1. Connect the case to a PC by USB cable, and use the WatchGas Suite PC software to
configure the calibration gas concentrations, alarm settings and other parameters. Be
sure to set the sensor type and gas concentration being used.
2. Connect a calibration gas cylinder to the Gas Inlet (2) using a demand-flow regulator
and Last-O-More hose. Insert bare 6 mm o.d. tubing directly into the beige inlet
quickconnect. To disconnect, push the quick connect in towards the box and pull the
tubing out.
CALIBRATION
1. Turn on the WatchGas UNI and place the instrument face-down into the cradle.
2. If the STATUS LED is off, press the CAL button until the LED turns green.
3. Push CAL to initiate calibration. The Unit LED should blink green during the calibration.
4. If the calibration is successful, Unit LED will be green, otherwise red.
5. The calibration report will be saved in the internal storage.
6. To cancel the calibration, hold the CAL button until the unit LED turns off.
7. To power off, hold the CAL button until the status LED turns off.
WatchGas | Klaverbaan 121 | 2908 KD Capelle aan den IJssel | The Netherlands | www.watchgas.com | info@watchgas.com
WatchGas | Klaverbaan 121 | 2908 KD Capelle aan den IJssel | The Netherlands | www.watchgas.com | info@watchgas.com
UNI MONO DOCK
Quick Start Guide

Advertisement

loading
Need help?

Need help?

Do you have a question about the UNI MONO DOCK and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for WatchGas UNI MONO DOCK

  • Page 1 Quick Start Guide SETUP AND INSTALLATION 1. Connect the case to a PC by USB cable, and use the WatchGas Suite PC software to configure the calibration gas concentrations, alarm settings and other parameters. Be sure to set the sensor type and gas concentration being used.
  • Page 2 Right click on a log to generate a certificate. WARNING Follow adequate training and read instructions before use and maintenance of this device! Visit www.WatchGas.com for more info. WatchGas | Klaverbaan 121 | 2908 KD Capelle aan den IJssel | The Netherlands | www.watchgas.com | info@watchgas.com...