Page 3
Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 This user manual was prepared for the user’s convenience when using JPI’s DynaVue Duo DR/Fluoroscopy system. We highly recommend you read the manual thoroughly before operating the system for your safety and proper operation.
5) The owner should ensure continuously that power can supply product specified voltage and current to the system. 6) If the product does not operate properly or failure occurred, the operator should contact to JPI HEALTHCARE field service representative immediately.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 1.2 Hazards 1) Radiation This system includes an X-Ray generator. Operator should follow the safety instructions and always wear protective shield when operating the system.
Page 9
X-Ray. Back up the database regularly. 5) Authorized Access Only Only personnel authorized by JPI with password can access the Service Mode for safe installation and usage of the system. DynaVue Duo Service Manual...
Page 10
10) Maintenance Only authorized personnel designated by JPI should perform maintenance work to the system. DynaVue Duo Service Manual 10/104 Page...
Page 11
Only authorized personnel by JPI HEALTHCARE should perform the installation and provide support/service of the system. Only authorized person, designated by JPI HEALTHCARE, should perform a calibration after initial installation or replacing generator, X-Ray tube and/or detector. DynaVue Duo Service Manual...
Page 12
The workstation will be locked down to prevent any damage to the operating system and software control for your digital hardware. JPI Healthcare, Inc. has implemented physical and technical safeguards for all workstations that access electronic protected health information that will restrict access to authorized users.
Mechanical stand, and a Detector. It is equipped with 4-way floating table which is controlled by the foot brake built into the system. This system supports both DR and Fluoroscopic function, and it provides the tools for diagnosis, intervention and/or operations. [DynaVue Duo system block diagram] 2.2 Detector Specification Technical specification Detector Technology...
Page 14
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Active Pixel Area 3,052 x 3,052 1,516 x 1,516 Screen Pixel Pitch, µm 140 μm 280 μm Spatial Resolution, lp/mm...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.1 Under Column Place the table on a flat ground and lock the wheels. Screw the four M8x20 wrench bolt (with S/W and F/W) to assemble the Middle support and the Under support on the back of the body.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.2 Upper Column 1) Align the holes in the Upper column with the pins in the Under column.
Page 20
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3) Screw the three M4x12 wrench bolt to (1) and screw the two M5x12 wrench bolt to (2) to assemble the Column Support bracket as shown in figure below.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.3 Monoblock 1) Insult the Monoblock to Upper column. 2) Screw the four M4x8 Flat head Screw to assemble the Collimator Block (1).
Page 22
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 4) Assemble the Left(Point1), Right(Point2) Monoblock bracket on both sides. Screw the Six M4x10 Wrench bolt to point 3 and 4.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.4 Harness 1) Place the cable along the back side of the column as shown in the figure below.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.5 Column Cover 1) Connect three USB Cable to Camera. 2) Align the column Cover and Screw Ten M4x8 Flat Head Screw.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.6 Under Cover Assemble the Under Cover (1) to Collimator Block (2). Adjust the position to secure it.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Collimator 1) Loosen the four Nut (2) on the upper part of Collimator (1) to move the fixing plate.
Page 27
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 4) After assembling the collimator (5) to the collimator bracket, move the blade (6) inward to fix it, and tighten the four nuts (7).
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Upper Cover 1) Align the Upper cover (1) according to the Assemble frame of Under cover (2). 2) Screw ten M4x20 Flat Head screw to point (3), (4).
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.11 Connector 3.11.1 PC Connector 1) Connect cables from the System to PC (1),(2),(3). 2) LAN cable (4) must be connected to embedded LAN port.
Page 32
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3.11.2 Power and Switches 1) Green circular connectors for hand switch. 2) Gray circular connectors for foot switch.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 4. Set up 4.1 Pre-check for software installation 1) Check the window status by using the Window Update function and make sure it is up to date before the installation.
Page 34
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5) Change the IP address to communicate with the detector. 6) Network driver updates as latest version, click the “Driver update”.
Page 35
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 8) Select “Jumbo Frame” and set as “Disable” 9) Select “Receive Buffers” and set as Max. (the Max value is depended on the the LAN Card)
Page 36
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 10) Click the Power Management tab and uncheck “Allow the computer to turn off this device to save power”.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Set the value ‘0’ in the Turn off hard disk after. 4.2 Software installation ⚫ ExamVueDuo_RF S/W: RAD and FLO image acquisition software 4.2.1 ExamVueDuo_RF installation...
Page 38
(C:\ExamVue Duo). ExamVue Duo icon will also appear on the desktop. 4.2.2 Set DynaVue Information 1) Open the X-RayConsole.exe (C:\ExamVue Duo\Generator\IMD_Duo\X-RayConsole.exe) 2) Input the production information of the DynaVue Duo. ➔ The user must enter the information before proceeding to the next window. DynaVue Duo Service Manual...
Page 39
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3) Press “SAVE” button. DynaVue Duo Service Manual 39/104 Page...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5. Calibration and Alignment 5.1 Alignment between Detector and Collimator 5.1.1 Set X-Ray condition settings 1) Run the “X-Ray Console.exe” in "C:\ExamVue Duo\Generator\IMD_Duo\X-Ray Console.exe”.
Page 41
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5.1.2 Run Calibration software 1) Run “NMS.exe” in “C:\ExamVue Duo\Detector\NeoQ\NMS” as an “Administrator” at first. 2) Check the IP address registered (192.168.2.200) and then click “Connect”...
Page 42
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5.1.3 Check Image Area 1) Push the Foot switch to exposure the X-Ray 2) Confirm an image obtained from “NMS”.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5.2 X-Ray Calibration ⚫ X-Ray Console S/W: ⚫ Set the X-RAY condition. ⚫ Shoot X-RAY and check the feedback from I/O board.
Page 44
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 ➔ New Data: Data is obtained by adjusting the input data from 10mA to 200mA ⚫ Acquiring of reference data of the equipment.
Page 45
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 9) When the calibration is complete, the resulting data sets are automatically saved in the path below. ⚫ IMD data by KVP: C:\ExamVue Duo\Generator\IMD_Duo\InputData\ FLUOROSCOPY 5.2.2...
Page 46
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 6) Select the Calibration mode. 7) After setting it to 60 KVP, 10MA, record INPUT VALUE. 8) Take an X-Ray using the left foot switch.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 11) “10” INPUT VALUE - Calculates “7” INPUT VALUEs. ex) 4150 - 4350 = -200 12) Enter the value obtained from number 11 in the GLUBAL OFFSET item and press SAVE.
Page 48
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3) Remove any Object or Phantoms on the Detector. 4) Click “Calibration Setup” menu. 5) For the Dark Calibration, click the “Dark” to get the Dark image from the detector. The Dark Image will save as Dark_average value.raw (The acquired Dark frame will be generated in the...
Page 49
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Out of range 7) Click the “Delete” button to delete the Dark image if it’s out of average range (±5) 8) Acquire new Dark image again and repeat the above 6) 9) Acquire the 5 Dark images as shown the below figure.
Page 50
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 “C:\ ExamVue Duo\Detector\NeoQ\Data\Calibration\RAD”). 12) Check the pixel value of the Bright image. It should be within from 17,000 to 23,000.
Page 51
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 PASS FAIL An Image is evenly loaded in all areas Empty spaces are visible in the image. FLUOROSCOPY – FULL MODE 5.3.2...
Page 52
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5) For the Dark Calibration, click the “Dark” to get the Dark image from the detector. The Dark Image will save as Dark_average value.raw (The acquired Dark frame will be generated in the...
Page 53
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 10) For the Bright Calibraiton, Set the X-Ray condition (kVp, mA, msec) ⚫ kVp: 60, mA: 10 11) Shoot X-Ray (Foot Switch) to acquire Bright images.
Page 54
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 16) After getting 5 Bright images, click “BPM” button to make the BPM file. The “BPM” file will be generated in the “C:\ExamVue Duo\Detector\NeoQ\Data\Calibration\NOBIN\”.
Page 55
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 FLUOROSCOPY – HALF MODE 5.3.3 1) Run “NMS.exe”. 2) Check Shot Mode as “FLO” and “BIN”. 3) Remove any Object or Phantoms on the Detector.
Page 56
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Out of range 7) Click the “Delete” button to delete the Dark image if it is out of average range (±5) 8) Acquire new Dark image again and repeat the above 6).
Page 57
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 12) Check the pixel value of the Bright image. It should be within from 3,000 to 5,000. 13) Click the “Delete” button to delete it if it is out of average range.
Page 58
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 PASS FAIL An Image is evenly loaded in all areas Empty spaces are visible in the image. 5.3.4 Getting RAD and FLO image in ExamVue Duo [Emergency] ⇒...
Page 59
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 2) Click in “Control panel”, it will show FLO screen. 3) Press the left pedal of the foot switch, then it will show Fluoroscopy image in real time.
Page 60
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 4) Check the X-Ray condition. (kV and mA) 5) Press the left pedal of the Foot switch, then it will show Fluoroscopy image in real time.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Main Power 1) Check wall power can supply over 15A current. 2) Main Power On (O → I) and wait for “10 seconds”.
OK: V, Tuning: T, Repair: R, Replace: E 5.7 Inverter and Monoblock 1. Maintenance and service of the system should be performed by authorized personnel with proper training by JPI HEALTHCARE only. 2. Recommended maintenance schedule could vary depending on regulations of different countries.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 ⚫ Disassemble the thermal switch circuit of Monoblock’s X-Ray tube and check for irregularity. (Do not take the X-Ray shot) ⚫...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 1) Set the technique at 50kV, 10mA and 100ms then shoot 10 times with 15 second interval between each shot.
Page 66
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Troubleshooting Check List List Contents Remark When was the unit installed? When was the issue reported first? Model number of the system and generator used? Site Condition Please provide a detailed picture of all cables connected.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 6. Peripheral Devices 6.1. Live Camera In Device Manager, engineer can check whether the camera is connected or not using the following items.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 6.2. Infrared Sensor In Device Manager, engineer can check whether the Infrared sensor is connected or not using the following items.
Page 69
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 B. Run the icon below on the ToolBar in Exam. Thickness Thickness Measurement Calibration Progress Calibration C. Click the AUTO icon.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 7. Trouble Shooting 7.1. DynaVue System Symptom What It Means 1. Check power cord. Green light for On/Off Switch is not turning 2.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 7.2.3. Detector connection error Check if the LAN cable is connected correctly. Check the power of the detector.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 8. Life Expectancy of Components This list is a detector-oriented list. See below chart for life expectancy of major components used in the system.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 9. Part Lists 9.1 System Parts List Q’ty Part Name Specification Unit DYNAVUE DUO Detector RFA-1717DIC Detector AC Adapter...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 9.3 Harness Parts List Q’ty Part No. Part Name Specification H0-DNV-0001 Inlet to AVR Input 900mm H0-DNV-0002 Inlet to Earth Bar...
Page 77
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 H0-DNV-0031 I/O Board to SSR Signal 700mm H0-DNV-0032 I/O Board to Foot Switch 1,200mm H0-DNV-0033 Foot Switch to I/O Board...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 9.4 Mechanical Parts List Q’ty Part No. Part Name Body Ass’y A-C2-0100 Column Frame Ass’y AC2-0200 Monoblock Ass’y AC2-0300 IMD Inverter Sub Ass’y...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 11. Storage and Maintenance after Usage 11.1 Power Shut Down Make sure the system is completely stopped. If the system is being used, please wait until the work is finished then turn the power off.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 12. Appendix 1: Caution 12.1 Installation 12.1.1 Location 1) Stay away from contact with water or any other form of liquid.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 12.2 For Operator 1) The operator should be equipped with sufficient protective gear. 2) The operator should keep the proper distance away from the system to avoid unnecessary exposure to X-Ray.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 12.8 Etc. Check if the subject has a history of implanted pacemaker or heart defibrillator. Pulse X-Ray used in system is known to cause damage/malfunctioning of such equipment. If the subject does have such a precondition and do not expose directly over the equipment and minimize the time of exposure.
⑧ ⑨ ⑦ 1) Generator STATUS: Show the Generator’s status 2) Information: Show the DynaVue Duo information 3) INIT: Select Port, Connect, Disconnect 4) MODE: Select mode. (Normal, Calibration, RAD, FLO) 5) STATUS: Show the SW’s status 6) X-RAY CONDITION: Adjust X-Ray condition. (KVP, MA, MS and Input Value)
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 14.2 Main Function 1) Set the X-Ray condition. 2) Shoot X-Ray and check the return feedback from I/O board.
Page 89
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 7) The value added to IMD_Val and Offset is entered into IMD Generator. Ex) IMD_Val (4230) + Offset (0) = Input (4230) 8) Separate REF Data is used for “RAD”...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 14.4 Debug 1) This is for checking the Feedback from the generator. 2) Use “X-RAY SHOT” button to take a shot.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 7) After the X-RAY shot, user can check the return feedback on the “RETURN FEEDBACK”. 8) Refer to “RETURN FEEDBACK” and proceed with debugging while adjusting the IMD VALUE.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 2) For more detail information about Error and Warning statues, please refer to the “12.10 Error and Warning statues” section.
Page 93
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 “CAIBRATION REF DATA”. 4) Click the “CALIBRATION” to enter Calibration mode. DynaVue Duo Service Manual 93/104 Page...
Page 94
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 5) Choose “RAD” or “FLO”. 6) Input the IMD value instead of MA. 7) Shot the X-Ray with “X-RAY SHOT” button.
Page 95
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 9) Repeat 6 ~8 until desired “RETURN FEEDBACK” is released. 10) Save the “GLOBAL OFFSET” obtained through process 8.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 14.7 Maximum KW settings for each MS 1) Open “50KVP_RAD.csv” in “ ExamVue Generator IMD_Duo IMDData InputData ”...
Page 97
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 DL17 = ON (Stator driver command run) when the firmware commands the stator driver to run. DL18 = ON (Stator driver command brake) when the firmware commands the stator driver to brake.
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Warning and Error Status 14.9 Type Code Device Possible Causes Solution (W/E) Fault Check the detector connection ᅳ...
Page 99
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Check Detector connection status 14.9.1 Cause) Loss the connections between the Detector and the PC. Solution) Check the detector connection status.
Page 100
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 3) Check the Connection status with “COM RX HW IN”, “INV ON” and “FIL.OK” Check Generator calibration data 14.9.4...
Page 101
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 Check Inverter status 14.9.6 Cause) Solution) Please check the inverter of the generator. 1) Run X-Ray Console program 2) Check the “COM 1”...
Page 102
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 1) Run X-Ray Console program 2) Check the “COM 1” for Generator and “COM4” for I/O board and click “Connect”...
JPI Healthcare Co., Ltd. DOC. NO.: JDP-014-DynaVue Duo 608-HO, 28, DIGITAL-RO 33-GIL, GURO-GU Issued Date: 2024.03.26 SEOUL, 08377, KOREA Revision No. 1.3 S219 Board Not Programmed 14.9.11 S219 B/D, RED LED Status) ALL (2,4,5,6,7) = Blink Cause) Programmed Data Loss Solution) Check the input voltage.