Hawk Intelligent 2000-SAN Series Instruction Manual

Pressure and level transmitters, hart
Table of Contents

Advertisement

Quick Links

Overview
Gladiator
Acoustic Switch Series
A Higher Level of Performance
Instruction Manual ("HART")
"Intelligent"
Pressure and Level Transmitters
Series 2000 - SAN
Series 2000
Series CER - 2000
Series 2000 - Hydrobar
For more information, please visit >
www.hawkmeasurement.com

Advertisement

Table of Contents
loading

Summary of Contents for Hawk Intelligent 2000-SAN Series

  • Page 1 Overview Gladiator Acoustic Switch Series A Higher Level of Performance Instruction Manual (“HART”) “Intelligent” Pressure and Level Transmitters Series 2000 - SAN Series 2000 Series CER - 2000 Series 2000 - Hydrobar For more information, please visit > www.hawkmeasurement.com...
  • Page 2: Table Of Contents

    Series 2000 HART Overview Table of Contents 1. Introduction 8. Explanation Programming Points P101 to P114 1.1 Description Series 2000-SAN P101 Zero Adjustment (4 mA) 1.3 Description Peramic “S” (Series CER 2000) P102 Span Adjustment (20 mA) Without Test Pressure 1.2 Description Series 2000 P103 Cancel Mounting Position Effect 1.4 Description HYDROBAR (Series 2000 - Hydrobar)
  • Page 3: Introduction

    Overview Series 2000 HART Gladiator Acoustic Switch Series 1. Introduction The SERIES 2000 and SERIES 2000-SAN are solid-state pressure- and level transmitters based upon a bridge resistive silicon sensor, with a very high burst pressure. The sensor element is mounted in a stainless steel foot. A very strong stainless steel “flush”...
  • Page 4: Dimensional Drawings

    Overview Series 2000 HART Gladiator Acoustic Switch Series Overview 2. Dimensional Drawings 2.1 2000-SAN Parts Descriptions Material Cover SS 304 Pushbutton + Display Cover with venting SS 304 Venting PG9 Cable Gland O-ring (2 pcs) EPDM Electronics housing SS 304 Foot SS 316 Lock ring...
  • Page 5: 2000-Hydrobar

    Series 2000 HART Overview 2.4 2000-Hydrobar Parts Descriptions Material Cover SS 304 Push buttons + Display Cover with venting SS 304 Venting PG9 Cable Gland 2x O-ring EPDM Electronics housing SS 304 2x Connection to cable SS 316 Cable with venting tube Foot SS 316 Diaphragm protection cap...
  • Page 6: Installation Transmitter

    Series 2000 HART Overview 3. Installation transmitter The diaphragm of the transmitter is protected with a special protection cap. Protect the diaphragm until installation takes place. * DO NOT DAMAGE THE DIAPHRAGM. *. 3.1 Installing Weld-On Nipple A skilled machinist or certified welder should perform installation of the weld-on nipple. Weld Argon, MIG or TIG with the smallest welding pin.
  • Page 7: Installing Transmitter Series Cer-2000

    3.7 Installing Transmitter Series 2000-Hydrobar The diaphragms are protected by a special protection cap (PE). Do not remove this cap. When clogging is a problem, please contact HAWK for other solutions. DO NOT DAMAGE THE DIAPHRAGM. 3.8 Calibration All transmitters are fully calibrated at the factory, to the conditions stipulated in users order. When the buyer has not requested calibration, the transmitter will be calibrated at the maximum span.
  • Page 8: Remaining

    Series 2000 HART Overview 4. Remaining 4.1 Digital Local Indicator All transmitters from the Series 2000 are standard equipped with a digital display. In the standard execution the covers are “closed”. The three push buttons and the display are behind the cover (3). As an option a transparent cover can be delivered to achieve the display can be used as a local display in the process.
  • Page 9 The gaskets and seals for the series 2000 are successfully tested according to IEC 60079-0 for temperature endurance (heat and cold), resistance to impact and ingress protection by DEKRA Certification. In harsh process and environmental conditions the lifetime of gaskets and seals may be reduced. Hawk Measurement strongly advises to inspect these yearly and replace them when necessary.
  • Page 10: Special Instructions For Safe Use

    Series 2000 HART Overview 4.5 Special Instructions for Safe Use For use in a potentially explosive atmosphere caused by air/dust mixtures the following should be taken into consideration: • The dust layer may not exceed a thickness of 5 mm. •...
  • Page 11: Function Of Push Buttons

    Series 2000 HART Overview 5. Function of Push Buttons The Series 2000 can be programmed easily by use of the 3 front panel pushbuttons (See picture right). The display can show engineering units of: O, inH O, bar and PSI. The functions of the three pushbuttons will be explained below.
  • Page 12: Programming Points (P101 - P114)

    Series 2000 HART Overview 6. Programming Points (P101 - P114) The following points can be adjusted by means of the Programming points: three push buttons. P101 Zero adjustment (4 mA) For an explanation of these points see page 11 to 18 of P102 Span adjustment (20 mA) this manual.
  • Page 13: Explanation Programming Points P101 To P114

    Series 2000 HART Overview 8. Explanation Programming Points P101 to P114 Zero Adjustment (4 mA) P101 The transmitter as standard is adjusted to 4.00 mA at atmospheric pressure. It is also possible to adjust a zero- suppression or elevation. For example: zero elevation of 1 mWK. 1.
  • Page 14: P103 Cancel Mounting Position Effect

    Series 2000 HART Overview Cancel Mounting Position Effect P103 All transmitters are calibrated vertically. When a transmitter of the Series 2000 or 2000-SAN is installed horizontally, there will be a small “mounting effect” on the zero (4 mA). For example the transmitter shows 4.03 mA instead of 4.00 mA. This can be easily canceled with P103. In P103 there are three options: 1.
  • Page 15: P105 Reverse Output (20 - 4Ma)

    Series 2000 HART Overview Reverse Output (20 - 4mA) P105 The transmitter as standard is adjusted to 4-20 mA. Push on [PROG] and go to P105. Push h to change the output to 20-4 mA (Reverse output). Push [PROG] to confirm this. Now the transmitter will give 20 mA at atmospheric pressure Adjustment Samping (0 till 25 sec) P106...
  • Page 16: P110 Simulation Of Current (4 - 20 Ma)

    Series 2000 HART Overview Simulation of Current (4 - 20 mA) P110 The transmitter can be used as a simulator of a current between 4 - 20 mA. This can be done in P110 with the push buttons h and i. The user can perform a current simulation (Curr) or a pressure simulation (Unit).
  • Page 17: P111 Linearisation

    Series 2000 HART Overview Linearisation P111 0 = no linearisation 1 = cylindrical tank (horizontal) 2 = tank with bottom cone 3 = tank with spherical bottom As standard the transmitter is delivered without linearisation (= 0). However, for a horizontal tank or a tank with a bottom cone, a linearisation can be applied to achieve the current signal (mA) is equal to the level in the tank.
  • Page 18: Linearization Horizontal Tank (Elliptical)

    Series 2000 HART Overview Linearization horizontal tank (Elliptical) 1. Push [PROG] until “100” is shown on the display. 7. Enter H3 in meters if the transmitter is installed like in the left picture. Enter 0m when the transmitter is installed like 2.
  • Page 19: P112 Specific Density

    Series 2000 HART Overview Specific Density P112 If the specific gravity of the medium differs from 1 kg/dm , you can enter the real density of the medium in P112. When this option is used, in menu P102, the ‘true’ height of the tank must be entered. Write Protection P113 The Series 2000 with HART-Protocol can be protected for writing (Write Protection).
  • Page 20: Programming The Series 2000

    To be able to install the DTM Klay Series 2000 HART on your system, download the installation file. This file can be downloaded from the HAWK Instruments site or via the following link: www.hawkmeasure.com The installation can be started by opening the “DTM Klay Series 2000 HART V1-2-0-1.zip” and continue by opening the “DTM Klay Series 2000 HART V1-2-0-1.exe”...
  • Page 21: Specifications

    0,5 s to 25 sec (0,5 sec = Std. Damping) Protection Grade IP 66 IP 68 (for 2000-Hydrobar-cable submersed parts) Material Housing AISI 304 “wetted” part AISI 316 L For higher temperatures use other kind of pressure transmitters. Contact Hawk Measurement Systems for information.
  • Page 22: Precautions And Warnings

    The warranty is 1 year from purchase date. HAWK does not accept liability for consequential damage of any kind due to use or misuse of the Series 2000. Warranty will be given, to be decided by the manufacturer. Transmitter must be shipped prepaid to the factory on manufacturers authorization.