INSTRUCTION
MANUAL
IM-462
Revision 5
MediSpin-6
Centrifuge
Cat. No. 462 -- For 120 VAC, 60 Hz
Cat. No. 463 -- For 240 VAC, 50 Hz
MediSpin-12
Centrifuge
Cat. No. 464 -- For 120 VAC, 60 Hz
Cat. No. 465 -- For 240 VAC, 50 Hz

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MediSpin-6 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for IEC MediSpin-6

  • Page 1 INSTRUCTION MANUAL IM-462 Revision 5 MediSpin-6 Centrifuge Cat. No. 462 -- For 120 VAC, 60 Hz Cat. No. 463 -- For 240 VAC, 50 Hz MediSpin-12 Centrifuge Cat. No. 464 -- For 120 VAC, 60 Hz Cat. No. 465 -- For 240 VAC, 50 Hz...
  • Page 2: Table Of Contents

    This manual may not contain information on all changes that have occurred to the subject instrument since the manual issue date. It was prepared for use by IEC authorized factory-trained service or dealer personnel who are kept current through a program of service letters and bulletins and training seminars.
  • Page 3: Introduction

    1 INTRODUCTION The IEC MediSpin is a portable benchtop centrifuge designed for use in the physician's office, the small clinical laboratory, and in the educational field, to perform separations by centrifugal force. The unit consists of: • A baseplate with three rubber suction cup pads for stability.
  • Page 4: Installation

    2 INSTALLATION The unit requires minimal assembly before it can be used. After unpacking, place the unit on a clean, level surface. The surface must be level to ensure quiet, vibration-free operation. A rigid, stable location is important since an improperly loaded unit can vibrate or even move.
  • Page 5: Operation

    Cautions •The IEC MediSpin is not intended or recommended for use with chemical solvents that may weaken critical parts. Refrain from using prohibited chemicals (see Table 1) for cleaning and/or centrifugation.
  • Page 6: Loading

    Loading Unlatch and open the cover. Check that the rotor turns freely, and that there are no loose objects in the chamber. Load the rotor SYMMETRICALLY with 2,3,4 or 6 tubes in the six- place rotor (NOT 1 or 5); or 2,3,4,6,8,9,10 or 12 tubes in the twelve-place rotor (NOT 1, 5, 7 or 11).
  • Page 7: Maintenance

    4 MAINTENANCE Cleaning This section describes the proper method of cleaning your centrifuge. Always unplug the unit before cleaning. Release the latch, open the cover and unplug the unit. Check the chamber for spillage or broken glass. If present, mop up with a sponge dampened with water.
  • Page 8: Fuse Change

    Fuse Change To change the internal fuse(s), open the cover and remove the rotor and shields. Unplug the centrifuge, carefully place the machine on its cover and remove the baseplate (4 Phillips head screws). The fuse(s) have the following values: For 120 VAC (Cat.
  • Page 9: Service

    Phillips head screws which secure the baseplate. Lift the baseplate off. Interlock The IEC MediSpin has a dual action interlock. When pressed, the cover open button releases the mechanical latch and activates the safety interlock release (if no rotation is sensed). There are two switches sensing the state of the cover and locking solenoid.
  • Page 10: Timer

    Timer The IEC MediSpin has a built in 30 minute mechanical timer. It also has a hold mode for indefinite spins. The timer acts as a switch delivering power to the motor. To replace it, the knob must first be removed. The knob is pressed onto the timer shaft and may require plyers to remove.
  • Page 11: Warranty

    IEC wants you to be satisfied with the quality of your MediSpin centrifuge. We warrantee your IEC centrifuge for one year. We will repair or replace a product that fails, within this period from the date of its delivery, due to defects in material and workmanship, and we will ship you the repaired product or its replacement at our expense.
  • Page 12: Specifications

    6 SPECIFICATIONS Specification Cat. No.s 462 & 464 Cat. No.s 463 & 465 (60 Hz) (50Hz) Maximum Speed 3100 rpm 2700 rpm Maximum RCF 1228 xg 920 xg Tube Size (mm) O.D. x Length Min: 7x100 7x100 O.D. x Length Max: 16x133 16x133...
  • Page 13: Drawings

    10875 PCB Schematic 10859 Wiring Diagram Parts Identification These drawings were accurate at the time of publication. Changes do occur. If you have any questions regarding these drawings, please contact IEC Technical Support at (800) 843-1113 ext.2002. MediSpin Instruction Manual...
  • Page 16 Part Description IEC No. Cover Assembly 50324 Housing 50322 Baseplate 50326 Motor 50339 Motor Mounting Plate 50333 PC Board 44608 Latch 50328 Catch 50325 Latch Solenoid Bracket 50327 Return Spring 50329 Latch Arm 50330 Latch Pivot Pin 50331 Rubber Tip...
  • Page 17 MediSpin Instruction Manual...