Unico IP750 Series Operation Manual

Infinity microscope

Advertisement

Quick Links

IP755 Phase Contrast
INFINITY MICROSCOPE
OPERATION MANUAL
United Products & Instruments, Inc.
182 Ridge Road, Suite E, Dayton, NJ 08810, USA
Tel: 732-274-1155 Fax: 732-274-1151
Toll Free 800-588-9776
http://www.unicosci.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the IP750 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Unico IP750 Series

  • Page 1 IP755 Phase Contrast INFINITY MICROSCOPE OPERATION MANUAL United Products & Instruments, Inc. 182 Ridge Road, Suite E, Dayton, NJ 08810, USA Tel: 732-274-1155 Fax: 732-274-1151 Toll Free 800-588-9776 http://www.unicosci.com...
  • Page 2 Phase contrast objectives Phase condenser set at 40X Below are instructions to be used with UNICO® IP750 Series Microscope User’s Manual. For all information related to microscope setup, operation and maintenance please refer to the user’s manual that comes with the microscope.
  • Page 3 If you have purchased the complete Model IP755 phase microscope, the installation has been done at the UNICO facility. Just make sure the condenser is securely fastened and is raised to the highest position. You may need to do alignment checks. Please refer to Section 4 for “Condenser Alignment Check”...
  • Page 4 IP755 Phase Condenser and Phase Objectives Installed If you have purchased the non-phase model of the IP750 series microscope with bright field objectives and Abbe condenser and have just received the phase kit to convert it to a phase microscope, please follow the procedures listed below to set it up: 3.1.1 Lower the Abbe condenser by turning the substage knob located on the...
  • Page 5 Note: When changing the condenser settings, please use TWO hands to avoid pushing the condenser out of alignment. You must hold the turret condenser steady with one hand, while changing the setting with the other hand. This will prevent you from having to re-center after each objective/condenser switch.
  • Page 6 5. Phase Annulus Ring Alignment Check Note: 1. This must be checked after the condenser alignment has been checked and aligned. 2. Centering telescope is required. 5.1.1 Turn 10X phase objective into working position. Put a slide on the mechanical stage. Rotate the phase condenser disc to “BF” position. After bringing the slide into focus, remove the slide.
  • Page 7 5.2.4 Look through the centering telescope. Rotate the rods until the dark phase ring overlaps with the bright annulus ring. (Minor adjustments may only be required. Turn each rod slowly and see the direction the ring is moving. This can avoid moving the annulus ring to wrong direction). Release the rods when the alignment is done.
  • Page 8 Adjustment on Binocular head left eyepiece tube Reverse position Quadruple nosepiece Objectives Slide Holder Mechanical Stage Condenser focusing knob Condenser Coaxial Coarse and Fine Focus Knobs Koehler Illumination Illumination Diaphragm intensity rheostat adjustment wheel ® UNICO IP750 Series Compound Microscope...