  • Page 1 Senator S80L Handbook 80 GHz Radar (FMCW) Level Transmitter for liquids in narrow tanks with internal obstructions © HAWK 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 2 IMPRINT ::::::::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of Hawk Measurement Systems Pty Ltd. Copyright 2023 by Hawk Measurement Systems Pty Ltd www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 3: Table Of Contents

    3.9 How to turn or remove the display module (option) ............36 3.10 Weather protection....................... 37 3.10.1 How to attach the weather protection to the device............37 3.10.2 How to open the weather protection .................. 39 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 4 6.4.7 How to measure correctly in tanks with curved or conical bottoms ......... 104 6.4.8 How to make a filter to remove radar signal interference ..........105 6.5 Status messages and diagnostic data................106 www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 5 9.7 Field Device Tool / Device Type Manager (FDT / DTM) ..........149 9.7.1 Installation .......................... 149 9.7.2 Operation..........................149 9.8 Process Device Manager (PDM)................... 149 9.8.1 Installation .......................... 149 9.8.2 Operation..........................149 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 6 9.10.2 PDM menu tree (details for settings) ................155 10 Appendix 10.1 Accessories: general items..................158 10.2 Accessories: process connection kits for the level transmitter........ 159 10.3 Accessories: purging system adaptor................ 160 10.4 Glossary ........................161 www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 7: Safety Instructions

    (LCD display option) microcontroller Main and Support 4002815711_ Sensor 4004742604_ Table 1-2: Software history 1 If the device does not have the display module option, the module reference number is 4002905802a 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 8: Intended Use

    • Electromagnetic Compatibility Directive, • the safety part of the Low Voltage Directive and • ATEX Directive (Ex-approved devices) • NAMUR Recommendations NE 21, NE 43, NE 53 and NE 107 www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 9: Radio Approvals

    2 HVIN (Hardware Version Identification Number). This number gives the radar signal frequency (80GHZ = 80 GHz), the location of the device (T=TLPR or L=LPR) and the type of signal converter (compact (C)) TLPR device: HVIN: 80GHZ-T-C LPR device: HVIN: 80GHZ-L-C 3 CE sign 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 10 • installation and maintenance of the TLPR equipment shall be performed by professionally trained individuals only. For data about how to install EMI/RFI shielding gaskets, refer to the instructions supplied with this accessory. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 11 40°31'27" N 03°05'22" W Pico Veleta 37°03'58" N 03°23'34" W Sweden Onsala 57°23’45" N 11°55’35" E Table 1-4: Radio quiet zones: locations of radio astronomy sites (stations) in Europe and northern Eurasia 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 12: U.s.a

    • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. • Consult the dealer or an experienced radio/TV technician for help. The Product Marketing Name (PMN) of this device is "Senator S80x Radar series". www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 13 (T=TLPR or L=LPR) and the type of signal converter (compact (C)) TLPR device: HVIN: 80GHZ-T-C LPR device: HVIN: 80GHZ-L-C 3 FCC ID TLPR device: FCC-ID: Q6BFMCW80G74TA LPR device: FCC-ID: Q6BFMCW80G74LA 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 14: Canada

    The Director of the DRAO may be contacted at 250-497-2300 (tel.) or 250-497-2355 (fax). Alternatively, the Manager, Regulatory Standards, Industry Canada, may be contacted. The Product Marketing Name (PMN) of this device is "Senator S80x Radar series". www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 15 (T=TLPR or L=LPR) and the type of signal converter (compact (C)) TLPR device: HVIN: 80GHZ-T-C LPR device: HVIN: 80GHZ-L-C 3 IC number TLPR device: 1991D-FMCW80GX5T LPR device: 1991D-FMCW80GX5L 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 16: Safety Instructions From The Manufacturer

    The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 17: Product Liability And Warranty

    This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 18: Warnings And Symbols Used

    In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 19: Device Description

    2 Quick Start and supplementary instructions (if the device has the appropriate options) 3 Bar magnet 4 Display extractor (for removal of the optional display module) 5 Cover wrench (for removal of the device covers) 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 20: Device Description

    For more data about general items, refer to Accessories: general items on page 158 . For more data about process connections, refer to Accessories: process connection kits for the level transmitter on page 159 . www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 21: Visual Check

    Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate. INFORMATION! Compare the material references on the side of the process connection with the order. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 22: Nameplate (Example)

    13 WARNING! Hot surface. If the device is connected to tank that operates at high temperature, there is a risk of injury. 14 Communication protocol option (HART® etc.) www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 23: Installation

    3 Storage temperature range: -40...+85°C / -40...+185°F • Store the device in a dry and dust-free location. • Keep the converter out of the sunlight. • Store the device in its original packing. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 24: Transport

    • Protect the signal converter from direct sunlight. If necessary, install the weather protection accessory. • Do not subject the signal converter to heavy vibrations. The devices are tested for vibration and agree with IEC 60068-2-6. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 25: Pressure And Temperature Ranges

    1 If the process connection temperature is more than +150°C / +302°F, the device has a distance piece. For more data about the overall dimensions of the device, refer to the "Dimensions and weights section". For more data on pressure ratings, refer to Guidelines for maximum operating pressure on page 135. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 26: Recommended Mounting Position

    If there is a nozzle on the tank before installation, the nozzle must be a minimum of 200 mm / 7.9¨ from the tank wall. The tank wall must be flat and there must not be obstacles adjacent to the nozzle or on the tank wall. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 27 Figure 3-6: There is no maximum limit to the number of devices that can be operated in the same tank There is no maximum limit to the number of devices that can be operated in the same tank. They can be installed adjacent to other radar level transmitters. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 28: Tanks With Dish-Shaped And Conical Bottoms

    • Objects in the tank or pit. • Sharp corners that are perpendicular to the path of the radar beam. • Sudden changes in tank diameter in the path of the radar beam. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 29 Antenna type Beam angle Beam radius, x [mm/m] [in/ft] Lens, DN20 (¾¨) 15° Lens, DN25 (1¨) 10° Lens, DN40 (1½¨) 8° Lens, DN70 (3¨) 4° Table 3-2: Beam radius of the antenna 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 30: Process Connections

    Attach the weather protection accessory to the device if you install it more than 2000 m / 6560 ft above mean sea level. 3.7.2 Process connections Flange connections: installation procedure Figure 3-10: Flange connections: installation procedure Ød = nozzle diameter h = nozzle height www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 31 Antenna extensions for devices with flanges If the device has an antenna extension, this option extends the maximum nozzle height. The antenna extension has a length of 112 mm / 4.4¨. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 32: Equipment Needed

    26 ("Point the device in the correct direction"). • Tighten the connection to the correct torque (not more than 40 N·m / 29.5 lbf·ft). End of the procedure. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 33: Lpr Devices: Recommendations For Pits And Tanks Made Of Non-Conductive Materials

    If device is used in dusty conditions, we recommend that you seal the support. If dust is on the top of the tank and directly below the device, this can have an effect on the device performance. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 34: How To Attach The Purging System Adaptor To The Device

    Figure 3-14: How to attach the purging system adaptor to the device Equipment needed: • Device • Purging system adaptor (accessory) • ISO 228-1 (G) connection: Gasket (not supplied) • NPT connection: PTFE threaded seal tape (not supplied) www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 35 End of the procedure. INFORMATION! If it is necessary to send an order for the purging system adaptor, refer to Accessories: process connection kits for the level transmitter on page 159 . 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 36: How To Turn Or Remove The Display Module (Option)

    • Display extractor CAUTION! Disconnect the power supply. Follow this procedure: • Remove the housing cover with the cover wrench. • Find the two clips that hold the display module in the housing. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 37: Weather Protection

    1 Weather protection cover (with an R-clip to hold the cover on the clamp) 2 Device 3 Weather protection clamp (2 parts) 4 2 locking nuts 5 10-mm socket wrench (not supplied) The overall dimensions of the weather protection are on page 137. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 38 3 Lower the weather protection cover onto weather protection clamp until the hole for the lock is in the slot at the front of the cover. 4 Put the R-clip into the hole at the front of the weather protection cover. 5 End of the procedure. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 39: How To Open The Weather Protection

    1 Remove the R-clip from the hole at the front of the weather protection cover. 2 Remove the weather protection cover. 3 Lift the display screen cover. End of the procedure. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 40: Electrical Connections

    Check for the correct supply voltage printed on the nameplate. 4.2 General notes This chapter includes electrical connection data about devices with the 4...20 mA output and HART® communication options. These are 2-wire, loop-powered devices. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 41: Electrical Installation: Output Options With Cable Gland

    • Use the applicable electrical cables with the cable glands. • Make sure that the current is not more than 5 A or that there is a 5 A-rated fuse in the electrical circuit that energizes the device. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 42 • Cover wrench Procedure 1 Loosen the lock screw with a 3-mm Allen wrench. 2 Remove the cover stop. 3 Turn the cover counterclockwise with the cover wrench. 4 Remove the cover. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 43 1 Loosen the cable gland. Put the electrical wires into the cable entry. Loosen the terminal screws with a POZIDRIV® PZ1 screwdriver. Connect the electrical wires to the connector. 2 Tighten the terminal screws with a POZIDRIV® PZ1 screwdriver. 3 Tighten the cable gland. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 44 1 Put the cover on the housing 2 Turn the cover clockwise until it is fully engaged. 3 Attach the cover stop and lock screw. 4 Tighten the lock screw with a 3-mm Allen wrench. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 45: Electrical Installation: Output Options With An M12 Male Connector

    2 Resistor for HART® communication (typically 250 ohms) 3 Optional connection to the grounding terminal 4 Output: 12...30 V DC for a maximum output of 21.5 mA at the terminals 5 Device 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 46: Devices For Hazardous Locations

    • Close unused cable glands with dummy plugs ③ . The diameter of the outer sheath of the electrical cable (for the power supply and current output) must be 6…10 mm or 0.24…0.39¨. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 47: Networks

    1 Address of the device (0 for point-to-point connection) 2 4...20 mA + HART® 3 Resistor for HART® communication (typically 250 ohms) 4 Power supply 5 HART® converter 6 HART® communication software 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 48: Multi-Drop Networks

    1 Address of the device (each device must have a different address in multi-drop networks) 2 4 mA + HART® 3 Resistor for HART® communication (typically 250 ohms) 4 Power supply 5 HART® converter 6 HART® communication software www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 49: Start-Up

    • A connection to a system or PC with AMS™. You can download the Device Description (DD) file from the website. • A connection to a HART® Field Communicator. You can download the Device Description (DD) file from the website. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 50: Digital Display Screen

    C6.4.2 1st Value Variable (1st Meas. page) is set to "Level", then the device shows "Level" as the current output percentage in Normal mode (refer to item ① in the illustration). www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 51: Keypad Buttons

    [Up] Program mode: Increase value or change parameter Table 5-1: Functions of keypad buttons For more data on keypad functions, refer to Keypad functions on page 63. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 52 This will operate the device. How to operate the keypad buttons with a bar magnet Figure 5-4: How to operate the keypad buttons with a bar magnet Equipment needed • Bar magnet www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 53: Remote Communication With Pactware

    Refer also to the PACTware Consortium site at http://www.pactware.com. Figure 5-5: Screen from the PACTware™ user interface 1 DTM menu 2 Information for device identification 3 Installation parameters menu and menu items 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 54: Remote Communication With The Ams™ Device Manager

    • Store and read process data. • Store and read diagnostic status information. • Help plan preventive maintenance to reduce a plant's downtime to a minimum. You can download the DD file from our website. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 55: Operation

    2 One value and % bar graph of the scale for the measurement value 3 Two values 4 Two values and % bar graph of the scale for the top measurement value 5 Three values 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 56 Sensor value Volume Ullage volume Mass Ullage mass UllM Linearized distance Ldis or Distance Lin. Linearized level LLvl or Level Lin. Table 6-1: Abbreviations of measurement types used in Normal mode www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 57 Change screen (measurement pages 1 and 2, and the status [ ▼ ] [Down] message page) [ ▲ ] Change screen (measurement pages 1 and 2 and the status [Up] message page) Table 6-2: Functions of keypad buttons (Normal mode) 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 58 (flange face or thread stop) to the surface of the liquid. You cannot change this value in Program mode. Refer also to "Distance" in this table. Table 6-3: Measurement type definitions www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 59: Program Mode

    If you did not supply all installation data to the supplier before the delivery, the standard setup procedure in the Quick Setup menu is mandatory. INFORMATION! It is not possible to enter the D Service menu. This menu is for factory calibration and approved personnel only. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 60: Protection Of The Device Settings (Access Levels)

    • Push [>] to enter Program mode. • Push 2 × [ ▼ ], [>], 5 × [ ▼ ], [>], [ ▼ ] and [>] to go to the menu C7.2 Security. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 61 If you de-energize the device and then energize it again, the access level will go back to "User". If you do not touch the keypad for 5 minutes, the device goes back to Normal mode and the access level will go back to "User". 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 62: How To Get Access To The Quick Setup Menu

    • Push [>] to start the empty spectrum recording procedure. For more data, refer to Empty spectrum recording on page 95. Push [ ↲ ] at the end of each step of the procedure to continue to the next step. • End of the procedure. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 63: Keypad Functions

    Scroll up the menu list (for example, from menu C2 to menu C1). • Scroll up the sub-menu list (for example, from sub-menu C2.2 to sub-menu C2.1). Table 6-5: Functions of buttons for menu navigation 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 64: Button Description

    Down Scroll down the list Scroll up the list Table 6-6: Function of buttons in menu items that have a list of parameters www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 65 (this increases the value by a factor of 10). Table 6-7: Function of buttons in menu items that have values 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 66: Special Characters

    "Yes" or "No", push [ ↲ ] to accept (Yes) or reject (No) the new settings. If you push [ ↲ ] when the screen is set to "Yes" or "No", the display goes back to Normal mode. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 67: Menu Overview

    A4.2.1.12 Empty Spectrum Enable Table 6-8: Menu overview: A – Quick Setup 1 The device shows this menu item if you set the menu item A4.2.1.1 to "Partial, Average" or "Partial, Max" 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 68 1 You must enter the "Expert" password in A3 Login or C7.2.1 Login to find and use this menu 2 The device shows this menu item if you make a strapping table in menu C.3 Conversion www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 69 Current Output 1 C4.1.2 0% Range C4.1.3 100% Range C4.1.4 Current Out. Range C4.1.5 Error Function C4.1.7 Low / High Error Current C4.1.8.2 Trimming / 4mA Trimming C4.1.8.5 Trimming / 20mA Trimming 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 70 C5.1.4.1 HART Variables / Current Out. 1 Var. C5.1.4.2 HART Variables / HART sec./CO2 Var. C5.1.4.3 HART Variables / Tertiary Var. C5.1.4.4 HART Variables / Quaternary Var. C6 Display C6.1 Language C6.2 Backlight C6.3 Contrast www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 71 C7.1.8 Production Date C7.1.9 Calibration Date C7.2 Security C7.2.1 Login C7.2.2 Change Password C7.2.3 Reset Passwords C7.2.4 Unlock Extended Range C7.2.5 Unlock SIL C7.3 Errors C7.3.1 Message View C7.3.2.1 Error-Mapping / Sensor: Information 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 72: Function Description

    For more data, refer to Keypad functions on page 63 (values in menu items). Minimum access level to change the setting: Operator www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 73 95. Refer also to "How to make a filter to remove radar signal interference" on page 105. Minimum access level to do the procedure: Expert Table 6-11: A – Quick Setup menu 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 74 "Yes". Push [↲] again to start the test. The device goes to back to Normal mode after 1 hour. Minimum access level to do the procedure: Expert www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 75 "Yes". Push [↲] again to start the test. The device goes to back to Normal mode after 1 hour. Minimum access level to do the procedure: Expert 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 76 "Yes". Push [↲] again to start the test. The device goes to back to Normal mode after 1 hour. Minimum access level to do the procedure: Expert www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 77 This measurement data is shown with the units set in menu C7.5 Units. This menu item is only available if you set up a linearization table in menu item C3.1 Edit Table. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 78 Table 6-12: B – Test menu 1 The default value is the measured value at the time that you start the simulation www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 79 0.0…2000 mm / is for the high-temperature version of the 0.0...78.740¨ / device. Each segment is 105 mm / 4.1¨ long. 0.0...6.5617 ft Minimum access level to change the setting: Expert 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 80 If the gas is not 1.0, set the ε value to the ε value of the gas. Minimum access level to change the setting: Expert www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 81 If this function is not in operation, the device looks for the largest signal below the process connection. Minimum access level to change the setting: Expert 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 82 0.0 mm / 0.0¨ C3.2.2. Push [ ↲ ] to confirm the level value 0.0...100.0 mm / and go to C3.2.5 Conv. Value. 0.0...3937.0¨ Minimum access level to do the procedure: Expert www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 83 100% output to be 20 mA. You use reversed limits when you want the 0% output to be to be 20 mA and 100% output to be 4 mA. Minimum access level to change the setting: Expert 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 84 4…20 mA signal. If this menu item is set to "Off', this stops the 4…20 mA signal and starts HART® multi- drop mode. Minimum access level to change the setting: Expert www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 85 (Format: Year-Month-Day / YYYY-MM-DD) Minimum access level to change the setting: Operator C5.1.3.4 Cfg. Change This function counts the number of changes Read only — Counter to the HART® device settings 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 86 1st Value This changes the measurement type of the Level, Distance, Distance Variable first value on the measurement page. Sensor Value, Reflection, CO Percent Minimum access level to change the setting: User www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 87 This Sensor Value, Reflection, menu item is only available if you set C6.4.1 CO Percent Function to "Three Values". Minimum access level to change the setting: User 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 88 This menu item is only available if you set CO Percent C6.4.1 Function to "Two Values", "Two Values & Bar" or "Three Values". Minimum access level to change the setting: User www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 89 Read only — the device. The date format is: Year-Month- Day. C7.1.9 Calibration Date This the date that the manufacturer Read only — calibrated the device. The date format is: Year-Month-Day. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 90 C7.5.2.2 Offset Enter an offset value. — 0.0 m C7.5.2.3 Factor Enter a factor. Multiply the measured value — by this factor to change m (metres) to the custom length unit. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 91 C3 Conversion. "Mass" and "Ullage Mass" are available if you created a level-mass table in menu C3 Conversion. "CO Percent" is the current output shown as a percentage. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 92: Further Information On Device Configuration In Program Mode

    "Storage". If the surface of the Agitator or Process). product is disturbed, select "Process". If • [ ↲ ] to confirm. the surface of the product is agitated with vortexes and foam, select "Agitator". www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 93 Current Out. 1 Var. is set to "Distance", • [ ↲ ] to confirm. illustration ② shows the position of 0% output as a distance below the flange facing or thread stop of the process connection. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 94 3.5 mA (default value) if an error occurs. If this menu item is set to "High", the output current changes to 21.5 mA (default value) if an error occurs. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 95: Empty Spectrum Recording

    Installation on page 23. Values and parameters that can be changed are shown between the « ... » marks in the illustrations that follow. Push the keypad buttons in the correct sequence: 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 96 [ ↲ ] to confirm. Make sure that the tank is empty or it contains not more than the minimum quantity. Energize and operate all equipment that moves in the tank (e.g. agitators). www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 97: Hart® Network Configuration

    It can operate in either point-to-point or multi-drop mode. The device will communicate in multi- drop mode if you change the polling address. CAUTION! Make sure that the address for this device is different from others in the multi-drop network. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 98: Distance Measurement

    • Tank Height (C1.2) • Blocking Distance (C1.5) • Reference Offset (C1.10) • Display readings in Normal mode: 1st Meas. Page (C6.4) menu • Display readings in Normal mode: 2nd Meas. Page (C6.5) menu www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 99 5 100% Range (C4.1.3), if menu item C4.1.4 Current Out. Range is set to "4-20mA" or "3.8-20.5mA" 6 Non-measurement zone For more data about the menu items, refer to Function description on page 72 – table C. Full Setup menu. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 100: Level Measurement

    C1.11 Tank Bottom Offset when you give a level for the C4.1.2 0% Range and C4.1.3 100% Range settings. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 101 C6.4.2 and C6.5.2 (1st Value Variable) and menu C1 Install. Parameters. If you change C1.11 Tank Bottom Offset, then we recommend that you change C6.4.3 / C6.5.3 (0% Range) and C6.4.4 / C6.5.4 (100% Range) by the same quantity. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 102: How To Configure The Device To Measure Volume Or Mass

    The device will give more accurate volume readings if you give more conversion data in these areas: • Surfaces with curves. • Sudden changes in the cross section. Refer also to the illustration that follows: www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 103 "Yes" or "No", push [ ↲ ] to accept (Yes) or reject (No) the new settings. If you push [ ↲ ] when the screen is set to "Yes", the device will delete the data for the strapping table and go back to Normal mode. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 104: How To Measure Correctly In Tanks With Curved Or Conical Bottoms

    • Select Yes and push [ ↲ ]. The device will go back to Normal mode. INFORMATION! For more data on menu items, refer to Function description on page 72 – table C: Full Setup. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 105: How To Make A Filter To Remove Radar Signal Interference

    INFORMATION! For more data on empty spectrum scans, refer to Function description on page 72 – table A. Quick Setup menu (menu item A4.2). 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 106: Status Messages And Diagnostic Data

    If the device status changes, the device shows a status symbol in the top left corner of the display screen in Normal mode: Figure 6-11: Device status: Normal mode 1 Device status symbol (NAMUR NE 107) www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 107 1 Device status letter code (NAMUR NE 107). In this example, M = Maintenance required. 2 Description of error If the device shows an error message, refer to the table that follows for more data and to find a solution to a problem. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 108 Inconsistent Parameters Internal bus communication error. De-energize and then energize the device again. Sensor Comm. Error Internal bus communication error or hardware failure. If the message is shown again, tell the supplier. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 109 "Yes". Then go back to Normal mode and set the "Save Configuration?" page to "Yes". De- energize and then energize the device again. If the message is shown again, tell the supplier. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 110 Elec. Temp. Out Of Spec. The temperature of the sensor is Make sure that the device is not in the specified limits. operated in the permitted ambient temperature range. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 111 Empty Spectrum invalid The empty spectrum recording Do a new empty spectrum does not agree with the process recording. conditions at this time (e.g. the tank height was changed). 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 112 Table 6-17: Description of errors and corrective actions 1 DM = data manager 2 CO = current output 3 This error message is shown if the device is in SIL mode. CO = current output. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 113: Service

    O-rings are correctly greased or, if it is necessary, replaced. For more data about the replacement of the O-rings, speak or write to your supplier. Figure 7-1: Maintenance of the O-rings 1 Display cover 2 Terminal compartment cover 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 114: How To Clean The Top Surface Of The Device

    The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training. INFORMATION! For more precise information, please contact your local sales office. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 115: Returning The Device To The Manufacturer

    • to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances, • to enclose a certificate with the device confirming that it is safe to handle and stating the product used. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 116: Form (For Copying) To Accompany A Returned Device

    The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 117: Disassembly And Recycling

    Weight Total weight 2.36 kg / 5.20 lb Weight of parts Aluminium: (percentage of total weight) Stainless steel: Plastic: Electronics: Dimensions Volume 3.51 dm³ / 214.2 in³ Table 7-1: Product description 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 118: Parts List

    Screw stainless steel (33) EMC board — (34) O-ring EPDM (35) Cover aluminium (36) Washer stainless steel (37) Screw stainless steel (38) Housing aluminium (40) Sticker polyester Table 7-2: Parts list www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 119: Equipment Needed

    • 5.5-mm socket wrench (not supplied) • POZIDRIV® PZ1 screwdriver (not supplied) • Adjustable wrench (not supplied) The product does not contain harmful gases or substances. There are no special precautions for disassembly of the device. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 120 Noble / precious metal Silicon 0.202 0.445 — Signal converter 0.063 0.139 — Aluminium 0.013 0.029 — Table 7-3: Materials and components which must be removed and independently prepared for treatment www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 121 Senator S80L Materials and components which can be recycled Material Percentage of total weight Total weight [kg] [lb] Stainless steel 2.63 5.80 Aluminium Table 7-4: Materials and components which can be recycled 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 122: Technical Data

    4 Distance to product surface, where change in frequency is proportional to distance 5 Differential time delay, Δt 6 Differential frequency, Δf 7 Frequency transmitted 8 Frequency received 9 Frequency 10 Time www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 123: Measurement Modes

    "TBF FULL" AND "TBF AUTO" MODES It is important to enter the correct dielectric constant value in menu item C2.2 Epsilon R Product. If this value is incorrect, the device will not measure level accurately. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 124: Technical Data

    Display and user interface Display Backlit LCD display 128 × 64 pixels in 64-step greyscale with 4-button keypad Interface languages English, French, German, Italian, Spanish, Portuguese, Chinese (simplified), Japanese, Russian, Czech, Polish and Turkish www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 125: Operating Conditions

    Make sure that there are not any obstructions directly below the process connection for the device. For more data, refer to Installation on page 23. For dimensions and weights data, refer to Dimensions and weights on page 137. Dimensions and weights 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 126: Process Connections

    Standard flanges: NPS 3…8 in Class 150 RF and Class 300 RF; others on request Optional flange facing for standard flanges: FF (Flat Face) JIS B2220 80…200A in 10K RF; others on request Other Others on request www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 127: Input And Output

    Max. ±15 mm / 0.6¨ for the full temperature range Multi-drop operation Yes. Current output = 4 mA. Enter Program mode to change the polling address (1...63). Available drivers FC475, AMS, PDM, FDT/DTM 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 128: Approvals And Certification

    Ex ia tb [ia Da] IIIC T85°C...T*°C Db (Canada) – antenna suitable for zone 20 NEPSI Ex ia IIC T3...T6 Ga/Gb; Ex db ia IIC T3...T6 Ga/Gb; Ex ia IIIC T85°C...T*°C Da / Ex ia IIIC T85°C...T*°C Db; Ex ia tb IIIC T85°C...T*°C Db www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 129 Flange facing finish, R (min...max) [µm Rz] [µin – AARH] EN 1092-1 3.2…12.5 125…500 ASME B16.5 RF or FF 3.2...6.3 125...250 JIS B2220 3.2...6.3 125...250 Table 8-2: Process connection options: flange facing finish 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 130: Minimum Power Supply Voltage

    Figure 8-3: Minimum power supply voltage for an output of 21.5 mA at the terminals (Hazardous Location approval (Ex d / XP/NI)) X: Power supply U [V DC] Y: Current output load R [Ω] www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 131: Measuring Accuracy

    Y: Measuring accuracy [+yy inches / -yy inches] 1 3.94¨ INFORMATION! To calculate the accuracy at a given distance from the antenna, refer to Technical data on page 124 (measuring accuracy). 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 132 Y: Measuring accuracy [+yy inches / -yy inches] 1 3.94¨ INFORMATION! To calculate the accuracy at a given distance from the antenna, refer to Technical data on page 124 (measuring accuracy). www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 133 Y: Measuring accuracy [+yy inches / -yy inches] 1 1.97¨ 2 7.87¨ INFORMATION! To calculate the accuracy at a given distance from the antenna, refer to Technical data on page 124 (measuring accuracy). 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 134 Y: Measuring accuracy [+yy inches / -yy inches] 1 3.94¨ INFORMATION! To calculate the accuracy at a given distance from the antenna, refer to Technical data on page 124 (measuring accuracy). www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 135: Guidelines For Maximum Operating Pressure

    2 Process connection temperature, T [°C] 3 Process pressure, p [psig] 4 Process connection temperature, T [°F] 5 Threaded connection, G (ISO 228-1) 6 Flange connection, PN40 7 Flange connection, PN16 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 136 2 Process connection temperature, T [°C] 3 Process pressure, p [psig] 4 Process connection temperature, T [°F] 5 Threaded connection, NPT (ASME B1.20.1) 6 Flange connection, Class 300 7 Flange connection, Class 150 www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 137: Dimensions And Weights

    5.94 6.30 7.56 8.39 1.24 0.83 connection Table 8-4: DN20 / ¾¨ Lens antenna: Dimensions in inches 1 If the process temperature is more than +302°F, add 4.41¨ to this value 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 138 5.94 6.30 7.56 8.46 1.25 0.89 connection Table 8-6: DN25 / 1¨ Lens antenna: Dimensions in inches 1 If the process temperature is more than +302°F, add 4.41¨ to this value www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 139 2 If the process temperature is more than +302°F, add 4.41¨ to this value. If the device has the antenna extension option, add 4.41¨ to this value. 3 If the device has the antenna extension option, add 4.41¨ to this value 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 140 • The diameter of the outer sheath of the cable must be 7…12 mm or 0.28…0.47¨. • Cable glands for cQPSus-approved devices must be supplied by the customer. • A weather protection cover is available as an accessory with all devices. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 141 4 If the device has a G 3 process connection, then f = 0.92¨. If the device has a 3 - 8 NPT process connection, then f = 1.18¨. 5 If the process temperature is more than +302°F, add 4.41¨ to this value 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 142 Stainless-steel cover options Figure 8-21: Stainless-steel cover options 1 Signal converter housing with stainless-steel cover - without locking system 2 Signal converter housing with stainless-steel cover - with locking system www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 143 1 If the process temperature is more than +150°C / +302°F, the housing has a distance piece. For more data about the overall dimensions of the device, refer to the "Dimensions and weights" section. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 144 DN80 PN16 / B1 or NPS 3 Class 150 / RF flange and antenna extension (length 112 mm / 4.4¨) Table 8-14: Antenna option weights 1 NPS = Nominal Pipe Size. For more data about the dimensions of ASME flanges, refer to the ASME B16.5 standard. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 145: Description Of Hart Interface

    An FSK or HART® modem is included in field devices and manual control units. It is necessary to have an external modem for PC-supported workstations. The external modem is connected to the serial or USB interface. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 146: Software History

    9.3.1 Point-to-Point connection – analogue / digital mode Point-to-Point connection between the signal converter and the HART® Master. The current output of the device is passive. Also refer to Point-to-point connection on page 47. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 147: Multi-Drop Connection (2-Wire Connection)

    Device Description (DD) file is installed. System and software requirements for the Field Communicator • System card that includes the "Easy Upgrade Option" • Field Communicator Easy Upgrade Programming Utility 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 148: Operation

    AMS and operating using the local keyboard. The service menu parameters are not supported and simulation is only possible for current outputs. The online help for each parameter contains its function number as a reference to the local device display. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 149: Field Device Tool / Device Type Manager (Fdt / Dtm)

    This function number agrees with the function number in the device menus. Use the same procedure for the protection of parameters in the supervisor menu. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 150: Hart® Menu Tree For Ams

    Output Display Device Hart Calibration Service Information Device Diagnostics Device Status Actual Values Simulation Test/Reset Information Process Variables Measured Values Inputs/Outputs Table 9-4: Overview AMS menu tree (positions in menu tree) www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 151: Ams Menu Tree (Details For Settings)

    Function / 1st Value Variable / Page Format 1st Value / 2nd Value Variable / Format 2nd Value / 3rd Value Variable Format 3rd Value / 0% Range / 100% Range 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 152 Service Calibration Calibration Current Output 1 Sensor Manual Correction Offset / Manual Correction Factor / Correction Offset / Correction Factor Rd, Opt Correction Offset Extended Rd, Opt Correction Factor Extended www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 153: Device Diagnostics

    Opt, Rd / Ullage / Ullage Mass Inputs / Out / PV % Range / PV output current / SV / TV / QV Table 9-5: AMS menu tree (details for settings) 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 154: Hart® Menu Tree For Pdm

    Information Security Units Factory Default HART Identification & Info Service Calibration Calibration Sensor Table 9-6: Overview: Device Menu Overview: View Menu Measured Value Input / Outputs Table 9-7: Overview: View Menu www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 155: Pdm Menu Tree (Details For Settings)

    Current Output 1 Current Out 1 Variable / 0% Range / 100% Range / Current Out Range / Error Function / Low Error Current / High Error Current / D/A Trim PV 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 156 Manual Correction Offset / Manual Correction Factor / Correction Offset / Correction Factor Rd, Opt Correction Offset Extended / Correction Rd, Opt Factor Extended Table 9-9: PDM menu tree: device menu www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 157 / Electronic Revision / Field Device Revision / Software Revision / Hardware Revision / Electronics Serial / Production Date / Calibration Date / Operating Time Table 9-11: PDM menu tree: Diagnosis 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 158: Appendix

    Viator RS232 / HART converter XFDX060200 Viator USB / HART converter XFDX060300 ③ Display extractor, magnet and cover wrench XFDX060400 ④ — USB / 24 V DC power supply XFDX060500 Table 10-1: Accessories: general items www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 159: Accessories: Process Connection Kits For The Level Transmitter

    NPS 4 Class 150 (ASME) 3 - 8 NPT threaded connection DN150 PN2.5...PN40 (EN) / XF70000N21 option NPS 6 Class 150 (ASME) DN200 PN2.5...PN40 (EN) / XF70000N22 NPS 8 Class 150 (ASME) 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 160: Accessories: Purging System Adaptor

    Purging system adaptor for 1 /12 - 11.5 NPT transmitter process XF70000N68 connection Purging system adaptor for G 3 transmitter process connection XF70000069 Purging system adaptor for 3 - 8 NPT transmitter process XF70000N69 connection Table 10-3: Accessories: purging system adaptor www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 161: Glossary

    Height from the bottom of the tank (user-defined) to the surface of the top product (Tank height – distance). See the diagrams at the end of this section. Mass Total mass of tank contents. 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 162 TBF mode uses the reflection of the tank bottom to indirectly measure the level of the tank contents. Ullage volume Unfilled volume. See the diagrams at the end of this section. Volume Total volume of tank contents. www.hawkmeasure.com 04/2023 - Senator S80L Radar Manual v3.00 en...
  • Page 163 Figure 10-4: Measurement definitions: distance 1 Distance 2 Blocking distance 3 Flange facing 4 Gas (Air) 5 Tank height 6 Ullage volume or mass Figure 10-5: Measurement definitions: level 1 Level 2 Volume or mass 04/2023 - Senator S80L Radar Manual v3.00 en www.hawkmeasure.com...
  • Page 164 Hawk Measurement Systems (Head Office) 15 - 17 Maurice Court Nunawading VIC 3131, AUSTRALIA Phone: +61 3 9873 4750 Fax: +61 3 9873 4538 info@hawk.com.au Hawk Measurement 5010 Gateway Drive, Medina, OH 44256, USA Phone: +1 888-429-5538 Phone: +1 978-304-3000 Fax: +1 978-304-1462 info@hawkmeasurement.com...