange connection makes the lower pipe fitting of the wind direction sensor firmly fixed on the flange. Four mounting holes are opened on the circumference of the chassis, and bolts are used to tightly fix it on the bracket to keep the entire instrument on The best levelness ensures the accuracy of the wind direction data, the flange connection is convenient to use and can...
Page 9
the top surface, which should be facing the north direction. Chapter 3 485 Communication Protocol 3.1Basic communication parameters parameters content coding 8-bit binary Data bit 8-bit 无 Parity bit Stop bit 1-bit CRC lengthy cyclic code Wrong calibration 2400bps/4800bps/9600 bps can be set, the Baud rate factory default is 9600bps coding...
Page 10
memory data). Data area: The data area is the specific query data area, pay attention to the 16bits data high byte first CRC code: two-byte check code. Interrogation frame addre unctio Register Register Check Check code n code address length code high 址...
Page 11
3.4 Communication protocol example explanation 3.4.1 Read the wind speed value of device address 0x01 Interrogation frame Check Check addres unctio initial Data length code low code s code n code address high 0x01 0x03 0x00,0x01 0x00,0x01 0xD5 0xCA Reply frame Check uncti Wind...
Page 12
Reply frame Check addre uncti Wind Check Effective code direction code bytes High code code value position 0x01 0x03 0x02 0x00 0x8A 0xFD 0xA0 wind direction: 008A H (hexadecimal) = 138 => wind direction = south-southeast The wind direction sensor output value corresponds to the wind direction position Chapter 4 Analog Communication The analog sensor wiring is simple, only need to connect...
Page 13
the wire to the designated port of the device. The following figure shows the wiring method of the current sensor. When measuring, first adjust the multimeter to the "DC mA" position, and then connect the sensor's power cord (brown wire and black wire) to the power supply; the sensor's yellow ( The gray) colored wire means that the signal is connected to the red lead of the multimeter, and the other end of the black lead of the multimeter is connected to the black wire...
Page 14
Current value variable 20mA The formula is wind speed=(I-4)*3.75 The unit of I is mA. 4mA represents 0 point and 20mA represents the maximum range linear conversion. wind direction: Current value variable 20mA 360° The formula is wind direction=(I-4)*22.5 The unit of I is mA; 4mA represents 0 point and 20mA represents the maximum range linear conversion.
Page 15
(2) In the WeChat applet, scan the QR code ID of the device and add the device to the account to view the uploaded data of the device. If you use a third-party product that is not Jingxuntong. Data can be uploaded to our cloud server according to the report docking agreement of our cloud server.
Page 16
Figure Jingxun Cloud Management Interface The functions of Jingxun cloud server include: data center, equipment center, camera management, trigger management, monitoring large screen, system language selection and setting and other functions. "Data Center -> Map Display" can view the device binding location and device name in the form of a map.
Page 17
by the device at any time. "Device Center -> Device Management" can search device ID to view device data; click "Add Device" in the upper left corner to add devices, as shown in the following figure: Figure Jingxun Cloud Device Adding Interface Follow the prompts on the interface to add devices.
Page 18
(5) "Device Address" click the blue coordinate arrow on the right to enter the map, and then drag a blue arrow inside the map to locate the coordinates. After the coordinates are determined, click "confirm", and finally click " Add" to complete the addition of the device.
Page 19
and negative wiring of the product; and read the product manual or consult our company. Any error in the wiring will cause irreversible damage to the transmitter. 第18页...
Need help?
Do you have a question about the JXBS-3001-FSFX and is the answer not in the manual?
Questions and answers